11 Measurement System Analysis
In June 2010, the 4th edition of the measurement systems analysis manual was launched. This version brings significant changes compared to the third edition. In our opinion, the main changes focus on the criteria for analyzing results. However, we also had changes in statistical techniques. Below, we present a brief discussion of the main changes.
The MSA fourth edition presents some changes in relation to the third edition
The main changes refer to:
-
Calibration System;
-
Criteria and form of analysis of the trend and linearity of measurement systems;
-
Criteria for analyzing the RR;
-
Better interpretation and analysis of attributional systems (pass/no pass);
-
Use of alternative techniques to evaluate non-replicable measurement systems;
Below, we comment on the main changes in the fourth edition
-
On page 10 of the fourth edition, a specific topic about the calibration system was added. In summary, an organization must have an internal calibration laboratory or an external organization that controls and maintains elements of calibration events. The calibration system is part of the scope of the organization’s quality management system and must be included in the internal audit requirements. When the calibration event is carried out by an external supplier (commercial or not), it can (or must) be accredited in accordance with ISO/IEC 17025. When there is no accredited laboratory, the calibration service must be carried out by the equipment supplier.
-
On page 77 of the fourth edition, the results analysis criteria were changed (section D). First, a criterion was added to analyze the process of fixing and assembling the measuring device. The criteria for analyzing trend and linearity are similar, with some changes in the way the results are calculated and interpreted. The criteria for variability analysis (RR) have essentially changed:
When beginning an analysis of an organization’s measurement systems, it is useful to identify the priorities that measurement systems should initially focus on. Since the total (or final) variation is based on the combination of process and measurement system variation
$(\sigma_{total}=\sqrt{\sigma^2_{processo}+\sigma^2_{MSA}})$,
When SPC is being applied to control the process or collect data, and the control chart indicates that the process is under statistical control (stable) and the total variability is acceptable, the measurement system can be considered acceptable for use and does not require a separate reanalysis. If an out-of-control condition or nonconformity is detected, we must first analyze the measurement system.
Comments
If we have a CEP graph for a certain characteristic, which is stable and with good capacity, IT IS NOT NECESSARY TO APPLY MSA to evaluate the measurement system. Unless an out-of-control point or non-conformity is detected.
Next, we have the RR analysis table.
| RR | Decision | Comments |
|---|---|---|
| Below 10% | Measurement system generally considered acceptable | Recommended, especially useful when trying to order or classify parts or when tight process control is required. |
| Between 10% and 30% | May be accepted for some applications | The decision should first be based on, for example, the importance of the measurement application, cost of the measuring device, cost of rework or repair. The measuring system must be approved by the customer. |
| Above 30% | Considered unacceptable | Every effort must be made to improve the measurement system. This condition can be resolved by using an appropriate measurement strategy; for example, using the average of several measurements of the same characteristic of the same part in order to reduce the variability of the final measurement. |
The NDC analysis is the same, that is, the NDC must be greater than or equal to five. We have a small modification in the calculation of this index to avoid values equal to zero.
Caution: Using RR as the only index to evaluate a measurement system is not acceptable.
By applying the acceptance criteria as simple thresholds, we assume that the statistics are deterministic estimates of the variability of the measurement system (which they are not). Specifying cutoff values as criteria may lead to inappropriate behavior. For example, the supplier can be “creative” in finding a certain RR value, eliminating the main sources of variation (such as the part x operator interaction) or simply manipulating the study.
Comments
Unfortunately, this is a fact that occurs in many companies in Brazil. The customer imposes a RR below 10% and the supplier manipulates the data. WE SHOULD NOT HAVE A SIMPLE (SINGLE) CRITERIA FOR ALL MEASUREMENT SYSTEMS. Each application must be evaluated individually.
When we analyze the variation of a measurement system, it is important to look at each application individually, to know what is required and how this measurement will be used. For example: the required temperature measurement accuracy may be different for dissimilar applications. A room thermostat can regulate the temperature for human comfort and is cheap, but has an RR above 30%. This is acceptable for this application. But in a laboratory, where small temperature variations can impact test results, more sophisticated temperature measurement and control may be required. This thermostat will be more expensive and we will also require less variability (lower RR value).
Variable Measurement System Studies
1. Stability: Nothing has changed.
2. Trend (page 88): In this study, we made some changes. First, the independent sample method (Student’s t-test) was introduced to assess the trend. The mean and range method is not included in the fourth edition. As a criterion, we can analyze the P-value or the confidence interval. Second, to perform trend analysis, we need validate variability associated with repeatability (the standard deviation of the data),
$$ (porcentagem) VE = 100*\frac{\sigma_{repetibilidade}}{VT} $$
in which $\sigma_{repetibilidade}$ is the standard deviation of the data and the total variation $VT$ is based on process variation (preferred) or process tolerance divided by 6. If the $(porcentagem) VE$ is high (see table above), then the measurement system may be inadequate. Since trend analysis admits that repetitiveness is acceptable, continuing with the analysis may lead to a contradictory or erroneous result, that is, the analysis may indicate a statistically null trend, while its absolute value may exceed what is acceptable for the equipment.
Comment
Finally they removed the mean and range method. Only in the CEP graph method for trend analysis do we refer to the mean and range method, if we have evaluated stability with the graph $\overline{X}$ and R. Another point is the validation of repeatability before concluding about the trend. Here, in our opinion, it is better to carry out the RR study before the trend. By performing the RR we can validate the repeatability.
3. Linearity: We also need to validate the variability associated with repeatability before concluding about linearity.
4. Repeatability and Reproducibility: The three methods were maintained: amplitude, mean and amplitude and ANOVA. However, the ANOVA method is recommended (page 101), as it is more complete and flexible.
4.1 Amplitude Method: Nothing has been changed.
4.2 Mean and range method: First, the minimum number of pieces changed from 5 to 10 pieces. The main change is in determining the process variability. In general, we have four methods for determining process variation (page 121):
-
Current process variation
-
process variation obtained through parts used in the RR study;
-
use when the selected parts represent the expected process variation;
-
-
Variation of an alternative process
- use when we do not have a sufficient number of parts that represent the process, but there is a process whose process variation is similar;
-
Target value of Pp (or Ppk)
- use when we do not have a sufficient number of parts representing the process and we do not have a process with similar variation, or the new process is expected to have lower variability than the current process;
-
Tolerance
- when the measurement system is used for a type of process and the process has Pp less than 1;
Comment
One of the main points for determining the RR indices is the process variation. In general, the 10 parts selected for the RR study do not represent the process variation well. In this sense, the fourth edition emphasizes the use of process history, target Pp value or tolerance.
4.3 ANOVA: No changes. However, the same observations regarding the estimation of process variation that we made in the mean and amplitude method apply.
Measurement System Studies by attribute
Initially, emphasis was placed on determining the gray area (page 132). Consider an attribute measurement system that compares each part with specification limits, the system accepts the part if it is within specifications and rejects it otherwise (known as a pass-no-pass system). Like any measurement system, there is a gray area around specification limits in which the measurement system makes misclassifications.
Since we do not know, a priori, the gray area, we must carry out studies of the measurement system. However, to determine the risk areas around the specification limits, we need to choose approximately 25% of the parts “close” to the lower limit and 25% of the parts “close” to the upper limit. In cases where it is difficult to make such pieces, the team may decide to use a lower percentage, despite recognizing that this attitude may increase the variability of results. If it is not possible to make parts close to specification limits, the team should reconsider the use of an attribute measurement system for this process. For each characteristic, the parts must be measured by a variable measurement system with acceptable variability. When a characteristic cannot be measured by a variable measurement system (for example, visual), we use other means, such as expert classification. Three operators are chosen and each operator takes three measurements of each part.
Comment
Among the chosen parts (for example 50) that we use to carry out the study of an attribute measurement system, we must choose 25% (around 12) “close” to the lower specification limit and 25% (around 12) “close” to the upper specification limit.
Sample Size (page 140)
Another interesting point of the fourth edition in relation to attribute measurement systems is the sample size. How many parts should we use to carry out the attribute measurement system study? To the dismay of users, the answer is “enough”. The purpose of studying a measurement system (attribute or variable) is to know its properties. A sufficient number of samples must be selected to cover an expected range of operation. In the case of an attribute system, the region of interest is the gray areas. If the process capability is good, then a small sample may not contain many parts in the gray area. This means that a process with good capacity requires a larger sample.
In the example mentioned in the fourth edition, for a Pp=Ppk=0.5 (in which we expect 13% of non-conforming parts), 50 parts were selected to carry out the study of the measurement system by attribute.
An alternative to avoiding large samples is to choose the pieces directly in the gray area to ensure that the effect of the evaluator’s variability will be visualized.
Signal Detection Method (page 143)
An alternative procedure for evaluating an attribute measurement system has been given considerable emphasis. In the fourth edition, we have a much more detailed description than what we found in the third edition.
Analytical method
Some accounts have been corrected. For example, on page 146, the t-statistic value has been corrected.
In this module, we will present the main tools for measuring system analysis according to the automotive industry measurement system analysis manual (MSA fourth edition). Although we follow the automotive industry, the methods presented in this module apply to any measurement system.
Measurement System: the set of operations, procedures, measuring devices and other equipment, software and personnel used to assign a number to the characteristic being measured; the complete process used to obtain the measurements.
Below we find the topics with content about MSA:
11.1 Analysis of Measurement Systems
Measurement data has been used in many different ways. The decision whether or not to adjust a manufacturing process is based on measurement data. Measurement data or data derived therefrom are compared to Statistical Process Control limits. If this comparison indicates that the process is outside of statistical control, some type of adjustment must be made. Otherwise, the process may continue without adjustments.
Another important use of measurement data is in planning experiments. Experiment planning allows you to understand the effect of different factors that may vary within a process, such as: raw material, operating conditions, types of machine adjustments, among others. In this case, the analysis of the effect of these factors depends on measurement data from a part, for example. In general, studies that explore this relationship are called analytical studies by Dr. W. E. Deming. The analytical study is one that increases knowledge about the system of causes that affect the process. These studies are among the most important applications of measurement data, as they have recently led to better understanding of products and processes.
The benefits obtained from the use of data-driven procedures are directly determined by the quality of the measurement data used. If the quality is low, the benefit of the procedure will likely be low. Similarly, if the quality is high, the benefit should be high as well. Therefore, due care must be taken when obtaining the data, as the benefits arising from the use of these data must exceed the costs of obtaining them.
The quality of measurement data is related to the statistical properties of multiple measurements obtained by the measurement system. Suppose that a measurement system that is operating under stabilized conditions is used to collect several measurements of a certain characteristic of a part. If the measurements obtained are all “close” to the reference value, then the quality of the data will be said to be “high”. Likewise, if some or all measurements are “away” from the reference value, then the quality of the data will be considered low.
The statistical properties most used to characterize data quality are trends and variances. The property called trend refers to the location of the data in relation to the reference value, the property called variance refers to the dispersion of the data. However, other statistical properties, such as misclassification rate, may be appropriate in some cases, such as attribute measurement systems.
One of the most common reasons that generates low quality data is the large variation in the data. Much of the variation in a set of measurements is due to the interaction between the measurement system and its medium. If this interaction generates very high variation, the quality of the data may be so low that it will be useless.
Much of the work involved in managing measurement systems aims to monitor and control variations. Among other things, this means that we must know how the measurement system interacts with its environment, so that only data of acceptable quality is generated.
Below we find the topics on Analysis of Measurement systems:
11.1.1 Measurement System
The main point for analysis is to interpret the measurement system as a process. Therefore, it is important to emphasize that we are not simply evaluating the equipment, but the process in which we use the equipment, the method and the people to obtain the measurement result.
Measurement System
It is the collection of measuring instruments or devices, standards, operations, methods, fixtures, software, personnel, environment and premises used to quantify the unit of measurement or correct the assessment of a characteristic being measured; the complete process to obtain measurements.
Measurement is the set of operations aimed at determining the value of a quantity. These operations can be performed automatically. (ISO GUM, 2008).

Figure 1.1.1: Measurement system
The objective of a measurement is to determine the value of a quantity to be measured. This measurement begins with an appropriate specification of the quantity, measurement method and procedure.
Example 1.1.1
Consider a measuring system to measure the diameter of a tap connector with a tolerance of +/- 0.5 mm.

Figure 1.1.2: Measuring a tap connector
Before any statistical analysis we must obtain a good definition of the measurement system. Below, we present in a simplified way the measurement system for measuring the connector diameter.
Measurement system definition:
- Measuring equipment: digital caliper with 0.01mm resolution;
Note that the measuring equipment (caliper) has an adequate resolution for the characteristic we are going to measure, as we have a tolerance of +/- 0.5 mm, which corresponds to a range of 1 mm. When we divide the tolerance by 10, we obtain that the minimum accuracy required is 0.1 mm. As the digital caliper has a resolution of 0.01 mm, we concluded that it is suitable for carrying out such a measurement.
Measurement method:
- Position the caliper in the center of the connector;

Figure 1.1.3: measurement method
- Carry out the measurement
Measurement Error
Every measurement has imperfections that give rise to errors in the measurement result. Traditionally, an error is viewed as having two components, namely a random component and a systematic component.
An ideal measurement system would produce only “correct” measurements each time it is used. However, measurement systems with such a property do not exist. Measurement error is defined by:

Figure 1.1.4: Measurement error
Error is an idealized concept and errors cannot be known exactly. In practice, we associate a random variable (for example, the normal distribution) to represent the measurement error.

Figure 1.1.5: Measurement error
-
Uncertainty: σ;
-
RR: σ.
Don’t confuse σ with error!
In general, there is some confusion between the meaning of RR and measurement uncertainty. Measurement uncertainty is a term used internationally to describe the quality of a measurement. Until a few years ago, this term was often associated with the quality of measuring equipment. However, due to the importance of some tests, the concept of measurement uncertainty was extended to establish the quality of the tests.
Measurement uncertainty is a value associated with the measurement result that describes a range in which we expect to contain the true value of the measurement, with a certain level of confidence.
In the calibration process, in which we evaluate measuring equipment, measurement uncertainty corresponds to the standard deviation (or, multiple thereof) associated with the measuring equipment measurements obtained under ideal measurement conditions (calibration). In calibration, the equipment is compared with respect to a reference standard in a laboratory with controlled environmental conditions. Furthermore, we use a properly trained technician to carry out such comparison.
In a test, we must consider the measurement system as a process, in which we have equipment, methods, software, personnel, everything involved in the measurement. The RR aims to quantify the variability associated with measurements of the measurement system (equipment, method and personnel) obtained under real conditions of use of the measurement system.
Thus, we can interpret the uncertainty in a calibration as the uncertainty due to the equipment we use, while the RR characterizes the variations due to the interactions of the measurement system with the environment in which it is inserted (mainly, the method, people, environment, product, among others). In general, measurement uncertainty determines a confidence interval associated with the measurement result, while RR is concerned with evaluating sources of variation and determining the suitability of the measurement system to control the process or evaluate a product. When studying a test, we often use techniques such as RR to help determine measurement uncertainty.
Types of Errors
Two types of errors will be characteristic of this study:
Random Error
A random error is one that occurs unexpectedly and with intensity that damages our measurements. This error represents the small variations that occur in repeated measurements of a quantity. These variations are caused by environmental or spatial changes, variation due to measuring equipment, electrical interference, among others. Although it is not possible to compensate for random error, it can generally be reduced by increasing the number of observations or improving the measurement system technology (better environment, new equipment, or technician training). We interpret random error as a random variable with zero mean.
Systematic Error
A systematic error is one that occurs in all measurements with more or less the same intensity. Just like random error, systematic error cannot be eliminated, but it can often be reduced. Suppose that a systematic error originates from a recognized effect of an influencing quantity on a measurement result. If this effect can be quantified and is significant with respect to the required measurement accuracy, a correction or correction factor can be applied to compensate for the effect. We assume that, after this correction, the expectation or expected value of the systematic error is zero.
Below we present the Ishikawa (fishbone) diagram to describe the main components of measurement error:

Figure 1.1.6: Ishikawa diagram
Example 1.1.2
Description of the main components of measurement error for the tap connector diameter measurement system.

Figure 1.1.7: Ishikawa diagram for faucet connector
Requirements of a measurement system
A “poor quality” system could mask the real variation in the process or product, leading to wrong conclusions:

Figure 1.1.8: “Poor quality” system
There are certain fundamental properties that define a good measurement system:
-
An adequate discrimination or sensitivity. The measurement increment must be small enough to detect variations in the process or specification limits. The common rule is known as the rule of ten, which consists of defining the discrimination of the measurement system by dividing the tolerance (or process variation) into 10 parts.
-
The measurement system must be under statistical control. This means that under repeatability conditions, measurement system variations are due to common causes and not special causes.
-
For product control, measurement system variability should be small compared to specification limits. Compare measurement system variability with product tolerances.
-
For process control, measurement system variability must demonstrate effective and small resolution compared to manufacturing process variation. Compare measurement system variability with 6-sigma process variation and/or total variation.

Figure 1.1.9: Assessment of measurement error
11.1.2 Planning and Strategy
Not every characteristic of the process or product requires a detailed analysis like the one we are developing. For simple measurement systems, such as systems determined by calipers, micrometers or calipers, they often do not require detailed analysis. The basic rule for choosing the system to be evaluated is whether it is identified in the control plan or is important to determine whether or not the process or product is rejected. Another indication is the level of tolerance determined for the specific dimension and the criticality towards the customer. However,
common sense is the guide in any case.
Guidelines for measurement system analysis
- Discriminate the quantities listed in the control plans;
- Identify measurement systems
- Set priorities
- Customer
- Scrap
- Complexity
- Identify a cross-functional team
- For each prioritized measurement system:
- Develop a flowchart of the measurement process;
- Train those involved;
- Develop the Ishikawa diagram;
- Choose statistical tools;
- Create a tool application schedule;
- Document solutions and corrections;
- Institutionalize change.

Measurement Error Model

11.2 Replicable Measurement System
The replicable measurement system is one for which we can measure the same characteristic of a part several times without damaging it. For example, the tap connector diameter measurement system is replicable, as we can measure the same connector several times under repetitive conditions.
11.2.1 Stability
Definitions
Stability is the amount of total variation in the system trend over time in a given part or standard part. Before studying any statistical properties of the measurement system, let’s analyze the system’s ability to maintain such properties over time. The objective of stability is to evaluate:
-
The interaction of the measurement system and the environment;
-
Component wear;
-
Adjustment of devices and sensors.
Guidelines for replicable system
-
Select and identify a part;
-
Prepare data collection form and logbook (date, time, operator, measuring equipment);
-
Measure the piece periodically (daily, weekly, fortnightly or monthly) with 3 to 5 measurements at a time (rational subgroup);
-
After 20 or more rational subgroups, build the graph $ \overline{X} $ and R, as described below.
Table
| Graph Limits | $N^o$ of element. sample (n) | $A_2$ | $D_3$ | $D_4$ |
|---|---|---|---|---|
| Averages Chart $ \overline{X} $ | 2 | 1,880 | 0 | 3,267 |
| LSC = Upper Limit = $ \overline{X} + A_2R $ | 3 | 1,023 | 0 | 2,574 |
| LC = Central Limit = $ \overline{X} $ | 4 | 0,729 | 0 | 2,282 |
| LIC = Lower Limit = $ \overline{X} - A_2\overline{R} $ | 5 | 0,577 | 0 | 2,114 |
| R Amplitude Graph | 6 | 0,483 | 0 | 2,004 |
| LSC = Upper Limit = $D_4\overline{R}$ | 7 | 0,419 | 0,076 | 1,924 |
| LC = Central Limit = $\overline{R}$ | 8 | 0,373 | 0,136 | 1,864 |
| LIC = Lower Limit = $D_3\overline{R}$ | 9 | 0,337 | 0,184 | 1,816 |
| 10 | 0,308 | 0,223 | 1,777 |
Assessment Criteria
Analyze the graphs $ \overline{X} $ and R. Firstly the R graph and then the graph $ \overline{X} $:
-
Points outside the control limits.
-
7 or more consecutive increasing or decreasing points.
-
7 or more consecutive points above or below the midline.
If the graphics $ \overline{X} $ and R are out of control, investigate the causes and establish corrective actions.
-
If the process lacks stability, identify the causes and establish corrective action. Repeat the stability study;
-
If the process is stable, proceed with the study of the measurement system;
-
If it is not possible to stabilize the measurement process, carry out studies over time to identify long-term variations.
Discrimination of the measurement system in the stability study
Ability of the measurement system to reliably detect and indicate small variations in the quantity being measured.
One way to quantify the discriminating power is to express the smallest variation in quantity that the measurement system can detect.
Assessment criteria
Check that the R control chart does not show too many amplitudes equal to zero (above 30%). If this occurs, there is good evidence that the measuring equipment does not have adequate resolution for this measurement. In this case, carry out a critical analysis.
Example 2.1.1
The process technician must carry out a study on the stability of the measuring system to evaluate the diameter of a steel bar with a micrometer. The technician selected 1 piece, which was measured at a daily frequency and sub-group size 3, by an evaluator. The values are in Table 1.
| Date | Schedule | Measure 1 | Measure 2 | Measure 3 |
|---|---|---|---|---|
| Aug 6 | 09:15 | 4,202 | 4,201 | 4,202 |
| Aug/13 | 16:35 | 4,201 | 4,202 | 4,203 |
| Aug 20 | 14:13 | 4,199 | 4,198 | 4,200 |
| Aug 27 | 09:40 | 4,200 | 4,201 | 4,201 |
| 4/set | 15:28 | 4,200 | 4,201 | 4,200 |
| 11/Sep | 10:39 | 4,202 | 4,201 | 4,200 |
| 19/Sep | 15:10 | 4,200 | 4,201 | 4,200 |
| 25/Sep | 09:25 | 4,200 | 4,199 | 4,199 |
| 1/Oct | 15:40 | 4,198 | 4,199 | 4,199 |
| 8/Oct | 09:25 | 4,200 | 4,202 | 4,200 |
| Oct 16 | 16:10 | 4,202 | 4,203 | 4,203 |
| Oct 24 | 10:05 | 4,201 | 4,202 | 4,201 |
| 1/nov | 13:40 | 4,199 | 4,199 | 4,198 |
| 8/nov | 14:55 | 4,200 | 4,200 | 4,201 |
| Nov/14 | 11:00 | 4,199 | 4,198 | 4,199 |
| Nov 22 | 15:50 | 4,200 | 4,199 | 4,200 |
| Nov 29 | 09:42 | 4,201 | 4,201 | 4,200 |
| 7/dec | 08:20 | 4,199 | 4,200 | 4,199 |
| 12/dec | 15:30 | 4,200 | 4,201 | 4,199 |
| 20/dec | 11:05 | 4,199 | 4,199 | 4,200 |
| 28/dec | 15:30 | 4,201 | 4,200 | 4,199 |
| 4/jan | 16:00 | 4,200 | 4,200 | 4,202 |
| 10/jan | 15:15 | 4,203 | 4,204 | 4,203 |
| 15/jan | 16:00 | 4,204 | 4,203 | 4,203 |
Table 2.1.1: Steel bar diameters
From the data in the table we take the average of the values in the Average column and the average of the values in the Amplitude column and obtain $ \overline{\overline{X}} $ = 4.200486 and $ \overline{R} $ = 0,001292
Control limits are calculated as follows: Chart R
Since we have 3 elements in our sample, we obtain a value of $D_{3}$ = 0 and $D_{4}$ = 2.574, with this:
LSC = 2.574*0.001292 = 0.003325
LIC = 0*0.001292 = 0

Chart $ \overline{X} $
Sample size n = 3, $A_{2}$ = 1.023, we obtain the following control limits:
LSC = 4.200486+1.023*0.001292 = 4.201807
LIC = 4.200486 -1.023*0.001292 = 4.199165

Interpreting the stability study is not simple and requires good knowledge of the needs of the measurement system. In the case above, we observed that the measurement system presents some symptoms of instability, we have 8 consecutive points below the central line on the amplitude graph and several points outside the control limits on the average graph. However, we must evaluate the results very carefully. First, we observed that the average amplitude of 0.0012mm is very close to the equipment resolution and has an upper limit of 0.003 mm, which in general is similar to the maximum allowable error for evaluating the calibration certificate of this type of equipment. Second, the biggest difference between the average measurements of the metal bar was 0.004mm, in which we have contributions from the lack of homogeneity of the bar (ovalization).
In short, we have symptoms of instability in the application of the measuring system, however, this instability results in variations of a maximum of 0.004mm. Therefore, the acceptability or otherwise of the measurement system must take into account the application and the need for more or less careful measurements. Depending on the application of the measurement system, we may or may not accept this measurement system. If we reject this measurement system, we need an action plan to eliminate the special causes. Otherwise, we accept system instability as negligible.
Example 2.1.2
The process technician must carry out a study on the stability of the measuring system to evaluate the internal diameter of the compressor block in the machining process (Bronzing). The technician selected 1 piece, which was measured at a daily frequency and sub-group size 3, by an evaluator. The values are in Table 2.
| Date | Schedule | Measure 1 | Measure 2 | Measure 3 |
|---|---|---|---|---|
| 12/8/2011 | 08:00 | 85,011 | 85,012 | 85,012 |
| 12/8/2011 | 14:30 | 85,011 | 85,011 | 85,011 |
| 13/8/2011 | 02:40 | 85,010 | 85,011 | 85,010 |
| 13/8/2011 | 10:30 | 85,009 | 85,010 | 85,011 |
| 15/8/2011 | 08:25 | 85,010 | 85,011 | 85,011 |
| 15/8/2011 | 14:35 | 85,011 | 85,011 | 85,010 |
| 16/8/2011 | 02:10 | 85,010 | 85,011 | 85,010 |
| 16/8/2011 | 10:10 | 85,011 | 85,012 | 85,012 |
| 16/8/2011 | 15:30 | 85,009 | 85,009 | 85,009 |
| 17/8/2011 | 02:45 | 85,012 | 85,011 | 85,012 |
| 17/8/2011 | 17:30 | 85,010 | 85,009 | 85,009 |
| 18/8/2011 | 05:10 | 85,012 | 85,012 | 85,011 |
| 18/8/2011 | 10:30 | 85,011 | 85,011 | 85,010 |
| 18/8/2011 | 16:30 | 85,010 | 85,009 | 85,008 |
| 19/8/2011 | 21:15 | 85,009 | 85,010 | 85,009 |
| 22/8/2011 | 09:40 | 85,010 | 85,009 | 85,010 |
| 22/8/2011 | 18:00 | 85,011 | 85,011 | 85,011 |
| 22/8/2011 | 02:30 | 85,012 | 85,011 | 85,011 |
| 23/8/2011 | 17:40 | 85,009 | 85,009 | 85,009 |
| 23/8/2011 | 02:45 | 85,010 | 85,011 | 85,011 |
Table 2.1.2: Block internal diameters
From the data in the table we take the average of the values in the Average column and the average of the values in the Amplitude column and obtain $ \overline{\overline{X}} $= 85.0104 and $ \overline{R} $ = 0,0009
Control limits are calculated as follows: Chart R
Since we have 3 elements in our sample, we obtain a value of $D_{3}$ = 0 and $D_{4}$ = 2.574, with this:
LSC = 2.574*0.0009 = 0.00232
LIC = 0*0.0009 = 0

Chart $ \overline{X} $
Sample size n = 3, $A_{2}$ = 1.023, we obtain the following control limits:
LSC = 85.0104+1.023*0.0009 = 85.0114
LIC = 85.0104 -1.023*0.0009 = 85.0095

Interpretation
Note that we have several out-of-control points on the average graph, this means that we have special causes of variation acting in the measurement system. However, note that the points below are around 85.009 mm and the points above are around 85.0120 mm, which means a maximum variation of 3 microns. Therefore, the decision on the acceptability of this measurement system depends on the application of the measurement system.
11.2.2 Trend
Definitions
Trend is the difference between the average of measurements of a quantity and the reference value for the measured quantity, carried out by an evaluator using the same equipment and method
(image missing)
Trend Study Guidelines
-
Select a production item whose measurement falls within the central range of the process variation;
-
Determine the reference value of the chosen item in relation to a traceable standard. Here, we can use external laboratories, such as laboratories accredited by INMETRO (VR);
-
A rater, trained in the use of the measurement system being analyzed, measures the item 10 times or more (the MSA suggests 12);
-
Calculate the average of measurements (x) and the trend;
$$\text{Tendência} = \overline{X}-VR$$
Example 2.2.1
Let’s evaluate the tendency of a measurement system to measure the height of an “MP3 Player”, with a tolerance of 0.7 mm. This height is measured with an altimeter. An “MP3 Player” was selected (close to the nominal value) and, after 10 measurements carried out by a coordinate measuring machine, the reference value VR = 89.73 mm was determined. Next, the same “MP3 Player” was measured 12 times with the measurement system under analysis. The data are:
| Sample | Measurements |
|---|---|
| 1 | 89,77 |
| 2 | 89,79 |
| 3 | 89,77 |
| 4 | 89,78 |
| 5 | 89,74 |
| 6 | 89,72 |
| 7 | 89,72 |
| 8 | 89,75 |
| 9 | 89,74 |
| 10 | 89,77 |
| 11 | 89,78 |
| 12 | 89,74 |
| VR | 89,73 |
From these data we have
a) Average = 89.7558 mm
b) Trend = (average) - (reference value) = 89.7558 - 89.73 = + 0.0258 mm
We must interpret that on average, the values measured by this evaluator with this instrument and this method are higher than the reference value by 0.0258 mm.
To perform trend analysis, we need validate the variability associated with repeatability (the standard deviation of the data), for this we calculate
$$ (porcentagem) VE = \frac{s}{VT} * 100$$
in which s is the standard deviation of the data and the total variation VT is based on process variation (preferred) or process tolerance divided by 6. If the %VE is high (see table above), then the measurement system may be inadequate. Since trend analysis admits that repetitiveness is acceptable, continuing with the analysis may lead to a contradictory or erroneous result, that is, the analysis may indicate a statistically null trend, while its absolute value may exceed what is acceptable for the equipment. At this point, the MSA fourth edition suggests the same criteria used to analyze RR as criteria for analyzing %VE.
Criteria for evaluating the trend
a) Confidence Interval
The trend is acceptable at the level of significance if zero belongs to the confidence interval (1 - ) * 100% with limits:
$$LI = \hbox{Tendência}-t_{(n-1,1-\alpha/2)}\frac{s}{\sqrt{n}},$$
$$LS = \hbox{Tendência}+t_{(n-1,1-\alpha/2)}\frac{s}{\sqrt{n}}.$$
in which $ t_{(n-1,1-\frac{\alpha}{2})} $ corresponds to the quantile $ 1-\frac{\alpha}{2} $ from the t-Student distribution. Here, we use a significance level $ (\alpha) $ of 5%.
b) Hypothesis Testing
Equivalently, we can perform the following hypothesis test to evaluate the trend:
$$\hbox{Tendência}\neq 0.$$
To do this, note that the t statistic given by:
$$t=\frac{\hbox{Tendência}}{s/\sqrt{n}}\sim t_{n-1},$$
where s is the standard deviation of the measurements, n is the number of measurements and $t_{n-1}$ t-Student distribution with n-1 degrees of freedom.
Therefore, we obtain the following decision rule for a significance level α
-
Se |t| > $t_{(n-1;1-\alpha/2)}$ we reject $H_{0}$, that is, the trend is significant from a statistical point of view;
-
Se |t| ≤ $t_{(n-1;1-\alpha/2)}$ we do not reject $H_{0}$, that is, the trend is not significant from a statistical point of view.
Figure 2.2.1 illustrates the critical region of the test, that is, the values of t for which we reject $H_{0}$.

Figure 2.2.1: Rejection region.
c) P-value
The P-value represents the lowest level of significance for which we reject $ H_0 $. Therefore, for a significance level = 0.05 adopted, we reject $ H_0 $ if the P-value obtained is less than 0.05, while we do not reject $ H_0 $ if the P-value is greater than 0.05, we observe this fact in (Figure 2.2.2). For the t-test, the P-value is calculated in the form
$$p - valor = 2 x P(t_n -1 > |t|)$$
Therefore, we reject $H_{0}$ when the p-value is lower than the significance level $ \alpha $ we do not reject $H_{0}$.

Figure 2.2.2: P-value.
Example 2.2.2
Considering the data from Example 2.2.1 and a value of $ \alpha $ equal to 0.05, we have that:
n = 12
s = 0.02392
$t_{(n-1;1-\alpha/2)}$ = $t_{(11;0,975)}$ = 2,201.
In this example, we will use the process tolerance to determine the total variation (VT). When we divide the tolerance by 6, we obtain that
$$VT = \frac{0,7}{6}=0,117.$$
As s = 0.02392, we have that the %VE of the process is given by
$$(porcentagem)VE=100\frac{0,02392}{0,117}=20,05(porcentagem).$$
Since the %VE is low, we have acceptable variability.
Therefore, the 95% confidence interval for the trend is obtained through the lower (LI) and upper (LS) limits:
$$LI = \hbox{Tendência}-t_{(n-1,1-\alpha/2)}\frac{s}{\sqrt{n}}=0,0258-(2,201)\frac{0,02392}{\sqrt{12}}=0,0106,$$
$$LS = \hbox{Tendência}+t_{(n-1,1-\alpha/2)}\frac{s}{\sqrt{n}}=0,0258+(2,201)\frac{0,02392}{\sqrt{12}}=0,041.$$
Conclusion
As zero is not part of the range, the trend found (0.0258 mm) is significant at the 5% level.
Equivalently, we can perform the test for the hypothesis $H_{0}$ : Trend = 0. To test this hypothesis, we take
$$t=\frac{\hbox{Tendência}}{s/\sqrt{n}}=\frac{0,0258}{0,02392}=3,741.$$
As t = 3.741 > $t_{(n-1;1-\alpha/2)}$ = 2.201, we reject the hypothesis $H_{0}$, that is, the trend is non-zero at a 95% confidence level. The hypothesis $H_{0}$ It can also be evaluated by the p-value, which is calculated as:
$$p-valor = 2 x P(t_{n-1} > |t|) = 2 x (1 - P(t_{11}>3,741)) = 0,00325.$$
As the p-value is smaller than the $\alpha$ adopted (0.05), we reject $H_{0}$at the 5% significance level.
Interpretation of results
As always, we must know the application of the measurement system to carry out a good critical analysis. In this case, we have a trend of 0.0258 mm with an upper limit of 0.041 mm (confidence interval for the trend). The equipment used for measurement is a digital altimeter with a resolution of 0.01mm. This equipment has a measurement uncertainty of around 0.03mm with accuracy (manufacturer specification) also around 3 hundredths of a millimeter. Note that the accuracy proposed by the manufacturer is above the trend found and close to the upper limit (0.041 mm). Additionally, we have a tolerance for the part of 0.7 mm. Obviously, we have an opportunity for improvement as zero is outside the confidence interval. However, the need for an action plan depends on the need for more or less precise measurements, that is, it depends on the application of the measurement system.
See below the results obtained by the Action software for the same example.

Trend analysis
If the trend is relatively large, look for these possible causes:
-
Error in measuring the standard piece;
-
Worn components;
-
Measuring device made for wrong dimension;
-
Measuring device measuring wrong characteristic;
-
Improperly calibrated measuring device;
-
Measuring device used improperly by the evaluator.
Example 2.2.3
An engineer is evaluating a new measurement system to monitor a process. An analysis of the measurement system indicated that there should be no concern about linearity, as the range of interest is small. A single piece was chosen in such a way that it is close to the nominal value of the processes. The part was measured by a sophisticated measuring system to determine its reference value (reference value = 6). The part was then measured 15 times by an operator and the total process variability value is 2.5, this value will be used to validate the repeatability of the measurement system. The values are given in the table below.
| Trials | Measurement | Trend |
|---|---|---|
| 1 | 5,8 | -0,2 |
| 2 | 5,7 | -0,3 |
| 3 | 5,9 | -0,1 |
| 4 | 5,9 | -0,1 |
| 5 | 6,0 | 0,0 |
| 6 | 6,1 | 0,1 |
| 7 | 6,0 | 0,0 |
| 8 | 6,1 | 0,1 |
| 9 | 6,4 | 0,4 |
| 10 | 6,3 | 0,3 |
| 11 | 6,0 | 0,0 |
| 12 | 6,1 | 0,1 |
| 13 | 6,2 | 0,2 |
| 14 | 5,6 | -0,4 |
| 15 | 6,0 | 0,0 |
n = 15
s = 0.212
$$t_{(n-1; 1-\frac{\alpha}{2})}=t_{14;0,975}=2,144$$
$$VT =2,5.$$
As s = 0.212, we have that the %VE of the process is given by
$$(porcentagem)VE=100\frac{0,212}{2,5}=8,48(porcentagem).$$
Since the %VE is low, we have acceptable variability.
Therefore, the 95% confidence interval for the trend is obtained through the lower (LI) and upper (LS) limits:
$$LI = \hbox{Tendência}-t_{(n-1,1-\alpha/2)}\frac{s}{\sqrt{n}}=0,006667-(2,144)\frac{0,212}{\sqrt{15}}=-0,11075,$$
$$LS = \hbox{Tendência}+t_{(n-1,1-\alpha/2)}\frac{s}{\sqrt{n}}=0,006667+(2,144)\frac{0,212}{\sqrt{15}}=0,124079.$$
Conclusion
As zero is part of the range, the trend found (0.006667 mm) is not significant at the 5% level.
Equivalently, we can perform hypothesis testing to $H_{0}$ : Trend = 0. To test this hypothesis, we take
$$t=\frac{\hbox{Tendência}}{s/\sqrt{n}}=\frac{0,006667}{0,0547}=0,12178.$$
How $ t= 0,12178 < t_{(n-1;1-\frac{\alpha}{2})}=2,144, $ We do not reject the hypothesis $H_{0}$, that is, the trend is negligible at the 95% confidence level. The hypothesis $H_{0}$ It can also be evaluated by the p-value, which is calculated as:
$$p-valor = 2 x P(t_{n-1} > |t|) = 2 x (1 - P(t_{14}>0,12178)) = 0,9048.$$
As the p-value is greater than the $\alpha$ adopted (0.05), we do not reject $H_{0}$ at the 5% significance level.
See below the results obtained by the Action software for the same example.
Example 2.2.4
A dimensional measurement is controlled with a measuring system using a micrometer with a resolution of 0.002 mm. The specification of this dimensional characteristic is 13,000 ${\pm}$ 0.020mm. The evaluator wants to know whether the measurement system is biased. To do this, he chose a part whose measurement is close to the nominal value (13.000 mm), with the part being measured using a measuring system with a micrometric bench with a resolution of 0.0005 mm. Several measurements of the piece were taken with the micrometric bench, obtaining an average of 13.001 mm. Next, we take 12 readings of the same part with the measuring system using the micrometer.
| Readings |
|---|
| 13,002 |
| 13,002 |
| 13,004 |
| 13,002 |
| 13,004 |
| 13,002 |
| 13,004 |
| 13,002 |
| 13,000 |
| 13,000 |
| 13,000 |
| 13,002 |
See below the results obtained by the Action software for the same example.


