Measurement Uncertainty
On this page

Book

18 Measurement Uncertainty

148 min read

When reporting the result of measuring a physical quantity, it is mandatory that some quantitative indication of the quality of the result is given, so that those who use the measurement result can evaluate its reliability. Without this indication, measurement results cannot be compared, either among themselves or with reference values ​​provided in a specification or standard. Therefore, it is necessary to have an implemented procedure that is easy to understand and generally accepted to characterize the quality of the measurement result, that is, to evaluate and express its uncertainty.

The concept of measurement uncertainty will be used as a quantifiable attribute to determine the quality of a measurement system. In this module, we will discuss in detail the concept of measurement uncertainty and present various applications in the calibration of measuring equipment and in testing. Our text was based on the ISO GUM (2008) standard.

18.1 Fundamentals of Calculating Uncertainty in Measurement

Measurement

The objective of a measurement is to determine the value of the measurand, that is, the value of the specific quantity to be measured. Measurement begins with an appropriate specification of the quantity, measurement method and procedure.

In general, the result of a measurement is an approximation or estimate of the value of the quantity. Therefore, the measurement result is only complete if it is accompanied by the uncertainty of the estimate.

In practice, the specification or definition of the quantity is a consequence of the desired accuracy. To meet the required accuracy, the quantity must be specified in such a way that it has a single value for the associated practical purposes.

Example 1.1

Consider a 75mm rod where the required accuracy is micrometers. In this case, your specification must include temperature and pressure. On the other hand, if the length of the rod must be determined in millimeters, its specification does not require the definition of temperature and pressure.

In the vast majority of cases, the measurement result is determined through a series of readings obtained under repetitive conditions. Variations obtained in repeated readings are a consequence of factors that affect the reading results.

Furthermore, the mathematical model of the measurement, which transforms repeated readings into the measurement result, is critical as it includes factors that are not fully known. Thus, the variation obtained in repeated readings and the lack of information from the mathematical model contribute to the uncertainty of the measurement result.

Measurement

It is the set of operations aimed at determining the value of a quantity. These operations can be performed automatically.

Measure

It is an experimental process by which the momentary value of a physical quantity (quantity to be measured) is determined as a multiple and/or a fraction of a unit, established by a standard, and recognized internationally.

Measurement Uncertainty

Parameter associated with the measurement result, which characterizes the dispersion of values that can be attributed to the measurand.

18.1.1 Errors, Effects and Corrections

In general, a measurement has imperfections that give rise to an error in the measurement result. Traditionally, error is viewed as having two components, a random component and a systematic component. It is important to make it clear that error is an idealized concept and errors cannot be known exactly. A detailed discussion of measurement error can be found in Appendix D of the ISO GUM (2008) standard.

Random error originates from temporal or spatial variations and occurs unpredictably. The effects of such variations (from now on we will call random effects) are the cause of variations in repeated observations of the quantity. Although it is not possible to compensate for the random error of a measurement result, it can generally be reduced by increasing the number of observations. Your hope or expected value is zero.

Systematic error, like random error, cannot be eliminated, but it can often be reduced. If a systematic error originates from a recognized effect of an influencing quantity on a measurement result, hereinafter referred to as a systematic effect, the effect can be quantified and, if it is significant with respect to the required accuracy of the measurement, a correction or correction factor can be applied to compensate for the effect. Suppose that, after this correction, the expectation of the probability distribution associated with the systematic error is zero.

As we said, error is an idealized concept and errors cannot be known exactly. In practice, we associate a random variable (for example, the normal distribution) to represent the measurement error, as can be seen in Figure 1.1.1. Other probability distributions can also be used, for example, we associate the error due to the resolution of the measuring instrument with the rectangular (or uniform) distribution.

FIG_MSP_024-560x300

Figure 1.1.1: Representation of measurement error.

We often use the sample standard deviation of the mean of a series of observations to study the behavior of random error. However, the sample standard deviation from the mean is not the random error, although it is designated as such in some publications. It is, instead, a measure of uncertainty in the mean due to random effects. The exact value of the mean error, which arises from these effects, cannot be known. We must be very careful to distinguish between the terms “error” and “uncertainty”. They are not synonymous, on the contrary, they represent completely different concepts.

The uncertainty of a correction applied to a measurement result, to compensate for a systematic effect, is not systematic error. It is, rather, a measure of uncertainty of the result due to incomplete knowledge of the required value of the correction. The error arising from imperfect compensation of a systematic effect cannot be exactly known. The terms “error” and “uncertainty” should be used appropriately and care should be taken to distinguish one from the other. We assume that the result of a measurement has been corrected for all systematic effects recognized as significant and that every effort has been made to identify such effects.

Example 1.1.1

A correction due to the finite impedance of a voltmeter used to measure a potential difference (the quantity) across a high impedance resistor is applied to reduce the systematic effect on the measurement result arising from the charging effect of the voltmeter. However, the voltmeter and resistor impedance values, which are used to estimate the correction value and are obtained from other measurements, are uncertain. These uncertainties are used to evaluate the uncertainty component of the potential difference determination arising from the correction and thus the systematic effect due to the finite impedance of the voltmeter.

Often, measurement instruments and systems are adjusted based on measurement standards and reference materials to eliminate systematic effects. However, the uncertainties associated with these standards and materials must still be taken into consideration.

18.1.2 Measurement Uncertainty

The uncertainty of a measurement result reflects the lack of knowledge associated with the value of the quantity to be measured. The result of a measurement, even after correction due to the effects related to recognized systematic errors, is only an estimate of the value of the quantity due to the uncertainty arising from the effects of random errors and the imperfect correction of the result for effects due to systematic errors.

The result of a measurement (after correction) may, without realizing it, be very close to the value of the quantity (and thus have a negligible error), although it may have a large uncertainty. Therefore, the uncertainty of a measurement result should not be confused with the remaining unknown error.

In practice, there are many possible sources of uncertainty in a measurement, including:

a) Incomplete definition of magnitude;

b) Flaws in the definition of magnitude;

c) Non-representative sampling - The measured sample may not represent the defined quantity;

d) Inadequate knowledge of the effects of environmental conditions on measurement or imperfect measurement of environmental conditions;

e) Personal bias error when reading analogue instruments;

f) Finite resolution of the instrument or mobility threshold;

g) Inaccurate values ​​of measurement standards and reference materials;

h) Inaccurate values ​​of constants and other parameters obtained from external sources and used in the algorithm to obtain data;

i) Approximations and assumptions incorporated into the measurement method and procedure;

j) Variations in repeated observations of the quantity under apparently identical conditions.

These sources are not necessarily independent and some of the sources from (a) to (i) may contribute to source (j). Naturally, an unrecognized systematic effect cannot be taken into account when assessing the uncertainty of a measurement result, but it contributes to its error.

In some publications, the components of uncertainty are categorized as “random” and “systematic” and are associated with errors arising from random effects and known systematic effects, respectively. Such categorization of uncertainty components can become ambiguous when applied generically. For example, a “random” component of uncertainty in one measurement may become a “systematic” component of uncertainty in another measurement in which the result of the first measurement is used as input. By categorizing the methods of evaluating (or, calculating) the components of uncertainty, instead of doing so with the components themselves, we avoid such ambiguity. At the same time, this does not prevent designating individual components that have been evaluated by the two different methods into distinct groups, to be used for a particular purpose.

The recommendation of the ISO GUM (2008) standard consists of dividing the uncertainty components into two types, called “A” and “B”. These categories are applied to the uncertainty assessment method and have no relation to the words random or systematic. The uncertainty associated with the correction of an effect related to a systematic error can be obtained by a type A assessment or by a type B assessment.

The purpose of the Type A and Type B classification is to indicate the two different ways of evaluating the components of uncertainty and is for discussion only. The classification is not intended to indicate that there is any difference in the nature of the components. Both types of assessment are based on probability distributions and the uncertainty components resulting from each type are quantified by variances or standard deviations.

The estimated variance $ u^2 $, characterizing an uncertainty component obtained from a Type A assessment, is calculated from a series of repeated observations, through the sampling variance of the mean of the measurements $ s^2/n $. The estimated standard deviation $ u=s/\sqrt{n} $ is called Type A standard uncertainty. For an uncertainty component obtained by a Type B assessment, the estimated variance $ u^2 $ is evaluated using available knowledge, and the estimated standard deviation $ u $ is sometimes called Type B standard uncertainty.

Thus, a Type A standard uncertainty is obtained from a probability density function derived from the observation of a frequency distribution, while a Type B standard uncertainty is obtained from a supposed probability density function, based on the degree of credibility that an event will occur (often called subjective probability). Both approaches employ recognized interpretations of probability. A Type B assessment of an uncertainty component is usually based on a set of comparatively reliable information.

The standard uncertainty of the result of a measurement, when this result is obtained from values of other quantities, is called combined standard uncertainty and designated by $ u_c $. It is the estimated standard deviation, associated with the result, obtained from all variance and covariance components, regardless of how they were evaluated, using what is called the law of uncertainty propagation.

To meet the needs of some industrial and commercial applications, as well as health and safety requirements, an expanded uncertainty $ U $ is obtained by multiplying the combined standard uncertainty $ u_c $ by a coverage factor $ k $. The intended purpose for $ U $ is to provide a range around the result of a measurement that is expected to cover a large fraction of the distribution of values that could reasonably be attributed to the quantity. The choice of $ k $, which is generally in the range of 2 to 3, is based on the probability of coverage or required confidence level of the interval.

The coverage factor $ k $ should always be stated so that the standard uncertainty of the measured quantity can be recovered for use in calculating the combined standard uncertainty of other measurement results that may depend on that quantity.

If there is variation in all the quantities on which the result of a measurement depends, its uncertainty can be calculated by statistical means. However, since this is rarely possible in practice due to limited time and resources, the uncertainty of a measurement result is generally assessed when using a mathematical model of the measurement and the law of propagation of uncertainty. Thus, the assumption is implicit that a measurement can be modeled mathematically to the degree imposed by the accuracy required in the measurement.

Since the mathematical model may be incomplete, all relevant quantities must be varied to the greatest practical extent possible, so that the assessment of uncertainty can be based, as far as possible, on the observed data.

Whenever feasible, the use of empirical measurement models, based on quantitative data collected over time, and the use of verification standards and control charts that can indicate whether a measurement is under statistical control, should be part of the effort to obtain reliable uncertainty assessments. The mathematical model should always be revised when observed data, including the result of independent determinations of the same quantity, demonstrate that the model is incomplete. A well-designed experiment facilitates reliable assessments of uncertainty and is an important part of the art of measurement.

In order to decide whether a measurement system is working properly, the experimentally observed variability of its output values, as measured by its observed standard deviation, is often compared with the predicted standard deviation, obtained through the combination of the various components of uncertainty that characterize the measurement. In such cases, only those components (obtained from Type A or Type B assessments) that could contribute to the experimentally observed variability of these output values ​​should be considered.

Such analysis can be facilitated by bringing together those components that contribute to variability and those that do not into two separate and appropriately labeled groups.

In some cases, the uncertainty of a correction for a systematic effect does not need to be included in the assessment of the uncertainty of a measurement result. Although uncertainty has been evaluated, it can be ignored if its contribution to the combined standard uncertainty of a measurement result is negligible. If the value of the correction itself is insignificant relative to the combined standard uncertainty, it can also be ignored.

It often happens in practice, especially in the field of legal metrology, that equipment is tested through a comparison with a measurement standard and the uncertainties associated with the standard and the comparison procedure are negligible in relation to the required accuracy of the test. An example is the use of a set of well-calibrated mass standards to verify the accuracy of a commercial balance. In such cases, because the uncertainty components are small enough to be ignored, the measurement can be viewed as determining the error of the equipment under test.

Example 1.2.1

A high quality Zener voltage standard is calibrated by comparison with a Josephson effect voltage reference based on the conventional value of the Josephson constant recommended for international use by the CIPM. The relative combined standard uncertainty $ u_c(V_S)/V_S $ of the calibrated potential difference $ V_S $ is reported in terms of the conventional value, but $ u_c(V_S)/V_S $ is $ 4\times10^{-7} $ when $ V_S $ is reported in terms of the SI unit of potential difference, volt($ V $), because of the additional uncertainty associated with the SI value of the Josephson constant.

Gross errors in recording or analyzing data can introduce significant unknown error into the result of a measurement. Large gross errors can usually be identified by an appropriate review of the data. Small, gross errors may be masked by random variations, or may even appear as such. Uncertainty measures are not designed to take such errors into account.

The assessment of uncertainty is neither a routine task nor a purely mathematical task, it depends on detailed knowledge of the nature of the quantity and the measurement. The quality and usefulness of the uncertainty indicated for the result of a measurement therefore depend, ultimately, on the understanding, critical analysis and integrity of those who contribute to establishing its value.

Measurement result

We find the expression of an incomplete measurement result if it is not presented with the statement of the associated measurement uncertainty. The uncertainty of a result defines a range of values ​​around the average of the measurements, within which the true value of the quantity lies with an established level of confidence.

$${Resultado} = {Média~(das~medidas)} - {Erro~Sistemático} \pm {IM~(Incerteza)}$$

Although it is not yet generally understood and even sometimes unknown to some, we must note that among the portions shown in the expression of the result of a measurement, the MI (measurement uncertainty) is the most important, even than the average (of the measurements) and would deserve greater understanding and application.

Let’s look at an example in which a metrologist was asked to measure the dimensions of his metrology laboratory to prepare a layout, and he did not have a tape measure or any other means of measurement. In this we could use the standardized dimensions of the floor plates (for example Paviflex, 30 $ \times $ 30 cm) and after counting the number of plates on each side output a measurement result as follows: 4.0 $ \times $ 4.0 m $ \pm $ 0.15m.

Metrologically speaking, the result of your measurement is correct even if the requester was not satisfied with the IM presented and in this case he could propose a change in the measurement procedure used, such as the use of a tape measure.

From the same point of view, it would be wrong if the measurement were made, for example, with a measuring tape and the result presented was: 4.010 $ \times $ 4,047 m (without IM declaration).

Flow for Uncertainty Calculation

(image missing)

18.1.3 Assessment of Standard Uncertainty

In many cases, a quantity $ \mathbf{y} $ is not measured directly, but is determined as a function of $ n $ other quantities $ \mathbf{x_1, x_2,\dots, x_n} $, through a functional relationship $ \mathbf{f} $, which becomes

$$\mathbf{y = f (x_1, x_2,\dots, x_n)}.$$

The input quantities $ \mathbf{x_1, x_2,\dots, x_n} $, over which the output value $ \mathbf{y} $ depends, may be a measure or depend on other variables, including corrections and correction factors for systematic effects. The function $ \mathbf{f} $ can be determined experimentally, or exists only, as an algorithm that can be evaluated numerically.

The input quantities $ \mathbf{x_1, x_2,\dots, x_n} $ can be characterized as:

  • Quantity whose values and uncertainties are determined directly from the measurement. These values ​​and uncertainties can be obtained from a simple observation, repeated observations or judgments based on experience.

They may also involve determining corrections for instrument indication and corrections for influence quantities, such as: ambient temperature, barometric pressure and humidity;

  • Values and uncertainties, which are driven by a measurement of external sources, such as:

quantities associated with calibration of standards, certificates of reference materials and reference information obtained through manuals.

Example 1.3.1

To measure the volume, we can use the following method

$$Vol=\frac{Massa}{Densidade}$$

, where the quantity volume is obtained through the quantities mass and density.

The standard deviation estimate, associated with each input estimate $ \mathbf{x_i} $ is called \textbf{standard uncertainty} and indicated by $ \mathbf{u(x_i)} $.

The estimate of the standard deviation, associated with the estimate of the measurement result $ \mathbf{y} $, is called combined standard uncertainty and indicated by $ \mathbf{u_c(y)} $, and is determined by combining the standard uncertainties associated with the input estimates ($ \mathbf{x_i} $). Each input estimate $ \mathbf{x_i} $ and its associated uncertainty $ \mathbf{u(x_i)} $ are obtained by distributing the values of an input quantity ($ \mathbf{x_i} $).

The “Type A” assessment of measurement uncertainty is based on frequency distribution, while the “Type B” assessment is based on available information on the variability of the input quantity ($ \mathbf{x}_\mathbf{i} $).

Example 1.3.2

(NIS 3003, 1995) Calibration of a standard mass with a nominal value of 10kg of class M1, using a comparator. In this case, we obtain the unknown mass equation $ W_{X} $, by

$$W_{X} = W_{S} + D_{S} + \delta C + Ab + \Delta W.$$

In practice we do not apply corrections for this mass class and the comparator has unknown linearity. However, we associate uncertainties with these contributions.

Symbol Source of Uncertainty Type Limits Average
$ W_{S} $ Standard dough B ± 30mg (k=2) 10kg
$ D_{S} $ Standard mass drift B ± 15mg 0
$ \delta C $ Comparator linearity B ± 10mg 0
$ Ab $ Effect of air B ± 10mg 0
$ \Delta W $ Repetitiveness A

Example 1.3.3

Determine the uncertainty of the area of a circle whose diameter was measured experimentally using a measuring system called a caliper.

Diameter value obtained with the caliper with a resolution of 0.01 mm and expanded uncertainty U= 0.02 mm (k = 2):

Readings Diameter
1 10,28
2 10,26
3 10,28
4 10,3
5 10,28

The expression for calculating the area is given by

$$y=\frac{1}{4}\pi d^2,$$

Example 1.3.4

Determine the measurement uncertainty in the composition of two gauge blocks, which were measured by the same measuring system.

  • Block 1

Nominal dimension: 10 mm.

Expanded Uncertainty: $ u(x_1)=0,0077~mm $ for $ k = 2 $.

  • Block 2

Nominal dimension: 20 mm.

Expanded Uncertainty: $ u(x_2)=0,084~mm $ for $ k = 2 $.

The result of combining blocks can be expressed mathematically by

$$\mathbf{y = x_1 + x_2}$$

18.1.4 Type A Uncertainty

In the vast majority of cases, the best estimate for the expected value of a quantity that varies randomly and for which we have n independent readings k obtained under conditions of repeatability, corresponds to the arithmetic mean.

Thus, when estimating an input quantity $ x_i $ has been obtained from n measurements under repeatability conditions, the standard uncertainty $ \mathbf{u(x_i)} $ is obtained by estimating the variance of the mean. This is given by

$$s_{\overline{X}}=\frac{s}{\sqrt{n}},$$

where n number of measurements and s standard deviation corresponding to the n readings.

Example 1.4.1

Returning to Example 1.3.2. Considering the standard mass calibration process in the previous example, the evaluator carried out five measurements of the difference between the standard mass and the unknown mass. The results are below.

Reading 1 15mg
Reading 2 25mg
Reading 3 20 mg
Reading 4 13 mg
Reading 5 18mg
Average 18.20mg
Standard Deviation 4.66mg
Standard Deviation from the Mean 2.08 mg

Calculating Type A Uncertainty, we obtain:

$$ u(\varepsilon)=\frac{s}{\sqrt{n}}=\frac{4,66}{\sqrt{5}}=2,08~ mg $$

Example 1.4.1

Returning to Example 1.3.3.

Readings Diameter Area
1 10,28 82,99963
2 10,26 82,67699
3 10,28 82,99963
4 10,3 83,3229
5 10,28 82,99963
Average Readings 10,28 82,99976
Standard Deviation of Readings 0,014142 0,228364
Standard Deviation of Average Readings 0,006325 0,102128

For the Area quantity, Type A Uncertainty is given by:

$$ u(\varepsilon)=\frac{s}{\sqrt{n}}=\frac{0,228364}{\sqrt{5}}=0,102128~ mm^2 $$

18.1.5 Type B Uncertainty

For an estimate of an input quantity $ \mathbf{x_i} $, which was not obtained from repeated observations, the estimated variance $ \mathbf{u^{2}(x_i)} $ or the standard uncertainty $ \mathbf{u(x_i)} $ is evaluated by specific judgment based on all available information on the variability of $ \mathbf{x_i} $. In all this information, we include:

a) Previous measurement information;

b) Experience or general knowledge of the behavior and properties of relevant instruments and materials;

c) Manufacturer’s specifications;

d) Information from calibration certificates and other specifications;

e) Uncertainty conveyed by reference information obtained from manuals.

For convenience, $ \mathbf{u^{2}(x_i)} $ and $ \mathbf{u(x_i)} $ evaluated in this way are called Type B Variance and Type B Standard Uncertainty, respectively.

The purpose of using various available information for the assessment of Type B standard uncertainty is to seek insight based on experience and general knowledge, and is a skill that can be obtained with practice. It is recognized that a Type B uncertainty assessment can be as reliable as a Type A assessment, especially in the situation where the Type A assessment is based on the comparison of small numbers of statistically independent observations (ISO GUM, 2008).

The following are 4 available assumptions for the influence input quantities $ \mathbf{x_i} $, for the assessment of Type B standard uncertainty.

Case 1

If the estimate $ \mathbf{x_i} $ is taken from the manufacturer’s specification, calibration certificates, manuals or other sources, its standard uncertainty $ \mathbf{u(x_i)} $ it is simply the quoted value divided by the multiplier.

Example 1.5.1

Returning to example 1.3.2, the calibration certificate states that the standard mass with a nominal value of 10kg of class M1 has the uncertainty $ 30~ mg $ for the confidence level with k = 2.

The standard mass uncertainty is then

$$\mathbf{u(W_S)= \frac{30~mg}{2} =15~mg}.$$

The uncertainty of $ \mathbf{x_i} $ it is not necessarily reported as a multiple of a standard deviation, as discussed above. Instead, we may find a statement that the stated uncertainty has a 90%, 95%, or 99% confidence level.

Unless otherwise stated, you can assume that a normal distribution (or t-Student) will be used to calculate the stated uncertainty, and the standard uncertainty $ \mathbf{u(x_i)} $ can be found by dividing the declared uncertainty by a factor k, appropriate from the normal distribution.

Example 1.5.2

A calibration certificate states that the resistance of a standard resistor $ R_S $ of nominal value $ 10\Omega $ is $ 10,000742 \Omega $ at 23ºC and with uncertainty of $ 129 \mu\Omega $, defining an interval with a significance level of 99%.

The standard uncertainty is given by

$$u(R_S)= \frac{129}{2,58} = 50~\mu \Omega.$$

In this case, we use the Normal distribution table to determine the value of k.

Example 1.5.3

Returning to example 1.3.4.

The standard uncertainty we obtain $ u(x_i) $ of each block is obtained by dividing the expanded uncertainty by the factor k. So

$$u(x_1) = 0,077 / 2 =0,0385$$

$$u(x_2) = 0,084 / 2 =0,042$$

Example 1.5.4

Returning to example 1.3.3

The expanded uncertainty of the caliper, defined in its calibration certificate, is 0.02 with coverage factor k=2. In this way, the uncertainty inherited from the equipment is

$$u(d)=\frac{0,02}{2}=0,01~mm$$

Case 2

In some cases, it may be possible to estimate only the limits (upper limit $ \mathbf{a_+} $ and lower $ \mathbf{a_-} $) to $ \mathbf{x_i} $, for example, when the influencing quantity is the temperature variation. In this case, we consider that the probability that the value of $ \mathbf{x_i} $ is within the range $ \mathbf{a_-} $ until $ \mathbf{a_+} $, for all practical purposes, is equal to 1 and the probability that $ \mathbf{x_i} $ is outside this range is essentially zero. If there is no knowledge

specific about the possibility of the value $ \mathbf{x_i} $ is within the range, one can only admit that it is equally likely to be found everywhere within the range (a uniform or rectangular distribution).