Interpretation of Results
Since zero is outside the confidence interval we have a significant trend (of 0.001mm, with an upper limit of 0.00194 mm) at the 95% confidence level. The equipment used in the measurement system is a micrometer with a resolution of 0.002 mm with a measurement uncertainty of around 0.003 mm. Note that the equipment meets the MSA requirement of good discrimination, because when we divide the tolerance (0.04 mm) by 10 we obtain the maximum allowable error of 0.004 mm, both resolution and uncertainty are below 0.004 mm. In this situation, in which the upper limit for the trend is below the instrument’s resolution and the instrument presents adequate discrimination, we conclude that the detected trend is negligible.
11.2.3 Trend and Linearity
Linearity measures the variation in trend for different reference values in the range of interest. Linearity is assessed via the slope of the line formed by the different reference values in relation to the respective trend. The less inclined the straight line, the better the quality of the measurement system.
(image missing)
Guidelines
Guidelines for studying trend and linearity for a replicable system:
Select a sample of parts (at least 5) whose measurements are distributed along the range of interest;
Determine the reference values of the parts. Once again we can use laboratories accredited by INMETRO or an internal laboratory;
Appraiser using the measurement system must measure each piece at least 10 times (MSA suggests 12), in a random sequence;
Determine the trend for each measurement (Trend = Measurement Result - Reference Value);
Graph the (trend) x (reference value);
Assessment
To evaluate the trend and linearity, let’s take the adjustment of the trend in relation to the reference value:
Trend = a + b*(reference value) + Fit error
Determination Coefficient (R²): degree of adjustment of the straight line;
Intercept (a);
Slope (b);
We present some of the metrics used in the fourth edition of the MSA:
Linearity = |b|
%Linearity = |b|* (100%)
Linearity in relation to the measuring scale (nominal range amplitude): L(scale) = |b|*(max - min)
Linear Regression
The linear regression model is given by:
$T_{ij}$ = $a+bVR_i + ε_{ij}$ (2.3.1)
in which
- g : number of pieces (≥ 5);
- m : number of measurements per piece (≥ 12);
- $T_{ij}$: corresponds to the jth trend of the ith reference value (test specimen);
- $VR_i$: corresponds to reference value i;
- $ε_{ij}$ is a normal random variable with zero mean and standard deviation σ (independent);
- a and b are the parameters, which together define the regression line.
Estimates
In this sense, the MSA 4th edition proposes the following tools as criteria:
Test of regression coefficients;
Confidence band for the regression line.
Next, we will study both criteria. To facilitate the calculations, we have established the following table:

Table 2.3.1: Data entry and linearity calculations
Estimates
The trend and reference value averages are given by:
$$\overline{VR}=\frac{1}{g\times m}\sum^g_{i=1}\sum^m_{j=1}VR_i~~~~~~~~\overline{T}=\frac{1}{g\times m}\sum^g_{i=1}\sum^m_{j=1}T_{ij}$$
Basic notations
$$S_{xx}=\sum^g_{i=1}\sum^m_{j=1}(VR_i-\overline{VR})^2=\sum^g_{i=1}\sum^m_{j=1}VR^2_i-g\times m\times \overline{VR}^2~~~~~~~(2.3.2)$$
$$S_{yy}=\sum^g_{i=1}\sum^m_{j=1}(T_{ij}-\overline{T})^2=\sum^g_{i=1}\sum^m_{j=1}T^2_{ij}-g\times m\times \overline{T}^2~~~~~~~(2.3.3)$$
$$S_{xy}=\sum^g_{i=1}\sum^m_{j=1}(VR_{i}-\overline{VR})(T_{ij}-\overline{T})=\sum^g_{i=1}\sum^m_{j=1}VR_i~T_{ij}-\overline{VR}\sum^g_{i=1}\sum^m_{j=1}T_{ij}~~~~~~~(2.3.4)$$
The adjusted model is given by:
$$T_{ij}=\hat{a}+\hat{b}\times VR_i$$
for all i = 1,…,g and j = 1,…,m.
Least squares estimates $ \hat{a} $ and $ \hat{b} $ are given by:
$$\hat{a}=\overline{T}-\hat{b}~\overline{VR}~~~~~~~\hat{b}=\frac{S_{xy}}{S_{xx}}$$
R² is given by:
$$R^2=1-\frac{SQ_E}{S_{yy}}=\frac{(S_{xy})^2}{S_{xx}~S_{yy}}$$
that is, it is the ratio between the product ($S_{xy}S_{xy}$) by the product ($S_{xx}S_{yy}$). And we have,
$$SQ_E=S_{yy}-\hat{b}~S_{xy}$$
Methodologies
Next, we will present two methodologies to test the statistical significance of the regression coefficients.
- Simple Linear Regression Coefficients Test:
1.1) Test for Angular Coefficient
$$~b\neq 0.$$
Test Statistics
$$t^{*}=\frac{\hat{b}}{s(\hat{b})}~\sim~ t_{(g\times m-2)}~~~({sob Ho}),$$
in which
$$s^2(\hat{b})=\frac{QME}{S_{xx}},~~~~QME=\frac{SQ_E}{g*m-2}$$
The value of t* must be compared with a Student t-distribution with g*m-2 degrees of freedom for a given significance level $ \alpha $ (see t-Student distribution table).
- Se |t| ≤ t(g*m-2; 1-($ \alpha $/2)) we do not reject $H_0$, that is, we reject the hypothesis that the angular coefficient is significant;
- If t > t(g*m-2; 1-($ \alpha $/2)) we reject $H_0$, that is, we do not reject the hypothesis that the angular coefficient is significant.
Another way to define a criterion to evaluate the hypothesis test is the P-value.

1.2) Test for the intercept
$$a\neq 0$$
Test Statistics
$$t^{*} = \frac{\hat{a}}{s(\hat{a})}\sim t_{(g*m-2)}~~~ ({sob~Ho})$$
in which,
$$s^2(\hat{a})=\left(\frac{1}{g*m}+\frac{\overline{VR}^2}{S_{xx}}\right)\times QME$$
This value must be compared with a t - Student distribution with (g*(m-2)) degrees of freedom for a given significance level $ \alpha $ (see t-Student distribution table).
- Se |t| j ≤ t(g * m-2; 1-( $ \alpha $ /2)) we do not reject $H_0$, that is, we reject the hypothesis that the intercept is significant;
- Se |t| j > t(g * m-2; 1-( $ \alpha $ /2)) we reject $H_0$, that is, we do not reject the hypothesis that the intercept is significant.
Another way to define a criterion to evaluate the hypothesis test is the P-value. The P-value represents the lowest level of significance for which we reject $H_0$. Therefore, for a significance level $ \alpha $ = 0.05 adopted, we reject $H_0$ if the P-value obtained is less than 0.05, while we do not reject $H_0$ if the P-value is greater than 0.05.

Criterion: Trend and linearity are considered non-significant when we do not reject the hypotheses $H_0$ in the two tests performed above.
- Confidence Interval for regression line:
$$LI= \hat{a} + \hat{b} VR -\left[t_{(g*m-2;1-\alpha/2)}\sqrt{\left(\frac{1}{g*m}+\frac{(VR - \overline{VR})^2}{S_{xx}}\right)\times QME}\right],$$
$$LS= \hat{a} + \hat{b} VR +\left[t_{(g*m-2;1-\alpha/2)}\sqrt{\left(\frac{1}{g*m}+\frac{(VR - \overline{VR})^2}{S_{xx}}\right)\times QME}\right],$$
where VR represents the Reference Value at the point where we calculate the confidence interval.
Criterion: the trend line equal to zero must be completely contained within the above limits.
Note: The coefficient estimates and the confidence interval are described in the MSA Standard 4th edition page 97.
Example 2.3.1
The measurement system engineer was interested in determining the linearity of a measurement system. Five standard parts, which span the entire process variation range, were measured 15 times in the measurement laboratory to determine the reference value. In this case, the metrologist used a measuring instrument with a better resolution than the instrument normally used. After determining the reference value, an evaluator carried out 12 measurements of each standard part under actual conditions of use of the measuring system. The values are summarized in the Table below. Here, we have g=5 (number of pieces) and m=12 (readings on each piece).
| Measurements | 2,0 | Trend | 4,0 | Trend | 6,0 | Trend | 8,0 | Trend | 10,0 | Trend |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 2,7 | 0,7 | 5,1 | 1,1 | 5,8 | -0,2 | 7,6 | -0,4 | 9,1 | -0,9 |
| 2 | 2,5 | 0,5 | 3,9 | -0,1 | 5,7 | -0,3 | 7,7 | -0,3 | 9,3 | -0,7 |
| 3 | 2,4 | 0,4 | 4,2 | 0,2 | 5,9 | -0,1 | 7,8 | -0,2 | 9,5 | -0,5 |
| 4 | 2,5 | 0,5 | 5 | 1 | 5,9 | -0,1 | 7,7 | -0,3 | 9,3 | -0,7 |
| 5 | 2,7 | 0,7 | 3,8 | -0,2 | 6 | 0 | 7,8 | -0,2 | 9,4 | -0,6 |
| 6 | 2,3 | 0,3 | 3,9 | -0,1 | 6,1 | 0,1 | 7,8 | -0,2 | 9,5 | -0,5 |
| 7 | 2,5 | 0,5 | 3,9 | -0,1 | 6 | 0 | 7,8 | -0,2 | 9,5 | -0,5 |
| 8 | 2,5 | 0,5 | 3,9 | -0,1 | 6,1 | 0,1 | 7,7 | -0,3 | 9,5 | -0,5 |
| 9 | 2,4 | 0,4 | 3,9 | -0,1 | 6,4 | 0,4 | 7,8 | -0,2 | 9,6 | -0,4 |
| 10 | 2,4 | 0,4 | 4 | 0 | 6,3 | 0,3 | 7,5 | -0,5 | 9,2 | -0,8 |
| 11 | 2,6 | 0,6 | 4,1 | 0,1 | 6 | 0 | 7,6 | -0,4 | 9,3 | -0,7 |
| 12 | 2,4 | 0,4 | 3,8 | -0,2 | 6,1 | 0,1 | 7,7 | -0,3 | 9,4 | -0,6 |
Estimation of Example Parameters
Next, we will estimate the parameters and apply the methodology defined by MSA 4th Edition.

Table 2.3.2: Data table
First we need to determine the averages of the variables T (trend) and VR (reference value).
$$\overline{VR} = \frac{1}{g*m} \sum^g_{i=1}\sum^m_{j=1} VR_i = \frac{360}{5*12} = 6$$
$$\overline{T} = \frac{1}{g*m} \sum^g_{i=1}\sum^m_{j=1} T_{ij} = \frac{-3,2}{5*12} = -0,05333. $$
Thus, we find the empirical sums of squares.
$$S{xx}=\sum^g_{i=1}\sum^m_{j=1} VR_i^2 - g*m\overline{VR}^2 = 2640 - 60(6)^2 = 480~~~~~(2.3.5)$$
$$S_{yy}=\sum^g_{i=1}\sum^m_{j=1} T_{ij}^2 - g*m\overline{T}^2 = 11,82 - 60(-0,0533)^2 = 11,64955~~~~~~ (2.3.6)$$
$$S_{xy}=\sum^g_{i=1}\sum^m_{j=1} VR_i T_{ij} - \overline{VR}\sum^g_{i=1}\sum^m_{j=1} T_{ij}=-82,4 - (6)(-3,2) = -63,2~~~~~(2.3.7)$$
Next, we use empirical sums of squares to find estimates of the parameters of the regression line. Here, we have to
$$\hat{b} = \frac{S_{xy}}{S_{xx}} = \frac{-63,2}{480} = -0,13167.~~~~(2.3.8)$$
$$\hat{a}=\overline{T} - \hat{b}\overline{VR} = -0,053333 -(-0,13167)(6) = 0,736687~~~(2.3.9)$$
and
$$R^2=\frac{(S_{xy})^2}{S_{xx}S_{yy}}=\frac{(-63,2)^2}{480*11,6495}=0,714308~~~~(2.3.10)$$
The estimated straight line is given by:
$$\hat{T_{ij}}=\hat{a} + \hat{b}*VR_i$$
Estimated trend = 0.7366 - 0.13167 *(Reference Value)
With this, we obtain that the %linearity = 13.2%. This means that the trend varies on average by 13.2% of the study range (10-2=8) between the beginning and end of the range. We will apply the two criteria for evaluating linearity, the parameter tests and the confidence interval for the regression line. To do this, we calculate the mean squared error (QME) by:
$$SQ_E = S_{yy} - \hat{b}S_{xy} = 11,6495 - (-0,13167)(-63,2) = 3,3279$$
$$QME=\frac{SQ_E}{g*m-2} ~=\frac{3,3279}{58} = 0,057377$$
Case 1
Hypothesis testing: Initially, we will test the angular coefficient (b).
$$~b\neq0$$
The test statistics are given by:
$$t^{*} =\frac{\hat{b}}{s(\hat{b})}\sim t_{(g*m-2)}~~~({sob }~H_0)$$
in which the standard error associated with the slope estimate is calculated by:
$$s^2(\hat{b})=\frac{QME}{S_{xx}} ~= \frac{0,057377}{480} = 0,00011953$$
Therefore, the value of the t-Student statistic is given by:
$$t^{*}=\frac{\mid{\hat{b}}\mid}{s(\hat{b})}=\frac{\mid{-0,13167}\mid}{0,010932}=12,044$$
For a t-Student distribution with 58 degrees of freedom we find
t(g*m-2; 1-$ \alpha $/2) = 2,0017.
Since t* = 12.044 > 2.0017 we reject the hypothesis that linearity is not significant. Therefore, we conclude that linearity is significant at the 5% confidence level.
Next, we will perform the test for the intercept
$$~a\neq0$$
In this case, the test statistic is given by:
$$t^{*} =\frac{\hat{a}}{s(\hat{a})}\sim t_{(g*m-2)}~~~ ({sob }~H_0),$$
in which,
$$s^2(\hat{a})=\left(\frac{1}{g*m}+\frac{\overline{VR}^2}{S_{xx}}\right)\times QME =$$
$$=\left(\frac{1}{60}+\frac{(6)^2}{480}\right)\times 0,057377 =$$
$$=\left(0,016667+0,075\right)\times 0,057377 = 0,0052595.$$
With this, we obtain that
$$t^{*} =\frac{\mid{\hat{a}}\mid}{s(\hat{a})}=\frac{0,7366}{0,072522}=10,156$$
For a t-Student distribution with 58 degrees of freedom we find t(gm-2; 1-α/2) = 2.0017. Since t = 10.16 > 2.0017 we reject the hypothesis that the intercept is not significant. Therefore, we conclude that the intercept is significant at the 5% confidence level.
Conclusion
Since the slope (linearity) was considered significant, the measurement system shows significant linearity with 95% confidence ($ \alpha $ = 0,05).
Case 2
Confidence interval for the regression line. We already know that the formula is given by
$$LI= \hat{a} + \hat{b} VR -\left[t_{(58;0,075)}\sqrt{\left(\frac{1}{g*m}+\frac{(VR - \overline{VR})^2}{S_{xx}}\right)\times QME}\right]$$
$$LS= \hat{a} + \hat{b} VR +\left[t_{(58;0,075)}\sqrt{\left(\frac{1}{g*m}+\frac{(VR - \overline{VR})^2}{S_{xx}}\right)\times QME}\right]$$
Therefore, substituting the values and leaving only the reference value as a function, we have:
$$LI= 0,7367 -0,13177\ast VR - \left[2,0017\ast\sqrt{\left(\frac{1}{5*12} + \frac{(VR - 6)^2}{480}\right)\times 0,5738}\right]$$
$$LS= 0,7367 -0,13177\ast VR + \left[2,0017\ast\sqrt{\left(\frac{1}{5*12} + \frac{(VR - 6)^2}{480}\right)\times 0,5738}\right]$$
To find the limits of the reference values, we can perform the calculation simply using the expression above and substituting the value of $VR$. Let’s carry out the calculation in case $ VR=2 $:
$$LI_{VR=2}= 0,7367 -0,13177\ast 2 - \left[2,0017\ast\sqrt{\left(\frac{1}{5*12} + \frac{(2 - 6)^2}{480}\right)\times 0,5738}\right]=0,36611$$
$$LS_{VR=2}= 0,7367 -0,13177\ast 2 + \left[2,0017\ast\sqrt{\left(\frac{1}{5*12} + \frac{(2 - 6)^2}{480}\right)\times 0,5738}\right]=0,58055$$
| Reference Value (VR) | LI | LS |
|---|---|---|
| 2 | 0,36611 | 0,58055 |
| 4 | 0,13418 | 0,28581 |
| 6 | -0,11523 | 0,00857 |
| 8 | -0,39248 | -0,24085 |
| 10 | -0,68722 | -0,47278 |
Table 2.3.3: Range limits
See below the results obtained by the Action Stat software for the same example.


Example 2.3.2
As an application of a trend and linearity study, we will evaluate a measurement system to measure the temperature of an oven via an optical pyrometer. To do this, we will carry out a study by comparison with a standard element term. We take 5 temperature levels
| Standard | Measurements | VR | Tolerance |
|---|---|---|---|
| 1 | 748,8 | 750 | 100 |
| 1 | 749,8 | 750 | 100 |
| 1 | 748,8 | 750 | 100 |
| 1 | 748,8 | 750 | 100 |
| 1 | 748,8 | 750 | 100 |
| 1 | 748,8 | 750 | 100 |
| 1 | 747,7 | 750 | 100 |
| 1 | 747,7 | 750 | 100 |
| 1 | 747,7 | 750 | 100 |
| 1 | 748,7 | 750 | 100 |
| 1 | 749,7 | 750 | 100 |
| 1 | 750,7 | 750 | 100 |
| 2 | 848,8 | 850 | 100 |
| 2 | 848,8 | 850 | 100 |
| 2 | 848,8 | 850 | 100 |
| 2 | 847,2 | 850 | 100 |
| 2 | 847,2 | 850 | 100 |
| 2 | 847,2 | 850 | 100 |
| 2 | 846,1 | 850 | 100 |
| 2 | 846,1 | 850 | 100 |
| 2 | 846,2 | 850 | 100 |
| 2 | 846,3 | 850 | 100 |
| 2 | 847,3 | 850 | 100 |
| 2 | 848,3 | 850 | 100 |
| 3 | 946,9 | 950 | 100 |
| 3 | 946,9 | 950 | 100 |
| 3 | 946,9 | 950 | 100 |
| 3 | 945,8 | 950 | 100 |
| 3 | 944,8 | 950 | 100 |
| 3 | 944,8 | 950 | 100 |
| 3 | 943,6 | 950 | 100 |
| 3 | 943,6 | 950 | 100 |
| 3 | 943,6 | 950 | 100 |
| 3 | 945,1 | 950 | 100 |
| 3 | 946,1 | 950 | 100 |
| 3 | 947,1 | 950 | 100 |
| 4 | 1045,4 | 1050 | 100 |
| 4 | 1045,4 | 1050 | 100 |
| 4 | 1045,4 | 1050 | 100 |
| 4 | 1044,9 | 1050 | 100 |
| 4 | 1043,9 | 1050 | 100 |
| 4 | 1044,9 | 1050 | 100 |
| 4 | 1042 | 1050 | 100 |
| 4 | 1042 | 1050 | 100 |
| 4 | 1042 | 1050 | 100 |
| 4 | 1045,6 | 1050 | 100 |
| 4 | 1046,6 | 1050 | 100 |
| 4 | 1047,6 | 1050 | 100 |
| 5 | 1141,9 | 1150 | 100 |
| 5 | 1141,3 | 1150 | 100 |
| 5 | 1142,9 | 1150 | 100 |
| 5 | 1144,3 | 1150 | 100 |
| 5 | 1143,5 | 1150 | 100 |
| 5 | 1140,9 | 1150 | 100 |
| 5 | 1141,9 | 1150 | 100 |
| 5 | 1142,2 | 1150 | 100 |
| 5 | 1142,1 | 1150 | 100 |
| 5 | 1140 | 1150 | 100 |
| 5 | 1140,7 | 1150 | 100 |
| 5 | 1142,7 | 1150 | 100 |
See below the results obtained by the Action software for the same example


Interpretation
In this study, we performed a trend analysis of the measurement system along the scale of interest. Through the Trend Analysis table, we observe that zero is outside the confidence interval for all reference values. With this, we conclude that we have a significant trend across the scale of the measurement system. Furthermore, the measurement system also presents a significant linearity (P-value of 1.8*10^(-9)) of 0.016, which means that the trend presents an average variation of 1.6% throughout the range of interest (750 to 1150). In the graph, we observe that the zero point line (red line) is not inserted in the confidence band (blue lines), which reinforces the fact that the measurement system presents significant linearity at the 5% significance level.
However, to properly evaluate the trend and linearity of this measurement system, we need to better understand the measurement system and its application. In this case, we have a temperature measurement, on the scale of 750°C to 1150°C, carried out using an optical pyrometer positioned ten meters from the part. At this distance, the optical pyrometer presents an accuracy of 2% of the reading. For example, at 1150 degrees Celsius we have an accuracy of 0.02*1150=23 degrees Celsius. Note that the highest trend value, at the 1150 point, was 10 degrees Celsius. Therefore, the biggest tendency found is below the accuracy guaranteed by the equipment. Furthermore, we have a tolerance of 100 degrees Celsius and for safety reasons, we measure the part at a distance of 10 meters.
In short, depending on the importance of the application for the rest of the process, the measurement system can be considered acceptable for process control. If not, we must develop an action plan.
A point frequently commented on by users is the interpretation of R². In the particular case of studying the linearity of a measurement system, high values of R² may mean that we have had a good linear adjustment. In this situation, we conclude that linearity (or slope) is statistically significant. Therefore, in a linearity study it is interesting to have a low coefficient of determination R² (less than 0.2). In this example, we have an R² equal to 0.75, which is high.
11.2.4 Repeatability and Reproducibility
In this module, we will present a method to estimate the variability associated with the measurement system. As presented in the measurement systems analysis module, variability is decomposed into two terms:
Repetitivity - VE
Variation in measurements obtained by a single operator, using the same measuring equipment and method, when repeatedly measuring the same quantity of a single piece (test specimen).
Reproducibility - VO
Variation in averages obtained by different operators using the same measuring equipment to repeatedly measure the same quantity of a single piece (test specimen).
Repeatability and Reproducibility (RR)
It is the sum of variations due to lack of Repeatability and Reproducibility.
$$RR=\sqrt{(VE)^2+(VO)^2}$$

As in any analysis, planning is essential for us to carry out a good analysis of repeatability and reproducibility:
-
The technique to be used must be planned. For example, there are some measuring systems whose reproducibility effect is negligible, for example, for automatic measuring systems in which the operator’s only intervention is to press a button. In this case, we planned the study without reproducibility (MSA, page 73).
-
The number of operators, number of parts and number of replicas must be determined using the criteria:
- Measurement criticality, critical dimensions require more parts and/or replicas;
- When we deal with heavy or difficult-to-handle parts, we use fewer parts and more replicas;
- Customer requirement;
-
Operators must be chosen from among everyone who uses the measurement system;
-
The selection of parts is critical for a good RR analysis. This selection depends on the purpose of the measurement system and the availability of parts that represent the production process. On page 73, the MSA manual establishes the following purposes:
- Product Control: measurement systems whose results and decision criteria determine the conformity or non-conformity of the product with respect to specifications (100% inspection or sampling). In this case, the selected parts do not need to cover the entire process variation range.
- Process Control: measurement systems whose result and decision criteria determine the stability and/or capacity of the production process (CEP, Farol Graphic, Continuous Improvement). In this case, the availability of parts that cover the entire variation range of the production process is essential. Often, a complementary study to determine the process capability is required and recommended to evaluate the suitability of the measurement system for process control.
- The measuring instrument must have a discrimination of at least one tenth of the tolerance or variation of the production process.
- Ensure that the measurement method is measuring the correct dimension and that the method is correctly applied.
Internal product variability
In many measurement systems, internal (or inherent) part variation, such as out-of-roundness, can inflate our estimate of repeatability. Internal variability can be assessed and separated from repeatability. The procedure is more complex and basically consists of:
- Obtain the measurement at various points on the piece;
- Work with average values (or maximum and/or minimum values) for each piece;
- From these experiments, estimate the internal variability.
Internal variability should always be an object of study to improve the production process.
RR Index
Based on the purpose of the measurement system (Product Control or Process Control), we must establish an index to “facilitate” the interpretation of the RR:
Process Control: Compare the variability of the measurement system (RR) with the expected variation of the production process.
Product Control: Compare measurement system variability (RR) with product tolerance.
Inter-part variability (production process variability) - VP
It is the variation in measurements observed between the items produced by the process, that is, the variability observed in the parts. We emphasize that the variability between the parts can be obtained from a process capacity study or from the study itself to determine the RR.
Total variability
It is the sum of variations due to the measurement system and the process.
$$VT=\sqrt{RR^2+VP^2}$$
or
$$VT=\sqrt{VE^2+VO^2+VP^2}.$$
With this, we can calculate the RR index as follows:
Product Control
$$ (porcentagem) RR = \frac{6 RR}{Tolerancia} 100 (porcentagem).$$
Process Control
$$ (porcentagem) RR = \frac{RR}{VT} 100 (porcentagem).$$
The criteria for analyzing the RR are defined on page 78 of the MSA. For a measurement system whose purpose is to analyze a process, a general rule for accepting a measurement system is defined in the following table:
| RR | Decision | Comments |
|---|---|---|
| Below 10% | Measurement system generally considered acceptable | Recommended, especially useful when trying to order or classify parts or when tight process control is required. |
| Between 10% and 30% | May be accepted for some applications | The decision should first be based on, for example, the importance of the measurement application, cost of the measuring device, cost of rework or repair. The measuring system must be approved by the customer. |
| Above 30% | Considered unacceptable | Every effort must be made to improve the measurement system. This condition can be resolved by using an appropriate measurement strategy; for example, using the average of several measurements of the same characteristic of the same part in order to reduce the variability of the final measurement. |
Another statistic that evaluates the variability of a measurement system is number of distinct categories (NDC). This statistic indicates the category number into which process measurements can be divided. The NDC must be greater than or equal to five and is only applied to measurement systems whose purpose is to analyze a process (page 78 of the MSA). The NDC is defined by:
$$ NDC = 1,41 \frac{ VP}{RR}.$$
Caution: Using RR as the only index to evaluate a measurement system is not acceptable.
By applying the acceptance criteria as simple thresholds, we assume that the statistics are deterministic estimates of the variability of the measurement system (which they are not). Specifying cutoff values as criteria may lead to inappropriate behavior. For example, the supplier can be “creative” in finding a certain RR value, eliminating key sources of variation (such as part x operator interaction) or simply manipulating the study.
Comments: Unfortunately this is a fact that occurs in many companies in Brazil. The customer imposes a RR below 10% and the supplier manipulates the data. WE SHOULD NOT HAVE A SIMPLE (SINGLE) CRITERIA FOR ALL MEASUREMENT SYSTEMS. Each application must be evaluated individually.
When we analyze the variation of a measurement system, it is important to look at each application individually, to know what is required and how this measurement will be used. For example: the required temperature measurement accuracy may be different for dissimilar applications. A room thermostat can regulate the temperature for human comfort and is cheap, but has an RR above 30%. This is acceptable for this application. But in a laboratory, where small temperature variations can impact test results, more sophisticated temperature measurement and control may be required. This thermostat will be more expensive and we will also require less variability (lower RR value).
Guidelines for the study of RR - Replicable
As with any analysis, good planning must be carried out so that we can conduct the RR study.
-
Randomly select operators who use and know well the measurement system to be studied. In general we recommend three operators. If this is not feasible, use at least two operators. If the operator does not influence the measurement, do not evaluate reproducibility.
-
Use properly calibrated measuring equipment.
-
Select 5 to 15 production parts whose dimensions span the process variation range
- If the measurement system is used for processes with very different fields of variation, we recommend carrying out different RR studies.
- Whenever possible, try to obtain g = (number of pieces) X (number of operators) greater than 15. If this is not possible, increase the number of readings per piece.
-
Choose the method of conducting and analyzing the study.
- Analysis of Variance (ANOVA).
For measurement systems in which the measurement result is used to determine the “conformity” or “non-conformity” of the product with respect to a specification the selected samples do NOT need to cover the entire specification. Furthermore, the RR analysis must be carried out in relation to tolerance. In this case, we calculate the RR percentage in relation to the tolerance.
For the case where the measurement system controls the process, that is, the measurement result is used to determine the stability, trend and variability of the process, through a SPC study, we must use the process variation to evaluate the RR. In this case, we must choose the parts carefully, as they must cover the entire range of process variations. If possible, we should conduct an independent study to estimate process variability. Thus, we calculate the percentage RR in relation to the total variability.
Next, we describe the steps to apply the ANOVA method.
1st Step
- Select the parts in such a way that they represent the natural variation of the process. In general, we take parts from different production batches. Identify the parts.
2nd Step
- Select operators to involve all shifts. Operators must have training to use the measurement system. The number of operators times the number of parts must be greater than 15. If the operator does not influence the measurement, we choose only one operator and do not evaluate reproducibility.
3rd Step
- Each operator measures each part three or more times in random order.
4th Step
- Randomize measurements.
5th Step
| Operator | Ask | Measure 1 | Measure 2 | Measure 3 | R | $ \overline{Y} $ |
|---|---|---|---|---|---|---|
| A | 1 | $ Y_{1A1} $ | $ Y_{1A2} $ | $ Y_{1A3} $ | $ R_{1A} $ | $ \overline{Y_{1A}} $ |
| A | 2 | $ Y_{2A1} $ | $ Y_{2A2} $ | $ Y_{2A3} $ | $ R_{2A} $ | $ \overline{Y_{2A}} $ |
| A | 3 | $ Y_{3A1} $ | $ Y_{3A2} $ | $ Y_{3A3} $ | $ R_{3A} $ | $ \overline{Y_{3A}} $ |
| A | 4 | $ Y_{4A1} $ | $ Y_{4A2} $ | $ Y_{4A3} $ | $ R_{4A} $ | $ \overline{Y_{4A}} $ |
| A | 5 | $ Y_{5A1} $ | $ Y_{5A2} $ | $ Y_{5A3} $ | $ R_{5A} $ | $ \overline{Y_{5A}} $ |
| A | 6 | $ Y_{6A1} $ | $ Y_{6A2} $ | $ Y_{6A3} $ | $ R_{6A} $ | $ \overline{Y_{6A}} $ |
| A | 7 | $ Y_{7A1} $ | $ Y_{7A2} $ | $ Y_{7A3} $ | $ R_{7A} $ | $ \overline{Y_{7A}} $ |
| A | 8 | $ Y_{8A1} $ | $ Y_{8A2} $ | $ Y_{8A3} $ | $ R_{8A} $ | $ \overline{Y_{8A}} $ |
| A | 9 | $ Y_{9A1} $ | $ Y_{9A2} $ | $ Y_{9A3} $ | $ R_{9A} $ | $ \overline{Y_{9A}} $ |
| A | 10 | $ Y_{10A1} $ | $ Y_{10A2} $ | $ Y_{10A3} $ | $ R_{10A} $ | $ \overline{Y_{10A}} $ |
| B | 1 | $ Y_{1B1} $ | $ Y_{1B2} $ | $ Y_{1B3} $ | $ R_{1B} $ | $ \overline{Y_{1B}} $ |
| B | 2 | $ Y_{2B1} $ | $ Y_{2B2} $ | $ Y_{2B3} $ | $ R_{2B} $ | $ \overline{Y_{2B}} $ |
| B | 3 | $ Y_{3B1} $ | $ Y_{3B2} $ | $ Y_{3B3} $ | $ R_{3B} $ | $ \overline{Y_{3B}} $ |
| B | 4 | $ Y_{4B1} $ | $ Y_{4B2} $ | $ Y_{4B3} $ | $ R_{4B} $ | $ \overline{Y_{4B}} $ |
| B | 5 | $ Y_{5B1} $ | $ Y_{5B2} $ | $ Y_{5B3} $ | $ R_{5B} $ | $ \overline{Y_{5B}} $ |
| B | 6 | $ Y_{6B1} $ | $ Y_{6B2} $ | $ Y_{6B3} $ | $ R_{6B} $ | $ \overline{Y_{6B}} $ |
| B | 7 | $ Y_{7B1} $ | $ Y_{7B2} $ | $ Y_{7B3} $ | $ R_{7B} $ | $ \overline{Y_{7B}} $ |
| B | 8 | $ Y_{8B1} $ | $ Y_{8B2} $ | $ Y_{8B3} $ | $ R_{8B} $ | $ \overline{Y_{8B}} $ |
| B | 9 | $ Y_{9B1} $ | $ Y_{9B2} $ | $ Y_{9B3} $ | $ R_{9B} $ | $ \overline{Y_{9B}} $ |
| B | 10 | $ Y_{10B1} $ | $ Y_{10B2} $ | $ Y_{10B3} $ | $ R_{10B} $ | $ \overline{Y_{10B}} $ |
| C | 1 | $ Y_{1C1} $ | $ Y_{1C2} $ | $ Y_{1C3} $ | $ R_{1C} $ | $ \overline{Y_{1C}} $ |
| C | 2 | $ Y_{2C1} $ | $ Y_{2C2} $ | $ Y_{2C3} $ | $ R_{2C} $ | $ \overline{Y_{2C}} $ |
| C | 3 | $ Y_{3C1} $ | $ Y_{3C2} $ | $ Y_{3C3} $ | $ R_{3C} $ | $ \overline{Y_{3C}} $ |
| C | 4 | $ Y_{4C1} $ | $ Y_{4C2} $ | $ Y_{4C3} $ | $ R_{4C} $ | $ \overline{Y_{4C}} $ |
| C | 5 | $ Y_{5C1} $ | $ Y_{5C2} $ | $ Y_{5C3} $ | $ R_{5C} $ | $ \overline{Y_{5C}} $ |
| C | 6 | $ Y_{6C1} $ | $ Y_{6C2} $ | $ Y_{6C3} $ | $ R_{6C} $ | $ \overline{Y_{6C}} $ |
| C | 7 | $ Y_{7C1} $ | $ Y_{7C2} $ | $ Y_{7C3} $ | $ R_{7C} $ | $ \overline{Y_{7C}} $ |
| C | 8 | $ Y_{8C1} $ | $ Y_{8C2} $ | $ Y_{8C3} $ | $ R_{8C} $ | $ \overline{Y_{8C}} $ |
| C | 9 | $ Y_{9C1} $ | $ Y_{9C2} $ | $ Y_{9C3} $ | $ R_{9C} $ | $ \overline{Y_{9C}} $ |
| C | 10 | $ Y_{10C1} $ | $ Y_{10C2} $ | $ Y_{10C3} $ | $ R_{10C} $ | $ \overline{Y_{10C}} $ |
| Average: | $ \overline{R} $ | $ \overline{\overline{Y}} $ |
Table 2.4.1: Readings
11.2.4.1 ANOVA model: with interaction between part and operator
Analysis of variance (ANOVA) Cross Method
Analysis of variance (ANOVA) is a classic statistical technique that can be used to evaluate measurement error and other sources of variability in data pertinent to the measurement system. The two-factor crossover design with interaction is the classic RR model. Typically, the two factors are referred to as “parts” and “operators”. In this chapter we consider this first RR model, in which we model the experiment with balanced data and both factors are random.
The advantages of the ANOVA method compared to the Mean and Range method are:
- Able to handle different experimental structures;
- Estimate variance with more accuracy and precision;
- Extract more information about the data, such as the effect of the interaction between parts and evaluators;
The disadvantages are the complexity of the calculations and the need for basic knowledge of statistics so that the user can interpret the results.
Statistical Model
The model with two balanced factors and cross-effects with interaction is given by

For this model $ \mu $ is a parameter common to all treatments and represents the general average of the data, $ \alpha_{i} $ and $ \gamma_j $ is the effect due to the ith and jth level of the factor P (part) and O (operator) and are independent random variables with zero mean and variances $ \sigma^2_P $ and $ \sigma^2_O $ respectively and $ \tau_{ij} $ is the interaction between the factors P and O, which also has a normal distribution with zero mean and variance $ \sigma^2_I. $ The random variable $ \varepsilon_{ijk} $ corresponds to experimental random error, that is, the variability not explained by the model due to variations present in several sources not considered in the study. This has a normal distribution with zero mean and variance $ \sigma^2. $
In short,
$$\mu={média~geral~dos~dados;}$$
$$\alpha_{i}={efeito~do~nível~i~do~fator~peça;}$$
$$\gamma_{j}={efeito~do~nível~j~do~fator~operador;}$$
$$\tau_{ij}={efeito~do~nível~ij~da~interação~entre~peça~e~operador;}$$
$$\varepsilon_{ijk}={componente~aleatória~do~erro.}$$
Now, let’s develop the analysis of variance for the random effects model. From considerations of the data, we have:
$${soma das observações do nível i do fator peça;}$$
(there are missing elements in the equation above)
$$~{média das observações do nível i do fator peça;}$$
(there are missing elements in the equation above)
$${soma das observações do nível j do fator operador;}$$
(there are missing elements in the equation above)
$$~{média das observações do nível j do fator operador;}$$
(there are missing elements in the equation above)
$${soma das observações dos níveis i e j dos fatores peça e operador;}$$
(there are missing elements in the equation above)
$$~{média das observações dos níveis i e j dos fatores peça e operador;}$$
(there are missing elements in the equation above)
$$~{soma de todas as observações;}\quad \text{e}$$
(there are missing elements in the equation above)
$$ \mu ~{média geral das observações}$$
Furthermore, we assume that the error has a Normal distribution with zero mean and variance $ \sigma^{2}_{\varepsilon} $ and that the errors are mutually independent. With this, we have to
$$\varepsilon_{ij}\sim~N(0,\sigma^{2}_{\varepsilon}).$$
Now, for the purpose $ \alpha_i $, we assume that it has a Normal distribution with zero mean and variance $ \sigma^2_{P}. $ We also assume that the effects are mutually independent. So,
$$\alpha_{i}\sim~N(0,\sigma^{2}_{P}).$$
To that end $ \gamma_j $, we assume that it has a Normal distribution with zero mean and variance $ \sigma^2_{O}. $ We also assume that the effects are mutually independent. So,
$$\gamma_{j}\sim~N(0,\sigma^{2}_{O}).$$
Finally, we have to $ \tau_{ij} $, we assume that it has a Normal distribution with zero mean and variance $ \sigma^2_{I}. $ We also assume that the effects are mutually independent. So,
$$\tau_{ij}\sim~N(0,\sigma^{2}_{I}).$$
The analysis of variance for model (2.4.1.1) is obtained by decomposing the total variation $ Y_{ijk}-\overline{Y_{…}} $ as follows
$$SQT=\sum_{i=1}^{a}\sum_{j=1}^{b}\sum_{k=1}^{r}(Y_{ijk}-\overline{Y_{…}})^2=$$
$$=o~r\sum_{i=1}^{p}(\overline{Y_{i..}}-\overline{Y_{…}})^2 +p~r\sum_{j=1}^{o}(\overline{Y_{.j.}}-\overline{Y_{…}})^2 +r~\sum_{i=1}^{p}\sum_{j=1}^{o}(\overline{Y_{ij.}}-\overline{y_{i..}}-\overline{Y_{.j.}}+\overline{Y_{…}})^2+\sum_{i=1}^{p}\sum_{j=1}^{o}\sum_{k=1}^{r}(Y_{ijk}-\overline{Y_{ij.}})^2$$
$$=SQP+SQO+SQI+SQE$$
in which
$$SQT=\sum_{i=1}^{p}\sum_{j=1}^{o}\sum_{k=1}^{r}(Y_{ijk}-\overline{Y_{…}})^2$$
$$SQP=o~r\sum_{i=1}^{p}(\overline{Y_{i..}}-\overline{Y_{…}})^2$$
$$SQO=p~r\sum_{j=1}^{o}(\overline{Y_{.j.}}-\overline{Y_{…}})^2$$
$$SQI=r~\sum_{i=1}^{p}\sum_{j=1}^{o}(\overline{Y_{ij.}}-\overline{Y_{i..}}-\overline{Y_{.j.}}+\overline{Y_{…}})^2$$
$$SQE=\sum_{i=1}^{p}\sum_{j=1}^{o}\sum_{k=1}^{r}(Y_{ijk}-\overline{Y_{ij.}})^2 $$
Thus, we obtain the ANOVA table as follows
| Factor | Degrees of Freedom | Sum of Squares | Medium Squares | F |
|---|---|---|---|---|
| Factor Piece P | p-1 | $ SQP $ | $ QMP=\frac{SQP}{p-1} $ | $ F^\ast_{1}=\frac{QMP}{QMI} $ |
| Operator Factor O | o-1 | $ SQO $ | $ QMO=\frac{SQO}{o-1} $ | $ F^\ast_{2}=\frac{QMO}{QMI} $ |
| Interaction between Part and Operator PxO | (p-1)(o-1) | $ SQI $ | $ QMI=\frac{SQI}{(p-1)(o-1)} $ | $ F^\ast_{3}=\frac{QMI}{QME} $ |
| Repetitiveness | po(r-1) | $ SQE $ | $ QME=\frac{SQE}{po(r-1)} $ |
Now, we will briefly summarize the expected mean square values.
| Factor | Degrees of Freedom | Medium Squares | Expected Value of Mean Squares |
|---|---|---|---|
| Factor Piece P | p-1 | $ QMP $ | $ E(QMP)=\sigma^2_E+r\sigma^2_I+or\sigma^2_P $ |
| Operator Factor O | o-1 | $ QMO $ | $ E(QMO)=\sigma^2_E+r\sigma^2_I+pr\sigma^2_O $ |
| Interaction between Part and Operator PxO | (p-1)(o-1) | $ QMI $ | $ E(QMI)=\sigma^2_E+r\sigma^2_I $ |
| Repetitiveness | po(r-1) | $ QME $ | $ E(QME)=\sigma^2_E $ |
Table 2.4.1.1: Expected Values of Mean Squares.
Statistics:
| $ QMP= $ | $ or\displaystyle\sum^p_{i=1}\frac{(\overline{Y_{i..}}-\overline{Y_{…}})^2}{p-1} $ |
|---|---|
| $ QMO= $ | $ pr\displaystyle\sum^o_{j=1}\frac{(\overline{Y_{.j.}}-\overline{Y_{…}})^2}{o-1} $ |
| $ QMI= $ | $ r\displaystyle\sum^p_{i=1}\sum^o_{j=1}\frac{(\overline{Y_{ij.}}-\overline{Y_{i..}}-\overline{Y_{.j.}}+\overline{Y_{…}})^2}{(p-1)(o-1)} $ |
| $ QME= $ | $ \displaystyle\sum^p_{i=1}\sum^o_{j=1}\sum^r_{k=1}\frac{(Y_{ijk}-\overline{Y_{ij.}})^2}{po(r-1)} $ |
| $ \overline{Y_{i..}}= $ | $ \displaystyle\sum^o_{j=1}\sum^r_{k=1}\frac{Y_{ijk}}{or} $ |
| $ \overline{Y_{.j.}}= $ | $ \displaystyle\sum^p_{i=1}\sum^r_{k=1}\frac{Y_{ijk}}{pr} $ |
| $ \overline{Y_{ij.}}= $ | $ \displaystyle\sum^r_{k=1}\frac{Y_{ijk}}{r} $ |
| $ \overline{Y_{…}}= $ | $ \displaystyle\sum^p_{i=1}\sum^o_{j=1}\sum^r_{k=1}\frac{Y_{ijk}}{por} $ |
Table 2.4.1.2: Summary of mean squares and means for model (2.4.1.1).
Table 2.4.1.3 represents the point estimators of the model (2.4.1.1).
| Model Representation | Spot Estimator |
|---|---|
| $ \hat{\mu} $ | $ \overline{Y_{…}} $ |
| $ \hat{\sigma}^2_P $ or $ VP^2 $ | $ \displaystyle\frac{QMP-QMI}{or} $ |
| $ \hat{\sigma}^2_O $ or $ V^2_{oper} $ | $ \displaystyle\frac{QMO-QMI}{pr} $ |
| $ \hat{\sigma}^2_I $ or $ VI^2 $ | $ \displaystyle\frac{QMI-QME}{po} $ |
| $ \hat{\sigma}^2_E $ or $ VE^2 $ | $ QME $ |
Table 2.4.1.3: Summary of Point Estimators for the model (2.4.1.1).
Here, we perform the following tests
- Part factor effect test:
$$\sigma^{2}_{P}> 0$$
the appropriate test statistic is
$$F^{*}=\frac{QMP}{QMI}$$
and the decision rule at the significance level $ \alpha $ is

- Testing the effect of the operator factor:
$$\sigma^{2}_{O}> 0$$
the appropriate test statistic is
$$F^{*}=\frac{QMO}{QMI}$$
and the decision rule at the significance level $ \alpha $ is

- Testing the effect of interaction between part and operator:
$$\sigma^{2}_{I}> 0$$
the appropriate test statistic is
$$F^{*}=\frac{QMI}{QME}$$
and the decision rule at the significance level $ \alpha $ is

Next, we calculate the sources of variation from table (2.4.1.3) and divide them into steps as follows:
6th Step
- Perform graphical analysis: Build the control chart $ \overline{X} $ and S or $ \overline{X} $ and R.
7th Step
- Calculate Repeatability
$$VE=\hat{\sigma}_E=\sqrt{QME}~~~~~(2.4.1.5)$$
8th Step
- Calculate Reproducibility as
$$VO=\sqrt{\hat{\sigma}^2_O+\hat{\sigma}^2_I}=\sqrt{V_{oper}^2+VI^2},~~~~(2.4.1.6)$$
being $V_{oper}$ and VI given by the expressions
$$V_{oper}=\hat{\sigma}_O=\sqrt{\frac{QMO-QMI}{pr}~}~~~~(2.4.1.7)$$
$$VI=\hat{\sigma}_I=\sqrt{\frac{QMI-QME}{r}}~~~~(2.4.1.8)$$
9th Step
- Calculate the RR
$$RR=\sqrt{~VE^2~+~VO^2~}$$
10th Step
- Calculate variation between parts
$$VP=\hat{\sigma}_P=\sqrt{\frac{QMP-QMI}{or}~}~~~~(2.4.1.8)$$
11th Step
- Calculate the total variation
$$VT=\sqrt{~VE^2~+~VO^2~+~VP^2~}~~~~~(2.4.1.9)$$
12th Step
- Contribution % table
| Sources of Variation | Variance | % Contribution |
|---|---|---|
| Repetitiveness | (VE)² | $ \displaystyle\frac{VE^2}{VT^2}*100 $ |
| Operator | (Voper)² | $ \displaystyle\frac{V^2_{oper}}{VT^2}*100 $ |
| Part x Operator | (VI)² | $ \displaystyle\frac{VI^2}{VT^2}*100 $ |
| Ask | (VP)² | $ \displaystyle\frac{VP^2}{VT^2}*100 $ |
| RR | (RR)² | $ \displaystyle\frac{RR^2}{VT^2}*100 $ |
| Total | (VT)² | 100 |
Table 2.4.1.4: Contribution table - with interaction.
The values used to fill in Table 2.4.1.2 come from expressions (2.4.1.6) to (2.4.1.9)
13th Step
- Table of % in relation to the total variation and or % of tolerance
| Source of variation | Standard Deviation | % Total Variation | % Tolerance |
|---|---|---|---|
| Repetitiveness | VE | $ \displaystyle\frac{VE}{VT}*100 $ | $ \left(k\times\frac{VE}{TOL}\right)*100 $ |
| Reproducibility | GRANDMOTHER | $ \displaystyle\frac{VO}{VT}*100 $ | $ \left(k\times\frac{VO}{TOL}\right)*100 $ |
| Operator | Voper | $ \displaystyle\frac{V_{oper}}{VT}*100 $ | $ \left(k\times\frac{V_{oper}}{TOL}\right)*100 $ |
| Part x Operator | VI | $ \displaystyle\frac{VI}{VT}*100 $ | $ \left(k\times\frac{VI}{TOL}\right)*100 $ |
| Ask | VP | $ \displaystyle\frac{VP}{VT}*100 $ | $ \left(k\times\frac{VP}{TOL}\right)*100 $ |
| RR | RR | $ \displaystyle\frac{RR}{VT}*100 $ | $ \left(k\times\frac{RR}{TOL}\right)*100 $ |
Table 2.4.1.5: Table of % of total variation and/or % of tolerance - with interaction.
The values used to fill in Table 2.4.1.3 come from expressions (2.4.1.6) to (2.4.1.9). In Table 2.4.1.3, the normally chosen value for k is 5.15. However, for ease of interpretation of this index with the Pp index (e or Cp) it is recommended to use k=6.
Suggested rule for analyzing variations:
Tolerance: Measurement system applied in final inspections and receiving inspection;
Total variation: Measurement system used during the production process.
- %RR less than 10% → acceptable measurement system.
- %RR between 10% and 30% → marginal measurement system, which may be accepted depending on the situation, costs, etc.
- %RR greater than 30% → unacceptable measurement system, requiring improvement or replacement.
14th Step
The ndc represents the ability to discriminate categories of parts in a measurement system considering process variation. This index gives us the number of bands we can divide the process variation into. The ndc is given by the largest integer less than or equal to the value:
$$ndc=1,41\times\frac{VP}{RR}~~~~(2.4.1.10)$$
The automotive industry manual (MSA) presents as a criterion an ndc ≥ 5. This means that the measurement system is capable of identifying 5 different types of parts within the process variation field. If a measuring system is evaluated by tolerance, the ndc index should not be considered.
15th Step
- Perform a data analysis to evaluate the system characteristics.
If the repeatability is high compared to the reproducibility, the reasons may be:
- The measuring device needs maintenance;
- The measuring device must be redesigned to be more robust;
- The fixation or position for measurement needs to be improved;
- There is an excessive variation inherent to the piece.
If the reproducibility is high compared to the repeatability, then the possible causes could be:
- The operator needs to be better trained in how to use and read the measuring device;
- The markings on the measuring device dial are unclear;
- Some type of device may be needed to help the operator use the measuring device more consistently.
Example 2.4.1.1
Consider a measuring system to measure the outer diameter of the bearing. The quality engineer carried out an experiment with 10 parts, 3 operators and 3 repetitions for each operator and part. The data relating to this experiment are shown in the Table below.
| Ask | Measurement | Operator |
|---|---|---|
| 1 | 114,958 | 1 |
| 2 | 114,957 | 1 |
| 3 | 114,962 | 1 |
| 4 | 114,963 | 1 |
| 5 | 114,965 | 1 |
| 6 | 114,963 | 1 |
| 7 | 114,967 | 1 |
| 8 | 114,963 | 1 |
| 9 | 114,963 | 1 |
| 10 | 114,967 | 1 |
| 1 | 114,962 | 1 |
| 2 | 114,956 | 1 |
| 3 | 114,963 | 1 |
| 4 | 114,965 | 1 |
| 5 | 114,966 | 1 |
| 6 | 114,965 | 1 |
| 7 | 114,969 | 1 |
| 8 | 114,97 | 1 |
| 9 | 114,955 | 1 |
| 10 | 114,965 | 1 |
| 1 | 114,958 | 1 |
| 2 | 114,6 | 1 |
| 3 | 114,965 | 1 |
| 4 | 114,966 | 1 |
| 5 | 114,967 | 1 |
| 6 | 114,964 | 1 |
| 7 | 114,97 | 1 |
| 8 | 114,97 | 1 |
| 9 | 114,955 | 1 |
| 10 | 114,966 | 1 |
| 1 | 114,957 | 2 |
| 2 | 114,958 | 2 |
| 3 | 114,962 | 2 |
| 4 | 114,963 | 2 |
| 5 | 114,965 | 2 |
| 6 | 114,962 | 2 |
| 7 | 114,967 | 2 |
| 8 | 114,968 | 2 |
| 9 | 114,952 | 2 |
| 10 | 114,967 | 2 |
| 1 | 114,961 | 2 |
| 2 | 114,957 | 2 |
| 3 | 114,963 | 2 |
| 4 | 114,966 | 2 |
| 5 | 114,967 | 2 |
| 6 | 114,963 | 2 |
| 7 | 114,97 | 2 |
| 8 | 114,968 | 2 |
| 9 | 114,955 | 2 |
| 10 | 114,964 | 2 |
| 1 | 114,958 | 2 |
| 2 | 114,959 | 2 |
| 3 | 114,965 | 2 |
| 4 | 114,965 | 2 |
| 5 | 114,966 | 2 |
| 6 | 114,965 | 2 |
| 7 | 114,97 | 2 |
| 8 | 114,97 | 2 |
| 9 | 114,954 | 2 |
| 10 | 114,966 | 2 |
| 1 | 114,958 | 3 |
| 2 | 114,958 | 3 |
| 3 | 114,953 | 3 |
| 4 | 114,965 | 3 |
| 5 | 114,967 | 3 |
| 6 | 114,962 | 3 |
| 7 | 114,967 | 3 |
| 8 | 114,968 | 3 |
| 9 | 114,953 | 3 |
| 10 | 114,966 | 3 |
| 1 | 114,961 | 3 |
| 2 | 114,96 | 3 |
| 3 | 114,963 | 3 |
| 4 | 114,966 | 3 |
| 5 | 114,968 | 3 |
| 6 | 114,965 | 3 |
| 7 | 114,968 | 3 |
| 8 | 114,969 | 3 |
| 9 | 114,953 | 3 |
| 10 | 114,965 | 3 |
| 1 | 114,958 | 3 |
| 2 | 114,958 | 3 |
| 3 | 114,964 | 3 |
| 4 | 114,965 | 3 |
| 5 | 114,966 | 3 |
| 6 | 114,965 | 3 |
| 7 | 114,969 | 3 |
| 8 | 114,971 | 3 |
| 9 | 114,955 | 3 |
| 10 | 114,967 | 3 |
Table 2.4.1.6: Bearing internal diameter measurements
We begin our statistical analysis in step 5, as the previous steps refer to the choice of parts and operators and data collection.
5th Step
- Calculation of $ S^2_p $
With the help of Table 2.4.1.6, we will calculate
| Ask | Averages $ (\overline{Y_{i..}}) $ | Average of Averages $ (\overline{Y_{…}}) $ | Dev. quadratics $ (\overline{Y_{i..}}-\overline{Y_{…}})^2 $ |
|---|---|---|---|
| 1 | 114,9590 | 114,9631 | 0,00001699 |
| 2 | 114,9581 | 114,9631 | 0,00002511 |
| 3 | 114,9622 | 114,9631 | 0,00000081 |
| 4 | 114,9649 | 114,9631 | 0,00000312 |
| 5 | 114,9663 | 114,9631 | 0,00001031 |
| 6 | 114,9638 | 114,9631 | 0,00000043 |
| 7 | 114,9686 | 114,9631 | 0,00002952 |
| 8 | 114,9686 | 114,9631 | 0,00002952 |
| 9 | 114,9539 | 114,9631 | 0,00008525 |
| 10 | 114,9659 | 114,9631 | 0,00000765 |
| SUM = | 0,00020873 |
Therefore,
$$S_{p}^2=\frac{0,00020873}{10-1}=0,00002319$$
- Calculation of $ S^2_o $
| Operator | Averages $ (\overline{Y_{j..}}) $ | Average of Averages $ (\overline{Y_{…}}) $ | Dev. quadratics $ (\overline{Y_{j..}}-\overline{Y_{j..}})^2 $ |
|---|---|---|---|
| A | 114,9619 | 114,9631 | 0,00000158 |
| B | 114,9636 | 114,9631 | 0,00000023 |
| C | 114,9639 | 114,9631 | 0,00000060 |
| SUM = | 0,00000241 |
Therefore,
$$S_{o}^2=\frac{0,00000241}{3-1}=0,000001205.$$
- QME calculation
To calculate the QME, we must calculate the mean and variance within each box ij in Table 2.4.1.6, that is, for each combination of part i and operator j. In the Table below, we calculate the average values of the measurements within each cell.
| ASK | OPERATOR 1 | OPERATOR 2 | OPERATOR 3 |
|---|---|---|---|
| 1 | 114,9577 | 114,9613 | 114,9580 |
| 2 | 114,9577 | 114,9577 | 114,9590 |
| 3 | 114,9590 | 114,9630 | 114,9647 |
| 4 | 114,9637 | 114,9657 | 114,9653 |
| 5 | 114,9657 | 114,9670 | 114,9663 |
| 6 | 114,9623 | 114,9643 | 114,9647 |
| 7 | 114,9670 | 114,9690 | 114,9697 |
| 8 | 114,9663 | 114,9690 | 114,9703 |
| 9 | 114,9527 | 114,9543 | 114,9547 |
| 10 | 114,9667 | 114,9647 | 114,9663 |
Table 2.4.1.7: Average values within each cell
Using Tables 2.4.1.6 and 2.4.1.7, we can find the values of
$$S^2_{ij}$$
from formula (2.4.1.4). For the first piece, we have to
$$S^2_{11}=\frac{(114,958-114,9577)^2 + (114,957 - 114,9577)^2 + (114,958 - 114,9577)^2}{3-1}$$
$$=0,00000033$$
$$S^2_{12}=\frac{(114,962-114,9613)^2+(114,961-114,9613)^2+(114,961-114,9613)^2}{3-1}$$
$$=0,00000033$$
$$S^2_{13}=\frac{(114,958-114,9580)^2+(114,958-114,9580)^2+(114,958-114,9580)^2}{3-1}$$
$$=0$$
Proceeding in the same way for the other parts, we obtain Table 2.4.1.8 with values of
$$S^2_{ij},$$
| ASK | OPERATOR 1 | OPERATOR 2 | OPERATOR 3 |
|---|---|---|---|
| 1 | 0,00000033 | 0,00000033 | 0,00000000 |
| 2 | 0,00000033 | 0,00000433 | 0,00000100 |
| 3 | 0,00002700 | 0,00000000 | 0,00000033 |
| 4 | 0,00000133 | 0,00000033 | 0,00000033 |
| 5 | 0,00000133 | 0,00000100 | 0,00000033 |
| 6 | 0,00000033 | 0,00000133 | 0,00000033 |
| 7 | 0,00000000 | 0,00000100 | 0,00000033 |
| 8 | 0,00000833 | 0,00000100 | 0,00000033 |
| 9 | 0,00000033 | 0,00000133 | 0,00000033 |
| 10 | 0,00000033 | 0,00000133 | 0,00000033 |
Table 2.4.1.8: Values of $S^2_{ij}.$
Therefore, from formula (2.4.1.8) we obtain the value of QME as:
$$QME=\frac{0,00000033+0,00000033+0+\cdots+0,00000033+0,00000033+0,00000033}{10\times3}$$
$$=0,00000181$$
Proceeding in the same way as in the previous example we can obtain SQT as
$$SQT=\sum_{i=1}^{p}\sum_{j=1}^{o}\sum_{k=1}^{r}(Y_{ijk}~-~\overline{Y_{…}})^2=0,002135656.$$
With the values of $ S_{p}^2 $ and $ S_{o}^2 $ obtained above, we can calculate the sums of square
$$SQP=or(p-1)S_{p}^2=0,001878544$$
$$SQO=pr(o-1)S_{o}^2 = 0,000072289.$$
With the value obtained for QME, we can obtain SQE as
$$SQE = (r-1)(po)QME = 0,000108667.$$
From these sums of squares we can obtain the sum of squares and the mean square of the part and operator interaction as
$$SQI=SQT-SQP-SQO-SQE=0,000076156,$$
$$QMI=\frac{SQI}{(p-1)(o-1)} = 0,00000423086.$$
6th Step
Perform graphical analysis: Build the control chart $ \overline{X} $ and S to evaluate the means and amplitudes of each repetition per operator.
7th Step
Calculate repeatability as
$$VE=\sqrt{QME}=0,001345775$$
8th Step
From equations (2.4.1.7) and (2.4.1.8), we have
$$V_{oper}=\sqrt{S_{o}^2 - \frac{QMI}{pr}~}=\sqrt{0,000001064}=0,0010314$$
$$VI=\sqrt{\frac{QMI - QME}{r}}=\sqrt{0,0000008065}=0,0008981$$
With this, we have from equation (2.4.1.9) that reproducibility is given by
$$VO=\sqrt{~V_{oper}^2~+~VI^2~}=\sqrt{0,00000187}=0,001367615$$
9th Step
$$RR=\sqrt{VE^2 + VO^2}=\sqrt{0,00000368}=0,001918719$$
10th Step
$$VT=\sqrt{~VE^2~+~VO^2~+~VP^2~}=\sqrt{0,0000264}=0,005138413$$
13th Step
Table 2.4.1.7 shows the percentage of each standard deviation in relation to the total variation.
| Source of variation | Standard deviation | % |
|---|---|---|
| Repetitiveness | 0,001345775 | 26,19048299 |
| Reproducibility | 0,001367615 | 26,61550832 |
| Operator | 0,001031400 | 20,07234423 |
| Part x Operator | 0,000898100 | 17,47816582 |
| Ask | 0,004766740 | 92,76676839 |
| RR | 0,001918719 | 37,34068401 |
| Total | 0,005138413 | 100 |
Table 2.4.1.7: Percentage of total variation
14th Step
$${ndc}=1,41\times\frac{VP}{RR}=3$$
Results of this example obtained with the Action software:



Conclusion
The measurement system needs to be improved, as we have a %RR of 37.4% and a value of ndc=3, which is considered low. Note that in this case, both repeatability and reproducibility are high (above 26%) and we have interaction between parts and operators, this means that we have parts that are more complicated to measure than others. Furthermore, notice that in the R graph, operator A presents points outside the control limits, which tells us that he did not adequately understand the measurement procedure. In short, we have several opportunities for improvement.
Example 2.4.1.2
Example of the Measurement System Analysis Manual (MSA 4th edition) page 118. Data collection from the measurement device is presented in the following table.
| Operator | Ask | Measure |
|---|---|---|
| 1 | 1 | 0,29 |
| 1 | 2 | -0,56 |
| 1 | 3 | 1,34 |
| 1 | 4 | 0,47 |
| 1 | 5 | -0,8 |
| 1 | 6 | 0,02 |
| 1 | 7 | 0,59 |
| 1 | 8 | -0,31 |
| 1 | 9 | 2,26 |
| 1 | 10 | -1,36 |
| 1 | 1 | 0,41 |
| 1 | 2 | -0,68 |
| 1 | 3 | 1,17 |
| 1 | 4 | 0,5 |
| 1 | 5 | -0,92 |
| 1 | 6 | -0,11 |
| 1 | 7 | 0,75 |
| 1 | 8 | -0,2 |
| 1 | 9 | 1,99 |
| 1 | 10 | -1,25 |
| 1 | 1 | 0,64 |
| 1 | 2 | -0,58 |
| 1 | 3 | 1,27 |
| 1 | 4 | 0,64 |
| 1 | 5 | -0,84 |
| 1 | 6 | -0,21 |
| 1 | 7 | 0,66 |
| 1 | 8 | -0,17 |
| 1 | 9 | 2,01 |
| 1 | 10 | -1,31 |
| 2 | 1 | 0,08 |
| 2 | 2 | -0,47 |
| 2 | 3 | 1,19 |
| 2 | 4 | 0,01 |
| 2 | 5 | -0,56 |
| 2 | 6 | -0,2 |
| 2 | 7 | 0,47 |
| 2 | 8 | -0,63 |
| 2 | 9 | 1,8 |
| 2 | 10 | -1,68 |
| 2 | 1 | 0,25 |
| 2 | 2 | -1,22 |
| 2 | 3 | 0,94 |
| 2 | 4 | 1,03 |
| 2 | 5 | -1,2 |
| 2 | 6 | 0,22 |
| 2 | 7 | 0,55 |
| 2 | 8 | 0,08 |
| 2 | 9 | 2,12 |
| 2 | 10 | -1,62 |
| 2 | 1 | 0,07 |
| 2 | 2 | -0,68 |
| 2 | 3 | 1,34 |
| 2 | 4 | 0,2 |
| 2 | 5 | -1,28 |
| 2 | 6 | 0,06 |
| 2 | 7 | 0,83 |
| 2 | 8 | -0,34 |
| 2 | 9 | 2,19 |
| 2 | 10 | -1,5 |
| 3 | 1 | 0,04 |
| 3 | 2 | -1,38 |
| 3 | 3 | 0,88 |
| 3 | 4 | 0,14 |
| 3 | 5 | -1,46 |
| 3 | 6 | -0,29 |
| 3 | 7 | 0,02 |
| 3 | 8 | -0,46 |
| 3 | 9 | 1,77 |
| 3 | 10 | -1,49 |
| 3 | 1 | -0,11 |
| 3 | 2 | -1,13 |
| 3 | 3 | 1,09 |
| 3 | 4 | 0,2 |
| 3 | 5 | -1,07 |
| 3 | 6 | -0,67 |
| 3 | 7 | 0,01 |
| 3 | 8 | -0,56 |
| 3 | 9 | 1,45 |
| 3 | 10 | -1,77 |
| 3 | 1 | -0,15 |
| 3 | 2 | -0,96 |
| 3 | 3 | 0,67 |
| 3 | 4 | 0,11 |
| 3 | 5 | -1,45 |
| 3 | 6 | -0,49 |
| 3 | 7 | 0,21 |
| 3 | 8 | -0,49 |
| 3 | 9 | 1,87 |
| 3 | 10 | -2,16 |
(image missing)
(image missing)
(image missing)
11.2.4.2 ANOVA model: no interaction between part and operator
Statistical Model
The model with two balanced factors and cross-effects without interaction is given by

For this model $ \mu $ is a parameter common to all treatments and represents the general average of the data, $ \alpha_{i} $ and $ \gamma_j $ is the effect due to the ith and jth level of the factor P (part) and O (operator) and are independent random variables with normal distributions with zero mean and variances $ \sigma^2_P $ and $ \sigma^2_O $ respectively. The random variable $ \varepsilon_{ijk} $ corresponds to experimental random error, that is, the variability not explained by the model due to variations present in several sources not considered in the study. This also has a normal distribution with zero mean, but the variance is $ \sigma^2. $
In short,
$$\mu={média~geral~dos~dados;}$$
$$\alpha_{i}={efeito~do~nível~i~do~fator~peça;}$$
$$\gamma_{j}={efeito~do~nível~j~do~fator~operador;}$$
$$\varepsilon_{ijk}={componente~aleatória~do~erro.}$$
Now, let’s develop the analysis of variance for the random effects model. Based on data considerations, we have:
$${soma das observações do nível i do fator peça;}$$
(there are missing elements in the equation above)
$$~{média das observações do nível i do fator peça;}$$
(there are missing elements in the equation above)
$${soma das observações do nível j do fator operador;}$$
(there are missing elements in the equation above)
$$~{média das observações do nível j do fator operador;}$$
(there are missing elements in the equation above)
$$~{soma de todas as observações;}\quad \text{e}$$
(there are missing elements in the equation above)
$${média geral das observações.}$$
(there are missing elements in the equation above)
Furthermore, we assume that the error has a Normal distribution with zero mean and variance $ \sigma^{2}_{\varepsilon} $ and that the errors are mutually independent. With this, we have to
$$\varepsilon_{ij}\sim~N(0,\sigma^{2}_{\varepsilon}).$$
Now, for the purpose $ \alpha_i $, we assume that it has a Normal distribution with zero mean and variance $ \sigma^2_{P}. $ We also assume that the effects are mutually independent. So,
$$\alpha_{i}\sim~N(0,\sigma^{2}_{P}).$$
To that end $ \gamma_j $, we assume that it has a Normal distribution with zero mean and variance $ \sigma^2_{O}. $ We also assume that the effects are mutually independent. So,
$$\gamma_{j}\sim~N(0,\sigma^{2}_{O}).$$
The analysis of variance for model (2.4.2.1) is obtained by decomposing the toral variation $ Y_{ijk}-\overline{Y_{…}} $ as follows
$$SQT=\sum_{i=1}^{a}\sum_{j=1}^{b}\sum_{k=1}^{r}(Y_{ijk}-\overline{Y_{…}})^2=$$
$$=o~r\sum_{i=1}^{p}(\overline{Y_{i..}}-\overline{Y_{…}})^2 +p~r\sum_{j=1}^{o}(\overline{Y_{.j.}}-\overline{Y_{…}})^2 +\sum_{i=1}^{p}\sum_{j=1}^{o}\sum^r_{k=1}(\overline{Y_{ijk}}-\overline{Y_{i..}}-\overline{Y_{.j.}}+\overline{Y_{…}})^2$$
$$=SQP+SQO+SQE$$
in which
$$SQT=\sum_{i=1}^{p}\sum_{j=1}^{o}\sum_{k=1}^{r}(Y_{ijk}-\overline{Y_{…}})^2$$
$$SQP=o~r\sum_{i=1}^{p}(\overline{Y_{i..}}-\overline{Y_{…}})^2$$
$$SQO=p~r\sum_{j=1}^{o}(\overline{Y_{.j.}}-\overline{Y_{…}})^2$$
$$SQE=\sum_{i=1}^{p}\sum_{j=1}^{o}\sum^r_{k=1}(\overline{Y_{ijk}}-\overline{Y_{i..}}-\overline{Y_{.j.}}+\overline{Y_{…}})^2$$
Thus, we obtain the ANOVA table as follows
| Factor | Degrees of Freedom | Sum of Squares | Medium Squares | F |
|---|---|---|---|---|
| Factor Piece P | p-1 | $ SQP $ | $ QMP=\frac{SQP}{p-1} $ | $ F^\ast_{1}=\frac{QMP}{QME} $ |
| Operator Factor O | o-1 | $ SQO $ | $ QMO=\frac{SQO}{o-1} $ | $ F^\ast_{2}=\frac{QMO}{QME} $ |
| Repetitiveness | por-p-o+1 | $ SQE $ | $ QME=\frac{SQE}{por-p-o+1} $ |
Now, we will briefly summarize the expected mean square values.
| Factor | Degrees of Freedom | Medium Squares | Expected Value of Mean Squares |
|---|---|---|---|
| Factor Piece P | p-1 | $ QMP $ | $ E(QMP)=\sigma^2_E+or\sigma^2_P $ |
| Operator Factor O | o-1 | $ QMO $ | $ E(QMO)=\sigma^2_E+pr\sigma^2_O $ |
| Repetitiveness | por-p-o+1 | $ QME $ | $ E(QME)=\sigma^2_E $ |
Table 2.4.2.1: Expected Values of Mean Squares.
Statistics
| $ QMP= $ | $ or\displaystyle\sum^p_{i=1}\frac{(\overline{Y_{i..}}-\overline{Y_{…}})^2}{p-1} $ |
|---|---|
| $ QMO= $ | $ pr\displaystyle\sum^o_{j=1}\frac{(\overline{Y_{.j.}}-\overline{Y_{…}})^2}{o-1} $ |
| $ QME= $ | $ \displaystyle\sum^p_{i=1}\sum^o_{j=1}\sum^r_{k=1}\frac{(\overline{Y_{ijk}}-\overline{Y_{i..}}-\overline{Y_{.j.}}+\overline{Y_{…}})^2}{por-p-o+1} $ |
| $ \overline{Y_{i..}}= $ | $ \displaystyle\sum^o_{j=1}\sum^r_{k=1}\frac{Y_{ijk}}{or} $ |
| $ \overline{Y_{.j.}}= $ | $ \displaystyle\sum^p_{i=1}\sum^r_{k=1}\frac{Y_{ijk}}{pr} $ |
| $ \overline{Y_{…}}= $ | $ \displaystyle\sum^p_{i=1}\sum^o_{j=1}\sum^r_{k=1}\frac{Y_{ijk}}{por} $ |
Table 2.4.2.2: Summary of mean squares and means for model (2.4.2.1).
Table 2.4.2.3 represents the point estimators of the model (2.4.2.1).
| Model Representation | Spot Estimator |
|---|---|
| $ \hat{\mu} $ | $ \overline{Y_{…}} $ |
| $ \hat{\sigma}^2_P $ or $ VP^2 $ | $ \displaystyle\frac{QMP-QME}{or} $ |
| $ \hat{\sigma}^2_O $ or $ V^2_{oper} $ | $ \displaystyle\frac{QMO-QME}{pr} $ |
| $ \hat{\sigma}^2_E $ or $ VE^2 $ | $ QME $ |
Table 2.4.2.3: Summary of Point Estimators for the model (2.4.2.1).
Here, we perform the following tests
- Part factor effect test:
$$\sigma^{2}_{P}> 0$$
the appropriate test statistic is
$$F^{*}=\frac{QMP}{QME}$$
and the decision rule at the significance level $ \alpha $ is

- Testing the effect of the operator factor:
$$\sigma^{2}_{O}> 0$$
the appropriate test statistic is
$$F^{*}=\frac{QMO}{QME}$$
and the decision rule at the significance level $ \alpha $ is

Next, we calculate the sources of variation from table (2.4.2.3) and divide them into steps as follows:
6th Step
- Perform graphical analysis: Build the control chart $ \overline{X} $ and S or $ \overline{X} $ and R.
7th Step
- Calculate repeatability
$$VE=\hat{\sigma}_E=\sqrt{QME}~~~~(2.4.2.2)$$
8th Step
- Calculate reproducibility
$$VO=V_{oper}=\hat{\sigma}_O=\sqrt{\frac{QMO-QME}{pr}}~~~~~(2.4.2.3)$$
9th Step
- Calculate the RR
$$RR=\sqrt{\hat{\sigma}^2_E+\hat{\sigma}^2_O}=\sqrt{VE^2+VO^2}~~~~(2.4.2.4)$$
10th Step
- Calculate the variation between parts
$$VP=\hat{\sigma}_P=\sqrt{ ~\frac{QMP - QME}{or}}~~~~(2.4.2.5)$$
11th Step
- Calculate the total variation
$$VT=\sqrt{RR^2+VP^2}~~~~~(2.4.2.6)$$
12th Step
- Contribution % table
| Sources of Variation | Variance | % Contribution |
|---|---|---|
| Repetitiveness | (VE)² | $ \displaystyle\frac{VE^2}{VT^2}*100 $ |
| Reproducibility | (VO)² | $ \displaystyle\frac{VO^2}{VT^2}*100 $ |
| Ask | (VP)² | $ \displaystyle\frac{VP^2}{VT^2}*100 $ |
| RR | (RR)² | $ \displaystyle\frac{RR^2}{VT^2}*100 $ |
| Total | (VT)² | 100 |
Table 2.4.2.4: Contribution table - no interaction
The values used to fill in Table 2.4.2.4 come from expressions (2.4.2.2) to (2.4.2.6).
13th Step
- Table of % in relation to the total variation and or % of tolerance
The values used to fill in the following table (Table 2.4.2.5) come from expressions (2.4.2.2) to (2.4.2.6).
| Source of variation | Standard Deviation | % Total Variation | % Tolerance |
|---|---|---|---|
| Repetitiveness | VE | $ \displaystyle\frac{VE}{VT}*100 $ | $ \left(6*\frac{VE}{TOL}\right)*100 $ |
| Reproducibility | GRANDMOTHER | $ \displaystyle\frac{VO}{VT}*100 $ | $ \left(6*\frac{VO}{TOL}\right)*100 $ |
| Ask | VP | $ \displaystyle\frac{VP}{VT}*100 $ | $ \left(6*\frac{VP}{TOL}\right)*100 $ |
| RR | RR | $ \displaystyle\frac{RR}{VT}*100 $ | $ \left(6*\frac{RR}{TOL}\right)*100 $ |
Table 2.4.2.5: Table of % of total variation and/or % of tolerance - no interaction
Suggested rule for analyzing variations
Tolerance: Measurement system applied in final inspections and receiving inspection;
Total variation: Measurement system used during the production process.
- %RR less than 10% → acceptable measurement system.
- %RR between 10% and 30% → marginal measurement system, which may be accepted depending on the situation, costs, etc.
- %RR greater than 30% → unacceptable measurement system, requiring improvement or replacement.
14th Step
- Calculate the ndc (Number of distinct categories)
The ndc represents the ability to discriminate categories of parts in a measurement system considering process variation. This index gives us the number of bands we can divide the process variation into. The ndc is given by the largest integer less than or equal to the value:
$$1,41\times\frac{VP}{RR}~~~~~(2.4.2.7)$$
The automotive industry manual (MSA) presents as a criterion an ndc ≥ 5. This means that the measurement system is capable of identifying 5 different types of parts within the process variation field. If a measuring system is evaluated by tolerance, the ndc index should not be considered.
15th Step
- Perform a data analysis to evaluate the system characteristics.
If the repeatability is high compared to the reproducibility, the reasons may be:
- The measuring device needs maintenance;
- The measuring device must be redesigned to be more robust;
- The fixation or position for measurement needs to be improved;
- There is an excessive variation inherent to the piece.
If the reproducibility is high compared to the repeatability, then the possible causes could be:
- The operator needs to be better trained, as well as how to use and read the measuring device;
- The markings on the measuring device dial are unclear;
- Some type of device may be needed to help the operator use the measuring device more consistently.
Example 2.4.2.1
In a study carried out to analyze the efficiency of the radio height measurement system, we used the altimeter. 5 pieces were randomly selected from different production shifts and identified appropriately. Operators who have been previously trained, when picking up radios with this date code, activate a program that will store the readings in a database for RR. We obtain the data presented in Table 2.4.2.6

Table 2.4.2.6: Readings
5th Step
SQP calculation
| Ask | Averages $ (\overline{Y_{i..}}) $ | Average $ (\overline{Y_{…}}) $ | Dev. Quad $ (\overline{Y_{i..}}-\overline{Y_{…}})^2 $ |
|---|---|---|---|
| 1 | 10,11888889 | 10,426481 | 0,094612907 |
| 2 | 10,23222222 | 10,426481 | 0,037736473 |
| 3 | 10,37555556 | 10,426481 | 0,002593401 |
| 4 | 10,13444444 | 10,426481 | 0,08528535 |
| 5 | 10,81444444 | 10,426481 | 0,150515634 |
| 6 | 10,88333333 | 10,426481 | 0,208714054 |
| Sum | 0,579457819 |
$$S^2_p=\frac{0,579457819}{6-1}=0,1159$$
$$SQP = 3 \times 3 \times (6-1) \times 0,1159 = 5,2151$$
- QME calculation
SQE calculation
$$SQE=5,4086-5,21-0,037=0,156$$
With this, we obtain that
$$QME=\cfrac{0,1935}{6 \times 3 \times 3-6-3+1}=0,003388$$
6th Step
Perform graphical analysis: build the control chart $ \overline{X} $and R to evaluate the means and amplitudes of each repetition per operator.
7th Step
$$VE=\sqrt{QME}$$
$$=\sqrt{0,003388}$$
$$=0,0582$$
8th Step
$$VO=\sqrt{V^2_{oper}}$$
$$=\sqrt{0,0188}$$
$$=0,0293$$
9th Step
$$RR=\sqrt{VE^2}$$
$$=\sqrt{0,004}$$
$$=0,063$$
10th Step
$$VP=\sqrt{\cfrac{QMP-QME}{or}}$$
$$\sqrt{\cfrac{1,04-0,00338}{(3*3)}}$$
$$=0,339$$
11th Step
$$VT=\sqrt{RR^2 + VP^2}$$
$$=\sqrt{0,00424 + 0,3398}$$
$$=0,3460mm$$
13th Step - Table of % in relation to the total variation
| Source of variation | Standard Deviation | % Total Variation |
|---|---|---|
| Repetitiveness | 0,0582 | 16,81 |
| Reproducibility | 0,0293 | 8,46 |
| Ask | 0,3398 | 98,21 |
| RR | 0,0651 | 18,829 |
| Total | 0,3460 | 100,00 |
14th Step
$$ndc=1,41\times\frac{VP}{RR}$$
$$=1,41\times\frac{0,3398}{0,0651}$$
$$=7$$
Results of this example obtained with the Action software:



Conclusion: We have a good measurement system. The RR percentage is low (19%) and in addition, we have an ndc=7, which tells us that this system has a good ability to discriminate parts.
11.2.4.3 ANOVA model: without operator
Statistical Model
The one-factor random model with balanced data is and the deductions are similar to the ANOVA model with fixed effects.