Logo $ \mathbf{x_i} $ is the midpoint of the range, where:

$ \mathbf{x_i = \frac{(a_-+ a_+)}{2}} $, whose associated variance is given by

$$\mathbf{u^2(x_i) = \frac{(a_+-a_-)^2}{12}}.$$

If the difference between the limits $ \mathbf{(a_+ - a_-)} $, is represented by $ \mathbf{2a} $, that is, the limits are symmetric, then the equation for variance will be

$$\mathbf{u^2(x_i) = \frac{a^2}{3}}.$$

retangulo

$$u=\frac{base}{2\sqrt{3}}=\frac{2a}{2\sqrt{3}}=\frac{a}{\sqrt{3}}.$$

Example 1.5.5

A manual establishes that the value of the linear coefficient of thermal expansion of a standard steel block is determined by $ \alpha_S = 11,5 \times 10^{-6}~^{\circ} C^{-1} $ and that the “error” in this value must not exceed $ 2 \times 10^{-6}~^{\circ} C^{-1} $. Based on this limited information, it is reasonable to assume that the

coefficient of thermal expansion belongs to the range: $ 9,5 \times10^{-6}~^{\circ} C^{-1} $ until $ 13,5 \times 10^{-6}~^{\circ} C^{-1} $, with probability 1. The standard uncertainty of the thermal expansion coefficient is given by

$$u(\alpha_S)=\frac{(2\times 10^{-6})}{\sqrt{3}}=1,2 \times 10^{-6}~^{\circ} C^{-1}.$$

Case 3

The upper and lower limits $ \mathbf{a_-~{e}~a_+} $ for an input quantity $ \mathbf{x_i} $ may not be symmetric, that is, if the lower limit is written as $ \mathbf{a_- = x_i - b_-} $ and the upper limit as $ \mathbf{a_+ = x_i + b_+} $, then $ \mathbf{b_- \neq~ b_+} $. In this case, $ \mathbf{x_i} $ not the center of the range $ \mathbf{(a_-,a_+)} $ and the probability distribution of $ \mathbf{x_i} $ cannot be uniform. However, there may not be enough information to choose an appropriate distribution and different models will lead to different expressions for the variance.

In the absence of such information, a simple approximation is

$$u^2(x_i)=\frac{(b_+ + b_-)^2}{12}=\frac{(a_+ + a_-)^2}{12}$$

which corresponds to the variance of the rectangular distribution with length $ \mathbf{b_- + b_+} $.

Example 1.5.6

Screenshot_1

Example 1.5.7

Returning to Example 1.3.2 of the standard mass calibration, we will estimate the Type B standard uncertainties.

Symbol Source of Uncertainty Limits Distribution Uncertainty
$ W_S $ Standard Dough $ \pm $ 30 mg Normal $ \frac{30}{2} = $15mg
$ D_S $ Standard mass drift $ \pm $ 15mg Rectangular $ \frac{15}{\sqrt{3}}= $8.66 mg
$ \delta_C $ Comparator linearity $ \pm $ 10 mg Rectangular $ \frac{10}{\sqrt{3}}= $5.77 mg
Ab Effect of air $ \pm $ 10 mg Rectangular $ \frac{10}{\sqrt{3}}= $5.77 mg

Example 1.5.8

Returning to example 1.3.3

The resolution of the caliper follows a rectangular distribution based on the resolution itself, which is 0.01 mm. Thus, the uncertainty due to resolution is

$$u({Res}_d)=\frac{{Res}_d}{2\sqrt{3}}=\frac{0,01}{2\sqrt{3}}=0,002887~mm$$

Case 4

In the cases above we do not have information about the value of the quantity $ X_i $, only that it is within the specified limits. Therefore, we assume that the values of the quantity are equiprobable within these limits, and that it has zero probability of being outside these limits. It is often more realistic to assume that values close to the specified limits are less likely than values close to the center. In this case, it is reasonable to change the triangular distribution. Assuming a triangular distribution for the quantity $ X_i $, we obtain as an average $ \mathbf{x_i = (a_+ + a_-) / 2} $ with associated uncertainty $ \mathbf{u^2(x_i) = a^2 / 6} $.

triangulo

So,

$$u=\frac{base}{2\sqrt{6}}=\frac{2a}{2\sqrt{6}}=\frac{a}{\sqrt{6}}.$$

18.1.6 Combined Standard Uncertainty

When the uncertainty of the measured result $ \mathbf{y} $ is obtained by combining the standard uncertainties of the input estimates $ \mathbf{x_1, x_2,\dots, x_N} $, this combined uncertainty of the estimate $ \mathbf{y} $ is represented by $ \mathbf{u_c(y)} $ and called combined standard uncertainty.

Input estimates $ x_1, x_2,\dots, x_N $, can be classified as quantities:

  • Statistically independent or uncorrelated;
  • Statistically dependent or correlated.

Uncorrelated input quantities

For statistically independent quantities, we consider series of measurements that were carried out with different measurement systems. In this case, the combined standard uncertainty $ \mathbf{u_c(y)} $ is the positive square root of the combined variance.

The expression to determine this combined standard uncertainty in the uncorrelated case is presented by

$$u^2_c(y)=\sum_{i=1}^N\left(\frac{\partial y}{\partial x_i}\right)^2u^2(x_i),~~~(1.6.1)$$

in which $ u(x_i) $ is the standard uncertainty associated with the input quantity$ _i $. The partial derivatives ($ \partial f/\partial x_i $) calculated at the point $ x_i $ are called sensitivity coefficients, as they describe how the estimate of y varies with small changes in the values of the estimates of the input quantities $ \mathbf{x_1,x_2,\dots, x_N} $.

Example 1.6.1

Returning to Example 1.3.2. When calibrating the standard mass, we obtain the following combined uncertainty

$$u_c(W_X)=\sqrt{u^2(W_s) + u^2(D_s) + u^2(\delta C) + u^2(Ab) + u^2(\Delta W)} =$$

$$=\sqrt{(15)^2 + (8,66)^2 + (5,77)^2 + (5,77)^2 + (2,08)^2} = 19,26~mg.$$

Example 1.6.1

Returning to Example 1.3.3, we obtain:

We admit that $ \frac{1}{4} $ and $ \pi $ are constants free of uncertainty or with negligible uncertainties, only the variable $ d $ is considered for uncertainty calculation. Firstly, we will calculate the sensitivity coefficient as follows

$$\frac{\partial y}{\partial d}=\frac{\pi~d}{2}$$

Thus, the combined uncertainty of the area is calculated as follows

$$u_c(y)~=\sqrt{\left(\frac{\pi d}{2}\right)^2~(u^2(d)+u^2({Res}_d))+u^2(\varepsilon)}=$$

$$=\sqrt{\left(\frac{3,1415\times 10,28}{2}\right)^2~(0,01^2+0,002887^2)+0,102128^2}=0,196669~mm^2$$

A second way to express uncertainty is as relative combined uncertainty and we calculate it as follows

$$\frac{ u^2_{c}(y)}{y^2}~=~\frac{\left(\frac{\pi~d}{2}\right)^2(u^2(d)+u^2({Res}_d))}{\left(\frac{\pi~d^2}{4}\right)^2}+\frac{u^2(\varepsilon)}{\left(\frac{\pi~d^2}{4}\right)^2}=$$

$$=\frac{4~(u^2(d)+u^2(Res_d))}{d^2}+\frac{16~u^2(\varepsilon)}{(\pi~d^2)^2}$$

Thus the relative uncertainty is expressed as

$$u_{cr}(y)=\frac{2~\sqrt{u^2(d)+u^2(Res_d)+\frac{4~u^2(\varepsilon)}{(\pi~d^2)^2}}}{d}$$

Substituting the values from the example, we obtain the relative combined uncertainty

$$u_{cr}(y)=\frac{2~\sqrt{0,01^2+0,002887^2+\frac{4~0,102128^2}{(\pi~10,28^2)^2}}}{10,28}=0,00237$$

From the relative combined uncertainty, we obtain the combined uncertainty of the Area in the form:

$$u_c(y)=u_{cr}(y)\times {Área}=0,00237\times 82,99963=0,196709~mm^2.$$

Correlated input quantities

For statistically dependent quantities, we consider the series of measurements that were carried out with the same measurement systems, that is, we consider the following mathematical model

$$y=x_i+x_j,~~~i=j=1,\dots,N$$

In this case, the estimated covariance must be considered as an additional contribution to uncertainty. The expression to determine this combined standard uncertainty in the correlated case is presented by

$$u^2_c(y)=\displaystyle\sum^N_{i=1}\sum^N_{j=1}\frac{\partial y}{\partial x_i}\frac{\partial y}{\partial x_j}u(x_i,x_j)=$$

$$=\underbrace{\sum^N_{i=1}\left[\frac{\partial y}{\partial x_i}\right]^2 u^2(x_i)}_{(I)}+\underbrace{2\sum^{N-1}_{i=1}\sum^N_{j=i+1}\frac{\partial y}{\partial x_i}\frac{\partial y}{\partial x_j}u(x_i,x_j)}_{(II)}~~(1.6.2)$$

in which, $ u(x_i,x_j)=u(x_j,x_i) $ is the correlated uncertainty, associated with the input quantities $ X_i $ and $ X_j $.

Thus, dividing and multiplying equation (1.6.1) by $ u(x_i)u(x_j), $ in (II) we obtain

Screenshot_2

in which $ \rho(x_i,x_j) $ is the degree of correlation between $ x_i $ and $ x_j $ with $ \rho(x_i,x_j)=\rho(x_j,x_i) $ and taking $ -1\leq \rho(x_i,x_j)\leq 1. $ Therefore, substituting equation (1.6.3) into equation (1.6.2), we obtain

$$u^2_c(y)=\sum^N_{i=1}\left[\frac{\partial y}{\partial x_i}\right]^2 u^2(x_i)+2\sum^{N-1}_{i=1}\sum^N_{j=i+1}\frac{\partial y}{\partial x_i}\frac{\partial y}{\partial x_j}\rho(x_i,x_j)u(x_i)u(x_j)~~(1.6.4)$$

If the variables $ x_i $ and $ x_j $ are independent, we have to $ \rho(x_i,x_j)=0, $ and equation (1.6.4) reduces to equation (1.6.1). We will take the extreme case in which $ \rho(x_i,x_j)=1, $ we obtain approximate equation of measurement uncertainty in the correlated case as follows

$$u^2_c(y)\approx\sum^N_{i=1}\left[\frac{\partial y}{\partial x_i}\right]^2 u^2(x_i)+2\sum^{N-1}_{i=1}\sum^N_{j=i+1}\frac{\partial y}{\partial x_i}\frac{\partial y}{\partial x_j}u(x_i)u(x_j)~~(1.6.5)$$

Example 1.6.2

Returning to example 1.3.4. Then, we obtain expression (1.6.5) as follows

$$u^2(y)=\frac{\partial y}{\partial x_1}u^2(x_1)+\frac{\partial y}{\partial x_2}u^2(x_2)+2\frac{\partial y}{\partial x_1}\frac{\partial y}{\partial x_2}u(x_1)u(x_2)~~(*)$$

From the expression in example 1.3.4, we obtain the sensitivity coefficients:

Screenshot_3

Thus, we obtain the expression (*) as follows

$$u^2(y)=1~u^2(x_1)+1~u^2(x_2)+2~u(x_1)u(x_2)=$$

$$=(u(x_1)+u(x_2))^2$$

Hence, we obtain the following equation

$$u(y)=u(x_1)+u(x_2)$$

Substituting the values, we have

$$u(y)=0,0385+0,042=0,0805~mm$$

18.1.7 Expanded Uncertainty

Although the combined uncertainty $ u_c(y) $ can be universally used to express the uncertainty of a measurement result (due to the needs of some industries and commercial applications, as well as requirements in health and safety areas) it is often necessary to present a measure of uncertainty that defines a range over the measurement result. In this case, uncertainty comprises a fraction of the distribution of values, which can be reasonably attributed to a measurand, called Expanded Uncertainty U. This requirement was recognized by the Working Group and Recommendations of the CIPM, INC (1981).

The expanded uncertainty U is obtained by multiplying the combined standard uncertainty $ u_c(y) $ by a factor k.

$$U = k \times u_c(y)$$

The value of the k factor is chosen based on the confidence level required for the interval. In general, k is used between 2 and 3. Therefore, for special applications, k can be determined according to the required level of confidence, according to the normal or t-Student distribution.

Namas (NIS 3003, 1995) recommends that the k factor be equal to 2 to calculate the expanded uncertainty. This value corresponds to approximately 95% confidence. However, if the contributions to the uncertainty related to repeatability are large compared to the other distributions and the number of repetitions is small, there is a possibility that the normal probability distribution is not suitable. In this case, the factor k=2 guarantees us a confidence level lower than 95%. Here, we must use the t-Student distribution to find the value of the k factor that guarantees 95%.

Rule: If the Type A uncertainty is less than half the combined uncertainty, we will use the factor $ k = 2. $ Otherwise, we must use the t-Student distribution to obtain the value of k that guarantees a 95% confidence interval. The ISO GUM standard [ver. 2008] recommends using the Welch-Satterwaite equation to calculate the degree of freedom, based on the degrees of freedom of each source of uncertainty. The formula for this calculation is given by

$$\upsilon_{eff}=\displaystyle\frac{u_c^4(y)}{\displaystyle\sum_{i=1}^N u_i^4(y)/\nu_i}= \frac{u_c^4(y)}{u_A^4(y)/\nu_A}=\left(\frac{u_c(y)}{u_A(y)}\right)^4 \nu_A,$$

in which $ u_i(y)=|\frac{\partial f}{\partial x_i}|u(x_i) $, $ \nu_i $ represents the degrees of freedom of the uncertainty factor i and $ \nu_A $ represents Type A degrees of freedom. For Type A uncertainty contributions, we consider as degrees of freedom the number of readings minus 1 times the number of calibration points. For the degrees of freedom referring to Type B uncertainty contributions, we will consider $ \upsilon_i $ equals infinity.

Example 1.4.1

Suppose that a measurement system with Type A uncertainty, based on 4 observations, has a value $ u_i(y) $ of 3.5 units. There are 5 other sources of Type B uncertainty that have very small estimated uncertainty, such that the combined uncertainty $ u_c(y) $ be equal to 5.7 units.

Since the Type A uncertainty is greater than half the combined uncertainty, we will use the t-Student distribution to determine the factor $ \mathbf{k} $. Using the Welch-Satterwaite equation, we have

$$\upsilon_{eff}=\frac{(5,7)^4}{((3,5)^4/(4-1))+ 0 + 0 + 0 + 0 + 0}=21,1.$$

Taking value of $ \upsilon_{eff} $ equal to 20, we obtain that k = 2.13.

Example 1.4.2

Returning to Example 1.3.2 of the standard mass calibration, note that the Type A uncertainty is less than half the combined uncertainty.

Let’s calculate the effective degrees of freedom

$$\upsilon_{eff}=\displaystyle\left(\frac{19,26}{2,08}\right)^4 (5-1)=29405,72$$

Thus, the coverage factor k will be 1.96.

Thus, the expanded uncertainty is given by

$$ U= 1,96 \times 19,26 mg = 37,7496~ mg .$$

Example 1.4.3

Returning to Example 1.3.3.

Let’s calculate the effective degrees of freedom

$$\upsilon_{eff}=\displaystyle\left(\frac{0,196669}{0,102128}\right)^4 (5-1)=55,00735$$

Thus, the coverage factor k will be 2.004045.

Therefore, the expanded uncertainty is given by

$$ U= 2,004045 \times 0,196669 =0,394133~mm^2 .$$

Example 1.4.4

Returning to Example 1.3.4.

We have no type A uncertainty, so the coverage factor k is 1.96;

Thus, the expanded uncertainty is given by

$$ U= 1,96 \times 0,0805 = 0,15778~mm .$$

18.1.8 Expression of the Measurement Result

In this section, we discuss the ways of presenting the measurement result. To make it easier for the user, we start by rounding the value of the expanded uncertainty according to the rule below.

Amount Rounding Rules

When we want to round a number so that it is expressed to a certain number of significant digits, we must apply conventional rounding rules.

Rule 1:

If the digit to the right of the last digit to be represented is less than 5, we just ignore the other digits to the right. Example: 3.14159265 to 3.14.

Rule 2:

If the digit to the right of the last digit to be represented is greater than 5, we add one unit to the last digit represented and disregard the other digits to the right. Example: 3.14159265 to 3.1416.

Rule 3:

If the right-hand digit to be represented is equal to 5, then the rounding must be such that the last digit represented after rounding must be even. Example: 3.14159265 to 3.142.

Digit numbers in measurement uncertainty

There is no well-defined rule for the number of digits that should be indicated for measurement uncertainty. In general, we use 2 significant figures, in addition to the leading zeros. In some cases, it may be necessary to use more significant digits to avoid rounding errors in subsequent calculations. In other cases, it is not possible to assign more than 1 digit for measurement uncertainty.

Summary:

  • Measurement uncertainty must be presented with 2 digits when the first digit in the uncertainty is 1 or 2.
  • Measurement uncertainty can be presented with 1 digit when the first digit of the uncertainty is 3 or greater.
  • Measurement uncertainty can be represented with 2 figures in any case.

According to the rules above, we present the examples:

Incorrect Correct
0.144 (mm) 0.14 (mm)
1,026 (s) 1 (s)
3.49 (mm) 3.5 (mm) or 3 (mm)
3.51 (mm) 3.5 (mm) or 4 (mm)
0.00514 (mm) 0.005 (mm) or 0.005 (mm)

Example 1.8.1

Returning to example 1.3.2

In this case, the measurement result is expressed in the form

$$10 000,000~g~+~0,018~g~\pm 0,04~g~(k=1,96),$$

i.e.

$$10 000,018~g~\pm~0,04~g~(k=1,96).$$

Example 1.8.2

Returning to example 1.3.3

In this case, the measurement result is expressed in the form

$$83,0~mm^2~\pm~0,4~mm^2~(k=2).$$

When we report the result of a measurement we must:

a) Provide a complete description of how the measurand Y is defined;

b) Express the measurement result: Y=y +/- U and provide the units of y and U;

c) Include the expanded relative uncertainty U / |y|, with y $ \neq $ 0, when appropriate;

d) Provide the value of k used to obtain U;

e) Provide the approximate confidence level associated with the interval y +/- U and explain how it was obtained;

18.1.9 Extreme Value Test (Grubbs)

This test is developed to check the presence of extreme values in sample observations. Extreme values ​​can be considered as manifestations of the random variability inherent in the data, or just an error in the calculation during data collection and even a hasty note by the operator.

There are numerous criteria for testing extreme values. In all of them, we develop numerical sample calculations (statistics) and compare them with a critical value based on the theory of random samples, to decide whether or not there is an observation considered an extreme value.

In the Grubbs test, we use the following statistic

$$Z~=~\frac{| x_i ~-~\overline{x} |}{s}$$

in which

  • $ x_i $: is a sample observation $ x_1, x_2, \cdots, x_n $;
  • $ \overline{x} $: is the sample mean;
  • $ s $: is the sample standard deviation.

This statistic tests the following hypotheses

Screenshot_4

(there are missing elements in the equation above)

We reject the hypothesis $ H_0 $, with significance level $ \alpha $, if $ Z > Z_c $. In which $ Z_c $ is a critical value based on the distribution of Z and is found in the table (see F. E. Grubbs (1969)) of values of $ \alpha $ single-tailed. In Table 1.9.1, we find some critical values for $ \alpha $ = 10%, 5%, 2.5%, 1% and 0.5%.

n 0,1 0,05 0,025 0,01 0,005
3 1,148 1,153 1,154 1,155 1,155
4 1,425 1,462 1,481 1,492 1,496
5 1,602 1,671 1,715 1,749 1,764
6 1,729 1,822 1,887 1,944 1,973
7 1,828 1,938 2,02 2,097 2,139
8 1,909 2,032 2,127 2,221 2,274
9 1,977 2,11 2,215 2,323 2,387
10 2,036 2,176 2,29 2,41 2,482
11 2,088 2,234 2,355 2,484 2,564
12 2,134 2,285 2,412 2,549 2,636
13 2,176 2,331 2,462 2,607 2,699
14 2,213 2,372 2,507 2,658 2,755
15 2,248 2,409 2,548 2,705 2,806
16 2,279 2,443 2,586 2,747 2,852
17 2,309 2,475 2,62 2,785 2,894
18 2,336 2,504 2,652 2,821 2,932
19 2,361 2,531 2,681 2,853 2,968
20 2,385 2,557 2,708 2,884 3,001
21 2,408 2,58 2,734 2,912 3,031
22 2,429 2,603 2,758 2,939 3,06
23 2,449 2,624 2,78 2,963 3,087
24 2,468 2,644 2,802 2,987 3,112
25 2,486 2,663 2,822 3,009 3,135
26 2,503 2,681 2,841 3,029 3,158
27 2,52 2,698 2,859 3,049 3,179
28 2,536 2,714 2,876 3,068 3,199
29 2,551 2,73 2,893 3,086 3,218
30 2,565 2,745 2,908 3,103 3,236
31 2,579 2,76 2,924 3,119 3,253
32 2,592 2,773 2,938 3,135 3,27
33 2,605 2,787 2,952 3,15 3,286
34 2,618 2,799 2,965 3,164 3,301
35 2,63 2,812 2,978 3,178 3,316
36 2,641 2,824 2,991 3,191 3,33
37 2,652 2,835 3,003 3,204 3,343
38 2,663 2,846 3,014 3,216 3,356
39 2,674 2,857 3,025 3,228 3,369
40 2,684 2,868 3,036 3,239 3,381
50 2,772 2,957 3,128 3,337 3,482
60 2,841 3,027 3,2 3,411 3,56
70 2,898 3,084 3,258 3,471 3,622
80 2,946 3,132 3,306 3,521 3,673
90 2,987 3,173 3,348 3,563 3,716
100 3,024 3,21 3,384 3,6 3,754
110 3,056 3,242 3,416 3,633 3,787
120 3,086 3,271 3,445 3,662 3,817
130 3,112 3,297 3,471 3,688 3,843
140 3,136 3,321 3,495 3,712 3,867

Table 1.9.1: Grubbs test table.

Example 1.9.1

Consider the following measurements:

Measurements
11,89896
11,9596
11,89856
11,91408
12,04252
12,1531
11,94553
11,8682
11,85949
12,13373
12,6

Let’s calculate the mean and standard deviation:

$$\overline{x}=\frac{11,89+11,95+\dots+12,6}{11}=12,02489$$

$$s=\sqrt{\frac{(11,89-12,024)^2+(11,95-12,024)^2+\dots+(12,6-12,024)^2}{11-1}}=0,215851$$

With that, let’s calculate the Grubbs test for point 11, using the following statistic

$$Z~=~\frac{| x_i ~-~\overline{x} |}{s}=\frac{|x_{11}-\overline{x}|}{s}=\frac{|12,6-12,02489|}{0,215851}=2,664396$$

How $ Z> 2,3547, $ therefore, this measurement is an extreme value (outlier).

Results of this example obtained with the Action software:

(image missing)

Exercise 1.9.1

Consider the following measurements in table 1.9.2 and calculate the mean, standard deviation and Grubbs test:

Measurements
9,988031
10,02081
9,997529
10,06985
9,995944
10,1367
9,936079
9,880081
9,99015
10,04604
11

Table 1.9.2: Measurements.

18.1.10 Comparison between Measurement Systems

We will present a technique for comparing two measurement systems. To illustrate, let’s consider an example.

Example 1.10.1

The diameter of a standard ring can be measured by two types of measuring systems. To compare these measuring systems, a standard ring was measured 5 times by each measuring system using the same operator. The results are below.

SM1 SM2
Average 15,601mm 15,603mm
Expanded Uncertainty 0.001mm 0.0015mm

Step 1

Calculate Statistics

$$EN~=~\frac{| MSM1~-~MSM2 |}{\sqrt{USM1^2~+~USM2^2}}$$

in which

  • MSM1: represents the average of measurement system 1;
  • MSM2: represents the average of measurement system 2;
  • USM1: represents the expanded uncertainty of measurement system 1;
  • USM2: represents the expanded uncertainty of measurement system 2.

With this, we have to

$$EN~=~\frac{| 15,601~-~15,603 |}{\sqrt{(0,001)^2~+~(0,0015)^2}}~=~\frac{0,002}{0,0018}~=~1,109.$$

Step 2

  • If EN ≤ 1, the two measurement systems are compatible;
  • If EN > 1, the two measurement systems are not compatible, that is, the measurement systems present significant differences.

As, in the example, EN=1.109 is greater than 1, we conclude that there is a significant difference between the two measurement systems.

Exercise

During the accreditation process of the engine’s net effective power testing laboratory, INMETRO required an inter-laboratory comparison study. With the engine at idle, a power test was carried out by the participating laboratory and a reference laboratory. Using the data presented in the table below, calculate the normalized error and make the appropriate conclusions.

Lab Ref Laboratory
Average 11,601 12,613
Expanded Uncertainty 0,3 0,4

18.2 Analysis and interpretation of the Calibration Certificate

Metrological Attestation generally includes calibration or verification, any necessary adjustments or repairs, recalibration, comparison with metrological requirements for the intended use of the equipment, as well as any necessary labeling or sealing. To achieve this, we need the suitability of the measuring equipment for your use to have been demonstrated and documented. In summary, we define metrological verification as the set of operations necessary to ensure that measuring equipment meets the requirements of its intended use. This section was prepared based on the ABNT NBR ISO 10012:2004 standard.