For this model $ \mu $ is a parameter common to all treatments and represents the general average of the data, $ \alpha_{i} $ is the effect due to the ith level of the factor. The random variable $ \varepsilon_{ij} $ corresponds to experimental random error, that is, the variability not explained by the model due to variations present in several sources not considered in the study.
In short,
$$\mu={média~geral~dos~dados;}$$
$$\alpha_{i}={efeito~i-ésimo~do~nível~i~do~fator~peça;}$$
$$\varepsilon_{ij}={componente~aleatória~do~erro.}$$
Now, let’s develop the analysis of variance for the random effects model. From considerations of the data, we have:
$${soma das observações do nível i do fator peça;}$$
(there are missing elements in the equation above)
$$~{média das observações do nível i do fator peça;}$$
(there are missing elements in the equation above)
$$~{soma de todas as observações;}\quad \text{e}$$
(there are missing elements in the equation above)
$${média geral das observações.}$$
(there are missing elements in the equation above)
Furthermore, we assume that the error has a Normal distribution with zero mean and variance $ \sigma^{2}_{\varepsilon} $ and that the errors are mutually independent. With this, we have to
$$\varepsilon_{ij}\sim~N(0,\sigma^{2}_{\varepsilon}).$$
Now, for the purpose $ \alpha_i $, we assume that it has a Normal distribution with zero mean and variance $ \sigma^2_{P}. $ We also assume that the effects are mutually independent. So,
$$\alpha_{i}\sim~N(0,\sigma^{2}_{P}).$$
The ANOVA technique is associated with partitioning the total data variability into components. The total sum of squares is defined as a measure of the total variability of the data,
$$SQT=\sum_{i=1}^{p}\sum_{j=1}^{r}(Y_{ij} - \overline{Y_{..}})^{2}=$$
$$=\sum_{i=1}^{p}\sum_{j=1}^{r}\left[(Y_{ij}-\overline{Y_{i.}})+(\overline{Y_{i.}}-\overline{Y_{..}})\right]^{2}=$$
$$=\sum_{i=1}^{p}\sum_{j=1}^{r}(Y_{ij}-\overline{Y_{i.}})^{2}+\sum_{i=1}^{k}\sum_{j=1}^{r}(\overline{Y_{i.}}-\overline{Y_{..}})^{2}=$$
$$=\sum_{i=1}^{p}\sum_{j=1}^{r}(Y_{ij}-\overline{Y_{i.}})^{2}+\sum_{i=1}^{p}r(\overline{Y_{i.}}-\overline{Y_{..}})^{2}$$
that is,
$$SQT=SQP+SQE.$$
Thus, we obtain the ANOVA table as follows
| Factor | Degrees of Freedom | Sum of Squares | Medium Squares | F |
|---|---|---|---|---|
| Factor Piece P | p-1 | $ SQP $ | $ QMP=\frac{SQP}{p-1} $ | $ F^\ast=\frac{QMP}{QME} $ |
| Repetitiveness | p(r-1) | $ SQE $ | $ QME=\frac{SQE}{p(r-1)} $ |
Now we show a brief summary of the mean squares expected values.
| Factor | Degrees of Freedom | Medium Squares | Expected Value of Mean Squares |
|---|---|---|---|
| Factor Piece P | p-1 | $ QMP $ | $ E(QMP)=\sigma^2_E+r\sigma^2_P $ |
| Repetitiveness | p(r-1) | $ QME $ | $ E(QME)=\sigma^2_E $ |
Table 2.4.3.1: Expected Values of Mean Squares.
Statistics
| $ QMP= $ | $ r\displaystyle\sum^p_{i=1}\frac{(\overline{Y_{i.}}-\overline{Y_{..}})^2}{p-1} $ |
|---|---|
| $ QME= $ | $ \displaystyle\sum^p_{i=1}\sum^r_{j=1}\frac{(\overline{Y_{ij}}-\overline{Y_{i.}})^2}{p(r-1)} $ |
| $ \overline{Y_{i.}}= $ | $ \displaystyle\sum^r_{j=1}\frac{Y_{ij}}{r} $ |
| $ \overline{Y_{..}}= $ | $ \displaystyle\sum^p_{i=1}\sum^r_{j=1}\frac{Y_{ij}}{pr} $ |
Table 2.4.3.2: Summary of mean squares and means for model (2.4.3.1).
Table 2.4.3.3 represents the point estimators of the model (2.4.3.1).
| Model Representation | Spot Estimator |
|---|---|
| $ \hat{\mu} $ | $ \overline{Y_{…}} $ |
| $ \hat{\sigma}^2_P $ or $ VP^2 $ | $ \displaystyle\frac{QMP-QME}{r} $ |
| $ \hat{\sigma}^2_E $ or $ VE^2 $ | $ QME $ |
Table 2.4.3.3: Summary of Point Estimators for the model (2.4.3.1).
Here, we perform the following tests
- Part factor effect test:
$$\sigma^{2}_{P}> 0$$
the appropriate test statistic is
$$F^{*}=\frac{QMP}{QME}$$
and the decision rule at the significance level $ \alpha $ is

Next, we calculate the sources of variation from table (2.4.3.3) and divide them into steps as follows:
6th Step
- Perform graphical analysis: Build the control chart $ \overline{X} $ and S or $ \overline{X} $ and R.
7th Step
- Calculate repeatability
$$VE=\hat{\sigma}_E=\sqrt{QME}~~~~(2.4.3.2)$$
8th Step
- Calculate the RR
$$RR=\sqrt{\hat{\sigma}^2_E}=\sqrt{VE^2}~~~~(2.4.3.3)$$
9th Step
- Calculate the variation between parts
$$VP=\hat{\sigma}_P=\sqrt{ ~\frac{QMP - QME}{r}}~~~~(2.4.3.4)$$
10th Step
- Calculate the total variation
$$VT=\sqrt{RR^2+VP^2}~~~~~(2.4.3.5)$$
11th Step
- Contribution % table
| Sources of Variation | Variance | % Contribution |
|---|---|---|
| Repetitiveness | (VE)² | $ \displaystyle\frac{VE^2}{VT^2}*100 $ |
| Ask | (VP)² | $ \displaystyle\frac{VP^2}{VT^2}*100 $ |
| RR | (RR)² | $ \displaystyle\frac{RR^2}{VT^2}*100 $ |
| Total | (VT)² | 100 |
Table 2.4.3.4: Contribution table - no interaction
The values used to fill in Table 2.4.3.4 come from expressions (2.4.3.2) to (2.4.3.5).
12th Step
- Table of % in relation to the total variation and or % of tolerance
The values used to fill in the following table (Table 2.4.3.5) come from expressions (2.4.3.2) to (2.4.3.5).
| Source of variation | Standard Deviation | % Total Variation | % Tolerance |
|---|---|---|---|
| Repetitiveness | VE | $ \displaystyle\frac{VE}{VT}*100 $ | $ \left(6*\frac{VE}{TOL}\right)*100 $ |
| Ask | VP | $ \displaystyle\frac{VP}{VT}*100 $ | $ \left(6*\frac{VP}{TOL}\right)*100 $ |
| RR | RR | $ \displaystyle\frac{RR}{VT}*100 $ | $ \left(6*\frac{RR}{TOL}\right)*100 $ |
Table 2.4.3.5: Table of % of total variation and/or % of tolerance - no interaction
Example 2.4.3.1
Consider the study carried out to analyze the efficiency of the measurement system for measuring the door dimensions of an excavator. The measuring system uses a CNC coordinate measuring machine. In this case, we consider that the operator does not influence the measurement, a fact that led us to consider only one operator and 15 parts in the analysis.
| Ask | Replica I | Replica II | Replica III | Average | Standard Deviation | Amplitudes |
|---|---|---|---|---|---|---|
| 1 | 461,28 | 461,5 | 461,2 | 461,327 | 0,1553 | 0,3 |
| 2 | 458,17 | 458,62 | 458,61 | 458,467 | 0,2569 | 0,45 |
| 3 | 460,57 | 460,28 | 460,32 | 460,39 | 0,1571 | 0,29 |
| 4 | 459,28 | 459,66 | 459,58 | 459,507 | 0,2 | 0,38 |
| 5 | 461,28 | 461,12 | 461,18 | 461,193 | 0,0808 | 0,16 |
| 6 | 460,25 | 460,68 | 460,28 | 460,403 | 0,24 | 0,43 |
| 7 | 458,82 | 458,95 | 458,66 | 458,81 | 0,1452 | 0,29 |
| 8 | 461,58 | 461,1 | 461,18 | 461,287 | 0,2571 | 0,48 |
| 9 | 459,36 | 459,52 | 459,57 | 459,483 | 0,1067 | 0,21 |
| 10 | 459,62 | 459,34 | 459,54 | 459,5 | 0,1442 | 0,28 |
| 11 | 461,38 | 461,57 | 461,53 | 461,4933 | 0,1001 | 0,19 |
| 12 | 458,67 | 459,03 | 458,98 | 458,8933 | 0,195 | 0,36 |
| 13 | 462,57 | 462,28 | 462,32 | 462,39 | 0,1571 | 0,29 |
| 14 | 459,58 | 459,66 | 459,28 | 459,5067 | 0,2003 | 0,38 |
| 15 | 461,76 | 461,12 | 461,15 | 461,3433 | 0,3611 | 0,64 |
Standard Deviation between the Averages of the pieces $ \mathbf{s_{p}}= 1,1847 $
$$QME = 0,03884$$
6th Step
Perform graphical analysis: build the control chart $ \overline{X} $ and R to evaluate the means and amplitudes of each repetition per operator.
7th Step
$$VE=\sqrt{QME}$$
$$=\sqrt{0,038}$$
$$=0,197$$
8th Step
$$VO = \sqrt{~V_{oper}^2}$$
$$=\sqrt{0}$$
$$=0$$
9th Step
$$RR=\sqrt{VE^2 + VO^2}$$
$$=\sqrt{0,038+0}$$
$$=0,197$$
10th Step
$$VP = \sqrt{~\frac{QMP - QME}{or}}$$
$$= s_{p}$$
$$=1,185$$
11th Step
$$VT = \sqrt{RR^2 + VP^2}$$
$$=\sqrt{0,197+1,184}$$
$$=1,20$$
13th Step - Table of % in relation to the total variation
TOTAL VARIATION AND/OR TOLERANCE
| Standard deviation | Total variation (%) | |
|---|---|---|
| Repetitiveness | 0,197073027 | 16,40808397 |
| Parts | 1,184794492 | 98,64468957 |
| Repeatability and reproducibility | 0,197073027 | 16,40808397 |
| Total | 1,201072757 | 100 |
14th Step
$$ndc=1,41 \times \frac{VP}{RR}$$
$$=1,41 \times \frac{1,18}{0,197}$$
$$=8$$
Results of this example obtained with the Action software:



Example 2.4.3.2
Example of the Measurement System Analysis Manual (MSA 4th edition) page 118. Data collection from the measurement device is presented in the following table.
| Operator | Ask | Measure |
|---|---|---|
| 1 | 1 | 0,29 |
| 1 | 2 | -0,56 |
| 1 | 3 | 1,34 |
| 1 | 4 | 0,47 |
| 1 | 5 | -0,8 |
| 1 | 6 | 0,02 |
| 1 | 7 | 0,59 |
| 1 | 8 | -0,31 |
| 1 | 9 | 2,26 |
| 1 | 10 | -1,36 |
| 1 | 1 | 0,41 |
| 1 | 2 | -0,68 |
| 1 | 3 | 1,17 |
| 1 | 4 | 0,5 |
| 1 | 5 | -0,92 |
| 1 | 6 | -0,11 |
| 1 | 7 | 0,75 |
| 1 | 8 | -0,2 |
| 1 | 9 | 1,99 |
| 1 | 10 | -1,25 |
| 1 | 1 | 0,64 |
| 1 | 2 | -0,58 |
| 1 | 3 | 1,27 |
| 1 | 4 | 0,64 |
| 1 | 5 | -0,84 |
| 1 | 6 | -0,21 |
| 1 | 7 | 0,66 |
| 1 | 8 | -0,17 |
| 1 | 9 | 2,01 |
| 1 | 10 | -1,31 |
| 2 | 1 | 0,08 |
| 2 | 2 | -0,47 |
| 2 | 3 | 1,19 |
| 2 | 4 | 0,01 |
| 2 | 5 | -0,56 |
| 2 | 6 | -0,2 |
| 2 | 7 | 0,47 |
| 2 | 8 | -0,63 |
| 2 | 9 | 1,8 |
| 2 | 10 | -1,68 |
| 2 | 1 | 0,25 |
| 2 | 2 | -1,22 |
| 2 | 3 | 0,94 |
| 2 | 4 | 1,03 |
| 2 | 5 | -1,2 |
| 2 | 6 | 0,22 |
| 2 | 7 | 0,55 |
| 2 | 8 | 0,08 |
| 2 | 9 | 2,12 |
| 2 | 10 | -1,62 |
| 2 | 1 | 0,07 |
| 2 | 2 | -0,68 |
| 2 | 3 | 1,34 |
| 2 | 4 | 0,2 |
| 2 | 5 | -1,28 |
| 2 | 6 | 0,06 |
| 2 | 7 | 0,83 |
| 2 | 8 | -0,34 |
| 2 | 9 | 2,19 |
| 2 | 10 | -1,5 |
| 3 | 1 | 0,04 |
| 3 | 2 | -1,38 |
| 3 | 3 | 0,88 |
| 3 | 4 | 0,14 |
| 3 | 5 | -1,46 |
| 3 | 6 | -0,29 |
| 3 | 7 | 0,02 |
| 3 | 8 | -0,46 |
| 3 | 9 | 1,77 |
| 3 | 10 | -1,49 |
| 3 | 1 | -0,11 |
| 3 | 2 | -1,13 |
| 3 | 3 | 1,09 |
| 3 | 4 | 0,2 |
| 3 | 5 | -1,07 |
| 3 | 6 | -0,67 |
| 3 | 7 | 0,01 |
| 3 | 8 | -0,56 |
| 3 | 9 | 1,45 |
| 3 | 10 | -1,77 |
| 3 | 1 | -0,15 |
| 3 | 2 | -0,96 |
| 3 | 3 | 0,67 |
| 3 | 4 | 0,11 |
| 3 | 5 | -1,45 |
| 3 | 6 | -0,49 |
| 3 | 7 | 0,21 |
| 3 | 8 | -0,49 |
| 3 | 9 | 1,87 |
| 3 | 10 | -2,16 |



11.2.4.4 Performance Curves
Purpose of the performance curve
The construction of a measuring device performance curve (GPC) aims to illustrate how the probability of acceptance (Pa) of a part varies depending on its reference value (VR). Once the measurement error has been characterized, that is, its mean and standard deviation are determined, it becomes possible to calculate the probability of acceptance of the part given the reference value. Admitting that the error is normally distributed with a mean equal to the trend (b) and the standard deviation (σ) equal to the RR, it is concluded that the measurement value of a part has a normal distribution with a mean equal to VR+b and standard deviation equal to the RR.
Probability of acceptance
Therefore, the probability of accepting a part with a reference value equal to VR is given by

with LIE and LSE being the Lower and Upper Specification Limits, respectively. We can rewrite this formula as
$$P_a=\Phi\left(\frac{LSE-(VR+b)}{{RR}}\right)-\Phi\left(\frac{LIE-(VR+b)}{{RR}}\right),~~~~~~(2.4.4.1)$$
where Φ is the cumulative distribution function of the standard normal N(0,1).
Example 2.4.4.1
Consider a rod production process whose specification limits are given by LIE = 0.6 mm and LSE = 1 mm. An analysis of the measurement system was carried out finding a bias b = 0.05 mm and RR = 0.05 mm. Determine the probability of accepting parts whose reference values (VR) are 0.5 mm, 0.7 mm, and 0.9 mm.
For the part with reference value VR = 0.5 mm, the probability of acceptance is given by
$$P_a=\Phi \left(\frac{LSE - (VR + b)}{RR}\right) - \Phi\left(\frac{LIE - (VR + b)}{RR}\right)$$
$$=\Phi \left(\frac{1,0 - (0,5 + 0,05)}{0,05}\right) - \Phi\left(\frac{0,6 - (0,5 + 0,05)}{0,05}\right)$$
$$=\Phi \left(9,0\right) - \Phi \left(1,0\right) = 1,0 - 0,84$$
$$=0,16.$$
For the part with reference value VR = 0.7 mm, the probability of acceptance is given by
$$P_a=\Phi \left(\frac{1,0-(0,7+0,05)}{0,05}\right)-\Phi\left(\frac{0,6-(0,7+0,05)}{0,05}\right)$$
$$=\Phi\left(5,0\right)-\Phi \left(-3,0\right)=0,999.$$
While for the part with reference value VR = 0.9 mm, the probability of acceptance is given by
$$P_a=\Phi\left(\frac{1,0-(0,9+0,05)}{0,05}\right)-\Phi\left(\frac{0,6-(0,9 + 0,05)}{0,05}\right)$$
$$=\Phi\left(1,0\right)-\Phi \left(-7,0\right)=0,84$$
By calculating this probability of acceptance for various values of VR, we obtain the GPC curve illustrated in Figure 2.4.4.1.

Figure 2.4.4.1: GPC curve.
Figure 2.4.4.2 illustrates the GPC curve for an ideal measurement system.

Figure 2.4.4.2: Ideal GPC curve.
11.2.4.5 Graphical analysis of RR
In this module we will discuss the main graphs associated with the study of repeatability and reproducibility. To do this, we used an experiment in graphical analysis with 3 operators, 5 parts and 3 measurements per operator for each part. Below we summarize the experiment in table (2.4.5.1).
| Operator | Ask | Measure 1 | Measure 2 | Measure 3 | R | $ \overline{Y} $ |
|---|---|---|---|---|---|---|
| A | 1 | $ Y_{1A1} $ | $ Y_{1A2} $ | $ Y_{1A3} $ | $ R_{1A} $ | $ \overline{Y_{1A}} $ |
| A | 2 | $ Y_{2A1} $ | $ Y_{2A2} $ | $ Y_{2A3} $ | $ R_{2A} $ | $ \overline{Y_{2A}} $ |
| A | 3 | $ Y_{3A1} $ | $ Y_{3A2} $ | $ Y_{3A3} $ | $ R_{3A} $ | $ \overline{Y_{3A}} $ |
| A | 4 | $ Y_{4A1} $ | $ Y_{4A2} $ | $ Y_{4A3} $ | $ R_{4A} $ | $ \overline{Y_{4A}} $ |
| A | 5 | $ Y_{5A1} $ | $ Y_{5A2} $ | $ Y_{5A3} $ | $ R_{5A} $ | $ \overline{Y_{5A}} $ |
| B | 1 | $ Y_{1B1} $ | $ Y_{1B2} $ | $ Y_{1B3} $ | $ R_{1B} $ | $ \overline{Y_{1B}} $ |
| B | 2 | $ Y_{2B1} $ | $ Y_{2B2} $ | $ Y_{2B3} $ | $ R_{2B} $ | $ \overline{Y_{2B}} $ |
| B | 3 | $ Y_{3B1} $ | $ Y_{3B2} $ | $ Y_{3B3} $ | $ R_{3B} $ | $ \overline{Y_{3B}} $ |
| B | 4 | $ Y_{4B1} $ | $ Y_{4B2} $ | $ Y_{4B3} $ | $ R_{4B} $ | $ \overline{Y_{4B}} $ |
| B | 5 | $ Y_{5B1} $ | $ Y_{5B2} $ | $ Y_{5B3} $ | $ R_{5B} $ | $ \overline{Y_{5B}} $ |
| C | 1 | $ Y_{1C1} $ | $ Y_{1C2} $ | $ Y_{1C3} $ | $ R_{1C} $ | $ \overline{Y_{1C}} $ |
| C | 2 | $ Y_{2C1} $ | $ Y_{2C2} $ | $ Y_{2C3} $ | $ R_{2C} $ | $ \overline{Y_{2C}} $ |
| C | 3 | $ Y_{3C1} $ | $ Y_{3C2} $ | $ Y_{3C3} $ | $ R_{3C} $ | $ \overline{Y_{3C}} $ |
| C | 4 | $ Y_{4C1} $ | $ Y_{4C2} $ | $ Y_{4C3} $ | $ R_{4C} $ | $ \overline{Y_{4C}} $ |
| C | 5 | $ Y_{5C1} $ | $ Y_{5C2} $ | $ Y_{5C3} $ | $ R_{5C} $ | $ \overline{Y_{5C}} $ |
| Average | $ \overline{R} $ | $ \overline{\overline{Y}} $ |
Table 2.4.5.1: Readings.
From the table (2.4.5.1) and the table of constants to calculate the control limits $ A_2,~D_3 $ and $ D_4 $ (see Appendix of CEP statistical content), we define the limits of the graphs:
- For averages:
Upper Control Limit:
$$LSC = \overline{\overline{Y}} + A_2 \ast \overline{R}=\overline{\overline{Y}}+1,023\ast \overline{R}$$
Central Line:
$$LC = \overline{\overline{Y}}$$
Lower Control Limit:
$$LIC = \overline{\overline{Y}} - A_2 \ast \overline{R}=\overline{\overline{Y}}-1,023\ast\overline{R}$$
- For amplitudes:
Upper control limit:
$$LSC = D_4 \ast \overline{R}=2,574\ast\overline{R}$$
Central Line:
$$LC = \overline{R}$$
Lower Control Limit:
$$LIC = D_3 \ast \overline{R}=0\ast\overline{R}=0$$
Next, we will study the most common cases that occur in an RR study.
Case 1: measurement system discrimination

Figure 2.4.5.1: Breakdown of the measurement system.
Conclusion
Figure 2.4.5.1 indicates that there is a lack of discrimination in the Measurement System. In general, the resolution of measuring equipment may be inadequate. According to the Automobile Industry Manual (MSA), most points on the Amplitude Graph (Figure 2.4.5.1) must be non-zero.
Case 2: R graph analysis - training

Figure 2.4.5.2: R graph analysis - training.
Conclusion
Figure 2.4.5.2 presents a problem related to training in the method. We may also have problems with the method used, perhaps the method requires a manual skill that not all people have.
Case 3: R graph analysis - method

Figure 2.4.5.3: R graph analysis - method.
Conclusion
When we have a graph of the amplitude as shown in Figure 2.4.5.3 we must check that the points (beyond the limits) between the evaluators do not correspond to the same part. If so, evaluate the piece. If it is not the same part, the method needs to be revised, as the evaluators are unable to reproduce it.
Case 4: repeatability analysis

Figure 2.4.5.4: Repeatability analysis.
Conclusion
According to Figure 2.4.5.4, the more points outside the control limits, the better. The amplitude between the control lines, LSC - LIC = $2A_2 \overline{R} $ reflects the repeatability of the measurement system, while the points reflect the variability between parts (production process). In this case, we compare the repeatability of the SM (control limits) with the variability of the production process (points on the graph). According to the automobile industry manual (MSA), when we analyze the RR by total variation, most of the points in the Average Graph (Figure 2.4.5.4) must be outside the control limits.
Case 5: reproducibility analysis

Figure 2.4.5.5: Reproducibility analysis.
Conclusion
In this case, we will compare the measurements between the evaluators. As the evaluators are measuring the same parts, the measurements must be similar, as shown in Figure 2.4.5.5. Therefore, the more parallel the line that joins the averages is in relation to the x-axis, the better.
Case 6: part interaction versus measurement system

Figure 2.4.5.6: Interaction between part and measurement system.
Conclusion
In the parts chart, see Figure 2.4.5.6, we evaluate the consistency of the Measurement System in relation to the parts using the following criteria:
a) The parts are distinct, therefore, the Measurement System must identify them. Therefore, the measurements of the parts cannot be aligned.
b) If one part varies more than the others (like part 2 for example), it means that the Measurement System had more difficulty in evaluating this part. Analyze the part and identify the cause.
Case 7: part versus evaluator interaction

Figure 2.4.5.7: Graph without interaction.

Figure 2.4.5.8: Graph with interaction.
We observed in Figure 2.4.5.7 that there is no interaction between evaluators and the part, since the measurements of the parts practically do not vary according to the evaluator. On the contrary, in Figure 2.4.5.8, we notice that the average of the measurements increases or decreases, depending on the evaluator. For example, the average of assessor C’s measurements is lowest for part 2 and highest for part 4.
11.3 Measurement System Analysis - Not Replicable
The non-replicable measurement system corresponds to measurement systems whose readings cannot be repeated on each part, we include systems where the parts undergo changes during testing or are destroyed. Below we present some analysis methods, which allow us to obtain information about the Variability of the process and its Stability.
11.3.1 Stability
In this module, we present a strategy for evaluating the stability of non-replicable measurement systems. Next, we present the sample division technique.
Guidelines
- Production process under statistical control;
- The part does not degrade during the experiment;
- Availability of a large number of parts;
- If possible, use reference standards that are representative of the process, rather than parts;
Analysis methods
1st Step:
To carry out this test it will be necessary to use a large quantity of parts (50 or more) whose characteristics such as raw material used, machine that produced them, process temperature and operator are as uniform as possible.
2nd Step:
We selected approximately half of the pieces and measured them in sequence and over a short period of time. Our goal is to determine the short-term variation of the measurement system. Through these measurements we created the CEP I-MR graph, with I being the individual values and MR being the moving amplitudes - with which we verify the variability of the process. The control limits of these graphs are calculated according to Table 3.1.1.
3rd Step:
The remaining pieces are measured periodically (daily, weekly, monthly), in sequence, and if possible at different times in cases where it is believed that this condition may be relevant to assess stability. Next, we prepare the I-MR graph with the same control limits obtained in step 2. If all points are within the control limits, we conclude that the process is stable.
This technique compares short-term variation (step 2), in which measurements are carried out in sequence and over a short period of time, with long-term variation (step 3), in which measurements are carried out over time. If the short-term and long-term variations are similar, we conclude that the system is stable.
| Graph Limits | Number of grouped readings (n) | $E_2$ | $D_3$ | $D_4$ |
|---|---|---|---|---|
| Chart of Individual Values (I) | 2 | 2,66 | 0 | 3,267 |
| LSC = Upper Limit =$ \overline{X} + E_2 \overline{R} $ | 3 | 1,77 | 0 | 2,574 |
| LC = Central Limit =$ \overline{X} $ | 4 | 1,46 | 0 | 2,282 |
| LIC = Lower Limit =$ \overline{X} -E_2 \overline{R} $ | 5 | 1,29 | 0 | 2,114 |
| R Amplitude Graph | 6 | 1,18 | 0 | 2,004 |
| LSC = Upper Limit = $D_4 \overline{R} $ | 7 | 1,11 | 0,076 | 1,924 |
| LC = Central Limit =$ \overline{R} $ | 8 | 1,05 | 0,136 | 1,864 |
| LIC = Lower Limit = $D_3 \overline{R} $ | 9 | 1,01 | 0,184 | 1,816 |
| 10 | 0,98 | 0,223 | 1,777 |
Table 3.1.1: Values of $E_2$, $D_3$ and $D_4$
Example 3.1.1
- Consider the tensile strength characteristic performed with steel specimens.
Objective:
Study the stability of the traction measurement system.
Experiment Description:
- Produce a very homogeneous batch of specimens (same run);
- The specimens do not degrade during the experiment;
- Availability of a large number of test specimens.
1st Step:
To carry out the experiment, we used 32 test specimens whose characteristics, such as the raw material used, the machine that produced them, the process and the evaluator, are as uniform as possible.
2nd Step:
Using 16 specimens (half of the batch), we carried out a stability study of the I-MR process with which we verified, through the control limits, the variability of the process. These 16 specimens are measured on the same day (Table 3.1.2).
3rd Step:
The remaining 16 pieces are measured periodically (weekly) at different times (Table 3.1.3). As a criterion for evaluating stability, we used the analysis of the I-MR control chart, with the same control limits obtained in the 2nd step.
- Tensile strength (MPA)
- Measurement of 16 specimens, carried out on the same day (Step 2);
- Evaluator: A
- Machine: Amsler
- Race: 0782697
- Date: 02/26/03
For the data in Table 3.1.2, we obtain the mean and mean amplitude values as
$ \overline{X} $= 1160,625
$ \overline{R} $= 22,533
Readings:
| 1164 | 1135 |
|---|---|
| 1171 | 1195 |
| 1181 | 1164 |
| 1155 | 1161 |
| 1159 | 1151 |
| 1119 | 1171 |
| 1119 | 1193 |
| 1166 | 1166 |
Table 3.1.2: Same day measurements
With this, we obtain the control limits for the individual measurements as
$$LIC=\overline{X}-E_2\times \overline{R}=1160,625-2,66\times22,53=1100,7$$
$$LC=\overline{X}=1160,625$$
$$LSC=\overline{X}+E_2\times\overline{R}=1160,625+2,66\times22,53=1220,5$$
and the control limits for the amplitude as
$$LIC=D_4\times\overline{R}=3,267\times\overline{R}=73,616$$
$$LC=\overline{R}=22,533$$
$$LSC=D_3\times\overline{R}=0\times\overline{R}=0$$
with the values of E2, D3 and D4 obtained from Table 3.1.2 with n = 2.
Below, we present graphs of individual values and moving ranges for the parts that were measured on the same day:
(image missing)
The short-term standard deviation estimate is given by
$$\sigma=\frac{\overline{R}}{d_2}$$
where the value of $d_2$ is obtained from the Table of $d_2$ with g > 15 (number of amplitudes to calculate $\overline{R}$) and m = 2 (number of measurements used to calculate each amplitude).
Therefore, we have that the estimated standard deviation is given by
$$\sigma=\frac{\overline{R}}{d_2}=\frac{22,53}{1,128}=19,97$$
In step 3, the remaining 16 specimens are measured weekly, with the results presented in Table 22.
- Measurement of 16 specimens, carried out weekly (Step 3);
- Evaluator: A
- Machine: Amsler
- Race: 0782697
Below, we present graphs of individual values and moving amplitudes for the parts that were measured over time:
| Date | Schedule | Resistance |
|---|---|---|
| 07/03/2003 | 10:30 | 1117 |
| 14/03/2003 | 15:35 | 1152 |
| 19/03/2003 | 12:35 | 1182 |
| 25/03/2003 | 08:25 | 1206 |
| 01/04/2003 | 10:45 | 1195 |
| 10/04/2003 | 07:35 | 1175 |
| 17/04/2003 | 08:55 | 1171 |
| 25/04/2003 | 11:05 | 1161 |
| 30/04/2003 | 09:05 | 1137 |
| 05/05/2003 | 13:00 | 1182 |
| 12/05/2003 | 14:20 | 1153 |
| 21/05/2003 | 12:50 | 1188 |
| 29/05/2003 | 09:10 | 1168 |
| 05/06/2003 | 09:35 | 1141 |
| 09/06/2003 | 11:00 | 1177 |
| 11/06/2003 | 09:00 | 1142 |
Table 3.1.3: Table with the remaining 16 measurements
Using the same control limits as the graph of individual values and moving amplitudes with measurements from the same day (16 specimens), we set up the graph of individual values and moving amplitudes for the measurements carried out weekly, as shown in Figure 15.

Figure 3.1.1: Graph of individual values and moving range, $ \overline{X} $= 1160;625 and $ \sigma $= 19;97.
Conclusion
When analyzing the graph of moving amplitudes, we observed that there is no point outside the control limits, the same happening for the graph of individual values. Therefore, we conclude that the measurement system is stable with respect to tensile strength.
11.3.2 Non-Replicable RR (Hierarchical Method)
In this module, we will analyze the variability (RR study) of non-replicable measurement systems. When carrying out the RR study of a replicable measurement system, we obtain an experimental arrangement as shown in Table 3.2.1. However, for a non-replicable measuring system this arrangement is not possible, as we cannot measure the same part several times. As an example, consider destructive welding testing where a welded nut is pulled off a part and the maximum amount of force before pullout is measured. The weld is destroyed in the process and therefore cannot be evaluated again. Therefore, we cannot measure the same part more than once and in this case, the arrangement described in Table 3.2.1 does not apply.
| Operator 1 | Operator 1 | Operator 1 | Operator 2 | Operator 2 | Operator 2 | |
|---|---|---|---|---|---|---|
| Ask | Part I | Part II | Part III | Part I | Part II | Part III |
| 1 | 1 | 1 | 1 | 1 | 1 | 1 |
| 2 | 2 | 2 | 2 | 2 | 2 | 2 |
| 3 | 3 | 3 | 3 | 3 | 3 | 3 |
| 4 | 4 | 4 | 4 | 4 | 4 | 4 |
| 5 | 5 | 5 | 5 | 5 | 5 | 5 |
| 6 | 6 | 6 | 6 | 6 | 6 | 6 |
| 7 | 7 | 7 | 7 | 7 | 7 | 7 |
| 8 | 8 | 8 | 8 | 8 | 8 | 8 |
| 9 | 9 | 9 | 9 | 9 | 9 | 9 |
| 10 | 10 | 10 | 10 | 10 | 10 | 10 |
Table 3.2.1: Readings.
The first step to take before tackling a non-replicable RR study is to ensure that all conditions encompassing the test are defined, standardized and controlled - operators must be similarly qualified and trained, lighting must be adequate and always controlled, work instructions must be detailed and operationally defined, environmental conditions must be controlled to an appropriate degree, equipment must be calibrated and adequately maintained, etc. Secondly, before making a study of a non-replicable measurement system, it is necessary to verify that the production process is stable and capable.
After this, since the piece cannot be reevaluated due to changes in its structure (or destruction), several similar (homogeneous) pieces must be chosen for the study and the assumption must be made that the pieces are identical (or similar). Sets of homogeneous parts are called lots. In this case, we have the experimental arrangement defined in Table 3.2.2, in which part 1 is treated as batch 1-1, 1-2, 1-3, 1-4, 1-5, 1-6, that is, they are different parts that are treated as if they were the same part. Therefore, the parts must be sampled consecutively (within the same production batch) and are identical (or similar) enough so that they can be treated as if they were the same part. If the process of interest does not satisfy this assumption, this method will not work. Generally if these parts are taken from production consecutively, under similar production conditions as much as possible, this requirement is met.
Study of the variability of the traction measurement system
| OPERATOR 1 | OPERATOR 1 | OPERATOR 1 | OPERATOR 2 | OPERATOR 2 | OPERATOR 2 | |
|---|---|---|---|---|---|---|
| Lot | Part 1 | Part 2 | Part 3 | Part 1 | Part 2 | Part 3 |
| 1A. . . 1F | 1-1 | 1-2 | 1-3 | 1-4 | 1-5 | 1-6 |
| 2A. . . 2F | 2-1 | 2-2 | 2-3 | 2-4 | 2-5 | 2-6 |
| 3A. . . 3F | 3-1 | 3-2 | 3-3 | 3-4 | 3-5 | 3-6 |
| 4A. . . 4F | 4-1 | 4-2 | 4-3 | 4-4 | 4-5 | 4-6 |
| 5A. . . 5F | 5-1 | 5-2 | 5-3 | 5-4 | 5-5 | 5-6 |
| 6A. . . 6F | 6-1 | 6-2 | 6-3 | 6-4 | 6-5 | 6-6 |
| 7A. . . 7F | 7-1 | 7-2 | 7-3 | 7-4 | 7-5 | 7-6 |
| 8A. . . 8F | 8-1 | 8-2 | 8-3 | 8-4 | 8-5 | 8-6 |
| 9A. . . 9F | 9-1 | 9-2 | 9-3 | 9-4 | 9-5 | 9-6 |
| 10A. . . 10F | 10-1 | 10-2 | 10-3 | 10-4 | 10-5 | 10-6 |
Table 3.2.2: Readings
In the experimental setup defined for replicable systems (see Table 3.2.1), all combinations of levels between factors (part and operator) are well defined. In this case, we have the crossing between all levels of all factors (cross factorial experiments).
On the other hand, in the experimental arrangement defined for non-replicable systems (see Table 3.2.2), the batch factor levels (similar parts) occur in combination with the operator factor levels, for example, part 1-1 was measured only by operator A. Such experimental arrangements are called hierarchical (“nested”). In figure 3.2.1, we present a hierarchical scheme applied to a non-replicable measurement system.