Requirements for intended use include:

  • Amplitude;
  • Resolution;
  • Maximum Permissible Error.

We present a diagram of the processes involved in metrological proof in Figure 2.1

ComprovaMetro

Figure 2.1: Metrological Proof Process for measuring equipment.

Metrological requirements are derived from requirements for the product. These requirements are necessary for both measuring equipment and measuring processes. These can also be expressed as maximum allowable errors, allowable uncertainty, range, stability, resolution, environmental conditions or operator skills.

The guidance is that we must specify the measurement processes and measurement equipment that are subject to the provisions of Standard ABNT NBR ISO 10012:2004. When deciding on the scope and extent of the measurement management system, we must take into account the risks and consequences of failure to comply with metrological requirements.

Note: The measurement management system consists of:

  • in the control of indicated measurement processes;
  • metrological proof of measuring equipment and necessary support processes (as shown in Figure 2.1).

Recalibration of measuring equipment is not necessary if the equipment is already in a valid calibration situation. The metrological attestation procedure may include methods for us to verify that measurement uncertainty and/or measuring equipment errors are within the permissible limits specified in the metrological requirements. Information pertinent to it must be readily available to the operator, including any limitations or special requirements.

Metrological Function

The metrological function must be defined by the organization. The organization’s Senior Management must ensure the availability of the necessary resources to establish and maintain the metrological function. The metrology function can be a single department or distributed throughout the organization.

The management of the metrological function must establish, document and maintain the measurement management system and continuously improve its efficiency.

Metrological Check Intervals

The guidance for the methods used to determine or change the intervals between metrological verifications is that we must describe them in documented procedures. We must critically analyze and adjust when necessary to ensure continued compliance with specified metrological requirements. To determine metrological verification intervals, we can use data obtained from calibration histories, metrological verification and advances in technology and knowledge. When using records using techniques such as Statistical Process Control (CEP), they can be useful in determining whether or not it is necessary to change metrological verification intervals.

According to (OIML D10), the calibration interval can be equal to the metrological verification interval.

Another important point is that each time we repair, adjust or modify measuring equipment that does not comply with the metrological verification range, it must be critically analyzed.

Control of equipment adjustments

To control equipment adjustments, we must take some precautions, such as:

  • Access to adjustment means and devices on proven measuring equipment, the position of which affects performance, we must seal or otherwise protect them in order to prevent unauthorized changes;
  • We must design or implement seals or safeguards in such a way that unauthorized changes are detected;
  • We must include actions to be taken when seals or safeguards are breached, damaged, circumvented or missing.

An important point is that we do not apply the requirement for sealing to adjustment means or devices that are intentionally positioned by the user without the need for external references, for example zero adjustments. It is also important to prevent unauthorized changes to computer programs and the organization’s procedures.

Decisions about sealing, controls or adjustments of sealing materials and seals, such as labels, welds, wires, paint, are normally left to the metrological function and the implementation of a sealing program is documented by it. But it is worth remembering that not all measuring equipment has the possibility of being sealed.

Records of the metrological verification process

An important process within metrological proof are the records of the metrological proof process, they must be dated and approved by a person authorized to attest to the correctness of the results, as appropriate and they must be maintained and available.

The minimum registration time depends on many factors, including customer requirements, statutory or regulatory requirements, and manufacturer liability. Records relating to measurement standards may need to be kept indefinitely. We must demonstrate in the metrological verification records whether each item of equipment meets the specified metrological requirements and must include, when necessary, the following:

  • Description and unique identification of the equipment manufacturer, type, serial number, etc.;
  • Date on which metrological verification was completed;
  • Result of metrological verification;
  • Fixed interval for metrological verification;
  • Identification of the metrological verification procedure;
  • Maximum permissible errors defined;
  • Relevant environmental conditions and statement of any necessary corrections;
  • Uncertainties involved in equipment calibration;
  • Details of any maintenance such as adjustments, repairs or modifications carried out;
  • Any limitations on use;
  • Identification of the people who carry out the metrological verification;
  • Identification of the people responsible for correcting the recorded information;
  • Unique identification (such as series number) of any calibration report or certificate and other pertinent documents;
  • Evidence of traceability of calibration results;
  • Metrological requirements for the intended use;
  • Calibration result after and where required prior to any adjustment, modification or repair.

The guidance according to ABNT NBR 10012:2004 is that calibration results are recorded so that the traceability of all measurements can be demonstrated and so that calibration results can be reproduced under conditions close to the original conditions. Sometimes the verification result is included in the calibration report or certificate where it is stated whether the equipment complies (or fails to conform) with the specified requirements. The requirements may be handwritten, or typed, or microfilmed, or electronic media, or magnetic media, or on another information medium.

The maximum allowable error can be determined by the metrological function or by reference to the published specifications of the measuring equipment manufacturer. It is important to know that the metrological function must ensure that only authorized people are allowed to generate, amend, issue or delete records.

Calibration Certificate Analysis

As we said, the metrological function must define a strategy to evaluate each measuring equipment, which is a mandatory requirement, according to item 5.5.2 of the ISO/IEC 17025 standard [9]. One of the most used ways is to define the maximum permissible error (EMP), through the product tolerance or manufacturer’s specifications, and compare it with the result of equipment calibration. For this, we take

$$EMP=\frac{\text{Menor tolerância medida}}{J},~~~ \text{com}~~J = (5; 15].$$

The most used is $ J=10 $. So

$$EMP=\frac{\text{Tolerância}}{10}.$$

EMP

Criterion

$$\max_i\left(\mid T_i \mid + U(i) \right) \leq EMP,\quad i=1,\dots,n$$

for every calibration point ($ i $ represents the calibration point and $ n $ the number of calibration points).

Metrological proof in the case where the EMP is a function of the readings is discussed below.

$$ EMP=\pm~(a + b\times leitura)$$

$$a=0,01$$

$$b=0,01$$

EMP1

Criterion

$ |T_i| + U(i) \leq EMP(i) $, for every calibration point ($ i $ represents the calibration point).

Example 2.1.1

Suppose we have a tolerance of 1 g for standard masses. After calibrating the masses, we obtained the following information from the calibration certificate. This information is presented in Table 2.1.1

Point (g) Trend (g) U(g) k
1000 0,009 0,015 2
1000 0,01 0,015 2
1000 0,016 0,015 2
1000 0,01 0,015 2
5000 -0,014 0,075 2
5000 -0,069 0,075 2
5000 -0,043 0,075 2
5000 0,025 0,075 2

Table 2.1.1: Calibration Certificate.

Considering J=10, we have

$$EMP=\frac{\text{Tolerância}}{10}=0,1~g.$$

Table 2.1.2 presents the approval criteria $ (|T|+U \leq EMP) $ for the eight standard masses. As we can see, two masses of 5 kg failed. As a result, the calibration certificate whose values ​​were presented in Table 2.1.1 is not approved.

Point (g) |T|+U Criterion
1000 0,024 Approved
1000 0,025 Approved
1000 0,031 Approved
1000 0,025 Approved
5000 0,089 Approved
5000 0,144 Failed
5000 0,118 Failed
5000 0,1 Approved

Table 2.1.2: Approval Criteria.

Example 2.1.2

We will apply the metrological proof in a graduated buret. Assume that the burette controls a 2 mL process tolerance for volume measurements. In Figure 2.1.3 we present the graduated buret calibration certificate.

(image missing)

Nominal value Average Correction |C| Expanded Uncertainty Coverage factor (k) Effective degrees of freedom (veff)
5 4,9077 0,0923 0,002 2,37 10
15 14,8280 0,1720 0,003 2,20 15
25 24,9328 0,0678 0,003 2,06 44

Table 2.1.3: Calibration certificate summary table.

Considering J=10, we have

$$EMP=\frac{\text{Tolerância}}{10}=\frac{2}{10}=0,2.$$

Table 2.1.3 presents the approval criteria $ (|C|+U \leq EMP) $ for the three standard volume measurements. As we can see, all volume ranges are approved.

Point (mL) |C|+U Criterion
5 0,0943 Approved
15 0,175 Approved
25 0,0708 Approved

Table 2.1.3: Approval Criteria.

TUR (Test Uncertainty Ratio)

The calibration process involves a comparison between the Measuring Equipment and a standard, having better metrological requirements. The comparison between the accuracy of the unit under test and the accuracy of the standard is known as the test accuracies ratio (TAR). However, this ratio does not consider other potential sources of error from the calibration process. The comparison between the accuracy of the unit under test and the estimated calibration uncertainty is known as a test uncertainty ratio (TUR). This relationship is more reliable, because it considers sources of error involved in the calibration process that TAR does not consider.

The test uncertainty ratio (TUR) is a measure of the ability of a given measuring instrument and/or process to meet a product (or process) specification. Thus, TUR is the ratio between the tolerance and/or specification of the product and the uncertainty present in the test of this specification or tolerance. Historically, a widely used rule is that the TUR must be at least 10:1. The higher the ratio, the better the test performance. Currently, a ratio of 4:1 or even 3:1 is considered acceptable in some cases. This is mainly due to the improved performance of manufacturing equipment. In many cases, we do not have equipment with a small enough uncertainty for a 10:1 TUR, or it is too expensive for the application.

There are two main applications for TUR:

  • The first is in the calibration of measuring instruments and equipment;
  • The second is in the inspection of manufactured components.

In general, we have the following equation for TUR:

$${TUR}=\frac{{Tolerância~ou~EMP}}{{Incerteza~Expandida}}$$

From this, we note that the TUR ratio compares the allowable variation for the measurand (the numerator) with the variability associated with the measurement of the measurand (the denominator).

TUR1_1

Figure 2.1.3: Relationship between the specification zone and the conformity zone.

Figure (2.1.3) shows the relationship between the specification zone and the conformity zone. If the “true value” of the measurand is within the specification zone, the specification is satisfied, otherwise the measurand is out of specification. However, we can never know the “true value” of the measurand as mentioned in module 2. In order to indicate whether the measurand is out of specification or not, we have to know the uncertainty in the measurement process. This is shown at the bottom of the horizontal line in the figure. If the measurand is in the compliance zone, we are confident that the true value is within the specification. Likewise, if the measurand is in the nonconformity zone, we are confident that the true value is out of specification. For the region of uncertainty shown between conformance and nonconformity, we do not have sufficient confidence to determine whether the part or product conforms or not.

TUR_0

Figure 2.1.4: Influence of TUR on the final result of a calibration.

As seen previously, due to the better performance of the equipment, we consider TUR ≥ 5 as an acceptable relationship when choosing equipment, and a TUR ≥ 3 when choosing the calibration standard, which ensures and allows that the uncertainty of the standard does not significantly interfere with metrological confirmation. Figure (2.1.4) illustrates the compromise of the standard’s uncertainty in relation to the maximum allowable error, depending on different TUR adopted.

18.3 Applications of Uncertainty Calculation in Calibrations

In the next modules, we will present some applications of Measurement Uncertainty for calibrations, to exemplify the calculation procedure and the initial steps before starting to calculate measurement uncertainty.

18.3.1 Calculation of Uncertainty of a Dial Dial

Consider the process of calibrating a 0.01 mm resolution watch. This calibration is performed by comparison using a clock calibrator. The calibrator has a resolution of 0.001mm with an expanded uncertainty of 0.001mm with k=2. The results are presented in Table 3.1.1

Screenshot_6

Table 3.1.1: Data table.

relogio

Measurement Method

Calibration corresponds to the difference between the watch and calibrator measurements

$$d = l - l_S$$

where

  • $ l $: represents the reading adjusted on the dial indicator;
  • $ l_S $: represents the reading obtained by the calibrator.

Mathematical Model

The deviations obtained present the following sources of uncertainty

$$d~=~\Delta_l~+~Res(Rel)~+~Res(Cal)~+~{Histerese}$$

So

$$l= d + l_s = ~\Delta_l~+~Res(Rel)~+~Res(Cal)~+~Histerese~ +~ l_s,$$

where,

  • $Δ_l$: represents the difference observed between the watch measurement and the repeatability calibrator measurement;
  • Res(Rel): clock resolution;
  • Res(Cal): calibrator resolution;
  • $l_s$: represents the contribution of the pattern;
  • Hysteresis: maximum difference between the measurement in advance and return.

Combined Uncertainty

Through the uncertainty propagation equation, we have that the expression of the combined uncertainty is given by

$$u_c(d)=\sqrt{~u^2(\Delta_l)~+~u^2(Res(Rel))~+~u^2(Res(Cal))~+~u^2(l_s)~+~u^2({Histerese})}$$

Calculation of Standard Uncertainty of Input Quantities

Next, we will calculate the uncertainty for each source.

Repetitiveness ($ \Delta_l $)

Type A uncertainty.

Let’s take the 0.5 mm point as an example. So, we have

$$u(\Delta_l)~=~\frac{s}{\sqrt{n}}=\frac{0,00121106}{\sqrt{6}}=0,000494413$$

in which

  • $ u(\Delta_l) $: represents Type A uncertainty;
  • $ s $: represents the standard deviation of the readings at the calibration point $ i $;
  • $ n $: represents the number of readings at the calibration point $ i $.

Inherited Pattern Uncertainty $ (u(l_s)) $

Distribution: Normal.

$$u(l_s)=\frac{U(L_s)}{k}=\frac{0,001}{2}=0,0005$$

Calibrator Resolution $ (Res(Cal)) $

Distribution: Rectangular.

$$u(Res(Cal))=\frac{Res(Cal)}{2\sqrt{3}}=\frac{0,001}{2\sqrt{3}}=0,000288675$$

Clock Resolution $ (Res(Rel)) $

Distribution: Rectangular.

$$u(Res(Rel))=\frac{Res(Rel)}{2\sqrt{3}}=\frac{0,01}{2\sqrt{3}}=0,00288675$$

Hysteresis Calculation

Reference Average Advance Average Return |Hysteresis|
0,1 0,101667 0,100333 0,0013333
0,5 0,503667 0,501667 0,002
0,7 0,701667 0,696667 0,005
1 0,999667 0,999667 0

For the 0.5 mm calibration point we have:

Screenshot_7

relogio1

Combined Uncertainty Calculation

$$u_c(d)=\sqrt{~u^2(\Delta_l)~+~u^2(Res(Rel))~+~u^2(Res(Cal))~+~u^2(l_s)~+~u^2({Histerese})}=$$

$$=\sqrt{(0,00049)^2~+~(0,002887)^2~+~(0,0002887)^2~+~(0,0005)^2~+~(0,00057)^2}$$

$$=0,00304$$

Effective Degrees of Freedom

$$\nu_{eff}~=~\left(\frac{u_c(d)}{u(\Delta_l)}\right)^4~\nu_A~=\left(\frac{0,00304}{0,00049}\right)^4~(6-1)=7151,07$$

Through the t-Student table, we find $ k=1,96 $

Expanded Uncertainty

$$U~=~k~\times~u_c(d)~=1,96\times~0,00304=0,00596$$

The summary of the measurement uncertainty calculation for the dial indicator is in the following table.

Screenshot_5

The measuring clock is used to measure, with a tolerance of 0.2mm. We will now carry out a study of the metrological proof.

Considering $ J=10 $, we have to

$$EMP=\frac{{Tolerância}}{10}=\frac{0,2}{10}=0,02~mm$$

$$|T|+U=|0,0026|+0,00596=0,00863\leq 0,02=EMP$$

Therefore, for the 0.5 mm point, it was approved.

18.3.2 Calculation of Uncertainty of an Outside Micrometer

Consider the process of calibrating a digital outside micrometer with a resolution of 0.001 mm. This calibration is carried out by comparison using a set of standard blocks. The results are presented in Table 3.2.1.

Screenshot_8

Table 3.2.1: Measurement data.

Measurement Method

Reading a pattern block corresponds to the difference between its lengths

$$ d = l - l_S $$

where,

  • l: represents the reading obtained by the micrometer;
  • lS: represents the length of the standard block for a temperature of 20ºC.

The correction due to temperature variation is given by

$$d = l (1+ \alpha~\theta )-l_S (1+\alpha_S~\theta_S)$$

where,

  • $ \alpha $ and $ \alpha_S $: correspond to the thermal expansion coefficients of the micrometer (scale) and the standard block used in the measurement, respectively;
  • $ \theta $ and $ \theta_S $: correspond to the temperature difference between the micrometer and the standard block, in relation to $ 20ordm;C $, respectively.

Mathematical Model

The reading of the block length obtained from the micrometer is given by

$$l=\frac{d~+~l_S(1 + \alpha_S~\theta_S)}{1+ \alpha~\theta}~~(3.2.1)$$

Multiplying and dividing by ($ 1 - \alpha~\theta $), we get

$$l=\frac{[d~+~l_S(1+\alpha_S~\theta_S)]}{1+\alpha~\theta}~\frac{(1-\alpha~\theta)}{(1-\alpha~\theta)}$$

$$=\frac{d~(1-\alpha~\theta)~+~l_S(1 + \alpha_S~\theta_S)(1-\alpha~\theta)}{1~-~(\alpha~\theta)^2}$$

Simplifying the above equation with respect to negligible terms, we can approximate the value of the standard block, for

$$l~=~d~+~l_S~+ l_S(\alpha_S~\theta_S~-~\alpha~\theta)$$

When denoting the difference between the temperatures of the micrometer and the standard block, for

$$\delta\theta~=~\theta~-~\theta_S$$

and the difference between the coefficients of thermal expansion of the micrometer and the gauge block, e.g.

$$\delta\alpha~=~\alpha~-~\alpha_S$$

we get that

$$l~=~d~+~l_S~-~l_S(\delta\alpha~\theta~+~\delta\theta\alpha_S~)$$

The deviations obtained present the following sources of uncertainty

$$d=\Delta_l~+~Res~+~Ep$$

in which,

  • $Δ_l$ : represents the difference observed in the length of the standard block (repetitivity); micrometer measurement - gauge block measurement;
  • Res: represents the micrometer resolution;
  • Ep: represents the parallelism of the micrometer.

Thus, the mathematical equation for block length is:

$$l~=~\Delta_l~+~Res~+~Ep~+ l_S~- l_S(\delta\alpha~\theta~+~\delta\theta~\alpha_S~)$$

Sources of uncertainty

  • Resolution (Res): The micrometer resolution is 0.001mm. This variable is considered to have a zero mean and variation limits defined by the equipment resolution;
  • Uncertainty inherited from the standard block ($l_S$): This variable has a mean given by the standard value with a standard deviation defined by the combined uncertainty, both declared in the calibration certificate;
  • Parallelism ($E_p$): The parallelism between faces is considered a variable with zero mean and variation limits of $ \pm~0,0003mm $ (result obtained via an optical parallel);
  • Temperature corrections: Before calibration, care is taken to ensure that the gauge block set and micrometer are at the same ambient temperature as the measuring room;

Screenshot_11

Furthermore, sources of uncertainty $ \theta,~\delta\alpha,~\delta\theta,~\alpha_S~ $ are assumed uncorrelated. Through equation (3.2.1), reading obtained from the micrometer for the length of the block, we observe that the estimated value for length is given by

$$l = ({Média~do~desvios}) + l_S$$

NOTE: This equation is the estimate of the micrometer measurement that is obtained by disregarding the variables that have a zero mean.

Combined Uncertainty Assessment (uc(l))

Using the uncertainty propagation equation, we have

$$u^2_c(l)~=~c^2_{l_S}~u^2(l_S)~+~c^2_d~u^2(d)~+~c^2_{\alpha_S}~u^2(\alpha_S)~+~c^2_\theta~u^2(\theta)~+~c^2_{\delta\alpha}~u^2(\delta\alpha)~+~c^2_{\delta\theta}~u^2(\delta\theta)$$

Where the sensitivity coefficients evaluated at the mean point are given by:

  • $ c_{l_S}~=~1~-~(\delta\alpha~\theta~+~\delta\theta~\alpha_S)~=~1 $;
  • $ c_d~=~1 $;
  • $ c_{\alpha_S}~=~-~l_S~\delta\alpha~=~0~({Média de}~ \delta\alpha~{ é igual a zero)} $;
  • $ c_{\theta}~=~-~l_S~\delta\theta~=~0~({Média de}~\delta\theta~{é igual a zero}) $;
  • $ c_{\delta\alpha}~=~-~l_S~\theta $;
  • $ c_{\delta\theta}~=~-~l_S~\alpha $.
  • Therefore, the combined uncertainty expression is given by

$$u_c(l)~=\sqrt{u^2(l_S)~+~u^2(\Delta_l)~+~u^2(Res)~+~u^2(Ep)~+~(~-~l_S~\theta)^2~u^2(\delta\alpha)~+~(~-~l_S~\alpha)^2~u^2(\delta\theta)}$$

Calculation of Standard Uncertainty of input quantities

Next, we will calculate the uncertainty for each source.

Reference standard (lS)

Distribution: Normal

Pattern Block Expanded Uncertainty (mm) Coverage Factor Combined Uncertainty
25 0,00008 2 0,00004
27,5 0,00007 2 0,000035
30,1 0,00009 2 0,000045
32,7 0,00008 2 0,00004
35,3 0,00008 2 0,00004
37,9 0,00008 2 0,00004
40 0,00009 2 0,000045
42,6 0,00009 2 0,000045
45,2 0,00009 2 0,000045
47,8 0,00008 2 0,00004
50 0,00009 2 0,000045

Table 3.2.2: Calculation of standard uncertainty.

Parallelism (Ep)

Distribution: Rectangular

$ u(Ep)=\frac{0,0006}{2\sqrt{3}}=0,000173 $

Micrometer Resolution (Res)

Distribution: Rectangular

$ u(Res)=\frac{0,001}{2\sqrt{3}}=0,000289 $

Difference between thermal expansion coefficients ($ \delta\alpha $)

Distribution: Rectangular

$ u(\delta\alpha)=\frac{3,98\times 10^{-6}}{2\sqrt{3}}=1,15\times 10^{-6} $

Difference between micrometer temperature and reference temperature $ t_0 = 20 {}^\circ C~(\theta) $

An average temperature of $ 19,9 {}^\circ C $, resulting in $ \theta=19,9 - 20 =-0,1 {}^\circ C $.

Temperature difference between the micrometer and the gauge block ($ \delta\theta $)

Distribution: Rectangular

$ u(\delta\theta)=\frac{0,3999}{2\sqrt{3}}=0,11547 $

Coefficient of thermal expansion ($ \alpha_S $)

Average of $ 11,5 \times 10^{-6}~{}^\circ C^{-1} $.

Calculation of uncertainty type A~($ u(\Delta_l) $

Screenshot_9

Table 3.2.3: Calculation of type A uncertainty.

micro

Calculation of Sensitivity Coefficients

Reference Value $ |i_S\times\theta| $ $ |l_S\times\alpha| $
25 $ 25\times 0,1=2,5 $ $ 25\times0,0000115=0,0002875 $
27,5 $ 27,5\times 0,1=2,75 $ $ 27,5\times0,0000115=0,0003163 $
30,1 $ 30,1\times 0,1=3,01 $ $ 30,1\times0,0000115=0,0003462 $
32,7 $ 32,7\times 0,1=3,27 $ $ 32,7\times0,0000115=0,0003761 $
35,3 $ 35,3\times 0,1=3,53 $ $ 35,3\times0,0000115=0,0004069 $
37,9 $ 37,9\times 0,1=3,79 $ $ 37,9\times0,0000115=0,0004369 $
40 $ 40\times 0,1=4 $ $ 40\times0,0000115=0,00046 $
42,6 $ 42,6\times 0,1=4,26 $ $ 42,6\times0,0000115=0,0004899 $
45,2 $ 45,2\times 0,1=4,52 $ $ 45,2\times0,0000115=0,0005198 $
47,8 $ 47,8\times 0,1=4,78 $ $ 47,8\times0,0000115=0,0005497 $
50 $ 50\times 0,1=5 $ $ 50\times0,0000115=0,000575 $

Table 3.2.4: Calculation of sensitivity coefficients.

Contribution to Uncertainty

Reference Value $ |l_S \theta|\times u(\delta\alpha) $ $ |l_S\alpha|\times u(\delta \theta) $
25 $ 2,5\times0,00000115=0,0000029 $ $ 0,0002875\times0,11547=0,000033 $
27,5 $ 2,75\times0,00000115=0,0000032 $ $ 0,0003163\times0,11547=0,000037 $
30,1 $ 3,01\times0,00000115=0,0000035 $ $ 0,0003462\times0,11547=0,00004 $
32,7 $ 3,27\times0,00000115=0,0000038 $ $ 0,0003761\times0,11547=0,000043 $
35,3 $ 3,53\times0,00000115=0,0000041 $ $ 0,0004059\times0,11547=0,000047 $
37,9 $ 3,79\times0,00000115=0,0000044 $ $ 0,0004359\times0,11547=0,00005 $
40 $ 4\times0,00000115=0,0000046 $ $ 0,00046\times0,11547=0,000053 $
42,6 $ 4,26\times0,00000115=0,0000049 $ $ 0,0004899\times0,11547=0,000057 $
45,2 $ 4,52\times0,00000115=0,0000052 $ $ 0,0005198\times0,11547=0,00006 $
47,8 $ 4,78\times0,00000115=0,0000055 $ $ 0,0005497\times0,11547=0,000063 $
50 $ 5\times0,00000115=0,0000058 $ $ 0,000575\times0,11547=0,000066 $

Table 3.2.5: Contribution to Combined Uncertainty

Combined Uncertainty ($ u_c(l) $)

Uncertainty Calculation Worksheet

The combined uncertainty is given by:

$$u_c(l)=\sqrt{u^2(l_S)~+~u^2(\Delta_l)~+~u^2(Res)~+~u^2(Ep)~+~(~-~l_S~\theta)^2~u^2(\delta\alpha)~+~(~-~l_S~\alpha)^2~u^2(\delta\theta)}$$

$$u_c(l)=\sqrt{0,00004^2+0,0003^2+0,00029^2+0,00017^2+(2,5)^2~(1,15\times10^{-6})^2+(0,00028)^2~0,11^2}$$

$$=0,0004766$$

VR $ u(l_S) $ $ u(\Delta_l) $ $ u(Res) $ $ u(Ep) $ $ |-l_S\thteta|\times u(\delta\alpha) $ $ |-l_S\thteta|\times u(\delta\theta) $ $ u_c(l) $
25 0,00004 0,00033 0,00029 0,00017 0,0000029 0,000033 0,00047
27,5 0,000035 0,00033 0,00029 0,00017 0,0000032 0,000037 0,00047
30,1 0,000045 0,00033 0,00029 0,00017 0,0000035 0,00004 0,00047
32,7 0,00004 0 0,00029 0,00017 0,0000038 0,000043 0,00034
35,3 0,00004 0 0,00029 0,00017 0,0000041 0,000047 0,00034
37,9 0,00004 0,00033 0,00029 0,00017 0,0000044 0,00005 0,00048
40 0,000045 0,00033 0,00029 0,00017 0,0000046 0,000053 0,00048
42,6 0,000045 0,00033 0,00029 0,00017 0,0000049 0,000057 0,00048
45,2 0,000045 0,00033 0,00029 0,00017 0,0000052 0,00006 0,00048
47,8 0,00004 0,00033 0,00029 0,00017 0,0000055 0,000063 0,00048
50 0,000045 0,00033 0,00029 0,00017 0,0000058 0,000066 0,00048

Table 3.2.6: Calculation of Combined Uncertainty - Per Point

In Table (3.2.6), we calculate the combined uncertainty for each calibration point.

Effective degrees of freedom

$$\nu_{eff}(l)~=~\left(\frac{u_c(l)}{u(\Delta_l)}\right)^4~\nu_A=\left(\frac{0,00047}{0,000333}\right)^4~(3-1)=8,36$$

Through the t-Student table we find k =2.31

Expanded Uncertainty (U(l))

$$U(l)~=~k~\times~u_c(l)=2,31\times 0,0004766=0,0011$$

Therefore, we summarize the results of the uncertainty calculation at the 25mm point in Table (3.2.7)

Screenshot_10

Table 3.6.7: Uncertainty Calculation Results for the 25 mm point.

Assessment of uncertainty

Suppose this equipment is used to measure a tolerance of 0.02 mm.

Considering J=10, we have

$$EMP=\frac{{Tolerância}}{10}=\frac{0,02}{10}=0,002~mm$$

$$|T|+U=|0,00033|+0,0011=0,00243\leq 0,002=EMP$$

Therefore, for the 25 mm point, it was approved.

18.3.3 Calculation of Uncertainty of a Manometer

Consider the process of calibrating a manometer in comparison with a standard manometer, as shown in figure 3.3.1.

manometro_makil

Figure 3.3.1: Pressure gauge.

The data relating to the equipment involved in calibration are given by:

Object to be calibrated

Manometer.

Resolution: $ 1 kgf/cm^2 $.

Indication Range (nominal range): $ 0 - 160 kgf/cm^2 $.

Reference standard

Standard pressure gauge.

Resolution: $ 0,1 kgf/cm^2 $

Expanded uncertainty - $ U = 0,1(porcentagem)~(k=2) $ (in relation to full scale).

Temperature during calibration: $ 22~\pm~1{}^\circ C $.

Indication range (nominal range): $ 0 - 200 kgf/cm^2 $.

Results

The bench was adjusted with the manometer (to be calibrated) and readings were taken with the standard. Table 3.3.1 presents the data.

Screenshot_12

Table 3.3.1: Data table for the pressure gauge.

manometro

Mathematical Equation

$${Desvio} = {Medida~do~Manômetro} - {Manômetro~padrão}$$

Sources of Uncertainty

  • Repetitiveness - Type A;
  • Manometer resolution;
  • Standard pressure gauge resolution;
  • Uncertainty inherited from the pattern $ U = 0,1(porcentagem)~(k=2) $.

Percentage taken in relation to full scale

$$U = \frac{U(unidade)}{200}~100(porcentagem).$$

So

$$U(unidade)~=~\frac{U((porcentagem))}{100}~200$$

$$u_{c}~=~\frac{U(unidade)}{k}$$

Hysteresis

$$U(hist)=|\overline{{Avanço}}-\overline{{Retorno}}|$$

$$u(hist)=\cfrac{U(hist)}{2\sqrt{3}}$$

manometro1

Results Obtained from Hysteresis:

manometro2

Combined Uncertainty

Through the uncertainty propagation equation, we have that the combined uncertainty expression for the manometer is given by

$$u_c(M)=\sqrt{~u^2(\Delta_l)~+~u^2(Res(man))~+~u^2(Res(padra))~+~u^2(h_{Pad})~+~u^2(Hist)}$$

in which,

$ u(\Delta_l) $: represents uncertainty due to repetitiveness;

$ u(Res(man)) $: represents the uncertainty due to the manometer resolution;

$ u(Res(padra)) $: represents the uncertainty due to the resolution of the standard manometer;

$ u(h_{Pad}) $: represents the uncertainty inherited from the standard pressure gauge;

$ u(Hist) $: represents uncertainty due to hysteresis.

Calculation of Standard Uncertainty of Input Quantities

Next, we will calculate the uncertainty for each source.

Repetitiveness ($ \Delta_l $)

Type A uncertainty.

$$u(\Delta_l) = \frac{s}{\sqrt{n}}$$

in which

$ u(\Delta_l) $: represents Type A uncertainty or uncertainty due to repetitiveness;

$ s $: represents the standard deviation of the measurement range;

n: represents the number of calibration points.

Inherited Uncertainty from the Standard Manometer $ (u(h_{Pad})) $

Distribution: Normal.

$ u(h_{Pad}) =\frac{0,1(porcentagem)\text{Fundo de Escala}}{2}=\frac{0,001\times 200}{2}=0,1 $

Standard Manometer Resolution $ (u(Res(padra))) $

Distribution: Rectangular.

$ u(Res(padra)) =\frac{0,1}{2\sqrt{3}}=0,0288 $

Manometer Resolution $ (u(Res(man))) $

Distribution: Rectangular.

$ u(Res(man)) =\frac{1}{2\sqrt{3}}=0,288 $

Hysteresis Calculation

Range Average Advance Average Return |Hysteresis|
15 14,867 14,9 0,033333
30 29,767 29,667 0,1
45 45,0667 44,667 0,4
60 60,067 59,767 0,3
75 75 74,667 0,333333
90 89,867 89,567 0,3
105 104,9 104,567 0,333333
120 119,867 119,733 0,133333
140 140,1 140,067 0,033333
160 160,233 160,233 0

30 range hysteresis $ kgf $:

$$U(hist)=|\overline{{Avanço}}-\overline{{Retorno}}|=|29,767-29,667|=0,1$$

$$u(hist)=\cfrac{U(hist)}{2\sqrt{3}}=\frac{0,1}{2\sqrt{3}}=0,0288$$

Calculation of Manometer Uncertainty

Combined Uncertainty

$$u_c(M)=\sqrt{~u^2(\Delta_l)~+~u^2(Res(man))~+~u^2(Res(padra))~+~u^2(h_{Pad})~+~u^2(Hist)}=$$

$$=\sqrt{(0,030731815)^2~+~(0,288675135)^2~+~(0,02887)^2~+~(0,1)^2~+~(0,028867)^2}=$$

$$=0,3097$$

Effective Degrees of Freedom

$$\nu_{eff}~=~\left(\frac{u_c(M)}{u(\Delta_l)}\right)^4~\nu_A=\left(\frac{0,3097}{0,0307}\right)^4~(6-1)=51600,84$$

Through the t-Student table, we find $ k =2 $

Expanded Uncertainty

$$U=k\times u_c(M)=2\times 0,3097=0,6194=0,62~kgf/cm^2$$

Screenshot_13

Expression of Uncertainty in Percentage

$$U = \frac{U(unidade)}{\overline{M}}~100(porcentagem)$$

for the 30 range $ kgf/cm^2 $ we have

$$U = \frac{0,619}{29,717}~100(porcentagem)=2,08(porcentagem)$$

The pressure gauge is used to measure a tolerance of $ 10kgf/cm^2 $. We will now carry out a study of the metrological proof.

Considering $ J=10 $, we have to

$$EMP=\frac{{Tolerância}}{10}=\frac{10}{10}=1~kgf/cm^2$$

$$|T|+U=|-0,283|+0,619=0,902\leq 1=EMP$$

Therefore, to the point of $ 30~kgf/cm^2 $, was approved.

18.3.4 Calculation of Uncertainty of a Digital Voltmeter

Consider calibrating a digital voltmeter by comparison.

Code: VO341.

Series: 006-C.

Description: digital voltmeter.

Smallest Div: $ 0,1~mV $.

Unit: $ mV $.

Reading range: $ 0 $ to $ 200~mV $.

Ambient temperature: 20 $ plusmn; 3^oC $.

Reference standard

HP Digital Multimeter - 3458$ ^A $.

Resolution: $ 0,01mV $.

Expanded uncertainty - $ U = \pm0,001(porcentagem) $ with $ k=2 $ (in relation to full scale).

Drift (instability) = $ \pm 0,002mV $.

Reading range: $ 0 $ to $ 200~mV $.

Results:

The bench was adjusted with the digital voltmeter (to be calibrated) and readings were taken with the standard. Table 3.4.1 presents the data.

Screenshot_14

Table 3.4.1: Data table.

voltimetro

Mathematical Model

$${Desvio} = {Voltagem~indicada~no~equipamento~em~calibração}~-~{Voltagem~do~padrão~de~referência}.$$

Uncertainty Worksheet

Screenshot_15

Combined Uncertainty

$$u_c(Volt)=\sqrt{~u^2(\Delta_l)~+~u^2(Res(Pad))~+~u^2(Res({Volt}))~+~u^2(Pad)~+~u^2({Drift})}=$$

$$=\sqrt{(0,0119)^2~+~(0,0028)^2~+~(0,0011)^2~+~(0,001)^2~+~(0,0288)^2}$$

$$=0,03139$$

Effective Degrees of Freedom

$$\upsilon_{eff}= \left(\frac{u_c(y)}{u_A(y)}\right)^4\nu_A =$$

$$= \left(\frac{0,03139}{0,0119}\right)^4(4-1) =145,22$$

Screenshot_17

Expanded Uncertainty

$$U(Volt)=u_c(Volt)\times k=$$

$$=0,03139\times 1,9765=0,062052$$

Thus, we obtain the following results:

Screenshot_16

Uncertainty assessment: Assume this equipment is used to measure a tolerance of 3$ mV $. We will critically analyze the ability of this equipment to measure such tolerance.

Criteria:

$$|T|+U\leq EMP$$

We set the Maximum Permissible Error (MPE) to J=10

$$EMP=\frac{{Tolerância}}{J}=\frac{3}{10}=0,3$$

$$|T|+U=|0,103|+0,062=0,19705\leq 0,3=EMP$$

Therefore, to the point of $ 40~mV $, was approved.

18.3.5 Calculation of Uncertainty of a Thermometer

Object to be calibrated

Thermometer.

Resolution: $ 0,5^oC $.

Indication Range (nominal range): $ 0~ a~ 100^oC $.

Reference standard

Standard thermometer.

Resolution: $ 0,1 $ºC

Expanded uncertainty: $ U = 0,25^oC~(k=2) $.

Indication range (nominal range): $ -40~ a~ 500^oC $.

Bath

Lack of Bath Homogeneity: $ \pm 0,05^o C $

Results

The bench was adjusted with the thermometer (to be calibrated) and readings were taken with the standard. Table 3.5.1 presents the data.

Screenshot_18

Table 3.5.1: Data table for the thermometer.

termometro

Mathematical Equation

$${Desvio} = {Medida~do~termômetro} - {termômetro~padrão}$$

Sources of Uncertainty

  • Repetitiveness - Type A;
  • Thermometer resolution;
  • Standard thermometer resolution;
  • Uncertainty inherited from the pattern;
  • Lack of Bath Homogeneity.

Combined Uncertainty

Through the uncertainty propagation equation, we have that the combined uncertainty expression for the thermometer is given by

$$u_c(M)=\sqrt{~u^2(\Delta_l)~+~u^2(Res(term))~+~u^2(Res(pad))~+~u^2(h_{Pad})~+~u^2(Banho)}$$

where

  • $ u(\Delta_l) $: represents uncertainty due to repetitiveness;
  • $ u(Res(term)) $: represents the uncertainty due to the resolution of the thermometer;
  • $ u(Res(pad)) $: represents the uncertainty due to the resolution of the standard thermometer;
  • $ u(h_{Pad}) $: represents the uncertainty inherited from the standard thermometer;
  • $ u(Banho) $: represents uncertainty due to the lack of homogeneity of the bath.

Calculation of Standard Uncertainty of Input Quantities

Next, we will calculate the uncertainty for each source.

Repetitiveness ($ \Delta_l $)

Type A uncertainty.

$$u(\Delta_l) = \frac{s}{\sqrt{n}}$$

to the range $ 25^oC $ we have

$$u(\Delta_l) =\frac{0,0577}{\sqrt{4}}=0,028$$

where

$ u(\Delta_l) $: represents Type A uncertainty or uncertainty due to repetitiveness;

$ s $: represents the standard deviation of the measurement range;

n: represents the number of calibration points.

Inherited Uncertainty from the Standard Thermometer $ (u(h_{Pad})) $

Distribution: Normal.

$ u(h_{Pad}) =\frac{0,25}{2}=0,125 $

Standard thermometer resolution $ (u(Res(pad))) $

Distribution: Rectangular.

$ u(Res(pad)) =\frac{0,1}{2\sqrt{3}}=0,028 $

Thermometer resolution $ (u(Res(term))) $

Distribution: Rectangular.

$ u(Res(term)) =\frac{0,5}{2\sqrt{3}}=0,144 $

Calculation of thermometer uncertainty

Screenshot_19

Combined uncertainty

$$u_c(Term)=\sqrt{(0,028)^2~+~(0,028)^2~+~(0,028)^2~+~(0,125)^2~+~(0,144)^2}=0,1973$$

Effective degrees of freedom

$$\nu_{eff}~=~\left(\frac{u_c(Term)}{u(\Delta_l)}\right)^4~\nu_A~=$$

$$=~\left(\frac{0,195789}{0,028}\right)^4~(4-1)~=6556,68$$

Through the t-Student table, we find $ k =1,96 $

Expanded Uncertainty

$$U~=~k~\times~u_c(Term)~=1,96\times 0,1973=0,3869$$

The thermometer is used to measure a tolerance of $ 5ordm;C $. We will now carry out a study of the metrological proof.

Considering $ J=10 $, we have to

$$EMP=\frac{{Tolerância}}{10}=\frac{5}{10}=0,5^oC$$

$$|T|+U=|0,05|+0,3869=0,43693\leq 0,5=EMP$$

Therefore, to the point of $ 25^oC $, was approved.

18.3.6 Calculation of uncertainty of a type N thermocouple at 1000ºC

Consider the calibration process of a type N thermocouple. This calibration is carried out by comparison using two type R thermocouples in a horizontal oven at a temperature of 1000ºC. The electromotive forces (emf) generated by the thermocouples are measured by a digital voltmeter using a selector switch. The thermocouple references are at 0ºC and the thermocouple to be calibrated is connected to the reference point using compensation cables.

To calculate the uncertainty, we will divide it into two steps, first we calculate the uncertainty of the temperature of the hot junction of the thermocouple to be calibrated, then the voltage between the thermocouple wires and the cold junction at 0ºC during calibration.

Uncertainty due to hot junction temperature

Next, we present the hot junction temperature measurement equation.

$$t_X=t_S\left(V_{iS}+\delta V_{iS1}+\delta V_{iS2}+\delta V_R-\frac{\delta t_{0S}}{C_{S0}}\right)+\delta t_D+\delta t_F+\delta t_S$$

$$\cong t_S(V_{iS})+C_S~\delta V_{iS1}+C_S~\delta V_{iS2}+C_S~\delta V_R-\frac{C_S}{C_{S0}}\delta t_{0S}+\delta t_D+\delta t_F+\delta t_S$$

in which

  • temperature of the reference thermometer in terms of voltage with the cold junction at 0ºC (calibration certificate) in °C. From the certificate we have that U(tS)=0.3ºC with k=2;
  • voltmeter indication in μV;
  • voltage corrections obtained from voltmeter calibration in μV. From the certificate we have that U=2 μV with k=2;
  • voltage corrections due to the limited resolution of the voltmeter in μV. Here we use a 4 1/2-digit microvoltmeter in the 10 mV range, resulting in resolution limits of ±0.5 μV;
  • voltage correction due to contact effects of the inverting switch in μV. The residual parasitic offset voltages due to the switch contacts were estimated to be zero within the limits of ±2 μV;
  • temperature corrections due to the deviation of the reference temperature in relation to the value in ºC. The set point temperature of each thermocouple is known to be 0ºC within the limits ±0.1ºC;
  • sensitivity of thermocouples in terms of voltage at the measuring temperature of 1000ºC in ºC/μV, in this case the value is 0.077ºC/μV;
  • sensitivity of thermocouples in terms of voltage at the reference temperature at 0ºC in ºC/μV;
  • variation in reference thermometer values ​​since their last calibration due to drift in ºC. From previous calibrations the drifts were estimated to be zero within the limits of ±0.3 ºC;
  • temperature correction due to non-uniformity of the oven temperature in °C. In this case, deviations from temperature inhomogeneity in the measuring region are within ±1 ºC;
  • standard deviation grouped between the 1st reference, 2nd reference and the thermocouple to be calibrated.

First, let’s calculate the pooled standard deviation $ t_S, $ To do this, observe the following table.

Screenshot_20

Converting voltage into temperature we have

Screenshot_21

Thus, the pooled standard deviation is given by

$$s_p=u(t_S)=\sqrt{\frac{(n_1-1)s^2_1+(n_2-1)s^2_2+(n_3-1)s^2_3}{n_1+n_2+n_3-3}}=0,1ordm;C$$

Below, we present the data for the uncertainty calculation:

Screenshot_22

Therefore, the combined uncertainty is given by:

$$u^2_c(t_X)=u^2(t_S)+\left(\frac{\partial t_X}{\partial \delta V_{iS1}}\right)^2 u^2(\delta V_{iS1})+\left(\frac{\partial t_X}{\partial \delta V_{iS2}}\right)^2 u^2(\delta V_{iS2})+\left(\frac{\partial t_X}{\partial \delta V_{R}}\right)^2 u^2(\delta V_{R})+$$

$$+\left(\frac{\partial t_X}{\partial \delta t_{0S}}\right)^2 u^2(\delta t_{0S})+\left(\frac{\partial t_X}{\partial \delta t_{D}}\right)^2 u^2(\delta t_{D})+\left(\frac{\partial t_X}{\partial \delta t_{F}}\right)^2 u^2(\delta t_{F})+u^2(\delta t_S)=$$

$$=0,1^2+0,077^2 1^2+0,077^2 0,29^2+0,077^2 1,15^2+(-0,407)^2 0,058^2+1^2 0,173^2+1^2 0,577^2+0,15^2$$

in which

$$\left(\frac{\partial t_X}{\partial \delta V_{iS1}}\right)=\left(\frac{\partial t_X}{\partial \delta V_{iS2}}\right)=\left(\frac{\partial t_X}{\partial \delta V_{R}}\right)=C_S=0,077$$

$$\left(\frac{\partial t_X}{\partial \delta t_{0S}}\right)=\frac{-C_S}{C_{S0}}=-\frac{0,077}{0,189}=-0,407$$

$$\left(\frac{\partial t_X}{\partial \delta t_{D}}\right)=\left(\frac{\partial t_X}{\partial \delta t_{F}}\right)=1$$

Therefore, the combined uncertainty is given by

$$u_c(t_X)=0,641$$

Thus, the expanded uncertainty is given by:

$$U(t_X)=k\times u_c(t_X)=2\times 0,641=1,282$$

Since we only have type B uncertainty, we adopt $ k=2. $

Below, we present a summary of the uncertainty calculation.

Screenshot_23

Uncertainty due to the emf of the thermocouple to be calibrated

Below, we present the equation for measuring the voltage between the thermocouple wires with the cold junction (reference) at 0ºC, during calibration.

$$V_X\cong V_X(t_X)+\frac{\Delta t}{C_X}-\frac{\delta t_{0X}}{C_{X0}}$$

$$=V_{iX}+\delta V_{iX1}+\delta V_{iX2}+\delta V_R+\delta V_{LX}+\frac{\Delta t}{C_X}-\frac{\delta t_{0X}}{C_{X0}}$$

in which

  • voltmeter indication in μV;
  • voltage corrections obtained from voltmeter calibration in μV. From the certificate we have that U=2 μV with k=2;
  • voltage corrections due to the limited resolution of the voltmeter in μV. Here we use a 4 1/2-digit microvoltmeter in the 10 mV range, resulting in resolution limits of ±0.5 μV;
  • voltage correction due to contact effects of the inverting switch in μV. The residual parasitic offset voltages due to the switch contacts were estimated to be zero within the limits of ±2 μV;
  • temperature corrections due to the deviation of the reference temperature in relation to the value in ºC. The set point temperature of each thermocouple is known to be 0ºC within the limits ±0.1ºC;
  • sensitivity of thermocouples in terms of voltage at the measuring temperature of 1000ºC in ºC/μV, in this case the value is 0.026ºC/μV;
  • sensitivity of thermocouples in terms of voltage at the reference temperature at 0ºC in ºC/μV, in this case the value is 0.039ºC/μV;
  • thermocouple temperature calibration point;
  • deviation of the calibration point temperature in relation to the oven temperature, in ºC;
  • voltage correction due to compensation cables (range 0 to 40ºC). Voltage differences between cables and thermocouple wires are within ±5 μV limits.

First, let’s calculate the repeatability.