Figure 3.2.1: Sketch of a hierarchical experiment.
Statistical Model
We denote by $ Y_{ijk} $ the value of the measurement of the kth part, from operator i to part j.
$$Y_{ijk}=\mu + \alpha_{i}+\beta_{j(i)}+ \epsilon_{(ij)k}$$
in which,
$ \mu $ = constant
$ \alpha_{i} $ = random effect due to the operator, $ i = 1,\dots,a $ (number of operators). We assume that $ \alpha_i $ has a normal distribution with zero mean and variance $ \sigma_{\alpha}^2 $.
$ \beta_{j(i)} $ = random effect due to the j-th hierarchical batch under the ith operator,$ j = 1, \dots, b $ (number of lots).
$ \epsilon_{(ij)k} $ = error associated with each observation, hierarchical in relation to operators and batches, $ k = 1,\dots, r $ (number of parts, replicas).
The analysis of variance for the above statistical model is obtained by decomposing the total variance $ Y_{ijk} - \overline{Y_{…}} $ as follows
$$\underbrace{Y_{ijk} -\overline{Y_{…}}}_{\text{Variação Total}}=\underbrace{\overline{Y_{i..}}-\overline{Y_{…}}}_{\text{Efeito devido ao operador}}+\underbrace{\overline{Y_{ij.}}-\overline{Y_{i..}}}_{\text{Efeito específico da peça quando o operador está no i-ésimo nível}}+\underbrace{\overline{Y_{ijk}}- \overline{Y_{ij.}}}_{\text{Repetitividade}}$$
Thus, the break in variability is defined by:
$$SQTotal = SQA + SQB(A) + SQE$$
in which
$$SQTotal = \sum_{i}\sum_{j}\sum_{k}(Y_{ijk}-\overline{Y_{…}})^{2}$$
$$SQA = br\sum_{i}(\overline{Y_{i..}}-\overline{Y_{…}})^{2}$$
$$SQB(A)=r\sum_{i}\sum_{j}(\overline{Y_{ij.}}-\overline{Y_{i..}})^{2}$$
$$SQE = \sum_{i}\sum_{j}\sum_{k}(\overline{Y_{ijk}}-\overline{Y_{ij.}})^{2}=\sum_{i}\sum_{j}\sum_{k} \epsilon^{2}_{ijk}$$
In this case, the ANOVA table looks like this:
| Factor | GL | Sum of Squares | Medium Squares | F |
|---|---|---|---|---|
| Operator | $ a-1 $ | $ SQA $ | $ QMO=\frac{SQA}{a-1} $ | |
| Hierarchical Batch to Operator | $ a(b-1) $ | $ SQB(A) $ | $ QML(O)=\frac{SQB(A)}{a(b-1)} $ | |
| Repetitiveness | $ ab(r-1) $ | $ SQE $ | $ \frac{SQE}{ab(r-1)} $ | |
| Total | $ abr-1 $ | $ SQT $ |
In the hierarchical model with random factors (operator and part), we have the following values for the expected mean squares and the statistics associated with testing the influence of the factors.
| E(QM) | A: Random |
|---|---|
| E(QM) | B: Random |
| E(QMO) | $ \sigma^{2}+r\sigma_{\beta}^{2}+br\sigma_{\alpha}^{2} $ |
| E(QML(O)) | $ \sigma^{2}+r\sigma^{2}_{\beta} $ |
| E(QME) | $ \sigma^{2} $ |
| Appropriate Statistical Tests | |
| Test for | A: Random |
| Test for | B: Random |
| Factor A | $ \frac{QMA}{QMB(A)} $ |
| Factor B(A) | $ \frac{QMB(A)}{QME} $ |
Here, we perform the following tests
- Testing the effect of hierarchical batching on the operator:
$$\sigma^{2}_{\beta(\alpha)}> 0$$
the appropriate test statistic is
$$F^{*}=\frac{QML(O)}{QME}$$
and the decision rule at the significance level $ \alpha $ is

- Operator effect test:
$$\sigma^{2}_{\alpha}> 0$$
the appropriate test statistic is
$$F^{*} = \frac{QMO}{QME}$$
and the decision rule at the significance level $ \alpha $ is

Below, we present the calculations of the variance components:
- Repetitiveness
$$VE=\sqrt{QME}$$
- Variation due to operator
$$VO=\sqrt{\frac{QMO-QML(O)}{br}}$$
- Variation of batches assigned to the operator
$$VP =\sqrt{\frac{QML(O)-QME}{r}}$$
- Repeatability and Reproducibility
$$RR=\sqrt{VE^2 + VO^2}$$
- Total variation
$$VT=\sqrt{VP^2 + RR^2}$$
- Index table
| Source of Variation | Standard Deviation | % Variation | % Tolerance |
|---|---|---|---|
| Repetitiveness | VE | $ \frac{VE}{VT}*100 $ | $ (6*\frac{VE}{TOL})*100 $ |
| Reproducibility | GRANDMOTHER | $ \frac{VO}{VT}*100 $ | $ (6*\frac{VO}{TOL})*100 $ |
| Ask | VP | $ \frac{VP}{VT}*100 $ | $ (6*\frac{VP}{TOL})*100 $ |
| RR | RR | $ \frac{RR}{VT}*100 $ | $ (6*\frac{RR}{TOL})*100 $ |
Example 3.2.1
Consider the tensile strength characteristic performed with steel specimens.
Objective:
Evaluate the reproducibility and repeatability of the measurement system.
Description of the experiment
- Select 5 steel runs (p = 5), with little variability within runs and natural variability of the (production) process between runs;
- One bar was processed from each run;
- Each bar was divided into six parts;
- The parts of the test body were submitted to the measurement system, where Resistance was measured (2 evaluators (o = 2); 3 parts per evaluator (r = 3)).
- Unit: MPA;
- Tolerance: 130 MPA
The results are below:
- Assessor 1: Approval. 1.
- Assessor 2: Approval. 2.
| Race | Appraiser | Resistance |
|---|---|---|
| 1 | 1 | 1168 |
| 1 | 1 | 1170 |
| 1 | 1 | 1171 |
| 2 | 1 | 1179 |
| 2 | 1 | 1155 |
| 2 | 1 | 1159 |
| 3 | 1 | 1161 |
| 3 | 1 | 1179 |
| 3 | 1 | 1170 |
| 4 | 1 | 1190 |
| 4 | 1 | 1182 |
| 4 | 1 | 1197 |
| 5 | 1 | 1135 |
| 5 | 1 | 1150 |
| 5 | 1 | 1130 |
| 1 | 2 | 1142 |
| 1 | 2 | 1164 |
| 1 | 2 | 1177 |
| 2 | 2 | 1173 |
| 2 | 2 | 1175 |
| 2 | 2 | 1148 |
| 3 | 2 | 1184 |
| 3 | 2 | 1159 |
| 3 | 2 | 1182 |
| 4 | 2 | 1190 |
| 4 | 2 | 1188 |
| 4 | 2 | 1188 |
| 5 | 2 | 1139 |
| 5 | 2 | 1137 |
| 5 | 2 | 1151 |
Application
The sum of squares is given by the following formulas
$$SQTotal = (1168-1166,433)^{2}+(1170-1166,433)^2+\dots+(1151-1166,433)^{2}=10017,37$$
$$SQO = 5.3[(1166,4-1166,433)^{2}+(1166,467-1166,433)^{2}]=0,033$$
$$SQL(O) = 3[(1169,667-1166,4)^2 +(1164,333-1166,4)^{2}+\dots+(1142,333-1166,467)^{2}]=7608,667$$
$$SQE = (1168-1169,667)^{2}+(1170-1169,667)^{2}+\dots+(1151-1142,333)^{2}=2408,667$$
This way, we have the ANOVA table
Considering that the factors are random, we will analyze whether there is a significant difference between the evaluators, for this we use test statistics
$$F^{*}=\frac{QMO}{QML(O)}=\frac{0,033}{951,083}=0,00003469$$
At the level $ \alpha $ = 0.05 significance we have
$$F[0,95;1;8]=5.317655$$
How $ F^{*}< F $, we do not reject the hypothesis $ H_{0} $, therefore we conclude that there is no difference between the evaluators.
We now evaluate the variability between runs, using test statistics
$$F^{*}=\frac{QML(O)}{QME}=\frac{951,083}{120,433}=7,897$$
At the level $ \alpha $ = 0.05 significance we have
$$F[0,95;8;20]=2,447064$$
How $ F^{*}> F $, we reject the hypothesis $ H_{0} $, therefore we conclude that there is variability between the analyzed runs.
- Repetitiveness
$$VE=\sqrt{QME}=10,97$$
- Variation due to operator
$$VO=\sqrt{\frac{QMO-QML(O)}{br}}=0$$
- Variation of batches assigned to the operator
$$VP=\sqrt{\frac{QML(O)-QME}{r}}=16,64$$
- Repeatability and Reproducibility
$$RR=\sqrt{VE^2 + VO^2}=10,97$$
- Total variation
$$VT=\sqrt{VP^2 + RR^2}=19,93$$
- Standard deviation between runs: 16.64 MPA;
- Repeatability: variability due to the Measurement System plus bar variability (lack of bar homogeneity);
- Measurement system repeatability: 10.97 MPA;
- Evaluators do not significantly influence the variability of the measurement system;
- The variability between runs is significant in relation to repeatability.
Conclusions
The measuring system has high repeatability and negligible reproducibility. The repeatability of the measurement system presents a rate of 55% in relation to the total variation. Furthermore, through the average graph (Xbar), we observe that the dispersion of the points and the control lines (red lines) are very close (the majority of points are within the control lines). With this, we conclude that the repeatability of the measurement system and the variation of the production process (variability between bars) are similar, as a consequence we were unable to achieve continuous improvement through the data generated by this measurement system. Therefore, the measurement system is not suitable to control the process.
Results obtained by Action:


11.4 Attribute Measurement System
Definitions
An attribute measurement system (example: buffer calibrator) classifies the part as defective or not. One of the main objectives of analyzing attribute measurement systems is understanding and preventing classification errors. To this end, we propose the following steps to analyze an attribute measurement system:
- Critically analyze the measuring device calibration report;
- Prepare an internal standard for use, handling and interpretation of the measuring device. Train all users;
- Collect data for analysis of the measurement system;
- Calculate the degree of agreement in relation to the(s);
- Reference system: evaluate the design and applicability of the device;
- Operators: evaluate the consistency of the device and training in use and handling;
- Error rate: probability of the operator approving a part that does not meet specifications;
- False alarm rate: probability of the operator failing a part that meets specifications;
- Signal Detection Method: calculate the regions with the highest probability of failure of the measurement system by attribute. Estimate the variability of the measurement system.
Data collection
Data collection and analysis of the attribute measurement system is conducted according to the steps described below:
Step 1: Select a sufficient number of parts from which we have defective and non-defective parts. How many parts should we use to carry out the attribute measurement system study? To the dismay of users, the answer is “enough”. The purpose of studying a measurement system (attribute or variable) is to know its properties. A sufficient number of samples must be selected to cover an expected range of operation. In the case of an attribute system, the region of interest is the gray areas. If the process capability is good, then a small sample may not contain many parts in the gray area. This means that a process with good capacity requires a larger sample.
In the example mentioned in the fourth edition, for a Pp=Ppk=0.5 (in which we expect 13% of non-conforming parts), 50 parts were selected to carry out the study of the measurement system by attribute.
An alternative to avoiding large samples is to choose the pieces directly in the gray area to ensure that the effect of the evaluator’s variability will be visualized.
Consider an attribute measurement system that compares each part with specification limits, the system accepts the part if it is within specifications and rejects it otherwise (known as a pass-no-pass system). Like any measurement system, there is a gray area around specification limits in which the measurement system makes misclassifications.
Since we do not know, a priori, the gray area, we must carry out studies of the measurement system. However, to determine the risk areas around the specification limits, we need to choose approximately 25% of the parts “close” to the lower limit and 25% of the parts “close” to the upper limit. In cases where it is difficult to make such pieces, the team may decide to use a lower percentage, despite recognizing that this attitude may increase the variability of results. If it is not possible to make parts close to specification limits, the team should reconsider the use of an attribute measurement system for this process. For each characteristic, the parts must be measured by a variable measurement system with acceptable variability. When a characteristic cannot be measured by a variable measurement system (for example, visual), we use other means, such as expert classification.
Step 2: Select operators, at least 2. The suggestion in the fourth edition is that we should choose three operators.
Step 3: Each operator must measure each part at least three times.
Step 4: For each pair of operators, present the data as follows.
| OP-B | OP-B | ||
|---|---|---|---|
| OP-A | 1 - C | 2 - NC | Total |
| 1 - C | $ o_{11} $ | $ o_{12} $ | $ o_{1} $ |
| 2 - NC | $ o_{21} $ | $ o_{22} $ | $ q_{1} $ |
| Total | $ o_{2} $ | $ q_2 $ | $ n $ |
Here, we use the following notation:
1: part classified as approved;
2: piece classified as rejected;
- $o_{11}$: number of classifications approved for evaluator A and approved for evaluator B;
- $o_{12}$: number of classifications approved for evaluator A and disapproved for evaluator B;
- $o_{21}$: number of failed classifications for evaluator A and approved for evaluator B;
- $o_{22}$: number of failed classifications for evaluator A and disapproved for evaluator B;
- $o_{1}$=$o_{11}$+$o_{12}$: total approvals from evaluator A;
- $q_{1}$=$o_{21}$+$o_{22}$: total number of rejections by evaluator A;
- $o_{2}$=$o_{11}$+$o_{21}$: total approvals from evaluator B;
- $q_{2}$=$o_{12}$+$o_{22}$: total number of rejections by evaluator B;
- n: total measurements per evaluator =$o_{1}$+$q_{1}$=$o_{2}$+$q_{2}$
Step 5: Calculate the observed proportions
| OP-B | OP-B | ||
|---|---|---|---|
| OP-A | 1 - C | 2 - NC | Total |
| 1 - C | $ P_{11}=\cfrac{o_{11}}{n} $ | $ P_{12}=\cfrac{o_{12}}{n} $ | $ PA_1=\cfrac{o_1}{n} $ |
| 2 - NC | $ P_{21}=\cfrac{o_{11}}{n} $ | $ P_{22}=\cfrac{o_{22}}{n} $ | $ PA_2=\cfrac{q_1}{n} $ |
| Total | $ PB_1=\cfrac{o_2}{n} $ | $ PB_2=\cfrac{q_2}{n} $ | 1 |
Step 6: Calculate expected proportions
| OP-B | OP-B | |
|---|---|---|
| OP-A | 1 - C | 2 - NC |
| 1 - C | $ E_{11}= PA_1* PB_1 $ | $ E_{12}= PA_1* PB_2 $ |
| 2 - NC | $ E_{21}= PA_2* PB_1 $ | $ E_{22}= PA_2* PB_2 $ |
Step 7: Calculate the Kappa coefficient
$${Kappa}=\cfrac{(P_{11}+P_{22})-(E_{11}+E_{22})}{1-(E_{11}+E_{22})}$$
Step 8: Criterion: the Kappa coefficient must be greater than 0.75. If the Kappa is less than 0.75, we say that there is not good agreement between the evaluators.
Example 4.1.1
Consider a measurement system, in which 50 parts were selected, 3 evaluators with three measurements per part (each evaluator). The data is presented in the Table below. Here we consider 0 Non-Conforming and 1 Conforming. For this characteristic we have a lower engineering limit of 0.45 (LIE) and an upper engineering limit of 0.55 (LSE).
| Ask | Operator 1 | Operator 2 | Operator 3 | Standard | Reference Value |
|---|---|---|---|---|---|
| 1 | 1 | 1 | 1 | 1 | 0,476901 |
| 1 | 1 | 1 | 1 | 1 | 0,476901 |
| 1 | 1 | 1 | 1 | 1 | 0,476901 |
| 2 | 1 | 1 | 1 | 1 | 0,509015 |
| 2 | 1 | 1 | 1 | 1 | 0,509015 |
| 2 | 1 | 1 | 1 | 1 | 0,509015 |
| 3 | 0 | 0 | 0 | 0 | 0,576459 |
| 3 | 0 | 0 | 0 | 0 | 0,576459 |
| 3 | 0 | 0 | 0 | 0 | 0,576459 |
| 4 | 0 | 0 | 0 | 0 | 0,566152 |
| 4 | 0 | 0 | 0 | 0 | 0,566152 |
| 4 | 0 | 0 | 0 | 0 | 0,566152 |
| 5 | 0 | 0 | 0 | 0 | 0,57036 |
| 5 | 0 | 0 | 0 | 0 | 0,57036 |
| 5 | 0 | 0 | 0 | 0 | 0,57036 |
| 6 | 1 | 1 | 1 | 1 | 0,544951 |
| 6 | 1 | 1 | 0 | 1 | 0,544951 |
| 6 | 0 | 0 | 0 | 1 | 0,544951 |
| 7 | 1 | 1 | 1 | 1 | 0,465454 |
| 7 | 1 | 1 | 0 | 1 | 0,465454 |
| 7 | 1 | 1 | 1 | 1 | 0,465454 |
| 8 | 1 | 1 | 1 | 1 | 0,502295 |
| 8 | 1 | 1 | 1 | 1 | 0,502295 |
| 8 | 1 | 1 | 1 | 1 | 0,502295 |
| 9 | 0 | 0 | 0 | 0 | 0,437817 |
| 9 | 0 | 0 | 0 | 0 | 0,437817 |
| 9 | 0 | 0 | 0 | 0 | 0,437817 |
| 10 | 1 | 1 | 1 | 1 | 0,515573 |
| 10 | 1 | 1 | 1 | 1 | 0,515573 |
| 10 | 1 | 1 | 1 | 1 | 0,515573 |
| 11 | 1 | 1 | 1 | 1 | 0,488905 |
| 11 | 1 | 1 | 1 | 1 | 0,488905 |
| 11 | 1 | 1 | 1 | 1 | 0,488905 |
| 12 | 0 | 0 | 0 | 0 | 0,559918 |
| 12 | 0 | 0 | 1 | 0 | 0,559918 |
| 12 | 0 | 0 | 0 | 0 | 0,559918 |
| 13 | 1 | 1 | 1 | 1 | 0,542704 |
| 13 | 1 | 1 | 1 | 1 | 0,542704 |
| 13 | 1 | 1 | 1 | 1 | 0,542704 |
| 14 | 1 | 1 | 1 | 1 | 0,454518 |
| 14 | 1 | 1 | 0 | 1 | 0,454518 |
| 14 | 0 | 1 | 0 | 1 | 0,454518 |
| 15 | 1 | 1 | 1 | 1 | 0,517377 |
| 15 | 1 | 1 | 1 | 1 | 0,517377 |
| 15 | 1 | 1 | 1 | 1 | 0,517377 |
| 16 | 1 | 1 | 1 | 1 | 0,531939 |
| 16 | 1 | 1 | 1 | 1 | 0,531939 |
| 16 | 1 | 1 | 1 | 1 | 0,531939 |
| 17 | 1 | 1 | 1 | 1 | 0,519694 |
| 17 | 1 | 1 | 1 | 1 | 0,519694 |
| 17 | 1 | 1 | 1 | 1 | 0,519694 |
| 18 | 1 | 1 | 1 | 1 | 0,484167 |
| 18 | 1 | 1 | 1 | 1 | 0,484167 |
| 18 | 1 | 1 | 1 | 1 | 0,484167 |
| 19 | 1 | 1 | 1 | 1 | 0,520496 |
| 19 | 1 | 1 | 1 | 1 | 0,520496 |
| 19 | 1 | 1 | 1 | 1 | 0,520496 |
| 20 | 1 | 1 | 1 | 1 | 0,477236 |
| 20 | 1 | 1 | 1 | 1 | 0,477236 |
| 20 | 1 | 1 | 1 | 1 | 0,477236 |
| 21 | 1 | 1 | 0 | 1 | 0,45231 |
| 21 | 1 | 0 | 1 | 1 | 0,45231 |
| 21 | 0 | 1 | 0 | 1 | 0,45231 |
| 22 | 0 | 0 | 1 | 0 | 0,545604 |
| 22 | 0 | 1 | 1 | 0 | 0,545604 |
| 22 | 1 | 0 | 0 | 0 | 0,545604 |
| 23 | 1 | 1 | 1 | 1 | 0,529065 |
| 23 | 1 | 1 | 1 | 1 | 0,529065 |
| 23 | 1 | 1 | 1 | 1 | 0,529065 |
| 24 | 1 | 1 | 1 | 1 | 0,514192 |
| 24 | 1 | 1 | 1 | 1 | 0,514192 |
| 24 | 1 | 1 | 1 | 1 | 0,514192 |
| 25 | 0 | 0 | 0 | 0 | 0,599581 |
| 25 | 0 | 0 | 0 | 0 | 0,599581 |
| 25 | 0 | 0 | 0 | 0 | 0,599581 |
| 26 | 0 | 0 | 0 | 0 | 0,547204 |
| 26 | 1 | 0 | 0 | 0 | 0,547204 |
| 26 | 0 | 0 | 1 | 0 | 0,547204 |
| 27 | 1 | 1 | 1 | 1 | 0,502436 |
| 27 | 1 | 1 | 1 | 1 | 0,502436 |
| 27 | 1 | 1 | 1 | 1 | 0,502436 |
| 28 | 1 | 1 | 1 | 1 | 0,521642 |
| 28 | 1 | 1 | 1 | 1 | 0,521642 |
| 28 | 1 | 1 | 1 | 1 | 0,521642 |
| 29 | 1 | 1 | 1 | 1 | 0,523754 |
| 29 | 1 | 1 | 1 | 1 | 0,523754 |
| 29 | 1 | 1 | 1 | 1 | 0,523754 |
| 30 | 0 | 0 | 0 | 0 | 0,561457 |
| 30 | 0 | 0 | 0 | 0 | 0,561457 |
| 30 | 0 | 1 | 0 | 0 | 0,561457 |
| 31 | 1 | 1 | 1 | 1 | 0,503091 |
| 31 | 1 | 1 | 1 | 1 | 0,503091 |
| 31 | 1 | 1 | 1 | 1 | 0,503091 |
| 32 | 1 | 1 | 1 | 1 | 0,50585 |
| 32 | 1 | 1 | 1 | 1 | 0,50585 |
| 32 | 1 | 1 | 1 | 1 | 0,50585 |
| 33 | 1 | 1 | 1 | 1 | 0,487613 |
| 33 | 1 | 1 | 1 | 1 | 0,487613 |
| 33 | 1 | 1 | 1 | 1 | 0,487613 |
| 34 | 0 | 0 | 0 | 0 | 0,449696 |
| 34 | 0 | 0 | 1 | 0 | 0,449696 |
| 34 | 1 | 1 | 1 | 0 | 0,449696 |
| 35 | 1 | 1 | 1 | 1 | 0,498698 |
| 35 | 1 | 1 | 1 | 1 | 0,498698 |
| 35 | 1 | 1 | 1 | 1 | 0,498698 |
| 36 | 1 | 1 | 1 | 1 | 0,543077 |
| 36 | 1 | 1 | 0 | 1 | 0,543077 |
| 36 | 0 | 1 | 1 | 1 | 0,543077 |
| 37 | 0 | 0 | 0 | 0 | 0,409238 |
| 37 | 0 | 0 | 0 | 0 | 0,409238 |
| 37 | 0 | 0 | 0 | 0 | 0,409238 |
| 38 | 1 | 1 | 1 | 1 | 0,488184 |
| 38 | 1 | 1 | 1 | 1 | 0,488184 |
| 38 | 1 | 1 | 1 | 1 | 0,488184 |
| 39 | 0 | 0 | 0 | 0 | 0,427687 |
| 39 | 0 | 0 | 0 | 0 | 0,427687 |
| 39 | 0 | 0 | 0 | 0 | 0,427687 |
| 40 | 1 | 1 | 1 | 1 | 0,501132 |
| 40 | 1 | 1 | 1 | 1 | 0,501132 |
| 40 | 1 | 1 | 1 | 1 | 0,501132 |
| 41 | 1 | 1 | 1 | 1 | 0,513779 |
| 41 | 1 | 1 | 1 | 1 | 0,513779 |
| 41 | 1 | 1 | 1 | 1 | 0,513779 |
| 42 | 0 | 0 | 0 | 0 | 0,566575 |
| 42 | 0 | 0 | 0 | 0 | 0,566575 |
| 42 | 0 | 0 | 0 | 0 | 0,566575 |
| 43 | 1 | 1 | 1 | 1 | 0,46241 |
| 43 | 0 | 1 | 1 | 1 | 0,46241 |
| 43 | 1 | 1 | 0 | 1 | 0,46241 |
| 44 | 1 | 1 | 1 | 1 | 0,470832 |
| 44 | 1 | 1 | 1 | 1 | 0,470832 |
| 44 | 1 | 1 | 1 | 1 | 0,470832 |
| 45 | 0 | 0 | 0 | 0 | 0,412453 |
| 45 | 0 | 0 | 0 | 0 | 0,412453 |
| 45 | 0 | 0 | 0 | 0 | 0,412453 |
| 46 | 1 | 1 | 1 | 1 | 0,493441 |
| 46 | 1 | 1 | 1 | 1 | 0,493441 |
| 46 | 1 | 1 | 1 | 1 | 0,493441 |
| 47 | 1 | 1 | 1 | 1 | 0,486379 |
| 47 | 1 | 1 | 1 | 1 | 0,486379 |
| 47 | 1 | 1 | 1 | 1 | 0,486379 |
| 48 | 0 | 0 | 0 | 0 | 0,587893 |
| 48 | 0 | 0 | 0 | 0 | 0,587893 |
| 48 | 0 | 0 | 0 | 0 | 0,587893 |
| 49 | 1 | 1 | 1 | 1 | 0,483803 |
| 49 | 1 | 1 | 1 | 1 | 0,483803 |
| 49 | 1 | 1 | 1 | 1 | 0,483803 |
| 50 | 0 | 0 | 0 | 0 | 0,446697 |
| 50 | 0 | 0 | 0 | 0 | 0,446697 |
| 50 | 0 | 0 | 0 | 0 | 0,446697 |
Table 4.1.1: MSA Attributes
Step 4: Frequency tables
| Op. B | Op. B | ||
|---|---|---|---|
| Op. A | 1-C | 2-NC | Total |
| 1-C | 97 | 3 | 100 |
| 2-NC | 6 | 44 | 50 |
| Total | 103 | 47 | 150 |
Table 25: Frequency table - A x B
| Op. C | Op. C | ||
|---|---|---|---|
| Op. A | 1-C | 2-NC | Total |
| 1-C | 92 | 8 | 100 |
| 2-NC | 7 | 43 | 50 |
| Total | 99 | 51 | 150 |
Table 26: Frequency table - A x C
| Op. C | Op. C | ||
|---|---|---|---|
| Op. B | 1-C | 2-NC | Total |
| 1-C | 94 | 9 | 103 |
| 2-NC | 5 | 42 | 47 |
| Total | 99 | 51 | 150 |
Table 27: Frequency table - B x C
Next, we calculated the Kappa coefficient for all combinations of evaluators. Initially, let’s calculate the observed proportions (Step 5):
| Op. B | Op. B | ||
|---|---|---|---|
| Op. A | 1-C | 2-NC | Total |
| 1-C | 0,647 | 0,02 | 0,667 |
| 2-NC | 0,04 | 0,293 | 0,333 |
| Total | 0,687 | 0,313 | 1 |
Table 28: Table of observed proportions - A x B
| Op. C | Op. C | ||
|---|---|---|---|
| Op. A | 1-C | 2-NC | Total |
| 1-C | 0,613 | 0,053 | 0,667 |
| 2-NC | 0,0467 | 0,287 | 0,333 |
| Total | 0,66 | 0,34 | 1 |
Table 29: Table of observed proportions - A x C
| Op. C | Op. C | ||
|---|---|---|---|
| Op. B | 1-C | 2-NC | Total |
| 1-C | 0,627 | 0,06 | 0,687 |
| 2-NC | 0,033 | 0,28 | 0,313 |
| Total | 0,66 | 0,34 | 1 |
Table 30: Table of observed proportions - B x C
Next, we calculate the expected proportions (Step 6):
| Op. B | Op. B | |
|---|---|---|
| Op. A | 1-C | 2-NC |
| 1-C | 0,458 | 0,209 |
| 2-NC | 0,229 | 0,104 |
Table 31: Table of expected proportions - A x B
| Op. C | Op. C | |
|---|---|---|
| Op. A | 1-C | 2-NC |
| 1-C | 0,44 | 0,227 |
| 2-NC | 0,22 | 0,113 |
Table 32: Table of expected proportions - A x C
| Op. C | Op. C | |
|---|---|---|
| Op. B | 1-C | 2-NC |
| 1-C | 0,453 | 0,233 |
| 2-NC | 0,207 | 0,107 |
Table 33: Table of expected proportions - B x C
Finally, we calculate the Kappa between raters (Step 7):
$${Kappa_{AB}}=\cfrac{(P_{11}+P_{22})-(E_{11}+E_{22})}{1-(E_{11}+E_{22})}=\cfrac{(0,647+0,293)-(0,458+0,104)}{1-(0,458+0,104)}=0,863$$
Conclusion
As Kappa is greater than 0.75 (0.863), we conclude that there is agreement between evaluators A and B.
$${Kappa_{AC}}=\cfrac{(P_{11}+P_{22})-(E_{11}+E_{22})}{1-(E_{11}+E_{22})}=\cfrac{(0,613+0,287)-(0,44+0,113)}{1-(0,44+0,113)}=0,776$$
Conclusion
As Kappa is greater than 0.75 (0.776), we conclude that there is agreement between evaluators A and C.
$${Kappa_{BC}}=\cfrac{(P_{11}+P_{22})-(E_{11}+E_{22})}{1-(E_{11}+E_{22})}=\cfrac{(0,627+0,28)-(0,453+0,107)}{1-(0,453+0,107)}=0,788$$
Conclusion
As Kappa is greater than 0.75 (0.788), we conclude that there is agreement between evaluators B and C.
Calculating the Kappa coefficient with respect to the standard
Frequency tables (Step 4):
| REF | REF | ||
|---|---|---|---|
| Op. A | 1-C | 2-NC | Total |
| 1-C | 97 | 3 | 100 |
| 2-NC | 5 | 45 | 50 |
| Total | 102 | 48 | 150 |
Table 34: Frequency table - A x Standard
| REF | REF | ||
|---|---|---|---|
| Op. B | 1-C | 2-NC | Total |
| 1-C | 100 | 3 | 103 |
| 2-NC | 2 | 45 | 47 |
| Total | 102 | 48 | 150 |
Table 35: Frequency table - B x Standard
| REF | REF | ||
|---|---|---|---|
| Op. C | 1-C | 2-NC | Total |
| 1-C | 93 | 6 | 99 |
| 2-NC | 9 | 42 | 51 |
| Total | 102 | 48 | 150 |
Table 36: Frequency table - C x Standard
Let’s now calculate the observed proportions (Step 5):
| REF | REF | ||
|---|---|---|---|
| Op. A | 1-C | 2-NC | Total |
| 1-C | 0,647 | 0,02 | 0,667 |
| 2-NC | 0,033 | 0,3 | 0,333 |
| Total | 0,68 | 0,32 | 1 |
Table 37: Table of observed proportions - A x Standard
| REF | REF | ||
|---|---|---|---|
| Op. B | 1-C | 2-NC | Total |
| 1-C | 0,667 | 0,02 | 0,687 |
| 2-NC | 0,013 | 0,3 | 0,313 |
| Total | 0,68 | 0,32 | 1 |
Table 38: Table of observed proportions - B x Standard
| REF | REF | ||
|---|---|---|---|
| Op. C | 1-C | 2-NC | Total |
| 1-C | 0,62 | 0,04 | 0,66 |
| 2-NC | 0,06 | 0,28 | 0,34 |
| Total | 0,68 | 0,32 | 1 |
Table 39: Table of observed proportions - C x Standard
Now the expected proportions (Step 6):
| REF | REF | |
|---|---|---|
| Op. A | 1-C | 2-NC |
| 1-C | 0,453 | 0,213 |
| 2-NC | 0,227 | 0,107 |
Table 40: Table of expected proportions - A x Standard
| REF | REF | |
|---|---|---|
| Op. B | 1-C | 2-NC |
| 1-C | 0,467 | 0,220 |
| 2-NC | 0,213 | 0,1 |
Table 41: Table of expected proportions - B x Standard
| REF | REF | |
|---|---|---|
| Op. C | 1-C | 2-NC |
| 1-C | 0,449 | 0,211 |
| 2-NC | 0,231 | 0,109 |
Table 42: Table of expected proportions - C x Standard
Finally, we calculate the Kappa between the raters and the standard (Step 7):
Rater A and the standard:
$${Kappa}=\cfrac{(P_{11}+P_{22})-(E_{11}+E_{22})}{1-(E_{11}+E_{22})}=\cfrac{(0,647+0,3)-(0,453+0,107)}{1-(0,453+0,107)}=0,879$$
Rater B and the standard:
$${Kappa_{AB}}=\cfrac{(P_{11}+P_{22})-(E_{11}+E_{22})}{1-(E_{11}+E_{22})}=\cfrac{(0,667+0,3)-(0,467+0,1)}{1-(0,467+0,1)}=0,923$$
C Evaluator and the standard:
$${Kappa_{AB}}=\cfrac{(P_{11}+P_{22})-(E_{11}+E_{22})}{1-(E_{11}+E_{22})}=\cfrac{(0,62+0,28)-(0,449+0,109)}{1-(0,449+0,109)}=0,774$$
Some definitions
In this Section, we will continue the analysis of the measurement system. We intend to evaluate classification errors in relation to compliant and non-compliant parts. We have
- Error rate: is the probability of the evaluator classifying a piece as Approved, when in fact this piece is Failed;
$$P(Ap_{Aval}| Rep_{Pad})=\cfrac{P(Ap_{Aval}\cap Resp_{Pad})}{P(Rep_{Pad})}$$
- False alarm rate: is the probability of the evaluator classifying a part as Failing, when in fact this part is Approved.
$$P(Rep_{Aval}| Ap_{Pad})=\cfrac{P(Rep_{Aval}\cap Ap_{Pad})}{P(Ap_{Pad})}$$
- Agreement rate: is the probability of the evaluator classifying a piece according to the reference system (and/or himself).
where:
$Ap_{Aval}$ = Measurements that the evaluator considers approved;
$Rep_{Aval}$ = Measurements that the evaluator considers failed;
$Ap_{Pad}$ = Measurements that the reference system considers approved;
$Rep_{Pad}$ = Measurements that the reference system considers failed.
So:
$$P(Rep_{Pad})=\cfrac{{Total~de~medições~que~o~sistema~referência~considera~reprovada}}{{Nordm;~total~de~medições}}$$
$$P(Ap_{Pad})=\cfrac{{Total~de~medições~que~o~sistema~referência~considera~aprovada}}{{Nordm;~total~de~medições}}$$
$$P(Ap_{Aval}\cap~Rep_{Pad})=\cfrac{{Total~de~medições~aprovado~pelo~avaliador~e~reprovada~pelo~sistema}}{{Nordm;~total~de~medições}}$$
$$P(Rep_{Aval}\cap~Ap_{Pad})=\cfrac{{Total~de~medições~reprovada~pelo~avaliador~e~aprovada~pelo~sistema}}{{Nordm;~total~de~medições}}$$
Like this:
- Error Rate:
$$P(Ap_{Aval}| Rep_{Pad})=\cfrac{P(Ap_{Aval}\cap Rep_{Pad})}{P(Rep_{Pad})}=$$
$$=\cfrac{{Total~de~medições~aprovadas~pelo~avaliador~e~reprovada~pelo~sistema~de~referência}}{{Total~de~medições~que~o~sistema~de~referência~considera~reprovada}}$$
- False alarm rate:
$$P(Rep_{Aval}| Ap_{Pad})=\cfrac{P(Rep_{Aval}\cap Ap_{Pad})}{P(Ap_{Pad})}=$$
$$=\cfrac{{Total~de~medições~reprovadas~pelo~avaliador~e~aprovada~pelo~sistema~de~referência}}{{Total~de~medições~que~o~sistema~de~referência~considera~aprovada}}$$
- Agreement rate:
$$P({Concordância})=\cfrac{{Nordm;~de~medições~em~que~não~há~discordância}}{{Total~de~medições}}$$
Steps for calculation
To calculate the probability of failure and the false alarm rate, we propose the following steps:
Step 8: Consider the observed frequency table, as in step 4. However, we take the observed frequency table of each operator in relation to the reference value, in the form
| Reference | Reference | ||
|---|---|---|---|
| Appraiser | 1-C | 2-NC | Total |
| 1-C | $o_{11}$ | $o_{12}$ | $o_{1}$ |
| 2-NC | $o_{21}$ | $o_{22}$ | $q_{1}$ |
| Total | $o_{2}$ | $q_{2}$ | n |
Here, we use the following notation:
1: part classified as approved;
2: piece classified as failed.
$o_{11}$ : number of classifications approved for the evaluator and approved for the reference;
$o_{12}$ : number of classifications approved for the evaluator and disapproved for the reference;
$o_{21}$ : number of disapproved classifications for the evaluator and approved for the reference;
$o_{22}$ : number of failed ratings for the evaluator and failed ratings for the reference;
$o_{1}$ = $o_{11}$ + $o_{12}$: total approvals from the evaluator;
$q_{1}$ = $o_{21}$ + $o_{22}$: total number of evaluator rejections;
$o_{2}$ = $o_{11}$ + $o_{21}$: total number of approvals for the reference;
$q_{2}$ = $o_{12}$ + $o_{22}$: total number of rejections of the reference;
n : total measurements = $o_{1}$ + $q_{1}$ = $o_{2}$ + $q_{2}$
Step 9: Calculate the indices:
9.1 - Failure rate
$${Taxa~de~Falha}=\cfrac{o_{12}}{q_2}$$
9.2 - Probability of false alarm
$${Falso~Alarme}=\cfrac{o_{21}}{o_2}$$
9.3 - Agreement
$${Concordância}=\cfrac{{Número~de~peças~com~medições~concordantes~com~a~referência}}{{Número~total~de~peças}}$$
Criteria for analysis
Next, we present the criteria for analyzing the failure rate, probability of false alarm and agreement (Table 43).
| Decision Measurement System | Concordance | Error Rate | False Alarm Rate |
|---|---|---|---|
| Acceptable to the evaluator | ≥ 90% | ≤ 2% | ≤ 5% |
| Marginal acceptable to the evaluator, may require improvement | ≥ 80% | ≤ 5% | ≤ 10% |
| Unacceptable to the evaluator, requires improvement | < 80% | > 5% | > 10% |
Table 43: Decision criteria
Example 4.1.2
Returning to Example 4.1.1, we will calculate the Error and False Alarm Rates for each Evaluator.
Let’s build tables with the number of measurements for each operator:
Assessor A
| Reference | Reference | ||
|---|---|---|---|
| Assessor A | Approve | Fail | Total |
| Approve | 97 | 3 | 100 |
| Fail | 5 | 45 | 50 |
| Total | 102 | 48 | 150 |
Table 44: Number of measurements by reference and by Evaluator A
Therefore, for evaluator A we have:
Step 9: Calculate the indices:
9.1 - Failure rate
$${Taxa~de~Falha}=\cfrac{o_{12}}{q_2}=\cfrac{3}{48}=0,0625~(6,25(porcentagem))$$
9.2 - Probability of false alarm
$${Falso~Alarme}=\cfrac{o_{21}}{o_2}=\cfrac{5}{102}=0,049~(4,9(porcentagem))$$
9.3 - Agreement
$${Concordância}=\cfrac{{Número~de~peças~com~medições~concordantes~com~a~referência}}{{Número~total~de~peças}}=\cfrac{42}{50}=0,84~(84(porcentagem))$$
Assessor B
| Reference | Reference | ||
|---|---|---|---|
| Assessor B | Approve | Fail | Total |
| Approve | 100 | 3 | 103 |
| Fail | 2 | 45 | 47 |
| Total | 102 | 48 | 150 |
Table 45: Number of measurements by reference and by Evaluator B
Therefore, for evaluator B we have:
Step 9: Calculate the indices:
9.1 - Failure rate
$${Taxa~de~Falha}=\cfrac{o_{12}}{q_2}=\cfrac{3}{48}=0,0625~(6,25(porcentagem))$$
9.2 - Probability of false alarm
$${Falso~Alarme}=\cfrac{o_{21}}{o_2}=\cfrac{2}{102}=0,0196~(1,96(porcentagem))$$
9.3 - Agreement
$${Concordância}=\cfrac{{Número~de~peças~com~medições~concordantes~com~a~referência}}{{Número~total~de~peças}}=\cfrac{45}{50}=0,9~(90(porcentagem))$$
Assessor C
| Reference | Reference | ||
|---|---|---|---|
| Assessor C | Approve | Fail | Total |
| Approve | 93 | 6 | 99 |
| Fail | 9 | 42 | 51 |
| Total | 102 | 48 | 150 |
Table 46: Number of measurements by reference and by Evaluator C
Therefore, for Evaluator C we have:
Step 9: Calculate the indices:
9.1 - Failure rate
$${Taxa~de~Falha}=\cfrac{o_{12}}{q_2}=\cfrac{6}{48}=0,0125~(12,5(porcentagem))$$
9.2 - Probability of false alarm
$${Falso~Alarme}=\cfrac{o_{21}}{o_2}=\cfrac{9}{102}=0,0823~(8,23(porcentagem))$$
9.3 - Agreement
$${Concordância}=\cfrac{{Número~de~peças~com~medições~concordantes~com~a~referência}}{{Número~total~de~peças}}=\cfrac{40}{50}=0,8~(80(porcentagem))$$
Example 4.1.2
| Assessors | Concordance | Error Rate | False Alarm Rate |
|---|---|---|---|
| A | 84% | 6,3% | 4,9% |
| B | 90% | 6,3% | 2% |
| C | 80% | 12,5% | 8,8% |
Raters table
Agreement between Evaluators




Agreement between Standard and each Evaluator





Conclusion
As all Kappa values are above 0.75, we conclude that the operators agree with each other and with the standard. However, operator error rates ($ op_A=6,25(porcentagem) $, $ op_B=6,25(porcentagem) $, $ op_C=12,5(porcentagem) $) are above that specified by the MSA manual, which is 5%. In this way, we have an opportunity for improvement.
Signal detection
An alternative to estimating the Gray Band (see Figure 18), a region in which the device has a high probability of misclassification, is to use Signal Detection theory.

Figure 18: Gray band.
Steps:
| Reference Value | Code | |
|---|---|---|
| 0,599581 | - | |
| 0,587893 | - | |
| 0,576459 | - | |
| 0,57036 | - | |
| 0,566575 | - | |
| $ d_{LSE} $ | 0,566152 | - |
| $ d_{LSE} $ | 0,561457 | x |
| $ d_{LSE} $ | 0,559918 | x |
| $ d_{LSE} $ | 0,547204 | x |
| $ d_{LSE} $ | 0,545604 | x |
| $ d_{LSE} $ | 0,544951 | x |
| $ d_{LSE} $ | 0,543077 | x |
| $ d_{LSE} $ | 0,542704 | + |
| 0,531939 | + | |
| 0,529065 | + | |
| 0,523754 | + | |
| 0,521642 | + | |
| 0,520496 | + | |
| 0,519694 | + | |
| 0,517377 | + | |
| 0,515573 | + | |
| 0,514192 | + | |
| 0,513779 | + | |
| 0,509015 | + | |
| 0,50585 | + | |
| 0,503091 | + | |
| 0,502436 | + | |
| 0,502295 | + | |
| 0,501132 | + | |
| 0,498698 | + | |
| 0,488905 | + | |
| 0,493441 | + | |
| 0,488184 | + | |
| 0,487613 | + | |
| 0,486379 | + | |
| 0,484167 | + | |
| 0,483803 | + | |
| 0,477236 | + | |
| 0,476901 | + | |
| $ d_{LIE} $ | 0,470832 | + |
| $ d_{LIE} $ | 0,465454 | x |
| $ d_{LIE} $ | 0,46241 | x |
| $ d_{LIE} $ | 0,454518 | x |
| $ d_{LIE} $ | 0,45231 | x |
| $ d_{LIE} $ | 0,449696 | x |
| $ d_{LIE} $ | 0,446697 | - |
| 0,437817 | - | |
| 0,427687 | - | |
| 0,412453 | - | |
| 0,409238 | - |
Table 4.1.2: Ordered Reference Values.
Step 1: Determine the tolerance (specification limits). For example from table 4.1.1, we have:
$$LIE: 0,450$$
$$LSE: 0,550$$
So, we have that the tolerance is given by:
$${tolerância}=LSE-LIE=0,100$$
This is the value to be used to calculate the RR.
Guidelines
- Se $ ppk~> ~1 $, compare the measurement system with the process;
- Se $ ppk~< ~1 $, compare the measurement system with the tolerance.
This “rule” is equivalent to comparing the measurement system with whichever is more restrictive (process or tolerance).
In example 4.1.1, the $ ppk=0,5 $. Since the process variation is greater than the tolerance, let’s compare the measurement system variation with the tolerance.
Step 2: Sort the data in descending order (largest to smallest) based on the reference values of the parts (see table 4.1.2).
Step 3: Identify the end and start points of the two gray areas. In table 4.1.2 it is shown in the code column.
| Signal | Description |
|---|---|
| $ + $ | Accepted with Full Agreement |
| $ - $ | Reject with Full Agreement |
| $ x $ | No Agreement |
Through the amplitude of the intervals referring to the gray areas, we determine their average, which will be compared with the tolerance or 6 sigma of the process variation (6 * standard deviation of the process).
Step 4: Referring to the chart above.
Be $ d_{LSE}= $ distance between the last piece accepted by all evaluators to the first piece rejected by all. This distance is equivalent to the gray region II around the $ LSE $ above.