Screenshot_24

$$u(V_{iX})=\frac{s}{\sqrt{n}}=\frac{2,449}{\sqrt{4}}=1,225$$

Below, we present the data for the uncertainty calculation:

Screenshot_25

Therefore, the combined uncertainty is given by:

$$u_c(V_X)=u^2(V_{iX})+\left(\frac{\partial V_X}{\partial \delta V_{iX1}}\right)^2 u^2(\delta V_{iX1})+\left(\frac{\partial V_X}{\partial \delta V_{iX2}}\right)^2 u^2(\delta V_{iX2})+\left(\frac{\partial V_X}{\partial \delta V_{R}}\right)^2 u^2(\delta V_{R})+$$

$$+\left(\frac{\partial V_X}{\partial \delta V_{LX}}\right)^2 u^2(\delta V_{LX})+\left(\frac{\partial V_X}{\partial \Delta t}\right)^2 u^2(\Delta t)+\left(\frac{\partial V_X}{\partial \delta t_{0X}}\right)^2 u^2(\delta t_{0X})=$$

$$=1,225^2+1^2~1^2+1^2~0,29^2+1^2~1,15^2+1^2~2,89^2+38,5^2~0,641^2+25,6^2~0,058^2$$

in which

$$\left(\frac{\partial V_X}{\partial \delta V_{iX1}}\right)=\left(\frac{\partial V_X}{\partial \delta V_{iX2}}\right)=\left(\frac{\partial V_X}{\partial \delta V_{R}}\right)=\left(\frac{\partial V_X}{\partial \delta V_{LX}}\right)=1$$

$$\left(\frac{\partial V_X}{\partial \Delta t}\right)=\frac{1}{C_{X}}=38,5$$

$$\left(\frac{\partial V_X}{\partial \delta t_{0X}}\right)=\frac{1}{C_{X0}}=25,6$$

Therefore, the combined uncertainty is given by

$$u_c(V_X)=24,94$$

Thus, the expanded uncertainty is given by:

$$U(V_X)=k\times u_c(V_X)=2\times 24,94=49,88$$

where k is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and confidence level of approximately 95.45%.

$$\nu_{eff}(V_X)=\frac{u^4(V_X)}{u^4(V_{iX})}(n-1)=\frac{24,94^4}{1,22^4}=515856,11$$

Below, we present a summary of the uncertainty calculation.

Screenshot_26

18.3.7 Calculation of uncertainty of a Caliper

Consider the process of calibrating a digital caliper with a range of 20 mm and a resolution of 0.01 mm. This calibration is carried out by comparison using a caliper calibrator and standard reference blocks. At each test level we obtain a calibration deviation ΔL which is the difference between indicated by the instrument to be calibrated and the conventionally true value of the reference standard. Given by the following equation:

$$d=L-L_s$$

in which

  • L: represents the reading obtained by the caliper;
  • $L_S$: represents the length of the standard block for a temperature of 20ºC.

Another determining factor in this measurement equation is temperature, so the correction due to temperature is given by:

$$d=L(1+\alpha\theta)-L_s(1+\alpha_s \theta_s)$$

in which,

  • $ \alpha $ and $ \alpha_s $: correspond to the thermal expansion coefficients of the caliper (scale) and the standard block used in the measurement, respectively;
  • $ \theta $ and $ \theta_s $: correspond to the temperature difference between the caliper and the standard block, in relation to 20ºC, respectively.

Mathematical Model

The reading of the block length obtained from the caliper is given by

$$L=\frac{d+L_s(1+\alpha_s \theta_s)}{(1+\alpha \theta)}\quad (*)$$

Multiplying and dividing by $ (1-\alpha_X\theta) $, we get

$$L=\frac{d+L_s(1+\alpha_s \theta_s)}{(1+\alpha \theta)}\frac{(1-\alpha\theta)}{(1-\alpha\theta)}=\frac{d(1-\alpha \theta)+L_S(1+\alpha_s\theta_s)(1-\alpha\theta)}{1-(\alpha\theta)^2}$$

Screenshot_29

Simplifying the above equation with respect to negligible terms, we can approximate the value of the standard block, for

$$L=d+L_S+L_S(\alpha_s\theta_s-\alpha\theta)$$

When denoting the difference between the temperatures of the caliper and the standard block, for

$$\delta\theta=\theta-\theta_s$$

and the difference between the coefficients of thermal expansion of the caliper and the gauge block, e.g.

$$\delta\alpha=\alpha-\alpha_s$$

we get that

$$L=d+L_s-L_s(\delta\alpha\times \theta+\delta\theta\times \alpha_s)$$

The deviations obtained present the following sources of uncertainty

$$d=\Delta_L+\text{Res}+E_p$$

in which,

ΔL: represents the difference observed in the length of the standard block (repeatability); caliper measurement - gauge block measurement;

Res: represents the resolution of the caliper;

Ep: represents the parallelism of the caliper.

Thus, the mathematical equation for block length is:

$$L=\Delta_L+\text{Res}+E_p+L_s-L_s(\delta\alpha\times \theta+\delta\theta\times \alpha_s)$$

Sources of uncertainty

  • Resolution (Res): The resolution of the caliper is 0.01mm. This variable is considered to have a zero mean and variation limits defined by the equipment resolution;
  • Uncertainty inherited from the pattern block (LS): This variable has a mean given by the standard value with a standard deviation defined by the combined uncertainty, both declared in the calibration certificate (U=0.0008 mm, k=2);
  • Parallelism (Ep): The parallelism between the faces is considered a variable with zero mean and variation limits of ±0.0025mm (result obtained as the maximum difference between the readings with the standard blocks at different points on the measuring rods);
  • Temperature corrections: Before calibration, care is taken to ensure that the gauge block set and micrometer are at the same ambient temperature as the measuring room;
  1. Temperature difference between the micrometer and the standard block (δθ): maximum difference of ±0.2ºC, with zero average;
  2. Thermal expansion coefficient of the standard block (αs): average of 11.5×10-6ºC-1 with a limit of ±2×10-6ºC-1;
  3. Difference between thermal expansion coefficients (δα): zero average with limits ±2×10-6ºC-1;
  4. Difference between the caliper temperature and the reference temperature tr=20ºC (θ): limits of ±0.2ºC.

Furthermore, the sources of uncertainty θ, δα, δθ and αs are assumed uncorrelated. Using equation (*), a reading obtained from the caliper for the length of the block, we observe that the estimated value for length is given by

$$L = (\text{Média do desvios}) + L_s$$

NOTE: This equation is the estimate of the caliper measurement that is obtained by disregarding the variables that have a mean of zero.

Combined Uncertainty Assessment (uc(L))

Using the uncertainty propagation equation, we have

$$u_c(L)=\sqrt{c^2_{L_s}~u^2(L_s)+c^2_d~u^2(d)+c^2_{\alpha_s}~u^2(\alpha_s)+c^2_\theta~u^2(\theta)+c^2_{\delta\alpha}~u^2(\delta\alpha)+c^2_{\delta\theta}~u^2(\delta\theta)}$$

where the sensitivity coefficients evaluated at the mean point are given by:

  • $ c_{L_s}=\dfrac{\partial L}{\partial L_s}=1-\underbrace{(\delta\alpha~\theta~+~\delta\theta~\alpha_S)}_{\text{desprezível}}=1 $;
  • $ c_d=\dfrac{\partial L}{\partial d}=1 $;
  • $ c_{\alpha_s}=\dfrac{\partial L}{\partial \alpha_s}=-L_s~\delta\alpha=0 $ (Average of $ \delta\alpha $ is equal to zero);
  • $ c_{\theta}=\dfrac{\partial L}{\partial \theta}=-L_s~\delta\theta=0 $ (Average of $ \delta\theta $ is equal to zero);
  • $ c_{\delta\alpha}=\dfrac{\partial L}{\partial \delta\alpha}=-L_s~\theta $;
  • $ c_{\delta\theta}=\dfrac{\partial L}{\partial \delta \theta}=-L_s~\alpha $.

Therefore, the combined uncertainty expression is given by

$$u_c(L)=\sqrt{u^2(L_s)+u^2(\Delta_l)~+~u^2(Res)~+~u^2(Ep)+(-L_s~\theta)^2~u^2(\delta\alpha)+(-L_s~\alpha)^2~u^2(\delta\theta)}$$

Calculation of Standard Uncertainty of input quantities

Next, we will calculate the uncertainty for each source.

Reference standard ($L_S$)

Distribution: Normal

$$u(L_s)=\dfrac{0,008}{2}=0,004$$

Parallelism ($E_p$)

Distribution: Rectangular

$$u(E_p)=\dfrac{0,005}{2\sqrt{3}}=0,002887$$

Caliper resolution (Res)

Distribution: Rectangular

$$u(Res)=\dfrac{0,01}{2\sqrt{3}}=0,0014$$

Difference between thermal expansion coefficients ($ \delta\alpha $)

Distribution: Rectangular

$ u(\delta\alpha)=\dfrac{4\times 10^{-6}}{2\sqrt{3}}=1,15\times 10^{-6} $

Difference between caliper temperature and reference temperature $t_0$=20ºC (θ)

An average temperature of 20.2ºC was recorded, resulting in θ=20.2-20 =0.2ºC.

Temperature difference between the caliper and the gauge block (δθ)

Distribution: Rectangular

$$u(\delta\theta)=\dfrac{0,4}{2\sqrt{3}}=0,115$$

Coefficient of thermal expansion ($ \alpha_s $)

Average of 11.5 x 10-6 ºC-1.

Calculation of type A uncertainty u(ΔL)

Next, we will calculate the repeatability (type A uncertainty)

Screenshot_27

Table 3.2.3: Calculation of type A uncertainty.

Type A uncertainty is calculated as follows:

$$u(\Delta L)=\dfrac{s}{\sqrt{n}}=\frac{0,004472}{\sqrt{5}}=0,002~\text{mm}$$

Now, let’s calculate by applying Action.

(image missing)

Combined uncertainty

Therefore, the combined uncertainty is given by

$$u^2_c(L)=u^2(L_s)+u^2(\Delta_l)+u^2(Res)+u^2(Ep)+(-L_s~\theta)^2~u^2(\delta\alpha)+(-L_s~\alpha)^2~u^2(\delta\theta)=$$

$$=0,004^2+0,002^2+0,005^2+0,0014^2+$$

Screenshot_30

So, we have to

$$u_c(L)=0,0069~\text{mm}$$

Effective degrees of freedom

$$\nu_{eff}(L)~=~\left(\frac{u_c(L)}{u(\Delta_L)}\right)^4~(n-1)=\left(\frac{0,0069}{0,002}\right)^4~(5-1)=555,92$$

Through the t-Student table we find k =2

Expanded Uncertainty (U(L))

$$U(L)=k\times u_c(L)=2\times 0,0069=0,014~\text{mm}$$

Therefore, we summarize the results of the uncertainty calculation at the 200 mm point below

Screenshot_28

18.3.8 Calculation of uncertainty of a measuring tape

Consider the calibration process of a tape measure compared to a graduated ruler, as shown in the figure.

(image missing)

Figure 3.13.1: Measurement scheme.

The data relating to the equipment involved in calibration are given by:

Object to be calibrated: steel tape measuring tape (5 m).

Resolution: $ 1 $ mm.

Indication Range (nominal range): $ 0 - 5000 $ mm.

Reference standard 1: graduated ruler.

Resolution: $ 0,01 $ mm.

Expanded uncertainty: $ U = 0,02~\text{mm}~(k=2). $

Reference standard 2: graduated magnifying glass.

Resolution: $ 0,01 $ mm.

Expanded uncertainty: $ U = 0,004~mm~(k=2). $

Temperature during calibration: $ 20~\pm~6{}^\circ C $.

Indication range (nominal range): $ 1000 - 5000 mm $

(image missing)

Figure 3.13.2: Measurement by comparison.

Measuring equation

$${Desvio} = \text{Medida da trena} - \underbrace{(\text{Medida da Régua graduada}+\text{Medida da Lupa graduada})}_{\text{P}=\text{medidas do padrão}}$$

Sources of Uncertainty

  • ε: Repetitiveness - Type A;
  • $Res_T$: Resolution of the tape measure;
  • $Res_R$: Resolution of the graduated ruler;
  • $Res_L$: Resolution of the graduated magnifying glass;
  • A: Uncertainty inherited from pattern 1 (graduated ruler);
  • L: Uncertainty inherited from pattern 2 (graduated magnifying glass);

In this application, let’s take the point at 2000 mm. Therefore, observe the following table.

Screenshot_32

Combined Uncertainty

Through the uncertainty propagation equation, we have the expression of the combined uncertainty for the tape as follows.

$$u_c(T)=\sqrt{u^2(\varepsilon)+u^2(\text{Res}_T)+u^2(\text{Res}_L)+u^2(\text{Res}_R)+u^2(L)+u^2(R)}$$

Calculation of Standard Uncertainty of Input Quantities

Next, we will calculate the uncertainty for each source.

Repeatability (ε)

Type A uncertainty.

$$u(\varepsilon)=\frac{s}{\sqrt{n}}=\frac{0,010954}{\sqrt{5}}=0,004899$$

in which

s: represents the standard deviation of the measurement range;

n: represents the number of calibration points.

Inherited Uncertainty of the Standard Graduated Ruler (u(R))

Distribution: Normal.

$$u(R)=\frac{0,02}{2}=0,01$$

Inherited Uncertainty of the Standard Graduated Magnifier (u(R))

Distribution: Normal.

$$u(R)=\frac{0,004}{2}=0,002$$

Resolution of the Standard Graduated Ruler (u($Res_R$)) and the Standard Graduated Magnifier (u($Res_L$))

Distribution: Rectangular.

$$u(Res_R)=u(Res_L)=\frac{0,01}{2\sqrt{3}}=0,00289$$

Measure Resolution (u($Res_T$))

Distribution: Rectangular.

$$u(Res_T)=\frac{1}{2\sqrt{3}}=0,289$$

Therefore, the combined uncertainty is given by:

$$u_c(T)=\sqrt{0,289^2+0,00289^2+0,00289^2+0,01^2+0,002^2+0,004899^2}=0,289$$

Now, let’s calculate the effective degrees of freedom as follows:

$$\nu_{eff}(T)=\frac{u^4_c(T)}{u^4(\varepsilon)}(n-1)=\frac{0,289^2}{0,004899^4}*4=48392892$$

Therefore, the expanded uncertainty is given by:

$$U(T)=k\times u_c(T)=2\times 0,289=0,58~\text{mm}$$

Screenshot_31

18.4 Applications of Uncertainty Calculation in Tests

In the next modules, we will present some applications of Measurement Uncertainty for tests, to exemplify the calculation procedure and the initial steps before starting to calculate measurement uncertainty.

18.4.1 Calculation of uncertainty for Hardness Measurement System

Hardness measurement system: Brinell (2.5/62.5)

Consider the system for measuring the hardness of a toothed part (photo 4.1.1) on the Brinell scale (2.5/62.5). When we measure the hardness of the part at a certain point, we cause a change in the part around this point and consequently, we cannot repeat the hardness measurement in this region. Thus, we can interpret the hardness measurement system as a non-replicable system. However, in some applications, there is a region in the part where the hardness is homogeneous and in this case, we can use this region to evaluate the hardness measurement, considering this system as replicable. In summary, depending on the characteristics of the part to be measured, we can interpret the hardness measurement system as replicable or not.

In the application we will make below, we have a very homogeneous region in relation to hardness, a fact that allows us to consider the measurement system as replicable. Although we have a region with homogeneous hardness, it is not constant and this variation (even if small) will add to the repeatability of the measuring system.

In calculating uncertainty for the hardness measurement system, we will adopt as sources of uncertainty the hardness meter, the hardness resolution, the repeatability of the test, and the reproducibility of the R&R study. Below we present the uncertainty calculation flowchart.

(image missing)

Next, we present a description of the measurement system.

Phase 1

Description of the Measurement System

We use this phase to carry out a review of the measurement system and planning for the application of the MSA. Here, it is important to define the team responsible for the analysis, the sectors that apply the measurement system and the need for measurement. For example, the part we measure the hardness of (Photo 5.7.1) mainly affects the safety of the end user. Therefore, it is important that we have a reliable measurement system.

(image missing)

Figure 4.1.1: photo of the part (Extruded Aluminum material).

For the hardness characteristic, the customer specified a tolerance of 4.5 HB. For this application, we have a minimum required accuracy of 0.45 HB.

So that the team can better understand the measurement system and its application, we carry out a scenario analysis using the following questions:

  • Importance/impact on the product?
  • Customer Complaints?
  • Failure History?
  • Is it a special feature?
  • Environmental Conditions?
  • Is the measurement system used for CEP?

Phase 2

Measurement Method:

Phase I prepares us to make a good description of the measurement system. Therefore, we gathered the team to describe the measurement system and prepare the Ishikawa diagram.

Description of hardness measurement system: Brinell (2.5/62.5)

  1. Select the scale on the equipment
  2. Clean the part
  3. Position the durometer
  4. Apply load
  5. Perform reading

We also define the measurement system parameterization variables

  • Variable: Hardness;
  • Description: measurement performed at 2.5/62.5 HB;
  • Specification: 4.5 HB tolerance.

Another important point for analyzing the measurement system is the Ishikawa diagram, or cause and effect diagram. With the entire team together (at this stage it is important to have operators present), we brainstormed on the main causes of variation in the measurement system. The Ishikawa diagram is essential for defining the critical points and statistical techniques that we will use to evaluate the measurement system.

Cause and Effect Diagram (Ishikawa)

(image missing)

One of the critical items of a measurement system is the equipment used to carry out the measurements. The equipment is calibrated by appropriate laboratories and the calibration certificates must be critically evaluated against previously defined criteria. In this application, the hardness tester (Reicherter model 187.5/2.5) has a resolution of 0.1 HB, is calibrated annually and has a maximum allowable error of 0.9 HB. With this, we guarantee the required accuracy of 0.45 HRC. Below, we present a photo of the durometer.

(image missing)

With the description of the measurement system and the Ishikawa diagram, we plan the application of statistical techniques. In most processes, we basically perform three analyses: stability, RR and trend and linearity. Next, we describe the planning for applying statistical analyses.

Phase 3

Study Planning.

  • Stability: We select 1 piece to be measured over time. In this application, we will make three rational subgroups of measurements per day for a week (18 subgroups).
  • Trend: The same piece selected for stability will be used for the trend study. In this case, we set the reference value in a reliable measuring system, in this example we use the customer’s measuring system.
  • Repetitiveness: 10 pieces from “distinct” production batches, with 3 measurements per piece carried out in random order.
  • Reproducibility: we selected 3 operators.
  • Linearity: Not applicable, as the measuring range is small.

After data collection, we will perform statistical analysis.

Stability

SAMPLE USED: EXTRUDED ALUMINUM

Subgroup Date Schedule Person Equipment Scale Measure 1 Measure 2 Measure 3
1 20/05/2013 morning A REIHERTER 2,5 / 62,5 74,7 74,3 73,9
2 23/05/2013 afternoon A REIHERTER 2,5 / 62,5 74,3 75,9 76,3
3 24/05/2013 afternoon B REIHERTER 2,5 / 62,5 77,5 74,7 75,5
4 04/06/2013 afternoon C REIHERTER 2,5 / 62,5 74,7 75,5 76,3
5 23/05/2013 morning A REIHERTER 2,5 / 62,5 76,7 75,1 75,9
6 27/05/2013 morning B REIHERTER 2,5 / 62,5 75,0 76,3 75,9
7 24/05/2013 morning C REIHERTER 2,5 / 62,5 76,3 75,9 75,9
8 28/05/2013 afternoon A REIHERTER 2,5 / 62,5 75,1 74,7 75,1
9 29/05/2013 afternoon B REIHERTER 2,5 / 62,5 75,5 75,1 74,3
10 06/06/2013 afternoon C REIHERTER 2,5 / 62,5 74,7 75,1 76,3
11 28/05/2013 morning A REIHERTER 2,5 / 62,5 75,9 75,1 75,9
12 07/06/2013 morning B REIHERTER 2,5 / 62,5 75,5 74,7 75,9
13 07/06/2013 morning C REIHERTER 2,5 / 62,5 73,2 75,5 74,7
14 04/06/2013 afternoon A REIHERTER 2,5 / 62,5 75,5 75,5 74,7
15 04/06/2013 afternoon B REIHERTER 2,5 / 62,5 75,6 75,1 75,5
16 07/06/2013 afternoon C REIHERTER 2,5 / 62,5 76,3 75,1 73,9
17 04/06/2013 morning A REIHERTER 2,5 / 62,5 75,1 75,5 75,9
18 10/06/2013 morning B REIHERTER 2,5 / 62,5 75,1 75,1 74,3
21 10/06/2013 morning C REIHERTER 2,5 / 62,5 76,3 73,9 73,2

Results of this example obtained with the Action software:

(image missing)

(image missing)

Some concluding suggestions

Stable system over time. All points are within the control lines.

RR - Repeatability and Reproducibility

Brinell Scale 2.5 / 62.5:

Ask Operator Measure
1 1 74,3
2 1 72,8
3 1 73,5
4 1 74,7
5 1 74,3
6 1 75,9
7 1 75,5
8 1 75,5
9 1 75,1
10 1 76,7
1 1 72,8
2 1 74,3
3 1 74,7
4 1 74,7
5 1 75,5
6 1 75,5
7 1 75,9
8 1 75,9
9 1 72,8
10 1 74,3
1 1 72,8
2 1 76,3
3 1 76,7
4 1 76,3
5 1 75,1
6 1 75,5
7 1 75,5
8 1 75,9
9 1 75,9
10 1 75,1
1 2 75,9
2 2 75,5
3 2 76,3
4 2 75,1
5 2 75,1
6 2 75,1
7 2 75,1
8 2 74,7
9 2 75,9
10 2 75,5
1 2 75,1
2 2 76,3
3 2 75,1
4 2 75,9
5 2 74,3
6 2 74,3
7 2 73,9
8 2 75,1
9 2 75,9
10 2 75,9
1 2 74,7
2 2 75,5
3 2 75,9
4 2 75,9
5 2 74,7
6 2 75,9
7 2 75,1
8 2 75,9
9 2 76,4
10 2 75,1
1 3 73,2
2 3 70,2
3 3 77,1
4 3 74,7
5 3 76,3
6 3 75,1
7 3 76,7
8 3 76,7
9 3 76,3
10 3 75,9
1 3 75,7
2 3 76,3
3 3 76,3
4 3 73,2
5 3 73,9
6 3 74,3
7 3 75,1
8 3 74,3
9 3 75,1
10 3 75,1
1 3 74,7
2 3 74,7
3 3 76,3
4 3 75,5
5 3 76,3
6 3 73,9
7 3 75,5
8 3 75,1
9 3 75,9
10 3 75,1

Results of this example obtained with the Action software:

(image missing)

(image missing)

(image missing)

Some concluding suggestions

Graphic analysis:

In the graph by operator we observe a similar behavior with regard to hardness measurement.

Variability analysis:

Operator variability is zero.

Therefore, the uncertainty due to reproducibility is zero.

Trend

Parts VR Measurements
1 72,7 75,1
2 72,7 74,7
3 72,7 75,5
4 72,7 75,5
5 72,7 74,7
6 72,7 75,1
7 72,7 75,9
8 72,7 74,7
9 72,7 74,3
10 72,7 75,1
11 72,7 73,9
12 72,7 75,9

(image missing)

(image missing)

Some concluding suggestions

As zero is not part of the confidence interval, we consider which measurement system presents a significant trend at the 95% confidence level.

Uncertainty calculation

Measuring equation

This section is dedicated to calculating uncertainty for hardness (equation 4.1.1). Below we present the measurement equation.

$$D=I+\text{Res}_I+R+\varepsilon~(4.1.1)$$

where:

  • $ D $: represents the hardness, in (HB);
  • $ I $: represents the hardness measuring instrument (durometer), in (HB). We have to $ I $ has a normal distribution with uncertainty $ u(I) $ and coverage factor k obtained via the durometer calibration certificate;
  • $ Res_I $: represents the resolution of the instrument. We have to $ u(Res_I) $ has a rectangular (or uniform) distribution over the interval $ \left(-\frac{{Res}_I}{2},\frac{{Res}_I}{2}\right). $ Therefore, the uncertainty due to the resolution will be given by:

$$u(Res_I)=\frac{{Res}_I}{2\sqrt{3}}$$

  • $ R $: represents reproducibility in $ (HB). $ We have to $ R $ has a normal distribution with uncertainty $ u(R) $ obtained in the Reproducibility test. Reproducibility represents the variability associated with the application of the measurement method by operators;
  • $ \varepsilon $: represents repetitiveness. We can assume that the distribution of the mean of measurements has a normal distribution. Therefore, we can estimate its uncertainty (standard deviation from the mean) as:

$$u(\varepsilon) = \frac{s}{\sqrt{n}},$$

in which $ s^2=\frac{1}{n-1}\displaystyle\sum_{i=1}^{n}(x_i -\overline{X})^2 $ is the sampling variance and $ \overline{X} = \frac{1}{n} \displaystyle\sum^{n_{i=1}} x_i $ is the sample mean.

Combined Uncertainty

The combined uncertainty for hardness is given by

$$u_c(D)=\sqrt{u^2(I)+u^2(\text{Res}_I)+u^2(R)+u^2(\varepsilon)}$$

Expanded Uncertainty

The expanded uncertainty for hardness is given by

$$U(D)=u_c(D)\times k$$

in which $ k $ is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and confidence level of approximately 95.45%.

$$\nu_{eff}(D)=\frac{u^4_c(D)}{u^4(\varepsilon)}(n-1)$$

Hardness test

Now let’s calculate the uncertainty for the hardness measurement system, for this, we carry out three measurements for the hardness test, the data is in the table.

Screenshot_33

We will calculate the average and repeatability as follows

$$\overline{X}=\frac{128+129+130}{3}=129\quad\text{e}\quad u(\varepsilon)=\frac{s}{\sqrt{3}}=\frac{1}{\sqrt{3}}=0,577$$

From the hardness calibration certificate, we obtain that the uncertainty due to the hardness meter is $ u(I)=0,61~HB $ and $ k=2 $ and resolution $ u(Res_I)=1~HB. $ From the RR study, we obtain negligible uncertainty. With this, we calculate the combined uncertainty for hardness as follows:

$$u_c(D)=\sqrt{u^2(I)+u^2(\text{Res}_I)+u^2(R)+u^2(\varepsilon)}=\sqrt{0,305^2+0,289^2+0,577^2+0^2}=0,71$$

The expanded uncertainty for hardness is given by

$$U(D) =u_c(D)\times k=0,71\times 2,78=1,97$$

in which $ k $ is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and confidence level of approximately 95.45%.

$$\nu_{eff}(D)=\frac{u^4_c(D)}{u^4(\varepsilon)}(n-1)=\frac{0,71^4}{0,577^4}(3-1)=4,58$$

Below, we present a summary of the uncertainty calculation for the hardness test on the Brinell scale (2.5/62.5 HB).

Screenshot_34

18.4.2 Calculation of Uncertainty for Lateral Compression of Cups

To calculate the uncertainty associated with the lateral compression test of cups, we carried out an initial survey of the possible sources of variability that affected the test. Figure 4.2.1 presents the Ishikawa diagram of the process.

(image missing)

Figure 4.2.1: Ishikawa graph.

Based on the diagram, some experiments were carried out to estimate the variability of related sources.

Stability Studies

Firstly, we need to attest to the stability of the measurement system. To do this, we take 100 measurements over 20 days, that is, each day we measure the measuring system 5 times. The measurements are presented in the following table:

SUBGROUP ID 1 2 3 4 5 6 7 8 9 10
TIME 09:30 14:30 09:30 14:30 09:30 14:30 09:30 14:30 09:30 14:30
DATE: DAY/MONTH April 25 April 26 April 27 April 28 April 29 30/Apr 1/may 2/may 3/may 4/may
$ x_1 $ 1,123 1,124 1,098 1,074 1,117 1,07 1,089 1,118 1,113 1,325
$ x_2 $ 1,18 0,947 1,185 1,14 1,057 1,066 1,173 0,968 1,095 1,19
$ x_3 $ 1,006 1,115 1,137 1,05 1,022 1,102 1,087 0,993 0,979 1,139
$ x_4 $ 1,07 1,047 0,989 1,101 1,046 1,31 1,135 1,041 1,092 1,213
$ x_5 $ 0,92 1,281 1,134 1,17 1,068 1,181 1,024 1,147 1,061 1,074
SUM VAL. INDIVIDUALS 5,3 5,51 5,54 5,54 5,31 5,73 5,51 5,27 5,34 5,94
AVERAGE $ \overline{X} $ 1,06 1,1 1,11 1,11 1,06 1,15 1,1 1,05 1,07 1,19
AMPLITUDE $ R $ 0,26 0,33 0,2 0,12 0,1 0,24 0,15 0,18 0,13 0,25
SUBGROUP ID 11 12 13 14 15 16 17 18 19 20
TIME 09:30 14:30 09:30 14:30 09:30 14:30 09:30 14:30 09:30 14:30
DATE: DAY/MONTH 5/may 6/may May 7 8/may 9/may 10/may May 11 May 12 May 13 May 14
$ x_1 $ 1,197 1,085 1,089 1,137 1,169 1,081 1,045 1,099 1,19 1,045
$ x_2 $ 1,101 1,108 1,149 1,188 1,085 1,051 1,093 1,004 1,246 1,082
$ x_3 $ 1,052 1,031 1,129 1,147 1,195 0,963 1,218 1,104 1,073 1,112
$ x_4 $ 1,194 1,143 1,19 1,183 1,103 1,212 1,171 1,162 1,074 1,219
$ x_5 $ 1,052 1,124 1,157 1,071 0,958 1,07 1,086 1,06 1,087 1,228
SUM VAL. INDIVIDUALS 5,6 5,49 5,71 5,73 5,51 5,38 5,61 5,43 5,67 5,69
AVERAGE $ \overline{X} $ 1,12 1,1 1,14 1,15 1,1 1,08 1,12 1,09 1,13 1,14
AMPLITUDE $ R $ 0,15 0,11 0,1 0,12 0,24 0,25 0,17 0,16 0,17 0,18

With this, we have the stability letter ($ \overline{X} ~{e}~R $) below:

copo1

copo2

Figure 4.2.1: Stability Chart.

18.4.2.1 Experiments

Experiment 1: Variation in the Cup

Here, we are interested in evaluating whether the variation in cup measurement is significant. This experiment follows the following procedure:

  • Sample a glass and measure three times at the same point;
  • Repeat the procedure for 10 glasses.

To obtain an estimate of the variability within the cup, we will use the 1-way ANOVA technique (Cup) - random effect.

Experiment 2: Variation due to measurement displacement

Here, we are interested in evaluating whether the measurement displacement in the same cup (radial and axial) is significant. This experiment follows the following procedure:

  • Sample a glass and measure three times at each point, moving in the height, radial and axial direction;
  • Repeat the procedure for three glasses.

To obtain an estimate of the variability due to displacement, we will use the variance components and 2-factor ANOVA (Cup and Displacement) techniques - random effect.

Experiment 3: Variation due to measurement angle

Here, we are interested in evaluating whether the angle formed between the mouth of the cup and the measuring base is significant. This experiment follows the following procedure:

  • Sample a glass and measure three times at each point, varying the number of blades positioned at the bottom of the glass when measuring from one point to the next;
  • Repeat the procedure for three glasses.

To obtain an estimate of the variability due to the angle, we will use the variance components and 2-way ANOVA (Cup and Displacement) techniques - random effect.

Experiment 4: Variation due to ambient temperature

Here, we are interested in evaluating whether the ambient temperature influences the measurement results. This experiment follows the following procedure:

  • Sample a glass and measure three times at each point and note the room temperature;
  • Repeat the procedure for three glasses.

To obtain an estimate of the variability due to temperature, we will use the simple linear regression technique.

Estimation of Sources of Uncertainty Obtained in Experiments for PS-type cups

Cup variability $ u_c(\epsilon_2) $

Here, we will use the 1-factor anova technique to estimate the variability in the same cup.

Table 4.2.1.1 presents the measurements taken on the PS cup.

Screenshot_35

Table 4.2.1.1: Resistance Measurements in the PS Cup.

The statistical model for this experiment is:

Screenshot_36

where:

  • $Y_{ij}$: represents the j-th measurement in the ith cup;
  • $α$ is the average of the measurements;
  • $β_i$ it is the effect of the glass;
  • $ε_{ij}$ is the replication error.

To calculate the $ SQ_T $, we will first calculate the sample variance of all data, that is,

$$S^2_g=\frac{(1,017 - 1,009)^2+(0,906 -1,009)^2+\dots+(1,053-1,009)^2+(1,005 - 1,009)^2}{30-1}=0,007$$

With this, we have to

$$SQ_T=(c * n - 1)~S^2_g=(10 * 3 - 1)~0,007=0,20538$$

To calculate the sum of squares due to the cup ($ SQ_C $), we will first calculate the sample variance of the mean column, that is,

$$S^2_c=\frac{(0,938 - 1,009)^2+(0,987-1,009)^2+\dots+(1,064 - 1,009)^2+(1,070 - 1,009)^2}{10-1}=0,003$$

With this, we have to

$$SQ_C =n~(c - 1)~S^2_c=3~(10 - 1)~0,003=0,09001$$

So, we have to

$$SQ_E =SQ_T - SQ_C=0,20538 - 0,09001=0,11537$$

The degrees of freedom are:

Therefore, the mean square is:

$$QM_C =\frac{SQ_C}{c - 1}=\frac{0,09001}{9}=0,010$$

$$QM_E=\frac{SQ_E}{c (n - 1)}=\frac{0,115371}{20}=0,005769$$

Figure 4.2.1.1 presents the summary of the variance analysis.

anova

Figure 4.2.1.1: Analysis of Variance (ANOVA).

Here, we have that p-value is given by:

$$P(F_{9;~20}>~1,73)=0,1463$$

Therefore, the estimate of variability within the cup is:

$$u_c(\epsilon_2)=0$$

Source of variation due to measurement offset $ u_c(\epsilon_3) $

In this application we again use an anova, however, a two-way ANOVA, that is, two factors.

Table 4.2.1.2 presents the measurements taken on the PS cup, the Point column represents the measurement point.

Cup Point Resistance in 10 mm (N)
1 1 1,079
1 2 1,024
1 3 1,062
1 4 1,005
1 5 1,035
2 1 1,066
2 2 0,888
2 3 0,962
2 4 1,005
2 5 1,045
3 1 1,119
3 2 0,984
3 3 1,107
3 4 1,069
3 5 1,042
4 1 1,126
4 2 1,022
4 3 1,149
4 4 1,09
4 5 1,073
5 1 1,182
5 2 1,078
5 3 1,025
5 4 1,168
5 5 1,013
Average 1,057

Table 4.2.1.2: Data Table regarding Measuring Point Displacement.

The statistical model for this experiment is:

Screenshot_37

where:

  • $Y_{ijk}$ represents the kth measurement in the jth Factor 2 in the ith cup;
  • $μ$ is the average of the measurements;
  • $α_i$ it is the effect of the glass;
  • $β_j$ is the effect of the Displacement Factor;
  • $ε_{ijk}$ is the replication error.

To calculate the sums of squares, we first need to calculate the following sample variances:

$$S^2_=\frac{1}{p~o~r~-~1}\sum^{p~o~r}_{l=1}(y_{l}~-~\overline{y})^2=0,00435$$

Screenshot_41

in which we know that

Screenshot_44

$$SQ_{\text{Total}}=(p~o~r - 1)~S^2_g = (24)~S^2_g = 0,104484$$

$$SQ_{\text{Copo}}=o~r~(p - 1)~S^2_p = 5~(4)~S^2_p=0,034567$$

$$SQ_{\text{Deslc}}=p~r~(o - 1)~S^2_o = 5~(4)~S^2_o=0,034983$$

$$SQ_{\text{Erro}}=SQ_{\text{Total}} - SQ_{\text{Deslc}} - SQ_{\text{Copo}}=0,034934$$

The degrees of freedom are:

Effect Degrees of Freedom
Cup $ p-1=4 $
Displacement $ o-1=4 $
Error $ por-p-o-1=16 $
Total $ por-1=24 $

Therefore, the mean square is:

$$QM_{\text{Copo}}=\frac{SQ_{\text{Copo}}}{p - 1}=\frac{0,034567}{4}=0,008642$$

$$QM_{\text{Deslc}}=\frac{SQ_{\text{Deslc}}}{o - 1}=\frac{0,034983}{4}=0,008746$$

$$QM_{\text{Erro}}=\frac{SQ_{\text{Erro}}}{por-p-o+1}=\frac{0,034934}{16}=0,002183$$

The table below presents the summary of the variance analysis.

(image missing)

Efeito

Here, we have that p-value is given by:

Cup $ P(F_{4;~16}~> 3,95)= 0,02029 $

Displacement $ P(F_{4;~16}> 4,00)=0,01943 $

Therefore, as the displacement is significant for the model, the uncertainty estimate associated with the displacement is given by:

$$u_c(\epsilon_3)=\sqrt{QM_{\text{Erro}}}=0,046$$

Source of variation due to measurement angle $ u_c(\epsilon_4) $

Here, we also use the two-factor anova technique. Table 4.2.1.3 presents the measurements taken on the PS cup.

Screenshot_38

Table 4.2.1.3: Data table regarding the angle of the measuring point.

The statistical model for this experiment is:

Screenshot_39

where:

  • $Y_{ijk}$ represents the kth measurement in the jth Factor 2 in the ith cup;
  • $μ$ is the average of the measurements;
  • $α_i$ it is the effect of the glass;
  • $β_j$ is the effect of the plate factor (angle);
  • $ε_{ijk}$ is the replication error.

To calculate the sums of squares, we first need to calculate the following sample variances:

Screenshot_42

in which we know that

Screenshot_43

The degrees of freedom are:

Effect Degrees of Freedom
Cup $ p-1=4 $
Plates $ o-1=2 $
Error $ por-p-o-1=8 $
Total $ por-1=14 $

Therefore, the mean square is:

$$QM_{\text{Copo}}=\frac{SQ_{\text{Copo}}}{p - 1}=\frac{0,1956}{4}= 0,0489$$

$$QM_{\text{Placa}}=\frac{SQ_{\text{Placa}}}{o - 1}=\frac{0,029243}{2}=0,0146$$

$$QM_{\text{Erro}}=\frac{SQ_{\text{Erro}}}{por-p-o+1}=\frac{0,016837}{8}=0,002105$$

The table below presents the summary of the variance analysis.

anova2

Efeito1

Figure 4.2.1.4: Variance Analysis Table

Here, we have that p-value is given by:

Cup $ P(F_{4;~9}> 12,76600)= 0,0002 $

Plates $ P(F_{1;~9}> 3,02895)=0,0178 $

Here we verify that the plates factor, that is, the measurement angle, are significant.

Therefore, the estimate of the uncertainty due to the angle is:

$$u_c(\epsilon_4) =\sqrt{QM_{\text{Erro}}}=0,046$$

Source of variation due to ambient temperature $ u_c(\epsilon_5) $

Here, we use the linear regression technique to determine the sources of uncertainty for the experiment. Table 4.2.1.4 presents resistance measurements, taking into account the ambient temperature value.

Screenshot_40

Table 4.2.1.4: Data table regarding Ambient Temperature.

First, we determine the means of the resistance variables ($ Y $) and temperature ($ X $). Easily, we get $ \overline{x}=23,744 $ and $ \overline{y} = 1,104 $. Thus, we can find the empirical sums of squares.

$$S_{xx}=\sum^n_{i=1}(x_i-\overline{x})^2= 23,7022$$

$$S_{yy}=\sum^n_{i=1}(y_i-\overline{y})^2= 0,03905$$

$$S_{xy}=\sum^n_{i=1}(x_i-\overline{x})(y_i-\overline{y})=-0,0272$$

Next, we use empirical sums of squares to find the parameter estimates of the regression line.

$$\hat{\beta_1}=\frac{S_{xy}}{S_{xx}} = \frac{-0,0272}{23,7022}=-0,00114757$$

$$\hat{\beta}_0=\overline{y} - \hat{\beta}_1\overline{x} = 1,104-0,0011475 \times 23,744 = 1,1312484$$

The adjusted model is given by:

$$\hat{Y}~=~1,1312484~ -~ 0,00114757~X$$

Figure 4.2.1.5 shows the relationship between the variables, as well as the adjusted line.

ajuste

Figure 4.2.1.5: Adjusted model graph.

The estimate of variability associated with the regression is given by:

$$QM_E=\frac{SQ_E}{n-2}$$

We have to $ SQ_E $ is the sum of squares due to random error, and is calculated by:

$$SQ_E=\sum^n_{i=1}(y_i-\overline{y})^2 - \hat{\beta_1}S_{xy}=\sum^n_{i=1} (y_i - 1,104)^2-(-0,0282)(-0,001147)=0,039$$

With this, we have to

$$QM_E=\frac{SQ_E}{n-2}=\frac{0,039}{7}=0,005574$$

Table 4.2.1.7 presents the coefficient values for the model presented, as well as the analysis of the significance of the coefficients.

regressao

Figure 4.2.1.7: Coefficients Table.

Here, we have that p-value is given by:

Temperature $ P(T_{7}>-0,075)=0,9424 $

We can see in Figure 4.2.1.7 that the p-value of the temperature coefficient presents a value above 5%. Therefore, we can conclude that temperature is not significant.

18.4.2.2 Calculation of Uncertainty for the Lateral Compression Test of Cups

$ \mathbf{\checkmark} $Measurement Equation

Expression 4.2.2 represents the measurement equation used to obtain the compressive strength.

$$R=\text{Herd} (CG)+\text{res}(CG)+\Delta+\varepsilon_2+\varepsilon_3+\varepsilon_4+\varepsilon_5 \quad(4.2.2)$$

in which,

  1. R: represents the compression resistance in the disposable cup (N);

  2. Res(CG): Represents the load cell resolution (N). We can assume that the resolution has a rectangular (or uniform) distribution in the interval (-Res(CG)/2,Res(CG)/2), in which Res(CG) is the bank’s resolution. Therefore, the uncertainty due to resolution will be given by: u(Res(CG))=Res(CG)/(2√3);

  3. Herd(CG): Represents the load cell (N). We have that Herd(CG) has a Normal distribution with uncertainty (u(Herd(CG))) and the coverage factor obtained via calibration certificate;

  4. Δ: Represents the random error due to the load cell (N). We can assume that the distribution of the mean of measurements (repeatability) has a Normal distribution.

Therefore, we can estimate its uncertainty (standard deviation) as u(Δ)=s/√n,

in which, $ s^2=\displaystyle\frac{1}{n-1}\sum_{i=1}^{n}(x_i -\overline{X})^2 $ is the sampling variance and $ \overline{X}=\displaystyle\frac{1}{n}\sum^{n_{i=1}}x_i $ is the sample mean;

  1. ε2: represents the random error due to variation in the same glass. Let’s consider that ε2~ N(0,σ2(cup)). To estimate σ2, we will use the variance components technique (1-way ANOVA - random effect), considering the data obtained in the following experiment:
  • Sample a glass and measure three times at the same point;
  • the procedure in 10 glasses;
  1. ε3: represents the random error due to displacement when carrying out measurements. Let’s consider that ε3~ N(0,σ2(displacement)). To estimate σ2, we will use the variance components technique (2-way ANOVA - random effect), considering the data obtained in the following experiment:
  • Sample a glass and measure three times at each point, moving in the height, radial and axial direction;
  • Repeat the procedure for three glasses;
  1. ε4: represents the random error due to the angle at the mouth of the cup. Let’s consider that ε4~ N(0, σ2(angle)). To estimate σ2, we will use the variance components technique (2-way ANOVA - random effect), considering the data obtained in the following experiment:
  • Sample a glass and measure three times at each point, varying the number of blades positioned at the bottom of the glass when measuring from one point to the next;
  • Repeat the procedure for three glasses;
  1. ε5: represents the influence of ambient temperature on the measurement results. Let’s consider that ε5~ N(0, σ2(temp)). To estimate σ2 we will use the simple linear regression technique:
  • This experiment follows the following procedure:
  • Sample a glass and measure three times at each point and note the room temperature;

Repeat the procedure for three glasses.

$ \mathbf{\checkmark} $Combined Uncertainty ($ u_{c}(R) $)

The combined uncertainty associated with resistance is given by:

$$u_c(R)=\sqrt{u^2(\text{Res}(CG))+u^2(\text{Herd}(CG))+u^2(\Delta)+u^2(\epsilon_2)+u^2(\epsilon_3)+u^2(\epsilon_4)+u^2(\epsilon_5)}~~(4.2.2.1)$$

in which $ u^2(.) $ is the contribution of each source of uncertainty. The expression (4.2.2.1) represents the combined uncertainty of the resistance of the cups.

$ \mathbf{\checkmark} $Expanded Uncertainty ($ U(R) $)

Here, the expression (4.2.2.2) represents the expanded uncertainty for the lateral resistance of the cups.

$$U(R)=k\times u_c(R)~~(4.2.2.2)$$

in which $ k $ (coverage factor) is the quantile of the t-Student distribution with $ \nu_{eff}(R) $ degrees of freedom and 95% confidence. Here, the degrees of freedom are given by:

$$\nu_{eff}(R)=\frac{u^4_c(R)}{u^4(\Delta)}(n-1) $$

Application

Cup Resistance
1 1,017
2 0,906
3 0,891
4 1,118
5 0,921
6 0,922
7 1,058
8 1,008
9 0,981
10 1,099
11 1,034
12 1,032
13 1,127
14 0,971
15 0,962
16 1,053
17 0,914
18 0,881
19 1,156
20 1,036
21 1,025
22 0,999
23 0,889
24 0,872
25 1,142
26 1,03
27 1,02
28 1,151
29 1,053
30 1,005
Standard Deviation 0,084155583

Repeatability is given by:

$$u(\Delta)=\frac{0,084155583}{\sqrt{30}}=0,002805186$$

Values obtained by experiments and calibration certificates

Sources Uncertainty (U) Coverage Factor (k)
Variability within the cup 0,075950 1
Measurement Offset 0,036228 1
Measuring Angle 0,049989 1
Ambient Temperature 0,074650 1
Load Cell Resolution 0,000098 3,464101615
Inherited from the Load Cell 0,008100 2
Repetitiveness 0,002805186 1

So we calculate it as follows:

$ \mathbf{\checkmark} $Combined Uncertainty ($ u_{c}(R) $)

The combined uncertainty associated with resistance is given by:

$$u_c(R)=\sqrt{u^2(\text{Res}(CG))+u^2(\text{Herd}(CG))+u^2(\Delta)+u^2(\epsilon_2)+u^2(\epsilon_3)+u^2(\epsilon_4)+u^2(\epsilon_5)}$$

$$=\sqrt{(2,83\times10^{-5})^2+0,00405^2+0,0028^2+0,0795^2+0,036^2+0,049^2+0,074^2}=$$

$$=0,1232$$

$ \mathbf{\checkmark} $Expanded Uncertainty ($ U(R) $)

$$U(R)=k\times u_c(R)=1,96\times 0,1232=0,2464$$

in which $ k $ (coverage factor) is the quantile of the t-Student distribution with $ \nu_{eff}(R) $ degrees of freedom and 95% confidence. Here, the degrees of freedom are given by:

$$\nu_{eff}(R)=\frac{u^4_c(R)}{u^4(\Delta)}(n-1) ==\frac{0,1232^4}{0,0028^4}(30-1) =\infty$$

So, we have to $ k=1,96 $

The summary of the calculations is presented in table 4.2.2.1

Screenshot_45

Table 4.2.2.1: Result of the Uncertainty Calculation for the Cup Lateral Compression Test.

18.4.3 Uncertainty Calculation for Radial Bar and Direction Terminal

OBJECTIVE

Develop the Uncertainty calculation procedure for Radial Bar and Direction Terminal for:

  • Radial Clearance in millimeters (mm);

Develop the Stability study for Radial Bar and Steering Terminal to:

  • Radial Clearance in millimeters (mm);

Develop the Repeatability and Reproductivity study for Radial Bar and Steering Terminal for:

  • Radial Clearance in millimeters (mm);

Stability Study

a. Select and identify a part;

b. Perform CEP testing

i. Sample size: 1 measurement;

ii. Frequency: 2 times a day;

iii. Carry out 20 subgroups;

iv. Logbook: Date, Time, Person, Ambient temperature;

v. NOTE: carry out the entire assembly of the system in each measurement;

After carrying out the experiment, we obtained the following results:

Subgroup Date Schedule Temp. Environment (ºC) Radial Clearance (mm)
1 27/7/2009 09:10 18 0,5439
2 28/7/2009 10:06 19 0,5655
3 29/7/2009 08:20 17 0,5414
4 30/7/2009 08:00 16 0,5444
5 30/7/2009 13:00 18 0,5397
6 31/7/2009 08:00 10 0,5664
7 31/7/2009 13:00 14 0,547
8 3/8/2009 08:00 15 0,5451
9 3/8/2009 13:00 17 0,5566
10 4/8/2009 08:00 13 0,5408
11 4/8/2009 13:00 18 0,544
12 5/8/2009 08:00 16 0,5435
13 5/8/2009 13:00 19 0,5365
14 6/8/2009 08:00 17 0,5335
15 6/8/2009 13:00 18 0,527
16 7/8/2009 08:00 16 0,5366
17 7/8/2009 13:00 17 0,5444
18 10/8/2009 08:00 15 0,5313
19 10/8/2009 13:00 17 0,5449
20 11/8/2009 08:00 14 0,5431

EstabFolgaRadial

EstabFolgaRadial1

Figure 4.3.1: Stability Chart.