Step 5:
Be,
$ d_{LIE}= $ distance between the last piece accepted by all evaluators in zone III to the first piece rejected by all evaluators in zone I.
$ d_i= $ distance between the last part accepted by all evaluators to the first part rejected by all (for each specification).
So,
$$d= \overline{d_i}$$
is an estimate of the width of the region II area and $ RR=6*\sigma_{RR} $
| Zone I | parts in agreement by all evaluators to be rejected |
|---|---|
| Zone III | parts in agreement by all evaluators to be accepted |
| Zone II | questionable parts without having 100% agreement, around the limit of each specification |
Returning to example 4.1.1, using a tolerance of 0.100, we have:
$$d_{LIE}=0,470832-0,446697=0,024135$$
$$d_{LSE}=0,566152-0,542704=0,023448$$
$$d=0,0237915$$
or the estimate of $ (porcentagem)RR $
$$(porcentagem)RR= \frac{0,0237915}{0,1} 100(porcentagem) = 24(porcentagem)$$
Example 4.1.3
Consider a measuring system (Calibrator), in which 50 parts were selected, 3 evaluators with three measurements per part (each evaluator). The data are presented in Table 6.26. Perform a study by Attribute with the Lower and Upper Specification Limits of 142.10 mm and 143.10 mm, respectively. Where NC is non-conforming and C is conforming.
| Operator | Operator | Operator | Value of | ||
|---|---|---|---|---|---|
| Ask | A | B | C | Standard | Standard |
| 1 | C | NC | NC | C | 143,06 |
| 1 | C | NC | NC | C | 143,06 |
| 1 | C | NC | NC | C | 143,06 |
| 2 | C | C | C | C | 142,68 |
| 2 | C | C | C | C | 142,68 |
| 2 | C | C | C | C | 142,68 |
| 3 | NC | NC | NC | NC | 145,28 |
| 3 | NC | NC | NC | NC | 145,28 |
| 3 | NC | NC | NC | NC | 145,28 |
| 4 | C | C | C | C | 142,54 |
| 4 | C | C | C | C | 142,54 |
| 4 | C | C | C | C | 142,54 |
| 5 | NC | NC | NC | NC | 143,28 |
| 5 | NC | NC | NC | NC | 143,28 |
| 5 | NC | NC | NC | NC | 143,28 |
| 6 | C | C | C | C | 142,8 |
| 6 | C | C | C | C | 142,8 |
| 6 | C | C | C | C | 142,8 |
| 7 | C | C | C | C | 142,28 |
| 7 | C | C | C | C | 142,28 |
| 7 | C | C | C | C | 142,28 |
| 8 | C | C | C | C | 143 |
| 8 | C | C | C | C | 143 |
| 8 | C | C | NC | C | 143 |
| 9 | C | C | C | C | 142,9 |
| 9 | C | C | C | C | 142,9 |
| 9 | C | C | C | C | 142,9 |
| 10 | C | C | C | C | 142,1 |
| 10 | NC | C | C | C | 142,1 |
| 10 | C | NC | NC | C | 142,1 |
| 11 | C | C | C | C | 142,3 |
| 11 | C | C | C | C | 142,3 |
| 11 | C | C | C | C | 142,3 |
| 12 | NC | NC | NC | NC | 142,08 |
| 12 | NC | NC | NC | NC | 142,08 |
| 12 | C | NC | NC | NC | 142,08 |
| 13 | NC | NC | NC | C | 142,13 |
| 13 | C | C | NC | C | 142,13 |
| 13 | C | NC | NC | C | 142,13 |
| 14 | NC | NC | NC | NC | 143,63 |
| 14 | NC | NC | NC | NC | 143,63 |
| 14 | NC | NC | NC | NC | 143,63 |
| 15 | NC | NC | NC | NC | 141,8 |
| 15 | NC | NC | NC | NC | 141,8 |
| 15 | NC | NC | NC | NC | 141,8 |
| 16 | C | NC | NC | C | 143,1 |
| 16 | C | NC | C | C | 143,1 |
| 16 | C | NC | C | C | 143,1 |
| 17 | C | C | C | C | 142,5 |
| 17 | C | C | C | C | 142,5 |
| 17 | C | C | C | C | 142,5 |
| 18 | NC | NC | NC | NC | 143,19 |
| 18 | NC | NC | NC | NC | 143,19 |
| 18 | NC | NC | NC | NC | 143,19 |
| 19 | C | C | C | C | 142,3 |
| 19 | C | C | C | C | 142,3 |
| 19 | C | C | C | C | 142,3 |
| 20 | C | C | C | C | 142,75 |
| 20 | C | C | C | C | 142,75 |
| 20 | C | C | C | C | 142,75 |
| 21 | C | C | C | C | 142,28 |
| 21 | C | C | C | C | 142,28 |
| 21 | C | C | C | C | 142,28 |
| 22 | C | C | NC | C | 142,94 |
| 22 | C | C | C | C | 142,94 |
| 22 | C | C | NC | C | 142,94 |
| 23 | NC | NC | NC | NC | 143,15 |
| 23 | NC | NC | NC | NC | 143,15 |
| 23 | NC | NC | NC | NC | 143,15 |
| 24 | C | NC | C | C | 143,08 |
| 24 | NC | NC | NC | C | 143,08 |
| 24 | C | C | NC | C | 143,08 |
| 25 | C | C | C | C | 142,11 |
| 25 | C | C | C | C | 142,11 |
| 25 | C | NC | NC | C | 142,11 |
| 26 | C | C | C | C | 143 |
| 26 | C | C | C | C | 143 |
| 26 | C | C | NC | C | 143 |
| 27 | NC | NC | NC | NC | 143,22 |
| 27 | NC | NC | NC | NC | 143,22 |
| 27 | NC | NC | NC | NC | 143,22 |
| 28 | NC | NC | NC | NC | 143,5 |
| 28 | NC | NC | NC | NC | 143,5 |
| 28 | NC | NC | NC | NC | 143,5 |
| 29 | C | C | C | C | 142,85 |
| 29 | C | C | C | C | 142,85 |
| 29 | C | C | C | C | 142,85 |
| 30 | C | C | C | C | 142,73 |
| 30 | C | C | C | C | 142,73 |
| 30 | C | C | C | C | 142,73 |
| 31 | C | C | C | C | 142,59 |
| 31 | C | C | C | C | 142,59 |
| 31 | C | C | C | C | 142,59 |
| 32 | C | C | C | C | 142,89 |
| 32 | C | C | C | C | 142,89 |
| 32 | C | C | C | C | 142,89 |
| 33 | C | C | NC | C | 143,02 |
| 33 | C | C | C | C | 143,02 |
| 33 | C | C | NC | C | 143,02 |
| 34 | NC | NC | NC | NC | 142 |
| 34 | NC | NC | NC | NC | 142 |
| 34 | NC | NC | NC | NC | 142 |
| 35 | C | C | C | C | 142,25 |
| 35 | C | C | C | C | 142,25 |
| 35 | C | C | C | C | 142,25 |
| 36 | C | C | C | C | 142,33 |
| 36 | C | C | C | C | 142,33 |
| 36 | C | C | C | C | 142,33 |
| 37 | C | C | NC | C | 143,05 |
| 37 | C | C | C | C | 143,05 |
| 37 | C | C | NC | C | 143,05 |
| 38 | C | C | C | C | 142,82 |
| 38 | C | C | C | C | 142,82 |
| 38 | C | C | C | C | 142,82 |
| 39 | C | C | C | C | 142,89 |
| 39 | C | C | C | C | 142,89 |
| 39 | C | C | C | C | 142,89 |
| 40 | C | C | C | C | 142,49 |
| 40 | C | C | C | C | 142,49 |
| 40 | C | C | C | C | 142,49 |
| 41 | NC | NC | NC | NC | 141,95 |
| 41 | NC | NC | NC | NC | 141,95 |
| 41 | NC | NC | NC | NC | 141,95 |
| 42 | C | C | NC | C | 142,13 |
| 42 | C | NC | NC | C | 142,13 |
| 42 | C | NC | NC | C | 142,13 |
| 43 | C | C | C | C | 142,45 |
| 43 | C | C | C | C | 142,45 |
| 43 | C | C | C | C | 142,45 |
| 44 | C | C | C | C | 142,68 |
| 44 | C | C | C | C | 142,68 |
| 44 | C | C | C | C | 142,68 |
| 45 | C | C | C | C | 142,26 |
| 45 | C | C | C | C | 142,26 |
| 45 | C | C | C | C | 142,26 |
| 46 | C | C | C | C | 142,27 |
| 46 | C | C | C | C | 142,27 |
| 46 | C | C | C | C | 142,27 |
| 47 | C | C | C | C | 142,31 |
| 47 | C | C | C | C | 142,31 |
| 47 | C | C | C | C | 142,31 |
| 48 | C | C | C | C | 142,68 |
| 48 | C | C | C | C | 142,68 |
| 48 | C | C | C | C | 142,68 |
| 49 | C | C | C | C | 142,59 |
| 49 | C | C | C | C | 142,59 |
| 49 | C | C | C | C | 142,59 |
| 50 | C | C | C | C | 142,25 |
| 50 | C | C | C | C | 142,25 |
| 50 | C | C | C | C | 142,25 |
Table 4.1.3: data table.
Results obtained by Software Action:










Conclusion
Evaluators 1 and 2 have Kappa values above 0.75, we conclude that operators 1 and 2 agree with each other and with the standard. However, operator 3 has poor agreement with the other evaluators and the operators' pattern and error rates ($ op_1=3,03(porcentagem) $, $ op_2=0(porcentagem) $, $ op_3=0(porcentagem) $) are below that specified by the MSA manual, which is 5%. In this way, we have an opportunity for improvement.
11.4.1 Analytical Method
A cheap and quick way for companies to control their processes and products is to use go/no go measurement systems. These systems classify parts as defective or not, without determining the numerical value of the quality characteristic under analysis. Despite being cheap and agile, these measurement systems are prone to failure, which can be crucial for companies' reliability in the eyes of their customers. The pass-not-pass measurement system determines whether the part characteristic falls within the range between the engineering limits.

Figure 4.1.1: Pass-no-pass measurement system.
Figure 4.1.1 illustrates the pass-no-pass measurement system.
LIE = Lower engineering limit;
LSE = Upper engineering limit.
To classify the part, the pass-not-pass measurement system presents two faces. The pass face is made with a few fractions of a quality characteristic unit above the LIE, while the non-pass face is made with a few fractions of a quality characteristic unit below the LSE, as shown in Figure 4.1.2.
(image missing)
Figure 4.1.2: Configuration of the pass-no-pass measurement system.
Here, we propose a logistic regression model, which associates the numerical value of the part with the probability that the measurement system will classify the part as approved (or not defective). Although the numerical value is determined by a reliable measurement system, it presents measurement errors. Therefore, it is important to incorporate these errors into the logistical model.

Figure 4.1.3: Classification ranges.
When a part is “close” to engineering limits, the pass-no-pass measurement system is susceptible to making misclassifications. Thus, according to Figure 4.1.3 we have three ranges:
Track I - The part considered failed by the reference measurement system and by the pass measurement system does not pass.
Track II - Region close to engineering limits that leads to classification errors by the pass-no-pass measurement system.
Track III - The part considered approved by the reference measurement system and by the pass measurement system does not pass.
11.4.1.1 Study of the attribute-type measurement system
Next, we present the experiment proposed by the MSA manual (2002) to evaluate a pass-not-pass measurement system via the analytical method. As the pass and non-pass sides are made together, in practice we evaluate only one of the sides. So, let’s take the upper engineering limits (LSE) side for analysis. Here, we perform the following experiment:
We select $ k $ parts in production that are around the LSE, see (Figure 4.1.1.1);

Figure 4.1.1.1: Part analyzed - car crankcase.
- Parts are measured in a reliable measuring system (reference). In this case, we determine the numerical value of the quality characteristic, see Figures 4.1.1.2 and 4.1.1.3;
(image missing)
Figure 4.1.1.2: Resetting the dial indicator using a standard ring.
(image missing)
Figure 4.1.1.3: Dial indicator.
- The same parts are evaluated by the pass-no-pass system (smooth buffer gauge) n times, see Figures 4.1.1.4 and 4.1.1.5.
(image missing)
Figure 4.1.1.4: Smooth buffer gauge - passing side.
(image missing)
Figure 4.1.1.5: Smooth buffer gauge - side does not fit.
- We select the pieces in such a way that the smallest (largest) piece passes (fails) in all evaluations.
To illustrate, we consider the pass-no-pass measuring system with a smooth buffer gauge that classifies parts as defective or not, see Figures 4.1.1.4 and 4.1.1.5. This measuring system is applied in the company to control the characteristic diameter of the fixing holes in the car crankcase (Figure 4.1.1.1), with a specification of LIE = $ 9,3~mm $, LSE = $ 9,7~mm $ and Nominal Value = $ 9,5~mm $. For a double limit measuring device, for convenience, only the upper limit will be analyzed (no-pass), making appropriate assumptions of linearity and uniformity of error (Figure 4.1.2.1).
11.4.1.2 Description of the study of the attribute-type measurement system
- We selected 8 parts from the production line that were sent to the dimensional measurement laboratory to determine the numerical value of the part’s quality characteristic, that is, the diameter value, measured by the dial indicator (Figure 4.1.1.2) reset by a standard ring (Figure 4.1.1.3).
- After measuring the pieces, they returned to the production line where they were measured n = 20 times (each piece) using the pass-not-pass measurement system (the smooth plug). The pieces are distributed as shown in Figure.
- After carrying out the experiment, we summarized the data according to Table 4.1.2.1.
- To facilitate trend analysis, we subtracted the LSE from the measurement result of the part using the reference measurement system, as shown in Table 4.1.2.2.
| Ask | Diameter | Approval | Total Measurements |
|---|---|---|---|
| 1 | 9,64 | 20 | 20 |
| 2 | 9,65 | 20 | 20 |
| 3 | 9,67 | 20 | 20 |
| 4 | 9,68 | 8 | 20 |
| 5 | 9,7 | 5 | 20 |
| 6 | 9,71 | 3 | 20 |
| 7 | 9,73 | 0 | 20 |
| 8 | 9,75 | 0 | 20 |
Table 4.1.2.1: Experiment data
| Ask | Diameter | Approval | Total Measurements |
|---|---|---|---|
| 1 | -0,06 | 20 | 20 |
| 2 | -0,05 | 20 | 20 |
| 3 | -0,03 | 20 | 20 |
| 4 | -0,02 | 8 | 20 |
| 5 | 0 | 5 | 20 |
| 6 | 0,01 | 3 | 20 |
| 7 | 0,03 | 0 | 20 |
| 8 | 0,05 | 0 | 20 |
Table 4.1.2.2: Standardized Data
Figures 4.1.2.1, 4.1.2.2 and 4.1.2.3 show the measurements of the parts in relation to the LSE.

Figure 4.1.2.1: Classification of parts in the pass - no pass measurement system.

Figure 4.1.2.2: Arrangement of parts in the pass - no pass measurement system around the LSE.

Figure 4.1.2.3: Difference between diameter and LSE.
11.4.1.3 Analytical method via linear model
Consider a piece $ P_i $ where we control a quality characteristic, for example, diameter. We denote by $ \pi_i $ the probability of the part being approved by the measurement system passes does not pass. In practice, it is important to relate the numerical value of the part quality characteristic ($ {x_i} $) with the probability of its acceptance by the measurement system passes does not pass. Thus, the piece $ P_i $ is sent to the laboratory for measurement of the Quality characteristic. In our application, the part $ P_i $ is measured by a dial indicator with a resolution of $ 0,01~mm $, where we subtract the LSE from all measurements, as shown in Table 4.1.2.2. In general, we take
$$\pi_i = g({x}_i)$$
for everyone $ i = 1,2,3,\cdots, k $ parts. In particular, the MSA(2002) takes the probit model,
$$\pi_i = \Phi \left(\frac{{x}_i - \mu}{b} \right).~~~~~~(4.1.3.1)$$
in which $ \Phi $ represents the cumulative distribution function of the standard normal distribution and
- $ \mu $: represents the average of the measurements of the parts subtracted from the LSE with the application of the pass-no-pass system (trend),
- $ \sigma^2 = b^2 $: variance in the measurements of parts when applying the pass-not-pass measurement system.
Thus, we can define the quantities of interest as the trend parameters ($ \mu $) and repeatability ($ \mid b^\prime \mid\sqrt{n} $), in which $ n $ is the number of applications of the pass-no-pass measurement system on each piece. Using the expression (4.1.3.1), we have
$$\left(\frac{{x}_i - \mu}{b} \right) = \Phi^{-1}(\pi_i),~~~~~(4.1.3.2)$$
and
$${x}_i = \mu + b\Phi^{-1}(\pi_i).~~~~~~(4.1.3.3)$$
The MSA manual (2002) proposes that the probabilities be estimated according to the expression (4.1.3.4) and that the least squares method be applied to estimate the trend and repetitiveness.

Figure 4.1.3.1: Estimation of the probability of part acceptance versus part value.
In Figure 4.1.3.1, we can visualize the measurement value of the part ($ {x}_i $) in relation to estimates of the probability of acceptance of each part. With these estimates, we apply simple linear regression models to estimate the parameters of interest.

From the results of our experiment (Table 4.1.2.2), we obtain estimates for the probability of acceptance of the part $ \pi_i^\prime $ and for the associated quantile $ \Phi^{-1}(\pi_i^\prime) $ according to Table 4.1.3.1.
| Ask | $ x_i $ | Approval | Total | $ \pi^\prime $ | $ \Phi^{-1}(\pi^\prime) $ |
|---|---|---|---|---|---|
| 1 | -0,06 | 20 | 20 | 0,975 | 1,959 |
| 2 | -0,05 | 20 | 20 | 0,975 | 1,959 |
| 3 | -0,03 | 20 | 20 | 0,975 | 1,959 |
| 4 | -0,02 | 8 | 20 | 0,425 | -0,189 |
| 5 | 0 | 5 | 20 | 0,275 | -0,597 |
| 6 | 0,01 | 3 | 20 | 0,175 | -0,934 |
| 7 | 0,03 | 0 | 20 | 0,025 | -1,959 |
| 8 | 0,05 | 0 | 20 | 0,025 | -1,959 |
Table 4.1.3.1: Observed data versus quantiles of the Normal distribution.
The least squares estimates are given by,

With this, we conclude that the estimate of the tendency of the measurement system passes does not pass is given by $ -0,008117 ~ mm $. Furthermore, the repeatability of the pass-no-pass measurement system is given by:
$$\sigma^{\prime}_{rep} = \mid b^\prime \mid\sqrt{n} = 0,02124 \times \sqrt{20} =0,095 ~mm.$$
Although the trend estimate is small in relation to the tolerance $ (0,4 mm) $, we obtain that the repeatability is high.

Figure 4.1.3.2: Adjusted regression = $ {x}(\pi) = - 0,008117 - 0,02124*\Phi^{-1}(\pi_i^\prime) $.
An important point for analysis is to test whether the trend of the measurement system passes does not pass is significant from a statistical point of view. For this, the MSA manual (2002) proposes a test to evaluate the trend,
$$\mu \neq 0$$
(there are missing elements in the equation above)
The test statistic proposed by MSA (2002) is given by, under $ H_0 $,
$$t_{0}^{*} = \frac{(k - 1)* \sqrt{n}* \mu^\prime}{\sigma^{^\prime}_{rep}}\approx t_{(n - 1)}.$$
In the application, we obtain a p-value of 0.007472 and therefore reject $ H_0 $ for $ \alpha = 0,05 $. Therefore, the pass-no-pass measurement system presents a significant bias of -0.008 mm. For a tolerance (Tol = LSE - LIE = 9.7 - 9.3 = 0.4 mm) of 0.4 mm we conclude that the trend represents 2%. In general, a trend of this order (2% of tolerance) is not bothersome. In our example, this trend was detected due to the metrological characteristics of the measurement system that measured the parts (resolution of 0.01 mm and uncertainty of 0.006 mm).
From expression (4.1.3.1) we calculate the estimate of the probability of the measuring system passing not passing approving the part. Thus, considering $ x_T = -0,06 $, we have to:
$$\pi^{\prime}(-0,06)=\Phi\left(\frac{(-0,06-(-0,008117))}{-0,02124}\right)$$
$$=\Phi(2,44295)$$
$$=0,9927163$$
Figure 4.1.3.3 presents the results of estimating the probability of the measurement system passing not passing to approve parts for different values of the quality characteristic of these parts. Next, we present the graph (Figure 4.1.3.4) so that we can visualize the probability of acceptance depending on the value of the quality characteristic of the part, in our example its diameter. This graph is important so that the engineering group can evaluate the range of part diameter values where the measurement system passes and cannot make classification errors.

Figure 4.1.3.3: Observed data versus probability of the part being approved.

Figure 4.1.3.4: Probability of the part being approved versus quality characteristic.
Example 4.1.3.1
An attribute measurement device is being used to measure a dimension that has a tolerance of $ \pm 0,010 $. The Measuring Device is used in 100% automatic, end-of-line inspection and is affected by repeatability and bias. To do the attribute measuring device study, 8 parts with standard measurements in intervals of 0.002 and -0.016 to -0.002 are passed through the measuring device 20 times each. The acceptance numbers for each part are in Table 4.1.3.1:
| $ x_i $ | Approval |
|---|---|
| -0,016 | 0 |
| -0,014 | 3 |
| -0,012 | 8 |
| -0,01 | 20 |
| -0,008 | 20 |
| -0,006 | 20 |
| -0,004 | 20 |
| -0,002 | 20 |
Table 4.1.3.2: Experiment data
Since there are two reference values with $ 1 \leq a \leq 19 $, at least four more pieces must be found. Consequently, it is necessary to pass parts with reference values at the midpoints of the existing intervals. These reference values and acceptance numbers are shown in Table 4.1.3.3:
| $ x_i $ | Approval |
|---|---|
| -0,015 | 1 |
| -0,013 | 8 |
| -0,011 | 16 |
Table 4.1.3.3: Continuation of experiment data.
Now we have five reference values with $ 1 \leq a \leq 19 $. The procedure requires that more than one part be found with $ 1 \leq a \leq 19 $. Consequently, the following part shown in Table 4.1.3.4 is evaluated:
| $ x_i $ | Approval |
|---|---|
| -0,0105 | 18 |
Table 4.1.3.4: Part evaluated.
We emphasize that these measurements are arranged around the Lower Limit. For example:

Figure 4.1.3.5: Classification of parts in the pass - no pass measurement system.
Now that the data collection criteria have been met, we calculate $ \pi_i^\prime $ and $ \Phi^{-1}(\pi_i^\prime) $, according to Table 4.1.3.5.
| $ X_i $ | Approval | Total | $ \pi^\prime $ | $ \Phi^{-1}(\pi_i^\prime) $ |
|---|---|---|---|---|
| -0,016 | 0 | 20 | 0,025 | -1,959 |
| -0,015 | 1 | 20 | 0,075 | -1,439 |
| -0,014 | 3 | 20 | 0,175 | -0,934 |
| -0,013 | 5 | 20 | 0,275 | -0,597 |
| -0,012 | 8 | 20 | 0,425 | -0,189 |
| -0,011 | 16 | 20 | 0,775 | 0,755 |
| -0,0105 | 18 | 20 | 0,875 | 1,150 |
| -0,01 | 20 | 20 | 0,975 | 1,959 |
| -0,008 | 20 | 20 | 0,975 | 1,959 |
Table 4.1.3.5: Calculations.
Plotting the regression graph, $ x_i $ in relation to $ \Phi^{-1}(\pi_i^\prime) $, we have to:

Figure 4.1.3.6: Adjusted regression = $ {x}(\pi) = - 0,0123 + 0,001735*\Phi^{-1}(\pi_i^\prime) $.
Therefore the equation is:
$$x_i = \mu^\prime + b*\Phi^{-1}(\pi_i^\prime)$$
$$x_i = - 0,0123 + 0,001735*\Phi^{-1}(\pi_i^\prime)$$
Like this:
$${Tendência} = \mu^\prime=-0,0123$$
To determine repeatability, we need to find the value corresponding to:
$ \pi_i^\prime = 0,995 $ and $ \pi_i^\prime = 0,005 $, then we have
For $ \pi_i^\prime = 0,995 $:
$$x_i = - 0,0123 + 0,001735*\Phi^{-1}(0,995)$$
$$= - 0,0123 + 0,001735*(2,5758)$$
$$=-0,007832$$
For $ \pi_i^\prime = 0,005 $:
$$x_i = - 0,0123 + 0,001585*\Phi^{-1}(0,005)$$
$$= - 0,0123 + 0,001735*(-2,5758)$$
$$=-0,01677$$
This way:
$$\sigma^\prime_{rep}=\frac{\pi_i^\prime(0,995)-\pi_i^\prime(0,005)}{\frac{5,15}{\sqrt{20}}}=\frac{-0,007832-(-0,01677)}{\frac{5,15}{4,472}}= \frac{0,00894}{1,1515}=0,007763$$
Or even,
$$\sigma^\prime_{rep}= b*\sqrt{n} = 0,001735*\sqrt{20}=0,007763$$
To check if the trend is significant, we do:
$$t_{0}^{*}=\frac{(9-1)*\sqrt{20}*\mid{Tendência}\mid}{\sigma^\prime_{rep}}=\frac{35,77*{Tendência}}{\sigma^\prime_{rep}}=\frac{-0,4401}{\sigma^\prime_{rep}}=56,71$$
How $ t_0^{*} = 56,71 > t_{0,025(19)} = 2,093 $, then we reject the null hypothesis, that is, the trend is different from zero.
Considering $ x_i = -0,016 $ The probability of the measuring device approving the part is given by
$$=\Phi \left(\frac{-0,016 -(-0,0123)}{0,001735}\right)$$
$$=\Phi(-1,9599)$$
$$=0,01655$$
The Figure presents the result of the probability of the measuring device approving the part for values
of the variable $ x_i $.

Figure 4.1.3.7: Results.
Taking the data from Figure 4.1.3.7, we have the graph of reference values versus the probability of the part being approved, as shown in Figure 4.1.3.8:

Figure 4.1.3.8: Chart Reference values versus the probability of the part being approved.
11.5 MSA applications
In this section, we will show some applications of MSA to measurement systems. All cases presented are real and were prepared by the client with the participation of the ESTATCAMP team.
11.5.1 Hardness Measurement System: HRC
Hardness measurement system: HRC
Consider the system for measuring the hardness of a toothed part (photo 6.1) on the HRC scale. When we measure the hardness of the part at a certain point, we cause a change in the part around this point and consequently, we cannot repeat the hardness measurement in this region. Thus, we can interpret the hardness measurement system as a non-replicable system. However, in some applications, there is a region in the part where the hardness is homogeneous and in this case, we can use this region to evaluate the hardness measurement, considering this system as replicable. In summary, depending on the characteristics of the part to be measured, we can interpret the hardness measurement system as replicable or not.
In the application we will make below, we have a very homogeneous region in relation to hardness, a fact that allows us to consider the measurement system as replicable. Although we have a region with homogeneous hardness, it is not constant and this variation (even if small) will add to the repeatability of the measuring system. Next, we present a description of the measurement system.
Phase 1: Description of the Measurement System
We use this phase to carry out a review of the measurement system and planning for the application of the MSA. Here, it is important to define the team responsible for the analysis, the sectors that apply the measurement system and the need for measurement. For example, the part we measure the hardness of (Photo 6.1) affects the safety of the end user. Therefore, it is important that we have a reliable measurement system.

Figure 6.1: photo of the part with toothed sector.
For the hardness characteristic, the customer specified a minimum value of 55 HRC. As we only have a lower limit, we have no tolerance in the process. In this case, accuracy can be defined through the criticality of the measurement in agreement with the customer. For this application, we set an accuracy of 0.5 HRC.
So that the team can better understand the measurement system and its application, we carry out a scenario analysis using the following questions:
Importance/impact on the product?
Customer Complaints?
Failure History?
Is it a special feature?
Environmental Conditions?
Is the measurement system used for CEP?
Phase 2: Measurement Method
Phase I prepares us to make a good description of the measurement system. Therefore, we gathered the team to describe the measurement system and prepare the Ishikawa diagram.
Hardness measurement system description: HRC
- Zero hardness tester with standard
- Configure with standard 91.44 HR15N
- Clean the part (wash/air)
- Position the durometer
- Bring the handpiece closer to the hardness tester/penetrant
- Apply load
- Perform reading
We also define the measurement system parameterization variables
- Variable: Hardness;
- Description: measurement performed with 15N and transformed to HRC;
- Specification: minimum 55 HRC.
Another important point for analyzing the measurement system is the Ishikawa diagram, or cause and effect diagram. With the entire team together (at this stage it is important to have operators present), we brainstormed on the main causes of variation in the measurement system. The Ishikawa diagram is essential for defining the critical points and statistical techniques that we will use to evaluate the measurement system.
Cause and Effect Diagram (Ishikawa)

One of the critical items of a measurement system is the equipment used to carry out the measurements. The equipment is calibrated by appropriate laboratories and the calibration certificates must be critically evaluated against previously defined criteria. In this application, the hardness tester (Mitutoyo ATK-F1000 - No. 014.00084) has a resolution of 0.1 HRC, is calibrated annually and has a maximum allowable error of 0.3HRC. With this, we guarantee the required accuracy of 0.5 HRC. Below, we present a photo of the durometer.

Figure 6.2: photo of the Durometer.

Figure 6.3: photo of the Durometer.

Figure 6.4: photo of the Durometer.
With the description of the measurement system and the Ishikawa diagram, we plan the application of statistical techniques. In most processes, we basically perform three analyses: stability, RR and trend and linearity. Next, we describe the planning for applying statistical analyses.
Phase 3: Study Planning
- Stability: We selected 1 piece to be measured over time. In this application, we will make three rational subgroups of measurements per day for a week (20 subgroups).
- Trend: The same piece selected for stability will be used for the trend study. In this case, we set the reference value in a reliable measuring system, in this example we use the customer’s measuring system.
- Repeatability: 10 pieces from “distinct” production batches, with 3 measurements per piece carried out in random order.
- Reproducibility: Not applicable, as the operator does not influence the measurement system.
- Linearity: Not applicable, as the measuring range is small.
After data collection, we will perform statistical analysis.
Stability
| Subgroup | Date | Schedule | Person | Equipment | Measure 1 | Measure 2 | Measure 3 |
|---|---|---|---|---|---|---|---|
| 1 | 31/5/2010 | 08:00 | A | Durometer | 61,0 | 62,3 | 59,5 |
| 2 | 31/5/2010 | 14:00 | A | Durometer | 59,5 | 61,3 | 61,8 |
| 3 | 31/5/2010 | 23:00 | B | Durometer | 60,0 | 59,2 | 62,5 |
| 4 | 1/6/2010 | 08:00 | A | Durometer | 61,8 | 58,6 | 58,2 |
| 5 | 1/6/2010 | 14:00 | A | Durometer | 63,0 | 59,0 | 60,8 |
| 6 | 1/6/2010 | 23:00 | B | Durometer | 59,1 | 60,0 | 61,7 |
| 7 | 2/6/2010 | 08:00 | A | Durometer | 62,3 | 58,6 | 61,3 |
| 8 | 2/6/2010 | 14:00 | A | Durometer | 61,8 | 62,3 | 63,5 |
| 9 | 2/6/2010 | 23:00 | B | Durometer | 60,8 | 61,3 | 61 |
| 10 | 3/6/2010 | 08:00 | A | Durometer | 62,3 | 59,0 | 60,8 |
| 11 | 3/6/2010 | 14:00 | A | Durometer | 60,8 | 58,2 | 60 |
| 12 | 3/6/2010 | 23:00 | B | Durometer | 56,8 | 59,5 | 60,8 |
| 13 | 4/6/2010 | 08:00 | A | Durometer | 58,6 | 59,0 | 56,5 |
| 14 | 4/6/2010 | 14:00 | A | Durometer | 60,0 | 62,3 | 61,8 |
| 15 | 4/6/2010 | 23:00 | B | Durometer | 61,3 | 61,0 | 62,3 |
| 16 | 7/6/2010 | 08:00 | A | Durometer | 61,0 | 59,5 | 60 |
| 17 | 7/6/2010 | 14:00 | A | Durometer | 61,8 | 63,0 | 59,1 |
| 18 | 7/6/2010 | 23:00 | B | Durometer | 62,3 | 61,8 | 60,8 |
| 19 | 8/6/2010 | 08:00 | A | Durometer | 58,6 | 59,0 | 60 |
| 20 | 8/6/2010 | 14:00 | A | Durometer | 58,6 | 62,3 | 61,3 |
| 21 | 8/6/2010 | 23:00 | B | Durometer | 60,8 | 60,0 | 60,8 |
Results of this example obtained with the Action software:


Some concluding suggestions:
Stable system over time. All points are within the control lines.
Trend
In the trend study, we will consider the total variation to validate the repeatability of the measurement system. Through a CEP application, we detected a variation of 10 HRC (standard deviation of the process).
| Measurements |
|---|
| 61,0 |
| 59,5 |
| 60,0 |
| 61,8 |
| 63,0 |
| 59,1 |
| 62,3 |
| 61,8 |
| 60,8 |
| 62,3 |
| 60,8 |
| 56,8 |
| 58,6 |
| 60,0 |
| 61,3 |
| 62,3 |
| 61,3 |
| 59,2 |
| 58,6 |
| 59,0 |
| 60,0 |
| 58,6 |
| 62,3 |
| 61,3 |
| 59,0 |
| 58,2 |
| 59,5 |
| 59,0 |
| 62,3 |
| 61,0 |
| 59,5 |
| 61,8 |
| 62,5 |
| 58,2 |
| 60,8 |
| 61,7 |
| 61,3 |
| 63,5 |
| 61 |
| 60,8 |
| 60 |
| 60,8 |
| 56,5 |
| 61,8 |
| 62,3 |
| VR | VT |
|---|---|
| 60,4 | 10 |
Results of this example obtained with the Action software:


Some concluding suggestions:
As zero is part of the confidence interval, we consider that the measurement system does not present a significant trend at the 95% confidence level.
RR - Repeatability and Reproducibility
| Parts | Measurements |
|---|---|
| 1 | 60,0 |
| 2 | 65,0 |
| 3 | 62,3 |
| 4 | 61,3 |
| 5 | 65,0 |
| 6 | 62,3 |
| 7 | 59,5 |
| 8 | 59,0 |
| 9 | 59,0 |
| 10 | 61,8 |
| 1 | 61,8 |
| 2 | 65,0 |
| 3 | 61,3 |
| 4 | 61,8 |
| 5 | 64,0 |
| 6 | 63,5 |
| 7 | 58,2 |
| 8 | 58,6 |
| 9 | 59,4 |
| 10 | 60,8 |
| 1 | 61,3 |
| 2 | 65,7 |
| 3 | 62,3 |
| 4 | 61,3 |
| 5 | 64,5 |
| 6 | 64,0 |
| 7 | 57,8 |
| 8 | 60,0 |
| 9 | 60,4 |
| 10 | 61,8 |
Results of this example obtained with the Action software:


Some concluding suggestions
Graphic analysis:
The system has good discrimination and repeatability and remains consistent between parts.
Index analysis:
Despite obtaining a RR of 29%, we must remember that we are dealing with a non-replicable measurement system (hardness). As we cannot measure the part at the same point, this causes a natural increase in repeatability, with the inclusion of the internal variation of the part.
Therefore, we conclude that the measurement system is compatible with the process.
11.5.2 Measurement of the actual position on a flange
Measuring system for actual position of an inlet flange
The real position measurement system is a replicable dimensional measurement system, as we can perform several measurements of the part without damaging it. We will develop an MSA study for a Real Position Measurement system for Inlet Flange.
Phase 1: Description of the Measurement System
We use this phase to carry out a review of the measurement system and planning for the application of the MSA. Here, it is important to define the team responsible for the analysis, the sectors that apply the measurement system and the need for measurement. For example, the flange we measure the actual position (Photo 6.1) affects the safety of the end user. Therefore, it is important that we have a reliable measurement system.

Figure 6.1: photo of the Flange.
For the actual position characteristic, the customer specified a maximum value of 0.8 mm for the maximum and 0.2 for the minimum. As we only have a lower limit, we have no tolerance in the process. In this case, accuracy can be defined through the criticality of the measurement in agreement with the customer. For this application, we set an accuracy of 0.02 mm.
So that the team can better understand the measurement system and its application, we carry out a scenario analysis using the following questions:
Importance/impact on the product?
Customer Complaints?
Failure History?
Is it a special feature?
Environmental Conditions?
Is the measurement system used for CEP?
Phase 2: Measurement Method
Phase I prepares us to make a good description of the measurement system. Therefore, we gathered the team to describe the measurement system and prepare the Ishikawa diagram.
Description of actual position measurement system: mm
- Acquire the updated part drawing from the customer
- Prepare the part (cleaning with cloth and alcohol)
- Remove burrs from the piece (file)
- Fix the part on the magnetic prism as indicated in the customer’s drawing
- Create the program as informed by the drawing
- Carry out the measurement (as requested in the drawing by the customer)
- As we have two real positions, mentioned in the drawing (0.10 (2 holes) and 0.25 (6 holes)), we use the minimum and maximum results of the two results
We also define the measurement system parameterization variables
- Variable: actual position measured in millimeters (mm);
- Specification: 0.2mm.
Another important point for analyzing the measurement system is the Ishikawa diagram, or cause and effect diagram. With the entire team together (at this stage it is important to have operators present), we brainstormed on the main causes of variation in the measurement system. The Ishikawa diagram is essential for defining the critical points and statistical techniques that we will use to evaluate the measurement system.
Cause and Effect Diagram (Ishikawa)

One of the critical items of a measurement system is the equipment used to carry out the measurements. The equipment is calibrated by appropriate laboratories and the calibration certificates must be critically evaluated against previously defined criteria. In this application, the coordinate measuring machine (Zeiss) has a resolution of 0.001 mm, is calibrated annually and has a maximum allowable error of 0.004 mm for the measuring point in question. With this, we guarantee the required accuracy of 0.02 mm. Below, we present a photo of the coordinate measuring machine.

Figure 6.2: photo of the Coordinate Measuring Machine.