After the stability study, we concluded that the system is stable over time.

18.4.3.1 RR Test (Repeatability and Reproductivity)

Now, we will describe the procedure and variables involved in studying RR

Experiment

a. We will select 3 pieces of different models;

b. We will select 2 operators;

w. Each operator will perform 3 measurements on each part;

For more details on the content on the study of RR, go to the MSA 4th edition RR module content.

Order Person Ask Radial Clearance
1 A 1 0,5439
2 A 3 0,584
3 B 2 0,5898
4 B 1 0,5437
5 B 1 0,5413
6 A 3 0,4594
7 A 2 0,586
8 B 3 0,4564
9 B 3 0,4488
10 A 1 0,5655
11 A 2 0,59
12 A 2 0,5697
13 A 3 0,4766
14 B 2 0,5545
15 B 2 0,5646
16 B 1 0,56
17 B 3 0,4681
18 A 1 0,5414

After the experiment, we obtain the following results:

RRFolgaRadial

RRFolgaRadial1

RRFolgaRadial2

From the RR study, we estimate the uncertainty due to Repeatability and Reproductivity (RR). So,

$$u(\varepsilon)=0,032$$

18.4.3.2 Measurement uncertainty for Radial Bar and Steering Terminal

This module describes the procedures and variables involved in calculating uncertainty.

Measurement Equation

$$F_r=I+{Res}_I+\varepsilon+\delta$$

where:

  • $F_r$: Radial Clearance in (mm);
  • I: Inherited from the Instrument (gap measuring pen) in (mm);
  • $Res_I$: Instrument Resolution in (mm);
  • ε: represents the random error due to repetitiveness and reproducibility;
  • δ: represents the random error due to repetitiveness.

Combined Uncertainty u(Fr)

The combined uncertainty for the radial clearance is given by

$$u_c(F_r)=\sqrt{u^2(I)+u^2({Res}_I)+u^2(\varepsilon)+u^2(\delta)}$$

Expanded Uncertainty U(Fr)

The expanded uncertainty for the radial clearance is given by

$$U(F_r)=k\times u_c(F_r)$$

in which,

k is the coverage factor, given by the quantile of the t-Student distribution with νeff (Fr) degrees of freedom and 95% confidence. and the degrees of freedom given by:

$$\nu_{eff}(F_r)=\frac{u^4_c(F_r)}{\frac{u^4(I)}{\infty}+\frac{u^4(Res_I)}{\infty}+\frac{u^4(\varepsilon)}{\infty}+\frac{u^4(\delta)}{n-1}}$$

Application

Radial clearance measurements

Screenshot_46

The uncertainty due to repetitiveness is given by:

$$u(\delta)=\frac{s}{\sqrt{n}}=\frac{0,0106}{\sqrt{6}}=0,0043$$

The uncertainty due to RR $ (u(\varepsilon)=0,032) $, was obtained in module 4.3.1

From the calibration certificate we obtain the instrument uncertainty and instrument resolution.

Therefore, the summary of uncertainties is in table 4.3.2.1

Screenshot_47

Table 4.3.2.1: Summary of uncertainties.

We will calculate the combined uncertainty as follows:

$$u_c(F_r)=\sqrt{u^2(I)+u^2({Res}_I)+u^2(\varepsilon)+u^2(\delta)}=$$

$$=\sqrt{0,0002^2+0,00014^2+0,032^2+0,0043^2}=0,032$$

From the combined uncertainty, we will calculate the expanded uncertainty as follows

$$U(F_r)=k\times u_c(F_r)=1,96\times 0,032=0,063$$

k is the coverage factor, given by the quantile of the t-Student distribution with νeff (FA) degrees of freedom and 95% confidence. and the degrees of freedom given by:

$$\nu_{eff}(F_r)=\frac{u^4_c(F_r)}{\frac{u^4(I)}{\infty}+\frac{u^4(Res_I)}{\infty}+\frac{u^4(\varepsilon)}{\infty}+\frac{u^4(\delta)}{n-1}}=$$

$$=\frac{0,0032^4}{\frac{0,0002^4}{\infty}+\frac{0,00014^4}{\infty}+\frac{0,032^4}{\infty}+\frac{0,0043^4}{6-1}}=15282,64$$

Therefore, the summary of the calculations is described in the table below:

Screenshot_48

in which

C.S.: sensitivity coefficient

Contr.: contribution.

GL: degrees of freedom.

18.4.4 Calculation of Uncertainty for Net Effective Power Otto cycle engine

In this module, we describe the uncertainty calculation regarding the net effective power test for Otto cycle engines (gasoline engines). The test is carried out in accordance with ISO 1585.

18.4.4.1 Torque

$ \mathbf{\checkmark} $Measurement Equation

Expression 4.4.1.1 represents the measurement equation used to obtain torque. In this case we have:

$$T = M ~ g ~ L ~ (1 - \delta_T) + ResB + \Delta + hist~~(4.4.1.1)$$

in which

  • M: Represents the Mass (kg). We have that M has a normal distribution with uncertainty (u(M)) and the coverage factor obtained via calibration certificate;
  • g: Represents the acceleration of gravity (m/s2). We have that g has a normal distribution with uncertainty (u(g)) and the coverage factor obtained via calibration certificate;
  • L: Represents the Arm Length (m). We have that L has a normal distribution with uncertainty (u(L)) and the coverage factor obtained via calibration certificate;
  • $ \delta_T $: Represents the influence of ambient temperature on the Arm length. To address this source of uncertainty, we will consider the worst-case situation. Consider $ \delta_T $ a random variable with rectangular distribution where the limits are given by:

$$LI_{\delta_T} = - \alpha \times \Delta_T ~~{e}~~LS_{\delta_T} = \alpha \times \Delta_T,$$

in which

  • $ \alpha $: Coefficient of thermal expansion of steel (11.5 x 10-6ºC-1);
  • $ \Delta_T $: Maximum temperature variation found in the laboratory. As the coefficient of thermal expansion has a unit of ºC-1 and the $ \Delta_T $ has as unit ºC, we conclude that $ \delta_T $ There is no unity. We can estimate the uncertainty due to the influence of temperature as:

$$u(\delta_T) = \frac{LS_{\delta_T} - LI_{\delta_T}}{2\sqrt{3}};$$

  • $Res_B$: Represents Bank Resolution (Nm). We can assume that the resolution has a rectangular (or uniform) distribution in the range $ \left(-\frac{ResB}{2},~\frac{ResB}{2}\right) $, where ResB is the bank’s resolution. Therefore, the uncertainty due to the resolution will be given by:

$$u(ResB) = \frac{ResB}{2\sqrt{3}};$$

  • $ \Delta $: Represents Repeatability (Nm). We can assume that the distribution of the mean of measurements has a normal distribution. Therefore, we can estimate its uncertainty (standard deviation) as

$$u(\Delta) = \frac{s}{\sqrt{n}},$$

in which $ s^2 = \frac{1}{n-1} \displaystyle\sum_{i=1}^{n} (x_i -\overline{X})^2 $ is the sampling variance and $ \overline{X} = \frac{1}{n}\displaystyle \sum^{n_{i=1}} x_i $ the sample mean;

  • hist: Represents hysteresis (Nm). We have that hysteresis has a rectangular distribution in the interval (MR, MA) if MR ≤ MA or (MA, MR) if MA < MR, em que MA: médias das medições no avanço e MR é a médias no retorno.

Logo, a sua incerteza será dada por:

$$u(hist) = \frac{\mid M_A - M_R \mid}{2\sqrt{3}}.$$

$ \mathbf{\checkmark} $Incerteza Combinada ($ u_{c}(T) $)

A incerteza combinada para a grandeza torque é dada por:

$$u^2_c(T)=\left(\frac{\partial T}{\partial M}\right)^2 u^2(M)+\left(\frac{\partial T}{\partial g}\right)^2 u^2(g) +\left(\frac{\partial T}{\partial L}\right)^2 u^2(L)+\left(\frac{\partial T}{\partial \delta_T}\right)^2 u^2(\delta_T)$$

$$+\left(\frac{\partial T}{\partial ResB}\right)^2 u^2(ResB)+ \left(\frac{\partial T}{\partial \Delta}\right)^2 u^2(\Delta)+ \left(\frac{\partial T}{\partial hist}\right)^2 u^2(hist)~~(4.4.1.2)$$

em que

$$\left(\frac{\partial T}{\partial M}\right)=g~L~(1-\delta_T)$$

$$\left(\frac{\partial T}{\partial g}\right)=M~L~(1-\delta_T)$$

$$\left(\frac{\partial T}{\partial L}\right)=M~g~(1-\delta_T)$$

$$\left(\frac{\partial T}{\partial \delta_T}\right)=M~g~L$$

$$\left(\frac{\partial T}{\partial ResB}\right)=1$$

$$\left(\frac{\partial T}{\partial \Delta}\right)=1$$

$$\left(\frac{\partial T}{\partial hist}\right)=1$$

em que $ u^2(.) $ é a contribuição de cada fonte de incerteza. A expressão 4.4.1.2 representa a incerteza combinada do torque.

$ \mathbf{\checkmark} $Incerteza Combinada Relativa ($ u_{r}(T) $)

A incerteza combinada relativa para o torque é dada por:

$$\frac{u^2_c(T)}{T^2}=\left[\frac{g ~ L ~ (1 - \delta_T)}{T}\right]^2 ~ u^2(M) + \left[\frac{M ~ L ~(1 - \delta_T)}{T}\right]^2 ~ u^2(g) + \left[\frac{M ~ g ~(1 -\delta_T)}{T}\right]^2 ~ u^2(L)$$

$$+ \left(\frac{- M ~ g ~L}{T}\right)^2 ~u^2(\delta_T)+ \frac{u^2(ResB)}{T^2} +\frac{u^2(\Delta)}{T^2} +\frac{u^2(hist)}{T^2}$$

sendo $ T = M~g~L~(1 - \delta_T) $. Fazendo $ u_{r} =\sqrt{\frac{u^2_c(T)}{T^2}} $, temos que:

$$u_r(T)=\sqrt{\frac{u^2(M)}{M^2}+\frac{u^2(g)}{g^2}+\frac{u^2(L)}{L^2}+\frac{u^2(\delta_T)}{(\delta_T-1)^2} + \frac{u^2(Res_B)}{T^2}+\frac{u^2(\Delta)}{T^2} + \frac{u^2(hist)}{T^2}}$$

$$=\sqrt{u_{r}^2(M) + u_{r}^2(g) + u_{r}^2(L) + u_{r}^2(\delta_T) + u_{r}^2(Res_B) + u_{r}^2(Hist)}~~(4.4.1.3)$$

A expressão 4.4.1.3 representa a incerteza combinada relativa do torque, sendo que $ u_{r}^2(.) $ é a incerteza de contribuição relativa para cada fonte padrão.

$ \mathbf{\checkmark} $Incerteza Expandida ($ U(T) $)

A expressão 4.4.1.4 representa a incerteza expandida relativa para o torque

$$U(T) = k*u_r(T)~~(4.4.1.4)$$

em que k (fator de abrangência) é o quantil da distribuição t-Student com $ \nu_{eff}(T) $ graus de liberdade e confiança de 95%. Os graus de liberdade são dados por:

$$\nu_{eff}(T) =\frac{(u_r(T))^4}{{\frac{u^4(\Delta)}{n - 1} +\frac{u^4(M)}{\infty} + \frac{u^4(g)}{\infty}+\frac{u^4(L)}{\infty} + \frac{u^4(\delta t)}{\infty}}+\frac{u^4(ResB)}{\infty} + \frac{u^4(hist)}{\infty}}$$

A Tabela 4.4.1.1 apresenta o resumo para o cálculo da incerteza da grandeza Torque.

Screenshot_49

C.S. Coeficiente de Sensibilidade

Contr. Contribuição

G.L. Graus de Liberdade

Como é realizado a calibração em vários pontos, teremos um cálculo de incerteza para o respectivo ponto. No

entanto, por determinação técnica, vamos expressar uma única incerteza para toda faixa de leituras (10 a 160 Nm).

Neste sentido, podemos considerar duas técnicas: determinar a máxima incerteza entre as já calculadas, ou ainda, considerar a incerteza expandida agrupada (expressão (4.4.1.5))

$$U_a(T) = 2~\sqrt{\frac{\displaystyle\sum_{i=1}^{p}\left(\frac{U_{i}(T)}{k_i}\right)^2 }{p} }~~(4.4.1.5)$$

em que

  • $ U_{i}(T) $: Representa a incerteza expandida para o i-ésimo ponto;
  • $ k_i $: Representa o fator de abrangência para o i-ésimo ponto;
  • $ p $: Representa o número de pontos de calibração.

Aplicação

A Tabela 4.4.1.2 apresenta todas a medições da calibração de torque.

Screenshot_50

Tabela 4.4.1.2: Medições para calibração de torque.

Aqui, o comprimento do braço é 0,59421 m e a aceleração da gravidade 9,7864598 m/s2. A calibração abrange a faixa de 10 a 160 Nm porém, vamos apresentar o cálculo de incerteza para o ponto de 10 Nm. Para isso, as informações necessárias são:

  • Incerteza expandida da massa de 2 kg (u(M2)) é 0,00021069, com k = 4,303 (valores declarado no certificado de calibração);
  • Incerteza devido a gravidade (u(g)) é 0,0000005. Aqui, o valor de k = 2 (valores declarado no certificado de calibração);
  • Incerteza expandida do comprimento do braço (u(L)) é 0,00018, com k = 2 (valores declarado no certificado de calibração);
  • Incerteza devido à dilatação do braço ($ u(\delta_T) $). Temos que:

$ \alpha = 11,5 ordm; \times 10^{-6} C^{-1} $ e

$ \Delta T = \max(T) - \min(T) = 28 - 20 = 8 ordm;C, $

Logo,

$$LI_{\delta_T}=- \alpha \times \Delta_T = -(11,5 \times 10^{-6}) \times 8 = -92 \times 10^{-6} ~~{e}$$

$$LS_{\delta_T}=\alpha \times \Delta_T = (11,5 \times 10^{-6}) \times 8 = 92 \times 10^{-6},$$

Portanto,

$$u(\delta_T)=\frac{92 \times 10^{-6} - (-92 \times 10^{-6})}{\sqrt{12}} = \frac{0,000184}{2\sqrt{3}}= 0,00005311 ~~{e}$$

$$\delta_T=\frac{92 \times 10^{-6} - (-92 \times 10^{-6})}{2} = 0;$$

  • Resolução do banco (u(ResB)) é 0,6 Nm;
  • Repetitividade ($ u(\Delta) $) é o desvio padrão da média. Para a faixa de 10 Nm temos 0,065 Nm;
  • Histerese (u(hist)) é 0,1 Nm.

Neste ponto (10 Nm), temos que:

$$T=M ~ g ~ L ~ (1 - \delta_T)$$

$$=2 \times 9,7864598 \times 0,59421 \times (1 - 0)$$

$$=11,63042.$$

A incerteza combinada relativa

$$u_r(T)=\sqrt{\frac{u^2(M)}{M^2} + \frac{u^2(g)}{g^2} + \frac{u^2(L)}{L^2} + \frac{u^2(\delta_T)}{(\delta_T-1)^2} + \frac{u^2(ResB)}{T^2} + \frac{u^2(\Delta)}{T^2} + \frac{u^2(hist)}{T^2}}$$

$$=\sqrt{\left(\frac{0,00021069 / 4,303}{2} \right)^2 + \left(\frac{0,0000005 / 2}{9,7864598}\right)^2 + \left(\frac{0,00018 / 2}{0,59421}\right)^2 }$$

$$\overline{+ \left( \frac{0,000184 / 2\sqrt{3}}{1} \right)^2 + \left(\frac{0,6 / 2\sqrt{3}}{11,63042} \right)^2 + \left(\frac{0,065 / 1}{11,63042}\right)^2 + \left(\frac{0,1 / 2\sqrt{3}}{11,63042}\right)^2}$$

$$=0,0161.$$

Os graus de liberdade são:

$$\nu_{eff}(T)=\frac{(u_r(T))^4}{\frac{u^4(\Delta) / M}{n - 1}}$$

$$=\frac{0,1880668^4}{\frac{(0,065 / 11,63042)^4}{3}}$$

$$=206,604.$$

Considerando os graus de liberdade efetivo e uma confiança de aproximadamente 95%, temos pela tabela da distribuição t-student, que o fator de abrangência é 1,968613. Com isso, a incerteza expandida relativa para a variável torque é dada por:

$$U(T)=k \times u_r(T)$$

$$=1,970855 \times 0,0161$$

$$=0,0317.$$

Screenshot_51

Tabela 4.4.1.3: Resumo do cálculo de incerteza para o Torque.

A Tabela 4.4.1.4 apresenta o resultado das incertezas expandidas e seus respectivos fatores de abrangência para todos os pontos de medição.

Pontos $ u_{ci} $ k $ U_i $
1 0,01609 1,97143 0,03171
2 0,00465 1,96982 0,00917
3 0,00225 2,00575 0,00451
4 0,00113 1,96049 0,00221
$ U_e $ Agrupada 0,016934

Tabela 4.4.1.4: Incertezas por ponto de medição.

Por determinação técnica, vamos expressar uma única incerteza para toda a faixa de leituras. Neste sentido, vamos considerar a incerteza expandida agrupada:

$$U(T)=2~\sqrt{\frac{\sum_{i=1}^{p}\left(\frac{U_{i}(T)}{k_i}\right)^2}{p}}$$

$$=2 \sqrt{\frac{\left(\frac{0,03171}{1,97143}\right)^2 + \ldots + \left(\frac{0,00221}{1,96049}\right)^2}{4}}$$

$$=0,016934.$$

Portanto, a incerteza expandida relativa para o torque, na faixa de 40 a 160 Nm, é 1,6934 % da leitura.

18.4.4.2 Potência

Nesta seção vamos apresentar o procedimento de cálculo de incerteza para a calibração de Potência. Note que o procedimento de calibração aqui empregado é feito de forma indireta, ou seja, a calibração da grandeza potência é estritamente dependente das calibrações da Rotação e Torque.

O método de medição, especificado pela indústria automobilística, para medir a potência do motor é o dispositivo dinamométrico com “célula de carga” (ver figura 4.4.2.1).

Motor

Figura 4.4.2.1: Dispositivo dinamométrico para medição da potência do motor.

$ \mathbf{\checkmark} $ Equação de Medição

$$Pot = \frac{Rot \times Torque}{\theta}$$

  • Pot: Representa o resultado da medição de Potência (kW);
  • Rot: Representa o resultado da medição de Rotação (rpm). Temos que Rot tem distribuição normal com incerteza (u(Rot)) e o fator de abrangência k;
  • Torque: Representa o resultado da medição de Torque (Nm). Temos que Torque tem distribuição normal com incerteza (u(Torque)) e o fator de abrangência determinados como no módulo (4.4.1 Torque).
  • $ \theta $: Representa uma constante. Logo, $ \theta $ não tem incerteza.

$ \mathbf{\checkmark} $Incerteza Combinada ($ u_{c}(Pot) $)

A incerteza combinada para a potência é dada por:

$$u^2_c(Pot)=\left(\frac{\partial Pot}{\partial Rot}\right)^2 u^2(Rot)+\left(\frac{\partial Pot}{\partial Torque}\right)^2 u^2(Torque)$$

$$=\left(\frac{Torque}{\theta}\right)^2 u^2(Rot) + \left(\frac{Rot}{\theta}\right)^2 u^2(Torque) $$

Logo,

$$u_c(Pot) = \sqrt{\left(\frac{Torque}{\theta}\right)^2 u^2(Rot) + \left(\frac{Rot}{\theta}\right)^2 u^2(Torque)}$$

$ \mathbf{\checkmark} $Incerteza Combinada Relativa ($ u_{r}(Pot) $)

Dividindo $ u_c^2(Pot) $ por $ Pot^2 $ obtemos a incerteza combinada relativa para a potência como:

$$u_{r}(Pot)=\sqrt{\left(\frac{1}{Rot}\right)^2 u^2(Rot) + \left(\frac{1}{Torque}\right)^2 u^2(Torque)}$$

$$=\sqrt{u_{r}^2(Rot) + u_{r}^2(Pot)}$$

Note que, como a calibração da grandeza potência depende das calibrações de Rotação e torque, as estratégia aqui aplicada será tomar a incerteza relativa agrupada para toda a faixa de leitura de ambas as grandezas de entrada, tendo assim uma única incerteza relativa para a grandeza potência referente a toda faixa de leitura.

$ \mathbf{\checkmark} $] Incerteza Expandida ($ U(Pot) $)

A expressão 4.4.2.1 representa a incerteza expandida relativa para a potência

$$U(Pot) = k \times u_r(Pot)~~(4.4.2.1)$$

em que k (fator de abrangência) é o quantil da distribuição t-Student com $ \nu_{eff}(Pot) $ graus de liberdade e confiança de 95%. Aqui, consideraremos os graus de liberdade igual a infinito, pois só temos fontes do tipo B. Logo, adotaremos k = 2.

A Tabela 4.4.2.1 apresenta o resumo para o cálculo da incerteza da potência, considerando as incertezas combinadas das grandezas de entrada.

Screenshot_52

C.S. Coeficiente de Sensibilidade

Contr. Contribuição

G.L. Graus de Liberdade

Tabela 4.4.2.1: Resumo do Cálculo de Incerteza para a Potência.

A Tabela 4.4.2.2 apresenta o resumo para o cálculo da incerteza da potência, considerando as incertezas combinadas relativas das grandezas de entrada.

Screenshot_53

C.S. Coeficiente de Sensibilidade

Contr. Contribuição

G.L. Graus de Liberdade

Tabela 4.4.2.2: Resumo do Cálculo de Incerteza para a Potência.

Aplicação

Agora vamos apresentar uma aplicação para o cálculo da incerteza expandida da variável potência. As informações necessárias para o cálculo são:

Incerteza expandida relativa herdada da Rotação (u(Rot)) é 0,000935 (ou 0,0935% do valor de lido), com fator de abrangência k = 2;

Incerteza expandida relativa herdada do Torque (u(Torque)) é 0,016933 (ou 1,6933% do valor lido), com fator de abrangência k = 2.

A incerteza combinada relativa para a potência é dada por:

$$u_{r}(Pot)=\sqrt{\left(\frac{u^2(Rot)}{Rot}\right)^2 + \left(\frac{u^2(Torque)}{Torque}\right)^2}$$

$$=\sqrt{u_r^2(Rot) + u_r^2(Torque)}$$

$$=\sqrt{\left(\frac{0,000935}{2}\right)^2 + \left(\frac{0,016933}{2}\right)^2}$$

$$=0,00848.$$

Como não temos incerteza do tipo A, vamos considerar que os graus de liberdade são infinito. Com isso, o fator de abrangência é k = 2. Portanto, a incerteza expandida relativa para a potência é dada por:

$$U(Pot)=k \times u_{r}(Pot)=2 \times 0,00848=0,01695.$$

Portanto, a incerteza relativa da potência para toda a faixa de leituras é 1,695% do valor lido.

Screenshot_54

Tabela 4.4.2.3: Resumo do Cálculo de Incerteza para a grandeza Potência.

18.4.4.3 Fator de Correção - Motor OTTO

$ \mathbf{\checkmark} $Equação de Medição

A expressão 4.4.3.1 representa a equação de medição utilizada para a obtenção do Fator de correção. Neste caso é dada por:

$$F_c=\left(\frac{99}{P_{as}}\right)^{1,2}~\left(\frac{T_{adm}}{298}\right)^{0,6}~~(4.4.3.1)$$

  • $P_{as}$: Representa a pressão do ar seco (kPa). Temos que $ P_{as} $ tem distribuição normal com incerteza (u($P_{as}$)) e fator de abrangência k.
  • $T_{adm}$: Representa a temperatura de admissão (Kelvin). Temos que $T_{adm}$ tem distribuição normal com incerteza u(Tadm) e fator de abrangência k.