Figure 6.3: photo of the Coordinate Measuring Machine.
Phase 3: Study Planning
- Stability: We selected 1 piece to be measured over time. In this application, we will make three rational subgroups of measurements per day for a week (20 subgroups).
- Trend: The same piece selected for stability will be used for the trend study. In this case, we set the reference value in a reliable measuring system, in this example we use the customer’s measuring system.
- Repeatability: 10 pieces from “distinct” production batches, with 3 measurements per piece carried out in random order.
- Reproducibility: Not applicable, as the operator does not influence the measurement system.
- Linearity: Not applicable, as the measuring range is small.
After data collection, we will perform statistical analysis.
Stability
| Minimum | Minimum | Minimum | Maximum | Maximum | Maximum | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Subgroup | Date | Schedule | Person | Equipment | Measure 1 | Measure 2 | Measure 3 | Measure 1 | Measure 2 | Measure 3 |
| 1 | 31/5/2010 | 07:30 | A | three-dimensional | 0,077 | 0,073 | 0,073 | 0,182 | 0,182 | 0,181 |
| 2 | 31/5/2010 | 13:00 | A | three-dimensional | 0,048 | 0,068 | 0,077 | 0,181 | 0,188 | 0,181 |
| 3 | 31/5/2010 | 16:00 | A | three-dimensional | 0,078 | 0,078 | 0,074 | 0,180 | 0,183 | 0,182 |
| 4 | 1/6/2010 | 07:00 | A | three-dimensional | 0,076 | 0,069 | 0,066 | 0,181 | 0,180 | 0,174 |
| 5 | 1/6/2010 | 13:10 | A | three-dimensional | 0,066 | 0,068 | 0,066 | 0,178 | 0,178 | 0,178 |
| 6 | 1/6/2010 | 16:15 | A | three-dimensional | 0,068 | 0,060 | 0,079 | 0,177 | 0,185 | 0,177 |
| 7 | 2/6/2010 | 07:30 | A | three-dimensional | 0,076 | 0,076 | 0,075 | 0,176 | 0,179 | 0,178 |
| 8 | 2/6/2010 | 13:25 | A | three-dimensional | 0,075 | 0,076 | 0,075 | 0,178 | 0,178 | 0,178 |
| 9 | 2/6/2010 | 16:00 | A | three-dimensional | 0,074 | 0,076 | 0,078 | 0,177 | 0,179 | 0,178 |
| 10 | 3/6/2010 | 07:10 | A | three-dimensional | 0,075 | 0,075 | 0,077 | 0,179 | 0,177 | 0,177 |
| 11 | 3/6/2010 | 13:30 | A | three-dimensional | 0,075 | 0,076 | 0,071 | 0,178 | 0,178 | 0,178 |
| 12 | 3/6/2010 | 16:00 | A | three-dimensional | 0,076 | 0,077 | 0,077 | 0,178 | 0,177 | 0,178 |
| 13 | 7/6/2010 | 09:00 | A | three-dimensional | 0,076 | 0,076 | 0,073 | 0,173 | 0,177 | 0,177 |
| 14 | 7/6/2010 | 13:20 | A | three-dimensional | 0,077 | 0,075 | 0,075 | 0,175 | 0,179 | 0,177 |
| 15 | 7/6/2010 | 16:30 | A | three-dimensional | 0,071 | 0,078 | 0,075 | 0,177 | 0,178 | 0,179 |
| 16 | 8/6/2010 | 07:15 | A | three-dimensional | 0,075 | 0,075 | 0,075 | 0,178 | 0,177 | 0,177 |
| 17 | 8/6/2010 | 14:00 | A | three-dimensional | 0,074 | 0,074 | 0,074 | 0,178 | 0,178 | 0,178 |
| 18 | 8/6/2010 | 16:30 | A | three-dimensional | 0,074 | 0,074 | 0,076 | 0,178 | 0,178 | 0,178 |
| 19 | 9/6/2010 | 07:30 | A | three-dimensional | 0,076 | 0,075 | 0,073 | 0,177 | 0,176 | 0,177 |
| 20 | 9/6/2010 | 13:40 | A | three-dimensional | 0,076 | 0,074 | 0,075 | 0,177 | 0,177 | 0,179 |
| 21 | 9/6/2010 | 16:00 | A | three-dimensional | 0,073 | 0,075 | 0,075 | 0,179 | 0,178 | 0,179 |
Results of this example obtained with the Action software:
Stability analysis for the minimum measurement


Stability analysis for maximum measurement


Some concluding suggestions
System unstable over time. Note that the instability points were generated at the beginning of the data collection process. We probably had problems cleaning the part. In this case, we must identify the problem (cleaning), take corrective action and carry out the study again.
RR - Repeatability and Reproducibility
- RR analysis for the minimum measurement
| Ask | Minimum Measure |
|---|---|
| 1 | 0,076 |
| 2 | 0,099 |
| 3 | 0,01 |
| 4 | 0,012 |
| 5 | 0,04 |
| 6 | 0,072 |
| 7 | 0,04 |
| 8 | 0,18 |
| 9 | 0,086 |
| 10 | 0,13 |
| 1 | 0,077 |
| 2 | 0,104 |
| 3 | 0,009 |
| 4 | 0,01 |
| 5 | 0,041 |
| 6 | 0,074 |
| 7 | 0,02 |
| 8 | 0,155 |
| 9 | 0,087 |
| 10 | 0,131 |
| 1 | 0,077 |
| 2 | 0,105 |
| 3 | 0,011 |
| 4 | 0,01 |
| 5 | 0,041 |
| 6 | 0,074 |
| 7 | 0,021 |
| 8 | 0,156 |
| 9 | 0,087 |
| 10 | 0,128 |
Results of this example obtained with the Action software:


- RR analysis for the maximum measurement
| Ask | Maximum Measure |
|---|---|
| 1 | 0,178 |
| 2 | 0,196 |
| 3 | 0,168 |
| 4 | 0,182 |
| 5 | 0,186 |
| 6 | 0,279 |
| 7 | 0,202 |
| 8 | 0,377 |
| 9 | 0,201 |
| 10 | 0,471 |
| 1 | 0,179 |
| 2 | 0,197 |
| 3 | 0,166 |
| 4 | 0,178 |
| 5 | 0,185 |
| 6 | 0,285 |
| 7 | 0,216 |
| 8 | 0,373 |
| 9 | 0,202 |
| 10 | 0,481 |
| 1 | 0,179 |
| 2 | 0,2 |
| 3 | 0,165 |
| 4 | 0,175 |
| 5 | 0,185 |
| 6 | 0,285 |
| 7 | 0,213 |
| 8 | 0,367 |
| 9 | 0,203 |
| 10 | 0,476 |
Results of this example obtained with the Action software:


Some Conclusion Suggestions
Graphic analysis:
The system has good discrimination and repeatability and remains consistent between parts.
Index analysis:
We obtained an RR of 3.61% for the maximum and 11.43% for the minimum. Therefore, we concluded that the variability of the measurement system is adequate for the process.
Trend
The part selected in the stability study was measured by the customer so that we could establish the reference value. We will validate the repeatability through the RR made previously.
- Trend analysis for the minimum measurement: VR=0.1mm and VP=0.051345.
| Data |
|---|
| 0,077 |
| 0,071 |
| 0,077 |
| 0,073 |
| 0,075 |
| 0,075 |
| 0,075 |
| 0,074 |
| 0,076 |
| 0,073 |
| 0,075 |
| 0,075 |
Results of this example obtained with the Action software:


- Trend analysis for maximum measurement: VR=0.2mm and VP=0.102732.
| Data |
|---|
| 0,177 |
| 0,178 |
| 0,178 |
| 0,177 |
| 0,177 |
| 0,179 |
| 0,177 |
| 0,178 |
| 0,178 |
| 0,177 |
| 0,179 |
| 0,179 |
Results of this example obtained with the Action software:


We observed that the measurement system presents a significant trend (zero does not belong to the confidence interval) at the 95% confidence level. In this case, the supplier and customer must make a decision about the measurement system.
11.5.3 Visual measurement system
Visual Measurement System
In this section, we present an analysis of a measurement system for defect analysis with visual inspection.
Phase 1: Description of the Measurement System
We use this phase to carry out a review of the measurement system and planning for the application of the MSA. Here, it is important to define the team responsible for the analysis, the sectors that apply the measurement system and the need for measurement. For example, the part we measure the hardness of (Photo 6.1) affects the safety of the end user. Therefore, it is important that we have a reliable measurement system.
So that the team can better understand the measurement system and its application, we carry out a scenario analysis using the following questions:
Importance/impact on the product?
Customer Complaints?
Failure History?
Is it a special feature?
Environmental Conditions?
Is the measurement system used for CEP?
Phase 2: Types of defects and use and handling procedures:
-
Type of defects: Scratch, Knocks, Bumps, Marks on the Face, Scrap (several defects, no recovery), Incomplete Thread, No Thread, Damaged Thread, No Chamfer, Burr.
-
Complete details when viewing defects.
Another important point for analyzing the measurement system is the Ishikawa diagram, or cause and effect diagram. With the entire team together (at this stage it is important to have operators present), we brainstormed on the main causes of variation in the measurement system. The Ishikawa diagram is essential for defining the critical points and statistical techniques that we will use to evaluate the measurement system.
Cause and Effect Diagram (Ishikawa)

Phase 3: Study Planning
- Repeatability: 50 pieces, 20 compliant (C) and 30 non-compliant (NC).
- Reproducibility: 50 pieces, 20 compliant(C) and 30 non-compliant(NC), 3 piece measurements per operator
| Ask | Operator A | Operator B | Standard |
|---|---|---|---|
| 1 | C | C | C |
| 1 | C | C | C |
| 1 | C | C | C |
| 2 | NC | NC | NC |
| 2 | NC | NC | NC |
| 2 | NC | NC | NC |
| 3 | NC | NC | NC |
| 3 | NC | NC | NC |
| 3 | NC | NC | NC |
| 4 | NC | NC | NC |
| 4 | NC | NC | NC |
| 4 | NC | NC | NC |
| 5 | NC | NC | NC |
| 5 | NC | NC | NC |
| 5 | NC | NC | NC |
| 6 | C | NC | C |
| 6 | C | NC | C |
| 6 | C | NC | C |
| 7 | NC | NC | NC |
| 7 | NC | NC | NC |
| 7 | NC | NC | NC |
| 8 | NC | NC | NC |
| 8 | NC | NC | NC |
| 8 | NC | NC | NC |
| 9 | C | C | C |
| 9 | C | C | C |
| 9 | C | C | C |
| 10 | C | C | C |
| 10 | C | C | C |
| 10 | C | C | C |
| 11 | NC | NC | NC |
| 11 | NC | NC | NC |
| 11 | NC | NC | NC |
| 12 | NC | NC | NC |
| 12 | NC | NC | NC |
| 12 | NC | NC | NC |
| 13 | C | NC | C |
| 13 | C | NC | C |
| 13 | C | NC | C |
| 14 | NC | NC | NC |
| 14 | NC | NC | NC |
| 14 | NC | NC | NC |
| 15 | C | C | C |
| 15 | C | C | C |
| 15 | NC | C | C |
| 16 | NC | NC | NC |
| 16 | NC | NC | NC |
| 16 | NC | NC | NC |
| 17 | C | C | C |
| 17 | C | C | C |
| 17 | C | C | C |
| 18 | C | C | C |
| 18 | C | C | C |
| 18 | C | C | C |
| 19 | C | C | C |
| 19 | C | C | C |
| 19 | C | C | C |
| 20 | NC | NC | NC |
| 20 | NC | NC | NC |
| 20 | NC | NC | NC |
| 21 | NC | NC | NC |
| 21 | NC | NC | NC |
| 21 | NC | NC | NC |
| 22 | C | C | C |
| 22 | C | C | C |
| 22 | NC | C | C |
| 23 | C | C | C |
| 23 | C | C | C |
| 23 | C | C | C |
| 24 | NC | NC | NC |
| 24 | NC | NC | NC |
| 24 | NC | NC | NC |
| 25 | NC | NC | C |
| 25 | NC | NC | C |
| 25 | NC | NC | C |
| 26 | NC | NC | NC |
| 26 | NC | NC | NC |
| 26 | NC | NC | NC |
| 27 | NC | NC | NC |
| 27 | NC | NC | NC |
| 27 | NC | NC | NC |
| 28 | NC | NC | NC |
| 28 | NC | NC | NC |
| 28 | NC | NC | NC |
| 29 | NC | NC | NC |
| 29 | NC | NC | NC |
| 29 | NC | NC | NC |
| 30 | C | C | C |
| 30 | C | C | C |
| 30 | C | C | C |
| 31 | C | NC | C |
| 31 | C | C | C |
| 31 | C | C | C |
| 32 | NC | NC | NC |
| 32 | NC | NC | NC |
| 32 | NC | NC | NC |
| 33 | NC | NC | NC |
| 33 | NC | NC | NC |
| 33 | NC | NC | NC |
| 34 | NC | NC | NC |
| 34 | NC | NC | NC |
| 34 | NC | NC | NC |
| 35 | C | C | C |
| 35 | C | C | C |
| 35 | C | C | C |
| 36 | NC | NC | NC |
| 36 | NC | NC | NC |
| 36 | NC | NC | NC |
| 37 | C | C | C |
| 37 | C | C | C |
| 37 | NC | C | C |
| 38 | NC | NC | NC |
| 38 | NC | NC | NC |
| 38 | NC | NC | NC |
| 39 | NC | NC | NC |
| 39 | NC | NC | NC |
| 39 | NC | NC | NC |
| 40 | NC | NC | NC |
| 40 | NC | NC | NC |
| 40 | NC | NC | NC |
| 41 | C | C | C |
| 41 | C | C | C |
| 41 | C | C | C |
| 42 | NC | NC | NC |
| 42 | NC | NC | NC |
| 42 | NC | C | NC |
| 43 | NC | NC | NC |
| 43 | NC | NC | NC |
| 43 | NC | NC | NC |
| 44 | NC | NC | NC |
| 44 | NC | NC | NC |
| 44 | NC | NC | NC |
| 45 | C | C | C |
| 45 | C | NC | C |
| 45 | C | NC | C |
| 46 | NC | NC | NC |
| 46 | NC | NC | NC |
| 46 | NC | NC | NC |
| 47 | NC | NC | NC |
| 47 | NC | NC | NC |
| 47 | NC | NC | NC |
| 48 | NC | NC | NC |
| 48 | NC | NC | NC |
| 48 | NC | NC | NC |
| 49 | C | C | C |
| 49 | C | C | C |
| 49 | C | C | C |
| 50 | C | C | C |
| 50 | C | C | C |
| 50 | C | C | C |
Results of this example obtained with the Action software:




Some concluding suggestions
All indices were very good, except for the false alarm from operator 2. Since the false alarm is associated with good parts that are scrapped, which does not cause harm to the customer, we considered the measurement system to be appropriate for the process.
11.5.4 Weighing of raw materials and PM
Weighing of raw material and MP
Consider the measurement system for weighing raw materials and MP (photo 5.4.1) in kilograms. When we measure the mass of raw material at a certain point, we can repeat the measurement process. Thus, we can interpret the raw material weighing measurement system as a replicable system. In the application we will make below, it allows us to consider the measurement system as replicable. Next, we present a description of the measurement system.
Phase 1: Description of the Measurement System
We use this phase to carry out a review of the measurement system and planning for the application of the MSA. Here, it is important to define the team responsible for the analysis, the sectors that apply the measurement system and the need for measurement. For example, the equipment to be measured is a precision scale (Photo 5.4.1). Dosing an incorrect quantity of raw material or PM impacts the quality, that is, the result of the product. Therefore, it is important that we have a reliable measurement system.
Figure 6.1: photo of the precision scale.
For scale characteristic, deviation +/- 1% for raw material, tolerance 2% (others) and + 0.5%, - 0.3% for Precious Metal. Thus, we have 0.8% tolerance. In this case, accuracy can be defined through the criticality of the measurement in agreement with the customer. For this application, we set a maximum allowable calibration error of 0.1%.
So that the team can better understand the measurement system and its application, we carry out a scenario analysis using the following questions:
Importance/impact on the product?
Customer Complaints?
Failure History?
Is it a special feature?
Environmental Conditions?
Is the measurement system used for CEP?
Phase 2: Measurement Method
Phase I prepares us to make a good description of the measurement system. Therefore, we gathered the team to describe the measurement system and prepare the Ishikawa diagram.
Description of raw material weighing measurement system.
- Make sure the scale is at zero
- Place the container used to quantify the material
- Check the tare of the container
- Tare the scale and quantify the material to be fractionated
- Check the value of the packaged material
We also define the measurement system parameterization variables
- Variable: Raw material or MP (kg);
- Description: measurement carried out in kilograms;
- Specification: minimum 0.001kg and maximum 10kg.
Another important point for analyzing the measurement system is the Ishikawa diagram, or cause and effect diagram. With the entire team together (at this stage it is important to have operators present), we brainstormed on the main causes of variation in the measurement system. The Ishikawa diagram is essential for defining the critical points and statistical techniques that we will use to evaluate the measurement system.
Cause and Effect Diagram (Ishikawa)
(image missing)
One of the critical items of a measurement system is the equipment used to carry out the measurements. The equipment is calibrated by appropriate laboratories and the calibration certificates must be critically evaluated against previously defined criteria. In this application, scale calibration is only approved when the error plus calibration uncertainty at each measured point does not exceed 0.1% in the range of 5 to 12kg and has a maximum allowable calibration error of 0.1%.
With the description of the measurement system and the Ishikawa diagram, we plan the application of statistical techniques. In most processes, we basically perform three analyses: stability, RR and trend and linearity. Next, we describe the planning for applying statistical analyses.
Phase 3: Study Planning
- Stability: We selected 3 samples to be measured over time. In this application, we will make three rational subgroups of measurements per day for two weeks (20 subgroups).
- Trend and Linearity: We will take 10 measurements in at least 5 points, trying to cover the full scale of use of the equipment. In this case, we set the reference value in a reliable measuring system, in this example we use the customer’s measuring system.
- Repetitiveness: 10 measurements distributed by 2 operators in at least 5 different points.
- Reproducibility: Not applicable, as the operator does not cause interference (double checking in normal process).
After data collection, we will perform statistical analysis.
Stability
| Reading (g) | Pattern (g) | Difference (g) |
|---|---|---|
| 456,3 | 456,1 | 0,20 |
| 456,4 | 456,2 | 0,20 |
| 456,4 | 456,1 | 0,30 |
| 456,1 | 456,2 | -0,10 |
| 456,2 | 456,6 | -0,40 |
| 456,2 | 456,1 | 0,10 |
| 456,2 | 456,2 | 0,00 |
| 456,2 | 456,1 | 0,10 |
| 456,2 | 456,2 | 0,00 |
| 456,1 | 456,1 | 0,00 |
| 456,2 | 456,2 | 0,00 |
| 456,2 | 456,2 | 0,00 |
| 456,1 | 456,1 | 0,00 |
| 456,0 | 456,1 | -0,10 |
| 456,1 | 456,1 | 0,00 |
| 456,2 | 456,1 | 0,10 |
| 456,2 | 456,1 | 0,10 |
| 456,0 | 456,1 | -0,10 |
| 456,1 | 456,1 | 0,00 |
| 456,0 | 456,0 | 0,00 |
Results of this example obtained with the Action software:


Some concluding suggestions
The system presents some points of instability in all ranges of use. However, in all ranges the system showed a small variation compared to the tolerance (0.8% of the specified value). With this, we conclude that the system presents stability consistent with the application.
Trend and Linearity
In the study of trend and linearity, we will consider the total variation to validate the repeatability of the measurement system. Through a CEP application, we detected a variation of 10 HRC (standard deviation of the process).
| Reading (g) | VR (g) | Tolerance |
|---|---|---|
| 456,35 | 456,15 | 3,6492 |
| 456,35 | 456,15 | 3,6492 |
| 456,45 | 456,15 | 3,6492 |
| 456,05 | 456,15 | 3,6492 |
| 455,75 | 456,15 | 3,6492 |
| 456,25 | 456,15 | 3,6492 |
| 456,15 | 456,15 | 3,6492 |
| 456,25 | 456,15 | 3,6492 |
| 456,15 | 456,15 | 3,6492 |
| 456,15 | 456,15 | 3,6492 |
| 919,315 | 919,215 | 7,35372 |
| 919,315 | 919,215 | 7,35372 |
| 919,215 | 919,215 | 7,35372 |
| 919,115 | 919,215 | 7,35372 |
| 919,515 | 919,215 | 7,35372 |
| 919,315 | 919,215 | 7,35372 |
| 919,215 | 919,215 | 7,35372 |
| 919,215 | 919,215 | 7,35372 |
| 919,415 | 919,215 | 7,35372 |
| 919,415 | 919,215 | 7,35372 |
| 2443,305 | 2443,105 | 19,54484 |
| 2443,605 | 2443,105 | 19,54484 |
| 2442,605 | 2443,105 | 19,54484 |
| 2443,105 | 2443,105 | 19,54484 |
| 2443,805 | 2443,105 | 19,54484 |
| 2443,305 | 2443,105 | 19,54484 |
| 2443,405 | 2443,105 | 19,54484 |
| 2443,505 | 2443,105 | 19,54484 |
| 2443,405 | 2443,105 | 19,54484 |
| 2443,305 | 2443,105 | 19,54484 |
| 5458,995 | 5457,895 | 43,66316 |
| 5459,095 | 5457,895 | 43,66316 |
| 5458,995 | 5457,895 | 43,66316 |
| 5458,495 | 5457,895 | 43,66316 |
| 5459,295 | 5457,895 | 43,66316 |
| 5458,995 | 5457,895 | 43,66316 |
| 5458,495 | 5457,895 | 43,66316 |
| 5458,695 | 5457,895 | 43,66316 |
| 5458,895 | 5457,895 | 43,66316 |
| 5458,895 | 5457,895 | 43,66316 |
| 7544,205 | 7542,805 | 60,34244 |
| 7544,505 | 7542,805 | 60,34244 |
| 7544,505 | 7542,805 | 60,34244 |
| 7544,105 | 7542,805 | 60,34244 |
| 7544,305 | 7542,805 | 60,34244 |
| 7544,405 | 7542,805 | 60,34244 |
| 7543,505 | 7542,805 | 60,34244 |
| 7543,905 | 7542,805 | 60,34244 |
| 7543,905 | 7542,805 | 60,34244 |
| 7543,905 | 7542,805 | 60,34244 |
| 10407,02 | 10406,32 | 83,25056 |
| 10407,22 | 10406,32 | 83,25056 |
| 10407,22 | 10406,32 | 83,25056 |
| 10407,12 | 10406,32 | 83,25056 |
| 10407,82 | 10406,32 | 83,25056 |
| 10408,12 | 10406,32 | 83,25056 |
| 10407,22 | 10406,32 | 83,25056 |
| 10407,32 | 10406,32 | 83,25056 |
| 10407,22 | 10406,32 | 83,25056 |
| 10407,22 | 10406,32 | 83,25056 |
Results of this example obtained with the Action software:


Some concluding suggestions
The measurement system presents significant linearity throughout the system’s range of use (450 to 10400) with a descriptive value of below 5% (P-value= 1.33775E-14). The system presents a linearity of 0.0001 in relation to the measurement result, which in percentage is given by 0.01% of the result. Note that the tolerance represents 0.8% of the specified value. Therefore, the detected linearity is small when compared to the tolerance. With this, we conclude that the measurement system presents a trend and linearity compatible with the process.
Critical Analysis Trend and Linearity: As the largest trend detected was at the 10000g point is below 1.8g, which is lower than the equipment acceptance criterion (0.1% of reading = 10g), we conclude that the measurement system meets the process requirements.
RR - Repeatability and Reproducibility
| Reading (g) | Pattern (g) | Difference (g) |
|---|---|---|
| 456,3 | 456,1 | -0,20 |
| 456,4 | 456,2 | -0,20 |
| 456,4 | 456,1 | -0,30 |
| 456,1 | 456,2 | 0,10 |
| 456,2 | 456,6 | 0,40 |
| 456,2 | 456,1 | -0,10 |
| 456,2 | 456,2 | 0,00 |
| 456,2 | 456,1 | -0,10 |
| 456,2 | 456,2 | 0,00 |
| 456,1 | 456,1 | 0,00 |
| 456,2 | 456,2 | 0,00 |
| 456,2 | 456,2 | 0,00 |
| 456,1 | 456,1 | 0,00 |
| 456,0 | 456,1 | 0,10 |
| 456,1 | 456,1 | 0,00 |
| 456,2 | 456,1 | -0,10 |
| 456,2 | 456,1 | -0,10 |
| 456,0 | 456,1 | 0,10 |
| 456,1 | 456,1 | 0,00 |
| 456,0 | 456,0 | 0,00 |
| Tolerance |
|---|
| 3,6492 |
Results of this example obtained with the Action software:

Thus, the repeatability index in relation to tolerance is given by
$$(porcentagem){Repe}=\frac{6\times {Repe}}{{Tol}}\times 100=\frac{6\times 0,143637}{3,6492}\times 100=23,6167(porcentagem).$$
Some concluding suggestions
Repeatability: Variability compatible with the process
Critical analysis Repeatability: The Repeatability value was calculated at several points in the working range of the measuring system, in which we detected values below 30% of the tolerance. However, the initial points present a greater variation (around 23%) and as we increase the point on the scale, the repeatability value decreases in percentage.
The 0.8% tolerance used in this stage of the process (metal fractionation) ensures that the fractionated material is within a 4% tolerance in the final stage of the process.
Therefore, we conclude that the system is suitable for the process.
11.5.5 Vehicle headlight dimensional measurement system
The procedure we will describe below is a replicable dimensional measurement system, as we can perform several measurements of the part without damaging it. We will develop an MSA study for a vehicle headlight dimensional measurement system.
Phase 1: Description of the Measurement System
We use this phase to carry out a review of the measurement system and planning for the application of the MSA. Here, it is important to define the team responsible for the analysis, the sectors that apply the measurement system and the need for measurement. For example, the size of the Headlight (Photo 5.5.1) affects the fit to the vehicle body. Therefore, it is important that we have a reliable measurement system.

Figure 5.5.1: Part fixed to the bench.
For the dimensional characteristic of the headlight, the customer specified a process tolerance value of 4 mm. In this case, accuracy can be defined through the criticality of the measurement in agreement with the customer. For this application, we set an accuracy of 0.4 mm.
So that the team can better understand the measurement system and its application, we carry out a scenario analysis using the following questions:
- Importance/impact on the product?
- Customer Complaints?
- Failure History?
- Is it a special feature?
- Environmental Conditions?
- Is the measurement system used for CEP?
Phase 2: Measurement Method
Phase I prepares us to make a good description of the measurement system. Therefore, we gathered the team to describe the measurement system and prepare the Ishikawa diagram.
Description of the headlight dimensional measurement system
- Fixing the Headlight to the Cage
- Fixing the Cage to the Bench
- Position the Dial Dial Indicator at the first measurement point
- Perform reading
- Position the Dial Indicator at the other points (2 to 12) and perform the measurement
We also define the measurement system parameterization variables
- Variable: measurement by comparing the size of the Lighthouse;
- Description: measurements by comparison over 12 measurement points;
- Specification: 4mm tolerance.
Another important point for analyzing the measurement system is the Ishikawa diagram, or cause and effect diagram. With the entire team together (at this stage it is important to have operators present), we brainstormed on the main causes of variation in the measurement system. The Ishikawa diagram is essential for defining the critical points and statistical techniques that we will use to evaluate the measurement system.
Cause and Effect Diagram (Ishikawa)

One of the critical items of a measurement system is the equipment used to carry out the measurements. The equipment is calibrated by laboratories accredited to INMETRO and the calibration certificates must be critically evaluated in relation to previously defined criteria. In this application, the dial indicator has a resolution of 0.01 mm with a maximum allowable error of 0.03 mm. With this, we guarantee the required accuracy of 0.4 mm. Below, we present a photo of the dial indicator.

Figure 5.5.2: Dial indicator positioned at point 1

Figure 5.5.3: Dial indicator positioned at point 6.
With the description of the measurement system and the Ishikawa diagram, we plan the application of statistical techniques. In most processes, we basically perform three analyses: stability, RR and trend and linearity. Next, we describe the planning for applying statistical analyses.
Phase 3: Study Planning
- Stability: We selected 1 piece to be measured over time. In this application, we will make three rational subgroups of measurements per day for a week (25 subgroups).
- Repeatability and Reproductivity: 5 pieces from “distinct” production batches, with 3 measurements per piece carried out in random order by each operator.
- Linearity: The same part selected for stability will be used to study linearity along the 12 measurement points. In this case, we set the reference value in a reliable measuring system, in this example we use the customer’s measuring system.
After data collection, we will perform statistical analysis.
Stability
In this application we will only give an example for point 1, with the other points being applied in a similar way.
| Date | Time | M1 | M2 | M3 |
|---|---|---|---|---|
| 21/01/2011 | 10:33:00 | 0 | -0,02 | -0,01 |
| 25/01/2011 | 10:25:00 | 0 | 0 | -0,02 |
| 31/01/2011 | 11:33:00 | 0 | 0 | -0,01 |
| 01/02/2011 | 10:55:00 | 0 | 0,01 | 0 |
| 02/02/2011 | 14:40:00 | 0 | -0,01 | 0 |
| 03/02/2011 | 11:54:00 | 0 | -0,01 | -0,01 |
| 04/02/2011 | 14:43:00 | 0,01 | 0,01 | 0 |
| 07/02/2011 | 14:38:00 | 0,03 | 0,03 | 0,01 |
| 08/02/2011 | 14:39:00 | 0 | 0,01 | 0,01 |
| 09/02/2011 | 14:35:00 | 0 | 0 | 0 |
| 15/02/2011 | 10:07:00 | -0,02 | -0,01 | -0,01 |
| 16/02/2011 | 14:17:00 | 0,01 | 0 | -0,01 |
| 17/02/2011 | 14:28:00 | -0,01 | -0,02 | -0,02 |
| 18/02/2011 | 14:40:00 | 0 | -0,01 | -0,01 |
| 21/02/2011 | 11:27:00 | 0 | -0,01 | -0,02 |
| 22/02/2011 | 11:02:00 | 0 | 0,01 | -0,01 |
| 23/02/2011 | 14:46:00 | 0,04 | 0,03 | 0,03 |
| 28/02/2011 | 14:03:00 | 0 | -0,01 | 0 |
| 01/03/2011 | 11:09:00 | 0 | 0 | 0 |
| 03/03/2011 | 10:22:00 | 0 | -0,01 | -0,01 |
| 04/03/2011 | 11:18:00 | -0,02 | -0,02 | -0,03 |
| 08/03/2011 | 11:34:00 | 0,01 | 0 | 0 |
| 09/03/2011 | 11:40:00 | 0 | 0 | 0,01 |
| 10/03/2011 | 13:28:00 | 0,02 | 0,02 | 0,01 |
| 11/03/2011 | 10:23:00 | 0,01 | 0,01 | 0,01 |
Results of this example obtained with the Action software:


Some concluding suggestions
- We observed instability at the 12 measurement points. However, the overall result is good, since the variation found is small (maximum variation of 0.07 mm) compared to the product tolerance.
- Critical Stability Analysis: we observed a maximum amplitude of 0.07 mm, which represents around 1.75 (percentage) of the tolerance, corresponding to a small variation in practical terms. Note that from the point of view of the instrument’s accuracy (for a centesimal dial indicator we have an accuracy of around 0.03 mm) the detected instability (variations of up to 0.07 mm) is significant. This variation is probably caused by the fixing device and the part itself.
Repeatability and Reproductivity (RR)
RR analysis for point 1:
| Ask | Operator | Measure |
|---|---|---|
| 1 | 1 | -0,95 |
| 2 | 1 | -0,41 |
| 3 | 1 | -0,47 |
| 4 | 1 | -0,53 |
| 5 | 1 | -0,8 |
| 1 | 2 | -0,87 |
| 2 | 2 | -0,42 |
| 3 | 2 | -0,5 |
| 4 | 2 | -0,53 |
| 5 | 2 | -0,76 |
| 1 | 3 | -0,9 |
| 2 | 3 | -0,42 |
| 3 | 3 | -0,48 |
| 4 | 3 | -0,49 |
| 5 | 3 | -0,83 |
| 1 | 1 | -0,89 |
| 2 | 1 | -0,4 |
| 3 | 1 | -0,51 |
| 4 | 1 | -0,57 |
| 5 | 1 | -0,79 |
| 1 | 2 | -0,87 |
| 2 | 2 | -0,42 |
| 3 | 2 | -0,54 |
| 4 | 2 | -0,59 |
| 5 | 2 | -0,82 |
| 1 | 3 | -0,91 |
| 2 | 3 | -0,43 |
| 3 | 3 | -0,54 |
| 4 | 3 | -0,61 |
| 5 | 3 | -0,81 |
Results of this example obtained with the Action software:



Some concluding suggestions
We observed variability compatible with the process. All RR indices are below 10% (page 78 of the MSA 4th edition standard) in relation to the product tolerance (4 mm).
Trend and Linearity
To perform the Linearity analysis, we will validate the variability associated with repeatability using the total variation obtained in the RR study.
| Point | Reference Value | History (Total Change) |
|---|---|---|
| 1 | -0,01 | 0,206806 |
| 2 | 0,03 | 0,355426 |
| 3 | 0,87 | 0,216434 |
| 4 | -0,81 | 0,441185 |
| 5 | -0,6 | 0,283815 |
| 6 | 0,05 | 0,229691 |
| 7 | 0,59 | 0,194622 |
| 8 | -0,98 | 0,195215 |
| 9 | -0,92 | 0,080622 |
| 10 | -0,47 | 0,148387 |
| 11 | 0,25 | 0,212301 |
| 12 | -0,52 | 0,157431 |
From the history we will carry out the study of Trend and Linearity, as in table 5.5.1
| Point | Measure | VR | History |
|---|---|---|---|
| 1 | -0,01 | -0,01 | 0,206806 |
| 1 | -0,01 | -0,01 | 0,206806 |
| 1 | -0,03 | -0,01 | 0,206806 |
| 1 | -0,03 | -0,01 | 0,206806 |
| 1 | -0,02 | -0,01 | 0,206806 |
| 1 | -0,02 | -0,01 | 0,206806 |
| 1 | -0,02 | -0,01 | 0,206806 |
| 1 | -0,03 | -0,01 | 0,206806 |
| 1 | -0,02 | -0,01 | 0,206806 |
| 1 | -0,02 | -0,01 | 0,206806 |
| 2 | -0,01 | 0,03 | 0,355426 |
| 2 | -0,02 | 0,03 | 0,355426 |
| 2 | 0 | 0,03 | 0,355426 |
| 2 | 0,02 | 0,03 | 0,355426 |
| 2 | -0,01 | 0,03 | 0,355426 |
| 2 | -0,01 | 0,03 | 0,355426 |
| 2 | -0,01 | 0,03 | 0,355426 |
| 2 | 0 | 0,03 | 0,355426 |
| 2 | 0 | 0,03 | 0,355426 |
| 2 | 0 | 0,03 | 0,355426 |
| 3 | 0,85 | 0,87 | 0,216434 |
| 3 | 0,84 | 0,87 | 0,216434 |
| 3 | 0,87 | 0,87 | 0,216434 |
| 3 | 0,88 | 0,87 | 0,216434 |
| 3 | 0,86 | 0,87 | 0,216434 |
| 3 | 0,84 | 0,87 | 0,216434 |
| 3 | 0,84 | 0,87 | 0,216434 |
| 3 | 0,86 | 0,87 | 0,216434 |
| 3 | 0,86 | 0,87 | 0,216434 |
| 3 | 0,86 | 0,87 | 0,216434 |
| 4 | -0,81 | -0,81 | 0,441185 |
| 4 | -0,8 | -0,81 | 0,441185 |
| 4 | -0,82 | -0,81 | 0,441185 |
| 4 | -0,82 | -0,81 | 0,441185 |
| 4 | -0,82 | -0,81 | 0,441185 |
| 4 | -0,81 | -0,81 | 0,441185 |
| 4 | -0,82 | -0,81 | 0,441185 |
| 4 | -0,81 | -0,81 | 0,441185 |
| 4 | -0,82 | -0,81 | 0,441185 |
| 4 | -0,81 | -0,81 | 0,441185 |
| 5 | -0,69 | -0,6 | 0,283815 |
| 5 | -0,7 | -0,6 | 0,283815 |
| 5 | -0,7 | -0,6 | 0,283815 |
| 5 | -0,7 | -0,6 | 0,283815 |
| 5 | -0,68 | -0,6 | 0,283815 |
| 5 | -0,68 | -0,6 | 0,283815 |
| 5 | -0,66 | -0,6 | 0,283815 |
| 5 | -0,67 | -0,6 | 0,283815 |
| 5 | -0,67 | -0,6 | 0,283815 |
| 5 | -0,67 | -0,6 | 0,283815 |
| 6 | 0,04 | 0,05 | 0,229691 |
| 6 | 0,03 | 0,05 | 0,229691 |
| 6 | 0,05 | 0,05 | 0,229691 |
| 6 | 0,04 | 0,05 | 0,229691 |
| 6 | 0,02 | 0,05 | 0,229691 |
| 6 | 0,03 | 0,05 | 0,229691 |
| 6 | 0,03 | 0,05 | 0,229691 |
| 6 | 0,04 | 0,05 | 0,229691 |
| 6 | 0,04 | 0,05 | 0,229691 |
| 6 | 0,04 | 0,05 | 0,229691 |
| 7 | 0,5 | 0,59 | 0,194622 |
| 7 | 0,55 | 0,59 | 0,194622 |
| 7 | 0,51 | 0,59 | 0,194622 |
| 7 | 0,53 | 0,59 | 0,194622 |
| 7 | 0,52 | 0,59 | 0,194622 |
| 7 | 0,56 | 0,59 | 0,194622 |
| 7 | 0,54 | 0,59 | 0,194622 |
| 7 | 0,55 | 0,59 | 0,194622 |
| 7 | 0,54 | 0,59 | 0,194622 |
| 7 | 0,54 | 0,59 | 0,194622 |
| 8 | -1,02 | -0,98 | 0,195215 |
| 8 | -0,96 | -0,98 | 0,195215 |
| 8 | -0,99 | -0,98 | 0,195215 |
| 8 | -0,96 | -0,98 | 0,195215 |
| 8 | -1,01 | -0,98 | 0,195215 |
| 8 | -0,96 | -0,98 | 0,195215 |
| 8 | -1 | -0,98 | 0,195215 |
| 8 | -0,96 | -0,98 | 0,195215 |
| 8 | -0,97 | -0,98 | 0,195215 |
| 8 | -0,96 | -0,98 | 0,195215 |
| 9 | -0,99 | -0,92 | 0,080622 |
| 9 | -0,98 | -0,92 | 0,080622 |
| 9 | -0,99 | -0,92 | 0,080622 |
| 9 | -0,99 | -0,92 | 0,080622 |
| 9 | -0,93 | -0,92 | 0,080622 |
| 9 | -0,96 | -0,92 | 0,080622 |
| 9 | -0,99 | -0,92 | 0,080622 |
| 9 | -0,99 | -0,92 | 0,080622 |
| 9 | -0,97 | -0,92 | 0,080622 |
| 9 | -0,96 | -0,92 | 0,080622 |
| 10 | -0,52 | -0,47 | 0,148387 |
| 10 | -0,51 | -0,47 | 0,148387 |
| 10 | -0,53 | -0,47 | 0,148387 |
| 10 | -0,52 | -0,47 | 0,148387 |
| 10 | -0,48 | -0,47 | 0,148387 |
| 10 | -0,51 | -0,47 | 0,148387 |
| 10 | -0,52 | -0,47 | 0,148387 |
| 10 | -0,49 | -0,47 | 0,148387 |
| 10 | -0,52 | -0,47 | 0,148387 |
| 10 | -0,51 | -0,47 | 0,148387 |
| 11 | 0,27 | 0,25 | 0,212301 |
| 11 | 0,28 | 0,25 | 0,212301 |
| 11 | 0,27 | 0,25 | 0,212301 |
| 11 | 0,28 | 0,25 | 0,212301 |
| 11 | 0,26 | 0,25 | 0,212301 |
| 11 | 0,27 | 0,25 | 0,212301 |
| 11 | 0,26 | 0,25 | 0,212301 |
| 11 | 0,27 | 0,25 | 0,212301 |
| 11 | 0,26 | 0,25 | 0,212301 |
| 11 | 0,27 | 0,25 | 0,212301 |
| 12 | -0,55 | -0,52 | 0,157431 |
| 12 | -0,54 | -0,52 | 0,157431 |
| 12 | -0,57 | -0,52 | 0,157431 |
| 12 | -0,56 | -0,52 | 0,157431 |
| 12 | -0,53 | -0,52 | 0,157431 |
| 12 | -0,54 | -0,52 | 0,157431 |
| 12 | -0,55 | -0,52 | 0,157431 |
| 12 | -0,55 | -0,52 | 0,157431 |
| 12 | -0,55 | -0,52 | 0,157431 |
| 12 | -0,54 | -0,52 | 0,157431 |
Table 5.5.1: Table for carrying out the Trend and Linearity study.
Results of this example obtained with the Action software:



Some concluding suggestions
- Trend and Linearity: The measurement system does not have significant linearity throughout the system usage points (1 to 12) as the p-value found is above 5% (P-value= 0.16).
- Critical analysis Trend and Linearity: Although we do not have significant linearity, most points present a significant trend from a statistical point of view (p-value below 5%). The largest trend detected (point 4) is -0.08 mm which is above the maximum allowable error for the dial indicator (approximately 0.03 mm). However, this maximum tendency represents 2% of the tolerance, which can be considered negligible from a practical point of view.
11.5.6 Measuring system for rivet height
The procedure we will describe below is a replicable measurement system, as we can take several measurements of the part without damaging it. We will develop an MSA study for a measurement system for tubular rivet height.
Phase 1: Description of the Measurement System
We use this phase to carry out a review of the measurement system and planning for the application of the MSA. Here, it is important to define the team responsible for the analysis, the sectors that apply the measurement system and the need for measurement. For example, the height of the rivet (Photo 5.6.1) guarantees a useful wear lining on a clutch disc, thereby preventing clutch slippage. Therefore, it is important that we have a reliable measurement system.

Figure 5.6.1: Photo of the clutch disc.
For the tubular rivet height characteristic, the customer specified a process tolerance value of 0.4 mm. In this case, accuracy can be defined through the criticality of the measurement in agreement with the customer. For this application, we set an accuracy of 0.04 mm.
So that the team can better understand the measurement system and its application, we carry out a scenario analysis using the following questions:
- Importance/impact on the product?
- Customer Complaints?
- Failure History?
- Is it a special feature?
- Environmental Conditions?
- Is the measurement system used for CEP?
Phase 2: Measurement Method
Phase I prepares us to make a good description of the measurement system. Therefore, we gathered the team to describe the measurement system and prepare the Ishikawa diagram.
Description of the tubular rivet height measurement system.
- Positioning the clutch disc on the bench
- Fixing the caliper base to the clutch disc gasket
- Carrying out the reading
We also define the measurement system parameterization variables
- Variable: measurement of the height of the clutch disc rod;
- Description: height measurements;
- Specification: 0.4mm tolerance. (+/- 0.2mm)
Another important point for analyzing the measurement system is the Ishikawa diagram, or cause and effect diagram. With the entire team together (at this stage it is important to have operators present), we brainstormed on the main causes of variation in the measurement system. The Ishikawa diagram is essential for defining the critical points and statistical techniques that we will use to evaluate the measurement system.
Cause and Effect Diagram (Ishikawa)

One of the critical items of a measurement system is the equipment used to carry out the measurements. The equipment is calibrated by laboratories accredited to INMETRO and the calibration certificates must be critically evaluated in relation to previously defined criteria. In this application, the caliper has a resolution of 0.01 mm and is calibrated annually. The maximum allowable caliper calibration error is 0.03 mm. With this, we guarantee the required accuracy of 0.04 mm. Below, we present a photo of the measurement of the height of the tubular pin.

Figure 5.6.2: Caliper positioning.

Figure 5.6.3: Measurement of the height of the tubular rivet.
With the description of the measurement system and the Ishikawa diagram, we plan the application of statistical techniques. In most processes, we basically perform three analyses: stability, RR and trend and linearity. Next, we describe the planning for applying statistical analyses.
Phase 3: Study Planning
- Stability: We selected 1 piece to be measured over time. In this application, we will make three rational subgroups of measurements per day for a week (20 subgroups).
- Repeatability and Reproducibility: 10 pieces from “distinct” production batches, with 3 measurements per piece carried out in random order by each operator. In this application, we choose 3 operators.
- Linearity: 5 pieces were selected (at different points), in which we defined the reference value in a reliable measuring system, in this example, we used the customer’s measuring system.
After data collection, we will perform statistical analysis.
Stability
In this application we will exemplify the application.

Results of this example obtained with the Action software:


Some concluding suggestions
- Measuring system stable over time.
- Critical Stability Analysis: in the amplitude graph it indicates that there is a lack of discrimination in the Measurement System, as we have many points equal to zero. In general, the resolution of measuring equipment may be inadequate.
Repeatability and Reproductivity (RR)
RR analysis for point 1:

Results of this example obtained with the Action software:



Some concluding suggestions
We observed variability compatible with the process.
Critical Analysis
The RR of the Measuring System is 0.03 mm, which is compatible with the instrument used (caliper). Furthermore, this RR represents 18% of the total variation and has an NDC index of 7 categories. We also observed that reproducibility is almost zero, which means that the impact of operators is small. In general, the variability of the measurement system is compatible with process control.
Trend and Linearity
To perform the Linearity analysis, we will validate the variability associated with repeatability using the total variation obtained in the RR study.

Table 5.5.1: Table for carrying out the Trend and Linearity study.
Results of this example obtained with the Action software:



Some concluding suggestions
Trend and Linearity
The measurement system presents non-significant linearity at the 5% level (P-value= 0.62). We also note that all trends were considered non-significant (see graph). With this, we conclude that the measurement system presents a negligible trend and linearity.
11.5.7 Torque measurement system applied with a digital torque meter
Torque measurement system applied with a digital torque wrench
Consider the torque measurement system applied with a digital torque meter (photo 5.7.1) on the Nm scale. When measuring the torque on a screw, we cause a change in the part around this point and consequently, we cannot repeat the torque measurement in this region. Thus, we can interpret the torque measurement system as a non-replicable system.
In the following application, we will adopt the following strategy:
- Choose a part type (MWM Euro 5 engine). As the measurement system is non-replicable, we measure the part only once. Therefore, we will use several engines of the same model. It is also important to choose some specific points on the engine for monitoring. We chose three torque application points.
- Determine the frequency of the study: we chose once a week.
Phase 1: Description of the Measurement System
We use this phase to carry out a review of the measurement system and planning for the application of the MSA. Here, it is important to define the team responsible for the analysis, the sectors that apply the measurement system and the need for measurement. For example, the part we measure torque on (Photo 5.7.1) affects end user safety. Therefore, it is important that we have a reliable measurement system.
(image missing)
Figure 5.7.1: photo of measuring torque applied to screws.
For the torque characteristic, we have a digital torque meter with a resolution of 0.1 Nm and a tolerance of ±30 Nm. For this application, we have an accuracy of 6 Nm.
So that the team can better understand the measurement system and its application, we carry out a scenario analysis using the following questions:
Importance/impact on the product?
Customer Complaints?
Failure History?
Is it a special feature?
Environmental Conditions?
Is the measurement system used for CEP?
Phase 2: Measurement Method
Phase I prepares us to make a good description of the measurement system. Therefore, we gathered the team to describe the measurement system and prepare the Ishikawa diagram.
Description of the torque measurement system applied with a digital torque meter: Nm
- Connect a torque transducer (standard equipment) to the screw
- Apply torque to the screw in a part (with the torque wrench)
- Record the torque value obtained from the torque wrench and standard equipment (torque transducer)
Another important point for analyzing the measurement system is the Ishikawa diagram, or cause and effect diagram. With the entire team together (at this stage it is important to have operators present), we brainstormed on the main causes of variation in the measurement system. The Ishikawa diagram is essential for defining the critical points and statistical techniques that we will use to evaluate the measurement system.
Cause and Effect Diagram (Ishikawa)
(image missing)
One of the critical items of a measurement system is the equipment used to carry out the measurements. The equipment is calibrated by appropriate laboratories and the calibration certificates must be critically evaluated against previously defined criteria. In this application, the digital torque wrench has a resolution of 0.1 Nm. With this, we guarantee the required accuracy of 6 Nm. Below, we present a photo of the digital torque wrench.
(image missing)
With the description of the measurement system and the Ishikawa diagram, we plan the application of statistical techniques. In most processes, we basically perform three analyses: stability, RR and trend and linearity. Next, we describe the planning for applying statistical analyses.
Phase 3: Study Planning
- Stability: We selected three engine torque points monitored over time (once a week). Analysis using the Xbar and R graph;
- Trend: Comparison with standard transducer. A trend analysis per point;
- Repeatability: three measurement points with 25 replicates;
- Reproducibility: We will not perform the reproducibility test;
- Note: all three analyzes were performed with the same data table.
After data collection, we will perform statistical analysis.
| Point 1 | Point 1 | Point 1 | Point 2 | Point 2 | Point 2 | Point 3 | Point 3 | Point 3 | |
|---|---|---|---|---|---|---|---|---|---|
| Date | Torque meter | Transducer | Trend | Torque meter | Transducer | Trend | Torque meter | Transducer | Trend |
| 1 | 351,9 | 351,9 | 0 | 316,5 | 317,06 | -0,56 | 359,3 | 361,87 | -2,57 |
| 2 | 353,3 | 362,89 | -9,59 | 334,2 | 337,68 | -3,48 | 344,3 | 346,74 | -2,44 |
| 3 | 357,2 | 363,04 | -5,84 | 316,4 | 319,66 | -3,26 | 347,5 | 351,51 | -4,01 |
| 4 | 347,2 | 352,79 | -5,59 | 326,8 | 335,18 | -8,38 | 342 | 343,31 | -1,31 |
| 5 | 355,2 | 364,86 | -9,66 | 324,1 | 327,88 | -3,78 | 348,7 | 352,49 | -3,79 |
| 6 | 340,7 | 342,22 | -1,52 | 329,3 | 332,2 | -2,9 | 355,5 | 359,46 | -3,96 |
| 7 | 352,8 | 353,9 | -1,1 | 326,9 | 336,69 | -9,79 | 351 | 355,63 | -4,63 |
| 8 | 347 | 355,29 | -8,29 | 326,1 | 326 | 0,1 | 356,5 | 360,32 | -3,82 |
| 9 | 354,6 | 357,91 | -3,31 | 327,6 | 327,52 | 0,08 | 353,3 | 352,64 | 0,66 |
| 10 | 347,2 | 350,48 | -3,28 | 339,6 | 339,51 | 0,09 | 350 | 360,18 | -10,18 |
| 11 | 348,8 | 354,02 | -5,22 | 331,1 | 337,11 | -6,01 | 349,3 | 355,24 | -5,94 |
| 12 | 351,1 | 359,25 | -8,15 | 325,4 | 327,68 | -2,28 | 350,9 | 355,52 | -4,62 |
| 13 | 343,1 | 343,6 | -0,5 | 341,8 | 345,09 | -3,29 | 355,2 | 356,71 | -1,51 |
| 14 | 352,2 | 359,02 | -6,82 | 320,5 | 329,73 | -9,23 | 357 | 361,48 | -4,48 |
| 15 | 357,2 | 364,09 | -6,89 | 314,7 | 315,64 | -0,94 | 341 | 342,29 | -1,29 |
| 16 | 354,6 | 355,24 | -0,64 | 330,5 | 334,77 | -4,27 | 347,3 | 347,94 | -0,64 |
| 17 | 350,8 | 357,62 | -6,82 | 346,2 | 348,61 | -2,41 | 346 | 356,14 | -10,14 |
| 18 | 346,6 | 347,68 | -1,08 | 318,4 | 320,18 | -1,78 | 353,1 | 357,9 | -4,8 |
| 19 | 347,1 | 353,13 | -6,03 | 322,5 | 329,13 | -6,63 | 342,7 | 348,82 | -6,12 |
| 20 | 353,5 | 361,27 | -7,77 | 336 | 339,81 | -3,81 | 345,6 | 347,74 | -2,14 |
| 21 | 353,3 | 356,44 | -3,14 | 321,8 | 326,13 | -4,33 | 347,5 | 354,7 | -7,2 |
| 22 | 357,2 | 359,42 | -2,22 | 334,9 | 338,32 | -3,42 | 352,2 | 357,54 | -5,34 |
| 23 | 353,4 | 353,26 | 0,14 | 341,3 | 347,58 | -6,28 | 353,3 | 355,86 | -2,56 |
| 24 | 353 | 354,13 | -1,13 | 335,4 | 339,97 | -4,57 | 353,8 | 358,23 | -4,43 |
| 25 | 347,9 | 348,44 | -0,54 | 319,8 | 324,13 | -4,33 | 357,9 | 358,85 | -0,95 |
Stability
| Week | T1 | T2 | T3 |
|---|---|---|---|
| 1 | 0 | -0,56 | -2,57 |
| 2 | -9,59 | -3,48 | -2,44 |
| 3 | -5,84 | -3,26 | -4,01 |
| 4 | -5,59 | -8,38 | -1,31 |
| 5 | -9,66 | -3,78 | -3,79 |
| 6 | -1,52 | -2,9 | -3,96 |
| 7 | -1,1 | -9,79 | -4,63 |
| 8 | -8,29 | 0,1 | -3,82 |
| 9 | -3,31 | 0,08 | 0,66 |
| 10 | -3,28 | 0,09 | -10,18 |
| 11 | -5,22 | -6,01 | -5,94 |
| 12 | -8,15 | -2,28 | -4,62 |
| 13 | -0,5 | -3,29 | -1,51 |
| 14 | -6,82 | -9,23 | -4,48 |
| 15 | -6,89 | -0,94 | -1,29 |
| 16 | -0,64 | -4,27 | -0,64 |
| 17 | -6,82 | -2,41 | -10,14 |
| 18 | -1,08 | -1,78 | -4,8 |
| 19 | -6,03 | -6,63 | -6,12 |
| 20 | -7,77 | -3,81 | -2,14 |
| 21 | -3,14 | -4,33 | -7,2 |
| 22 | -2,22 | -3,42 | -5,34 |
| 23 | 0,14 | -6,28 | -2,56 |
| 24 | -1,13 | -4,57 | -4,43 |
| 25 | -0,54 | -4,33 | -0,95 |
Results of this example obtained with the Action software:
(image missing)
(image missing)
Some concluding suggestions
Stable system over time. All points are within the control lines.
Trend
Let’s apply the trend study to the three engine torque points. Next, we have the study in the first point.
| Week | T1 |
|---|---|
| 1 | 0 |
| 2 | -9,59 |
| 3 | -5,84 |
| 4 | -5,59 |
| 5 | -9,66 |
| 6 | -1,52 |
| 7 | -1,1 |
| 8 | -8,29 |
| 9 | -3,31 |
| 10 | -3,28 |
| 11 | -5,22 |
| 12 | -8,15 |
| 13 | -0,5 |
| 14 | -6,82 |
| 15 | -6,89 |
| 16 | -0,64 |
| 17 | -6,82 |
| 18 | -1,08 |
| 19 | -6,03 |
| 20 | -7,77 |
| 21 | -3,14 |
| 22 | -2,22 |
| 23 | 0,14 |
| 24 | -1,13 |
| 25 | -0,54 |
Results of this example obtained with the Action software:
(image missing)
(image missing)
Now, let’s apply the trend study in point 2.
| Week | T2 |
|---|---|
| 1 | -0,56 |
| 2 | -3,48 |
| 3 | -3,26 |
| 4 | -8,38 |
| 5 | -3,78 |
| 6 | -2,9 |
| 7 | -9,79 |
| 8 | 0,1 |
| 9 | 0,08 |
| 10 | 0,09 |
| 11 | -6,01 |
| 12 | -2,28 |
| 13 | -3,29 |
| 14 | -9,23 |
| 15 | -0,94 |
| 16 | -4,27 |
| 17 | -2,41 |
| 18 | -1,78 |
| 19 | -6,63 |
| 20 | -3,81 |
| 21 | -4,33 |
| 22 | -3,42 |
| 23 | -6,28 |
| 24 | -4,57 |
| 25 | -4,33 |
Results of this example obtained with the Action software:
(image missing)
(image missing)
Finally, let’s apply the trend study in point 3.
| Week | T3 |
|---|---|
| 1 | -2,57 |
| 2 | -2,44 |
| 3 | -4,01 |
| 4 | -1,31 |
| 5 | -3,79 |
| 6 | -3,96 |
| 7 | -4,63 |
| 8 | -3,82 |
| 9 | 0,66 |
| 10 | -10,18 |
| 11 | -5,94 |
| 12 | -4,62 |
| 13 | -1,51 |
| 14 | -4,48 |
| 15 | -1,29 |
| 16 | -0,64 |
| 17 | -10,14 |
| 18 | -4,8 |
| 19 | -6,12 |
| 20 | -2,14 |
| 21 | -7,2 |
| 22 | -5,34 |
| 23 | -2,56 |
| 24 | -4,43 |
| 25 | -0,95 |
Results of this example obtained with the Action software:
(image missing)
(image missing)
Some concluding suggestions
At the three engine torque points, the zero line is not part of the confidence interval, the trends found are significant at the 5% significance level.
RR - Repeatability and Reproducibility
| Point | Trend |
|---|---|
| 1 | 0 |
| 1 | -9,59 |
| 1 | -5,84 |
| 1 | -5,59 |
| 1 | -9,66 |
| 1 | -1,52 |
| 1 | -1,1 |
| 1 | -8,29 |
| 1 | -3,31 |
| 1 | -3,28 |
| 1 | -5,22 |
| 1 | -8,15 |
| 1 | -0,5 |
| 1 | -6,82 |
| 1 | -6,89 |
| 1 | -0,64 |
| 1 | -6,82 |
| 1 | -1,08 |
| 1 | -6,03 |
| 1 | -7,77 |
| 1 | -3,14 |
| 1 | -2,22 |
| 1 | 0,14 |
| 1 | -1,13 |
| 1 | -0,54 |
| 2 | -0,56 |
| 2 | -3,48 |
| 2 | -3,26 |
| 2 | -8,38 |
| 2 | -3,78 |
| 2 | -2,9 |
| 2 | -9,79 |
| 2 | 0,1 |
| 2 | 0,08 |
| 2 | 0,09 |
| 2 | -6,01 |
| 2 | -2,28 |
| 2 | -3,29 |
| 2 | -9,23 |
| 2 | -0,94 |
| 2 | -4,27 |
| 2 | -2,41 |
| 2 | -1,78 |
| 2 | -6,63 |
| 2 | -3,81 |
| 2 | -4,33 |
| 2 | -3,42 |
| 2 | -6,28 |
| 2 | -4,57 |
| 2 | -4,33 |
| 3 | -2,57 |
| 3 | -2,44 |
| 3 | -4,01 |
| 3 | -1,31 |
| 3 | -3,79 |
| 3 | -3,96 |
| 3 | -4,63 |
| 3 | -3,82 |
| 3 | 0,66 |
| 3 | -10,18 |
| 3 | -5,94 |
| 3 | -4,62 |
| 3 | -1,51 |
| 3 | -4,48 |
| 3 | -1,29 |
| 3 | -0,64 |
| 3 | -10,14 |
| 3 | -4,8 |
| 3 | -6,12 |
| 3 | -2,14 |
| 3 | -7,2 |
| 3 | -5,34 |
| 3 | -2,56 |
| 3 | -4,43 |
| 3 | -0,95 |
Results of this example obtained with the Action software:
(image missing)
(image missing)
Some concluding suggestions
In the study, we obtained a RR= 2.9 Nm (corresponding to 29% of tolerance). The RR was considered satisfactory, as the system is not replicable, and this value is contaminated by product and standard variations.
11.5.8 Color measurement system
Consider the color measurement system applied to a car bumper (photo 5.8.1) on the scale of wavelength units. The color measurement system is a replicable dimensional measurement system, as we can perform several measurements on the part without damaging it.
Phase 1: Description of the Measurement System
We use this phase to carry out a review of the measurement system and planning for the application of the MSA. Here, it is important to define the team responsible for the analysis, the sectors that apply the measurement system and the need for measurement. For example, color checking ensures similarity with the car body. That’s why it’s important that we have a reliable measurement system.
(image missing)
Figure 5.8.1: Photo of the part.
For the color measurement feature, the customer specified a tolerance of ±0.5. So that the team can better understand the measurement system and its application, we carry out a scenario analysis using the following questions:
Importance/impact on the product?
Customer Complaints?
Failure History?
Is it a special feature?
Environmental Conditions?
Is the measurement system used for CEP?
Phase 2: Measurement Method
Phase I prepares us to make a good description of the measurement system. Therefore, we gathered the team to describe the measurement system and prepare the Ishikawa diagram.
Color measurement system description:
Lightly press the sensor on the surface of the painted part, wait for the measurement and record the values.
(image missing)
One of the critical items of a measurement system is the equipment used to carry out the measurements. The equipment is calibrated by appropriate laboratories and the calibration certificates must be critically evaluated against previously defined criteria. For this application, the colorimeter (X-Rite) has a resolution of 0.01 and is calibrated annually with a maximum allowable error of 0.1 L axis. This guarantees the required accuracy of 0.1.
(image missing)
Figure 5.8.2: Photo of color measurement.
Phase 3: Study Planning
- Stability: We selected 1 piece to be measured over time. In this application, we will make three rational subgroups of measurements per day for a week (20 subgroups).
- Trend: Not applicable, as it was not possible to obtain the reference value.
- Repeatability: 10 pieces from “distinct” production batches, with 3 measurements per piece carried out in random order.
- Reproducibility: We selected 3 operators to measure the 10 pieces.
- Linearity: Not applicable, as the measuring range is small.
After data collection, we will perform statistical analysis.
Stability
| Subgroup | Date | Schedule | Measure 1 | Measure 2 | Measure 3 |
|---|---|---|---|---|---|
| 1 | 14/08/2014 | 16:14 | 90,87 | 90,84 | 90,93 |
| 2 | 15/08/2014 | 00:44 | 90,96 | 90,80 | 90,94 |
| 3 | 15/08/2014 | 01:00 | 90,84 | 90,80 | 90,81 |
| 4 | 15/08/2014 | 05:45 | 90,91 | 90,97 | 90,85 |
| 5 | 15/08/2014 | 06:20 | 91,10 | 90,97 | 91,01 |
| 6 | 15/08/2014 | 15:05 | 90,75 | 90,99 | 90,94 |
| 7 | 15/08/2014 | 16:00 | 91,03 | 91,07 | 90,81 |
| 8 | 16/08/2014 | 00:44 | 90,84 | 90,96 | 90,80 |
| 9 | 16/08/2014 | 01:00 | 90,81 | 90,85 | 91,03 |
| 10 | 16/08/2014 | 06:10 | 91,00 | 90,87 | 91,05 |
| 11 | 16/08/2014 | 00:30 | 91,02 | 90,91 | 90,67 |
| 12 | 16/08/2014 | 05:30 | 90,97 | 91,04 | 90,63 |
| 13 | 18/08/2014 | 06:00 | 90,85 | 90,85 | 90,99 |
| 14 | 18/08/2014 | 15:38 | 90,86 | 91,02 | 90,83 |
| 15 | 18/08/2014 | 22:33 | 90,96 | 90,98 | 91,01 |
| 16 | 19/08/2014 | 00:58 | 90,93 | 90,78 | 90,82 |
| 17 | 19/08/2014 | 01:00 | 90,73 | 90,82 | 90,91 |
| 18 | 19/08/2014 | 05:50 | 90,88 | 90,94 | 90,95 |
| 19 | 19/08/2014 | 06:00 | 91,01 | 90,78 | 90,87 |
| 20 | 19/08/2014 | 15:20 | 91,00 | 90,97 | 90,70 |
Results of this example obtained with the Action software:
(image missing)
(image missing)
Some concluding suggestions
Stable measurement system over time.
RR - Repeatability and Reproducibility
| Ask | Operator | Measure |
|---|---|---|
| 1 | 1 | 91,250 |
| 2 | 1 | 91,160 |
| 3 | 1 | 91,010 |
| 4 | 1 | 91,260 |
| 5 | 1 | 91,320 |
| 6 | 1 | 91,160 |
| 7 | 1 | 91,480 |
| 8 | 1 | 91,240 |
| 9 | 1 | 91,290 |
| 10 | 1 | 91,300 |
| 1 | 2 | 91,350 |
| 2 | 2 | 91,190 |
| 3 | 2 | 91,100 |
| 4 | 2 | 91,280 |
| 5 | 2 | 91,380 |
| 6 | 2 | 91,150 |
| 7 | 2 | 91,510 |
| 8 | 2 | 91,290 |
| 9 | 2 | 91,240 |
| 10 | 2 | 91,330 |
| 1 | 3 | 91,240 |
| 2 | 3 | 91,160 |
| 3 | 3 | 90,980 |
| 4 | 3 | 91,250 |
| 5 | 3 | 91,320 |
| 6 | 3 | 91,190 |
| 7 | 3 | 91,510 |
| 8 | 3 | 91,260 |
| 9 | 3 | 91,320 |
| 10 | 3 | 91,310 |
| 1 | 1 | 91,320 |
| 2 | 1 | 91,110 |
| 3 | 1 | 91,050 |
| 4 | 1 | 91,310 |
| 5 | 1 | 91,330 |
| 6 | 1 | 91,120 |
| 7 | 1 | 91,460 |
| 8 | 1 | 91,180 |
| 9 | 1 | 91,330 |
| 10 | 1 | 91,260 |
| 1 | 2 | 91,280 |
| 2 | 2 | 91,160 |
| 3 | 2 | 91,040 |
| 4 | 2 | 91,250 |
| 5 | 2 | 91,410 |
| 6 | 2 | 91,110 |
| 7 | 2 | 91,460 |
| 8 | 2 | 91,250 |
| 9 | 2 | 91,270 |
| 10 | 2 | 91,360 |
| 1 | 3 | 91,190 |
| 2 | 3 | 91,170 |
| 3 | 3 | 91,000 |
| 4 | 3 | 91,200 |
| 5 | 3 | 91,310 |
| 6 | 3 | 91,120 |
| 7 | 3 | 91,470 |
| 8 | 3 | 91,240 |
| 9 | 3 | 91,370 |
| 10 | 3 | 91,300 |
| 1 | 1 | 91,280 |
| 2 | 1 | 91,070 |
| 3 | 1 | 90,960 |
| 4 | 1 | 91,310 |
| 5 | 1 | 91,280 |
| 6 | 1 | 91,150 |
| 7 | 1 | 91,430 |
| 8 | 1 | 91,270 |
| 9 | 1 | 91,350 |
| 10 | 1 | 91,230 |
| 1 | 2 | 91,270 |
| 2 | 2 | 91,100 |
| 3 | 2 | 91,020 |
| 4 | 2 | 91,230 |
| 5 | 2 | 91,360 |
| 6 | 2 | 91,160 |
| 7 | 2 | 91,460 |
| 8 | 2 | 91,270 |
| 9 | 2 | 91,320 |
| 10 | 2 | 91,280 |
| 1 | 3 | 91,280 |
| 2 | 3 | 91,120 |
| 3 | 3 | 91,060 |
| 4 | 3 | 91,220 |
| 5 | 3 | 91,360 |
| 6 | 3 | 91,150 |
| 7 | 3 | 91,570 |
| 8 | 3 | 91,220 |
| 9 | 3 | 91,310 |
| 10 | 3 | 91,300 |
Results of this example obtained with the Action software:
(image missing)
(image missing)
(image missing)
Some concluding suggestions
We analyzed the R&R and it presents a rate of around (24%), with an RR of 0.04 in color units. We note from the ANOVA graph and table that the interaction between part and operator is significant at the 5% significance level (p-value of 0.025), a fact not detected for operators (p-value of 0.28). Therefore, the variability associated with reproducibility is due to the interaction between part and operator. Graphically, we noticed good discrimination and repeatability remains consistent between pieces. Furthermore, the colorimeter has an uncertainty of 0.07 (calibration certificate), consequently the RR is below the accuracy guaranteed by the equipment. Therefore, although we obtained an RR rate of 24%, the RR obtained is acceptable from a practical point of view. Therefore, we consider the variability of the measurement system is appropriate to the process.
11.5.9 Tensile strength
Tensile Strength Measuring System
Consider the tensile strength measurement system on the kgf scale (figure 5.9.1). When measuring the tensile strength of a piece of wire (test specimen, figure 5.9.2) it breaks at the end of the test. Consequently we cannot repeat the measurement on the same piece of wire. In this case, the tensile strength measurement system is not replicable.
(image missing)
Figure 5.9.1: Katros machine for measuring traction.
(image missing)
Figure 5.9.2: Coil of yarn that will be used in the tensile strength study.
Phase 1: Description of the measurement system
For the tensile strength characteristic, we have the Traction Machine (L-109) (figure 5.9.1) with a resolution of 0.01 kgf. For this application, we have an accuracy of 0.125 kgf and the minimum for acceptance of the product is 30 kgf (Unilateral tolerance).
Furthermore, some characteristics are important to define, such as:
- The need for measurement is to guarantee the durability and resistance of the disc;
- The importance/impact on the product is that it does not meet the specified minimum resistance, this can impact the fragility of the disc;
- For the product in question, we have no customer complaints or failure history;
- The measurement system is not used for CEP;
- The measuring system is used for product control, so we will compare the measurement system variation with the product tolerance.
Phase 2: Measurement method
At this stage, we bring together the team to describe the measurement system and prepare the Ishikawa diagram.
The description of the tensile strength measurement system applied with the Traction Machine (L-109) (figure 5.9.1) is presented below:
- Connect the wire to the machine;
- Apply traction to the thread (with the L-109 machine);
- Record the traction value obtained on the L-109 machine.
Another important point for analyzing the measurement system is the Ishikawa diagram, or cause and effect diagram (figure 5.9.3). With the entire team together (at this stage it is important to have operators present), we brainstormed on the main causes of variation in the measurement system. The Ishikawa diagram is essential for defining the critical points and statistical techniques that we will use to evaluate the measurement system.
(image missing)
Figure 5.9.3: Cause and effect diagram (Ishiakawa)
Phase 3: Study Planning
We will apply the stability study, as well as the repeatability and reproducibility study. As the measurement system is non-replicable, we will use the following strategy:
- Stability: We will select 30 samples (pieces of yarn) of which 15 will be measured for tensile strength on the same day and the other 15 will be measured over 15 days (1 per day).
- Repeatability: We select 3 coils (from the same batch) and cut 15 samples per coil.
- Reproducibility: We select 3 operators who will each measure 15 samples from each coil (from the same batch).
Stability
We present below the data for the stability study:
| Short term | Short term | Long term | Long term |
|---|---|---|---|
| Race | Measure | Race | Measure |
| 1 | 59,80 | 1 | 62,95 |
| 2 | 60,75 | 2 | 65,25 |
| 3 | 56,05 | 3 | 65,05 |
| 4 | 61,65 | 4 | 64,85 |
| 5 | 56,90 | 5 | 64,40 |
| 6 | 62,55 | 6 | 64,75 |
| 7 | 57,50 | 7 | 62,20 |
| 8 | 59,80 | 8 | 60,30 |
| 9 | 62,90 | 9 | 64,20 |
| 10 | 65,70 | 10 | 65,75 |
| 11 | 61,35 | 11 | 62,40 |
| 12 | 63,70 | 12 | 59,00 |
| 13 | 62,45 | 13 | 59,50 |
| 14 | 62,95 | 14 | 57,55 |
| 15 | 61,80 | 15 | 61,40 |
Results of this example obtained with the Action software:
Short term:
(image missing)
(image missing)
Long term:
(image missing)
(image missing)
Some concluding suggestions
Stable system over time. All points are within the control lines.
Repeatability and reproducibility
The data used for the repeatability and reproducibility study are presented below:
| Operator 1 | Operator 1 | Operator 2 | Operator 2 | Operator 3 | Operator 3 |
|---|---|---|---|---|---|
| Coil | Measurements | Coil | Measurements | Coil | Measurements |
| 1 | 66,70 | 1 | 66,25 | 1 | 63,25 |
| 1 | 55,70 | 1 | 63,50 | 1 | 59,85 |
| 1 | 62,20 | 1 | 66,25 | 1 | 65,00 |
| 1 | 67,50 | 1 | 63,55 | 1 | 62,10 |
| 1 | 66,30 | 1 | 66,55 | 1 | 63,20 |
| 1 | 68,15 | 1 | 66,15 | 1 | 64,30 |
| 1 | 58,75 | 1 | 51,55 | 1 | 64,35 |
| 1 | 65,10 | 1 | 64,60 | 1 | 67,15 |
| 1 | 58,60 | 1 | 66,15 | 1 | 62,20 |
| 1 | 63,75 | 1 | 63,75 | 1 | 60,95 |
| 1 | 65,50 | 1 | 64,15 | 1 | 59,75 |
| 1 | 63,00 | 1 | 67,85 | 1 | 62,45 |
| 1 | 53,50 | 1 | 64,20 | 1 | 67,35 |
| 1 | 63,20 | 1 | 58,90 | 1 | 66,05 |
| 1 | 64,25 | 1 | 60,20 | 1 | 62,45 |
| 2 | 61,35 | 2 | 61,80 | 2 | 59,95 |
| 2 | 54,05 | 2 | 64,15 | 2 | 61,75 |
| 2 | 63,30 | 2 | 67,55 | 2 | 57,20 |
| 2 | 53,90 | 2 | 57,00 | 2 | 61,75 |
| 2 | 54,70 | 2 | 57,20 | 2 | 58,70 |
| 2 | 54,75 | 2 | 56,55 | 2 | 55,35 |
| 2 | 56,65 | 2 | 63,35 | 2 | 53,45 |
| 2 | 59,90 | 2 | 60,50 | 2 | 53,50 |
| 2 | 58,25 | 2 | 67,55 | 2 | 66,10 |
| 2 | 53,25 | 2 | 54,50 | 2 | 62,40 |
| 2 | 56,25 | 2 | 66,40 | 2 | 63,50 |
| 2 | 66,75 | 2 | 61,30 | 2 | 65,00 |
| 2 | 58,90 | 2 | 58,15 | 2 | 59,65 |
| 2 | 64,80 | 2 | 52,20 | 2 | 54,90 |
| 2 | 60,80 | 2 | 60,90 | 2 | 62,55 |
| 3 | 64,55 | 3 | 67,40 | 3 | 60,20 |
| 3 | 64,15 | 3 | 69,05 | 3 | 60,15 |
| 3 | 55,80 | 3 | 63,75 | 3 | 60,20 |
| 3 | 59,10 | 3 | 65,05 | 3 | 61,35 |
| 3 | 61,95 | 3 | 63,70 | 3 | 60,60 |
| 3 | 62,40 | 3 | 63,60 | 3 | 55,15 |
| 3 | 61,20 | 3 | 63,35 | 3 | 60,00 |
| 3 | 61,80 | 3 | 65,25 | 3 | 60,50 |
| 3 | 65,00 | 3 | 59,65 | 3 | 58,55 |
| 3 | 65,05 | 3 | 64,90 | 3 | 58,05 |
| 3 | 59,20 | 3 | 68,10 | 3 | 59,95 |
| 3 | 60,90 | 3 | 61,90 | 3 | 60,25 |
| 3 | 64,50 | 3 | 64,90 | 3 | 60,60 |
| 3 | 55,55 | 3 | 55,95 | 3 | 62,00 |
| 3 | 63,75 | 3 | 56,95 | 3 | 58,30 |
The results of this example obtained with the Action software:
(image missing)
(image missing)
(image missing)
(image missing)
(image missing)
(image missing)
(image missing)
(image missing)
Therefore, the minimum process limit is given by 41.14. Therefore, the chance of producing below 30 is less than 0.135%.
The following table shows the chance of producing below 30 for 10 different reading points:

(image missing)
11.6 Appendix
This section contains the tables used throughout the statistical content of the MSA 4th edition.
11.6.1 t-Student Distribution
Values $ t_c $ such that $ P(-t_c< t_n< t_c)1-\alpha $
| GL\ $\alpha$ | 0,20 | 0,10 | 0,05 | 0,02 |
|---|---|---|---|---|
| 1 | 3,078 | 6,314 | 12,706 | 31,821 |
| 2 | 1,886 | 2,920 | 4,303 | 6,965 |
| 3 | 1,638 | 2,353 | 3,182 | 4,541 |
| 4 | 1,533 | 2,132 | 2,776 | 3,747 |
| 5 | 1,576 | 2,015 | 2,571 | 3,365 |
| 6 | 1,440 | 1,943 | 2,447 | 3,143 |
| 7 | 1,415 | 1,895 | 2,365 | 2,998 |
| 8 | 1,397 | 1,860 | 2,306 | 2,896 |
| 9 | 1,383 | 1,833 | 2,262 | 2,821 |
| 10 | 1,372 | 1,812 | 2,228 | 2,764 |
| 11 | 1,363 | 1,796 | 2,201 | 2,718 |
| 12 | 1,356 | 1,782 | 2,179 | 2,681 |
| 13 | 1,350 | 1,771 | 2,160 | 2,650 |
| 14 | 1,345 | 1,761 | 2,145 | 2,624 |
| 15 | 1,341 | 1,753 | 2,131 | 2,602 |
| 16 | 1,337 | 1,745 | 2,120 | 2,583 |
| 17 | 1,333 | 1,740 | 2,110 | 2,567 |
| 18 | 1,330 | 1,734 | 2,101 | 2,552 |
| 19 | 1,328 | 1,729 | 2,093 | 2,539 |
| 20 | 1,325 | 1,725 | 2,086 | 2,528 |
| 21 | 1,323 | 1,721 | 2,080 | 2,518 |
| 22 | 1,321 | 1,717 | 2,074 | 2,508 |
| 23 | 1,319 | 1,714 | 2,069 | 2,500 |
| 24 | 1,318 | 1,711 | 2,064 | 2,492 |
| 25 | 1,316 | 1,708 | 2,060 | 2,485 |
| 26 | 1,315 | 1,706 | 2,056 | 2,479 |
| 27 | 1,314 | 1,703 | 2,052 | 2,473 |
| 28 | 1,313 | 1,701 | 2,048 | 2,467 |
| 29 | 1,311 | 1,699 | 2,045 | 2,462 |
| 30 | 1,310 | 1,697 | 2,042 | 2,457 |
| 40 | 1,303 | 1,684 | 2,021 | 2,423 |
| 60 | 1,296 | 1,671 | 2,000 | 2,390 |
| 120 | 1,289 | 1,658 | 1,980 | 2,358 |
| bigger | 1,282 | 1,645 | 1,960 | 2,326 |
11.6.2 d2 table

Note: The first row of each cell corresponds to the value of $ d_2 $ and the second line to the value of the degrees of freedom ($ \nu $). The line corresponding to the value > 15 corresponds to the $ d^i_2. $ The value of m corresponding to the number of readings, while the value of g corresponds to the number of amplitudes.
References
[1] BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML, Guide to the Expression of Uncertainty in Measurement, 1995.
[2] NBR ISO 10012, Measurement Management Systems - Requirements for measurement processes and measurement equipment, 04/2004.
[3] MSA, Measurement Systems Analysis, fourth edition.
[4] Brazilian version of reference document EA-4/02: Expression of measurement uncertainty in calibration, INMETRO.
[5] Montgomery, Douglas C., Design and Analysis of Experiments - Fourth Edition - Ed. John Wiley & Sons, 1997.
[6] ISO GUM (2008), Measurement data evaluation - Guide for expressing measurement uncertainty, 1st Brazilian edition.