$ \mathbf{\checkmark} $Incerteza Combinada ($ u_{c}(F_c) $)

A incerteza combinada para a grandeza Fator de Correção é

$$u_c(F_c)=\sqrt{ \left(\frac{\partial F_c}{\partial P_{as}}\right)^2~u^2(P_{as}) + \left(\frac{\partial F_c}{\partial T_{adm}}\right)^2~u^2(T_{adm})}$$

$$=\sqrt{\left[1,2\left(\frac{T_{adm}}{298}\right)^{0,6}\left(\frac{99}{P_{as}}\right)^{0,2}\left(\frac{99}{(P_{as})^2}\right)\right]^2u^2(P_{as})+\left[\frac{0,6}{298} \left(\frac{99}{P_{as}}\right)^{1,2}\left(\frac{298}{T_{adm}}\right)^{0,4} \right]^2~u^2(T_{adm})}$$

$ \mathbf{\checkmark} $Incerteza Expandida ($ U(F_c) $)

A expressão 4.4.3.2 representa a incerteza expandida para o Fator de Correção

$$ U(F_c) = k\times u_{c}(F_c)~~(4.4.3.2)$$

em que k (fator de abrangência) é o quantil da distribuição t-Student com $ \nu_{eff}(F_c) $ graus de liberdade e confiança de 95%. Como não temos incerteza do tipo A, vamos considerar que os graus de liberdade são infinito. Com isso, o fator de abrangência é k = 2.

A Tabela 4.4.3.1 apresenta o resumo para o cálculo da incerteza da grandeza Fator de correção.

Screenshot_55

C.S. Coeficiente de Sensibilidade

Contr. Contribuição

G.L. Graus de Liberdade

Tabela 4.4.3.1: Resumo do Cálculo de Incerteza para o Fator de Correção.

Aplicação

As informações necessárias para o cálculo são:

  • A incerteza expandida herdada de temperatura de admissão é 0,4908189ºC com fator de abrangência 2,03218340.

No ensaio, a temperatura de admissão não pode ultrapassar 25ºC. É este valor (25 ºC) que vamos utilizar para o cálculo do coeficiente de sensibilidade;

A incerteza expandida herdada da pressão do ar seco (Considerando um DewPoint) é 0,169187429kPa com fator de abrangência k=2. Aqui, vamos utilizar uma pressão igual a 91,25493838kPa.

Assim,

$$F_c=\left(\frac{99}{P_{as}}\right)^{1,2}~\left(\frac{T_{adm}}{298}\right)^{0,6}$$

$$=\left(\frac{99}{91,25493838}\right)^{1,2}~\left(\frac{273+25}{298}\right)^{0,6}$$

$$=1,10269291.$$

A incerteza combinada é:

$$u_{c}(F_c)=\sqrt{ \left(\frac{\partial F_c}{\partial P_{as}}\right)^2~u^2(P_{as}) + \left(\frac{\partial F_c}{\partial T_{adm}}\right)^2~u^2(T_{adm})}$$

$$=\sqrt{ \left(0,014500382\right)^2~\left(\frac{0,169187429}{2} \right)^2 + \left(0,002220187\right)^2~\left(\frac{0,4908189}{2,03218340} \right)^2}=$$

$$=0,001338723.$$

Como não temos incerteza do tipo A, vamos considerar que os graus de liberdade são infinito. Com isso, o fator de abrangência é k = 2.

Desta forma, a incerteza expandida para o fator de correção é:

$$U(F_c)=2 \times u_{c}(F_c)=2 \times 0,001338723=0,00267746.$$

Screenshot_56

Tabela 4.4.3.2: Resumo do Cálculo de Incerteza para a grandeza Fator de Correção.

Calcularemos a incerteza relativa do fator de correção do motor OTTO da seguinte forma:

$$U_r(F_c)=\frac{U(F_c)}{F_c}=\frac{0,002677446}{1,10269291}=0,0024281.$$

Portanto, a incerteza relativa do fator de correção do motor OTTO é 0,243% do valor lido.

OBS: As aplicações via Higroclip, Bulbo seco e Bulbo úmido são similares à esta acima, com a pressão do ar seco herdada conforme o equipamento utilizado.

18.4.4.4 Torque Corrigido

O procedimento aqui empregado será de forma a obtermos uma incerteza para toda a faixa de medição de torque corrigido. Desta forma:

$ \mathbf{\checkmark} $Equação de Medição

A expressão 4.4.4.1 representa a equação de medição utilizada para a obtenção do torque corrigido. Neste caso, temos:

$$T_c = T \times F_c~~(4.4.4.1)$$

em que

  • $T_c$: Representa o Torque Corrigido;
  • $T$: Representa o Torque. Temos que T tem distribuição normal com incerteza agrupada U(T) e fator de abrangência k;
  • $F_c$: Representa o Fator de Correção. Temos que $F_c$ tem distribuição normal com incerteza U($F_c$) e fator de abrangência k.

$ \mathbf{\checkmark} $Incerteza Combinada $ (u_{c}) $

A incerteza combinada para a grandeza torque corrigido é dada por:

$$u_{c}^2(T_c)=\left(\frac{\partial T_C}{\partial T} \right)^2 u^2(T) + \left(\frac{\partial T_C}{\partial F_c} \right)^2 u^2(F_c)$$

$$=F_c^2~ u^2(T) + T^2~ u^2(F_c)$$

Logo,

$$u_{c}^2(T_C) = \sqrt{F_c^2 u^2(T) + T^2 u^2(F_c)}~~(4.4.4.2)$$

em que $ u^2(.) $ é a contribuição de cada fonte de incerteza. A expressão 4.4.4.2 representa a incerteza combinada do torque corrigido.

$ \mathbf{\checkmark} $Incerteza Combinada Relativa ($ u_{r} $)

A incerteza combinada relativa para o torque corrigido é dada por:

$$\frac{u_{r}^2(T_C)}{T_C^2}=\frac{F_c^2}{T \times F_c} u^2(T) + \frac{T^2}{T \times F_c}u^2(F_c)$$

$$=\left( \frac{F_c}{T} \right)^2 u^2(T) + \left( \frac{T}{F_c} \right)^2 u^2(F_c)$$

$$=u_r^2(T) + u_r^2(F_c)$$

Assim, temos que:

$$u_{r}(T_c) = \sqrt{u_r^2(T) + u_r^2(F_c)}~~(4.4.4.3)$$

sendo $ u_{r}^2(.) $ a incerteza de contribuição relativa para cada fonte padrão. A expressão 4.4.4.3 representa a incerteza combinada relativa do torque corrigido.

$ \mathbf{\checkmark} $Incerteza Expandida ($ U(T_c) $)

A expressão 4.4.4.4 representa a incerteza expandida relativa para o torque corrigido

$$U(T_c) = k*u_{r}(T_c)~~(4.4.4.4)$$

em que k (fator de abrangência) é o quantil da distribuição t-Student com $ \nu_{eff}(T_c) $ graus de liberdade e confiança de 95%.Os graus de liberdade são dados por:

$$\nu_{eff}(T_c) =\frac{(u_r(T_c))^4}{\frac{u_r^4(T)}{\infty} + \frac{u_r^4(F_c)}{\infty}}$$

A Tabela 4.4.4.1 apresenta o resumo para o cálculo da incerteza da grandeza Torque Corrigido.

A Tabela 4.4.4.1 apresenta o resumo para o cálculo da incerteza do torque corrigido, considerando as incertezas combinadas das grandezas de entrada.

Screenshot_57

C.S. Coeficiente de Sensibilidade

Contr. Contribuição

G.L. Graus de Liberdade

Tabela 4.4.4.1: Resumo do Cálculo de Incerteza para o torque corrigido.

A Tabela 4.4.4.2 apresenta o resumo para o cálculo da incerteza do torque corrigido, considerando as incertezas combinadas relativas das grandezas de entrada.

Screenshot_58

C.S. Coeficiente de Sensibilidade

Contr. Contribuição

G.L. Graus de Liberdade

Tabela 4.4.4.2: Resumo do Cálculo de Incerteza para o torque corrigido.

Aplicação

As informações necessárias para o cálculo de incerteza do torque corrigido são:

  • A incerteza expandida relativa herdada do fator de correção é 0,002428, com k=2;
  • A incerteza expandida relativa agrupada herdada do torque é 0,016934, com k=2.

Assim,

$$u_{r}(T_C)=\sqrt{\left( \frac{u(T)}{T} \right)^2 +\left( \frac{u(F_c)}{F_c} \right)^2}$$

$$=\sqrt{\left( u_{r}(T) \right)^2 +\left( u_{r}(F_c) \right)^2}$$

$$=\sqrt{\left( \frac{0,016934}{2} \right)^2 + \left( \frac{0,002428}{2}\right)^2}$$

$$=\sqrt{(0,008467)^2 +(0,001214)^2}$$

$$=0,008553.$$

Como não temos incerteza do tipo A, vamos considerar que os graus de liberdade são infinito. Com isso, o fator de abrangência é k = 2. Desta forma, a incerteza expandida relativa para o torque corrigido é:

$$U(T_C)=2 \times u_{cr}(T_C)=2 \times 0,008553=0,0171.$$

Screenshot_59

Tabela 4.4.4.3: Resumo do Cálculo de Incerteza para a grandeza Torque Corrigido.

Portanto, a incerteza relativa do Torque Corrigido é 1,71% do valor lido.

18.4.4.5 Potência Corrigida

O procedimento aqui empregado será de forma a obtermos uma incerteza para toda a faixa de medição da potência corrigida. Desta forma:

$ \mathbf{\checkmark} $Equação de Medição

A expressão 4.4.5.1 representa a equação de medição utilizada para a obtenção da potência corrigida. Neste caso, temos:

$$P_c = P \times F_c~~(4.4.5.1)$$

em que

  • $P_c$: Representa a Potência Corrigida;
  • P: Representa a Potência. Temos que P tem distribuição normal com incerteza relativa U(P) e fator de abrangência calculado no módulo 4.4.2 Potência;
  • $F_c$: Representa o Fator de Correção. Temos que $ F_c $ tem distribuição normal com incerteza U($F_c$) e fator de abrangência calculado no módulo 4.4.3 Fator de Correção do Motor OTTO;

$ \mathbf{\checkmark} $Incerteza Combinada $ (u_{c}) $

A incerteza combinada para a grandeza Potência Corrigida é dada por:

$$u_{c}^2(P_c)=\left(\frac{\partial P_c}{\partial P} \right)^2 u^2(P) + \left( \frac{\partial P_c}{\partial F_c} \right)^2 u^2(F_c)$$

$$=F_c^2~ u^2(P) + T^2~ u^2(F_c)$$

Logo,

$$u_{c}^2(P_c)=\sqrt{F_c^2 u^2(P) + P^2 u^2(F_c)}~~(4.4.6.2)$$

em que $ u^2(.) $ é a contribuição de cada fonte de incerteza. A expressão 4.4.5.2 representa a incerteza combinada da Potência Corrigida.

$ \mathbf{\checkmark} $Incerteza Combinada Relativa ($ u_{r} $)

A incerteza combinada relativa para a Potência Corrigida é dada por:

$$\frac{u_{c}^2(P_c)}{P_c^2}=\frac{F_c^2}{P \times F_c} u^2(P) + \frac{P^2}{P \times F_c}u^2(F_c)$$

$$=\left( \frac{F_c}{P} \right)^2 u^2(P) + \left( \frac{P}{F_c} \right)^2 u^2(F_c)$$

Assim, temos que:

$$u_{r}(P_c)=\sqrt{\left( \frac{1}{P} \right)^2 u^2(P) + \left( \frac{1}{F_c} \right)^2 u^2(F_c)}$$

$$=\sqrt{u_r^2(P) + u_r^2(F_c)}~~(4.4.5.3)$$

sendo $ u_{r}^2(.) $ a incerteza de contribuição relativa para cada fonte padrão. A expressão 4.4.5.3 representa a incerteza combinada relativa da potência corrigida.

$ \mathbf{\checkmark} $Incerteza Expandida ($ U(P_c) $)

A expressão 4.4.5.4 representa a incerteza expandida relativa para a Potência Corrigida

$$U(P_c) = k*u_{r}(P_C)~~(4.4.5.4)$$

em que k (fator de abrangência) é o quantil da distribuição t-Student com $ \nu_{eff}(P_c) $ graus de liberdade e confiança de 95%. Os graus de liberdade são dados por:

$$\nu_{eff}(P_c) =\frac{(u_r(P_c))^4}{\frac{u_r^4(P)}{\infty} + \frac{u_r^4(F_c)}{\infty}}$$

A Tabela 4.4.5.1 apresenta o resumo para o cálculo da incerteza da grandeza Potência Corrigida.

Screenshot_60

C.S. Coeficiente de Sensibilidade

Contr. Contribuição

G.L. Graus de Liberdade

Tabela 4.4.5.1: Resumo do Cálculo de Incerteza para a Potência Corrigida.

A Tabela 4.4.5.2 apresenta o resumo para o cálculo da incerteza da potência corrigida, considerando as incertezas combinadas relativas das grandezas de entrada.

Screenshot_61

C.S. Coeficiente de Sensibilidade

Contr. Contribuição

G.L. Graus de Liberdade

Tabela 4.4.5.2: Resumo do Cálculo de Incerteza para a Potência corrigida.

Aplicação

As informações necessárias para o cálculo de incerteza da potência corrigida são:

A incerteza expandida relativa herdada do fator de correção é 0,0024281, com k=2;

A incerteza expandida relativa herdada da potência é 0,01695, com k=2.

Assim,

$$u_{r}(P_c)=\sqrt{\left(\frac{u(P)}{P} \right)^2 + \left( \frac{u(F_c)}{F_c} \right)^2}$$

$$=\sqrt{\left(u_{r}(P) \right)^2 +\left(u_{r}(F_c) \right)^2}$$

$$=\sqrt{\left(\frac{0,0024281}{2} \right)^2 + \left( \frac{0,01695}{2}\right)^2}$$

$$=\sqrt{(0,001214)^2 +(0,008475)^2}$$

$$=0,008561515.$$

Como não temos incerteza do tipo A, vamos considerar que os graus de liberdade são infinito. Com isso, o fator de abrangência é k = 2. Desta forma, a incerteza expandida relativa para a potência corrigida é:

$$U(P_C) =2 \times u_{cr}(P_C)=2 \times 0,0085561515=0,01712303.$$

Screenshot_62

Tabela 4.4.5.3: Resumo do Cálculo de Incerteza para a grandeza Potência Corrigida.

Portanto, a incerteza relativa do Potência Corrigida é 1,712% do valor lido.

18.4.4.6 Ensaio de Repetitividade e Reprodutibilidade

Nessa seção vamos realizar um ensaio de repetitividade e reprodutibilidade para potência efetiva líquida do motor ciclo OTTO. Neste ensaio, consideramos um experimento fatorial completo com os fatores Período (Manhã e Tarde) e Equipe (I e II) e 4 réplicas. este experimento foi realizado nos seguintes pontos de rotação: 1200, 2000, 3000, 3600, 4400, 5200, 5600, 6000 e 6400. Os dados para rotação de 6400 estão apresentados abaixo.

Potência Corrigida Período Equipe
51,4 Tarde I
51,9 Tarde I
51,4 Manha I
51,4 Manha I
51,3 Tarde II
51,3 Tarde II
51,7 Manha I
51,9 Manha I
52,1 Tarde I
52,2 Tarde I
52,3 Manha II
52,3 Manha II
52,1 Tarde II
51,4 Tarde II
52,1 Manha II
52,1 Manha II

Tabela 4.4.6.1: Dados no ponto de 6400 rpm.

A seguir apresentamos uma análise descritiva dos dados

Período Equipe Mínimo 1º Quartil Média Mediana 3º Quartil Máximo Desvio Padrão Amplitude
Manha I 51,4 51,4 51,6 51,6 51,9 51,9 0,24 0,5
Manha II 52,1 52,1 52,2 52,2 52,3 52,3 0,12 0,2
Tarde I 51,4 51,5 51,9 52,0 52,2 52,2 0,36 0,8
Tarde II 51,3 51,3 51,5 51,4 51,9 52,1 0,39 0,8

Tabela 4.4.6.2: Estatísticas descritivas.

O gráfico de interações é apresentado abaixo

RR_maua

Figura 4.4.6.1: Gráfico de Interações.

Através da técnica de ANOVA com efeitos aleatórios determinamos a repetitividade do método no ponto de 6400 rpm.

ANOVA_maua

Como podemos observar na tabela da ANOVA, tanto o fator período quanto o fator equipe foram considerados não significativos com $ \alpha=0,05. $ Neste caso a repetitividade foi de 0,38204 kW. De forma geral, temos a tabela 4.4.6.3.

Screenshot_63

Tabela 4.4.6.3: Repetitividade e Reprodutibilidade Relativo.

18.4.4.7 Incerteza do Ensaio Potência Efetiva Líquida

Nesta fase, vamos combinar a incerteza efetiva na calibração com a incerteza devido a repetitividade e reprodutibilidade do método de ensaio. Para isto, tomamos a seguinte equação de medição:

$ \mathbf{\checkmark} $Equação de Medição

A expressão 4.4.7.1 representa a equação de medição utilizada para a obtenção da potência efetiva líquida do motor ciclo OTTO. Neste caso temos:

$$PEL=P_c+\varepsilon~~(4.4.7.1)$$

em que

  • PEL: Representa a Potência efetiva do motor;
  • $P_c$: Representa a Potência corrigida. Temos que $P_c$ tem distribuição normal com incerteza relativa $U_r$($P_c$) e fator de abrangência calculado no módulo 4.4.6 Potência Corrigida;
  • ε: Representa a Repetitividade e Reprodutibilidade. Podemos supor que a $ \varepsilon $ tem distribuição normal com média zero e desvio padrão calculada no módulo Ensaio de RR.

$ \mathbf{\checkmark} $Incerteza Combinada ($ u_{c}(PEL) $)

A incerteza combinada relativa para a Potência efetiva do motor é dada por:

$$u_{cr}(PEL)=\sqrt{u^2_r(P_c)+u^2_r(\varepsilon)}~~(4.4.7.2)$$

sendo $ u_{r}^2(.) $ a incerteza de contribuição relativa para cada fonte padrão. A expressão (4.4.7.2) representa a incerteza combinada relativa da potência efetiva do motor.

$ \mathbf{\checkmark} $Incerteza Expandida ($ U(PEL) $)

A expressão 4.4.7.3 representa a incerteza expandida relativa para a Potência Efetiva Líquida.

$$U(PEL) = k*u_{cr}(PEL)~~(4.4.7.3)$$

em que k (fator de abrangência) é o quantil da distribuição t-Student com $ \nu_{eff}(PEL) $ graus de liberdade e confiança de 95%. Os graus de liberdade são dados por:

$$\nu_{eff}(PEL) =\frac{(u_{cr}(PEL))^4}{\frac{u_r^4(P_c)}{\infty} + \frac{u_r^4(\varepsilon)}{\infty}}$$

A Tabela 4.4.7.1 apresenta o resumo para o cálculo da incerteza da potência efetiva líquida do motor.

Screenshot_64

Tabela 4.4.7.1: Resumo do Cálculo de Incerteza para a Potência Efetiva Líquida.

Aplicação

As informações necessárias para o cálculo de incerteza da potência efetiva líquida do motor são:

A incerteza expandida relativa herdada da potência corrigida é 0,01712303, com k=2;

A incerteza expandida relativa herdada da potência é 0,00545.

Assim,

$$u_{r}(PEL)=\sqrt{\left(u_{r}(P_c) \right)^2 +\left(u_{r}(\varepsilon) \right)^2}$$

$$=\sqrt{\left(\frac{0,01712303}{2} \right)^2 + 0,00545^2}$$

$$=0,01015045.$$

Como não temos incerteza do tipo A, vamos considerar que os graus de liberdade são infinito. Com isso, o fator de abrangência é k = 2. Desta forma, a incerteza expandida relativa para a potência efetiva líquida do motor é:

$$U(PEL) =2 \times u_{cr}(PEL)=2 \times 0,01015045=0,0203.$$

Screenshot_65

Tabela 4.4.7.3: Resumo do Cálculo de Incerteza para a grandeza Potência efetiva Líquida.

Portanto, a incerteza relativa da Potência Efetiva Líquida do motor ciclo OTTO é 2,03% do valor lido.

18.4.5 Cálculo de Incerteza para Migração à partir de materiais

Esta aplicação é referente ao cálculo de incerteza para determinar a migração específica de um certo composto químico à partir de materiais, embalagens e equipamentos plásticos, por meio de sua extração em simulantes de alimentos, a uma determinada condição de temperatura e tempo de contato, e posterior análise quantitativa por cromatografia gasosa e detecção por ionização de chama com um padrão interno de referência.

A migração específica deste composto é determinada em simulante aquoso (água destilada, solução de ácido acético em água destilada ou soluções de etanol em água destilada). Após a etapa de extração e adição de padrão interno, a concentração do composto químico é determinada diretamente por cromatografia gasosa nos simulantes.

O ensaio é dividido em duas partes principais, a calibração do cromatógrafo e a medição da amostra. Para calibrar o cromatógrafo, vamos preparar soluções padrão e realizar a curva de calibração do equipamento.

18.4.5.1 Cálculo de Incerteza devido às Soluções

Primeiramente, para calibração do cromatógrafo, temos a fase de preparação das soluções que descrevemos a seguir:

  1. Preparação da solução Mãe do padrão primário (6 mg/mL), ou seja, pesamos 0,15 g em uma balança analítica do padrão primário em um balão volumétrico de 25 mL;

balao

  1. Preparação da solução Mãe do padrão secundário (6 mg/mL), ou seja, pesamos 0,6 g em uma balança analítica do padrão secundário em um balão volumétrico de 100 mL;

balao

  1. Preparação das soluções Filhas dos padrões primário e secundário. Em balões volumétricos de 25 mL os volumes adicionados com pipeta volumétrica de cada solução mãe de padrão primário (item 1) e padrão secundário (item 2);

Aplicacao_Migracao_0

  1. Preparação das soluções Netas dos padrões primário e secundário. Pipetar com auxílio de uma micropipeta volumétrica de 5 mL do simulante aquoso em uma série de frascos do amostrador de espaço-livre de 20 mL. Em seguida, adicionar com uma micropipeta volumétrica calibrada 100 μL de cada uma das soluções preparadas. Cada solução de ver preparada em duplicata. Assim, as soluções estarão prontas para análise no amostrador de espaço-livre e injeção no cromatógrafo a gás com detector de ionização de chama.

Aplicacao_Migracao1

Solução Mãe do composto químico

$ \mathbf{\checkmark} $Equação de Medição

A expressão (4.5.1.1) representa a equação de medição utilizada para a grandeza concentração mãe, neste caso temos:

$$S_{M1}~=~\left(\frac{M}{V(1~-~\alpha \Delta)}\right)\times P~~(4.5.1.1)$$

  • V: representa o Volume, mL. Temos que $ V $ tem distribuição normal com incerteza u(V) e o fator de abrangência dados pelo certificado de calibração;
  • $ \alpha $: representa o coeficiente de expansão volumétrico, ºC-1. Podemos supor que a $ \alpha $ tem distribuição retangular (ou uniforme) no intervalo $ \left(-\frac{\alpha}{2},~\frac{\alpha}{2}\right) $, em que $ \alpha $ é o coeficiente de expansão volumétrico. Logo a incerteza devido ao coeficiente de expansão volumétrico será dada por:

$$u(\alpha) = \frac{\alpha}{2\sqrt{3}};$$

  • $ \Delta $: representa a diferença entre a temperatura final e inicial ($ \Delta~=~t_f~-~t_i $), ºC;
  • M: representa a massa $ (g) $. Possui duas fontes de incerteza:

a) Herdada da balança, temos que $ M $ tem distribuição normal com incerteza $ u(M) $ e o fator de abrangência k.

b) ResM: Representa a Resolução da balança (g). Podemos supor que a resolução tem distribuição retangular (ou uniforme) no intervalo $ \left(-\frac{ResM}{2},~\frac{ResM}{2}\right) $, em que ResM é a resolução do balança. Logo a incerteza devida a resolução será dada por:

$$u(ResM) = \frac{ResM}{2\sqrt{3}}$$

  • P: representa a Pureza. No entanto, será considerada uma constante, ou seja, uma variável com incerteza associada desprezível;

$ \mathbf{\checkmark} $ Incerteza Combinada

$ u_c(S_{M1})=\sqrt{\left(\frac{\partial S_{M1}}{\partial M}\right)^2 (u^2(M)+u^2(Res_M))+\left(\frac{\partial S_{M1}}{\partial V}\right)^2 u^2(V)+\left(\frac{\partial S_{M1}}{\partial \alpha}\right)^2 u^2(\alpha)+ \left(\frac{\partial S_{M1}}{\partial \Delta} \right)^2 u^2(\Delta)} $

em que,

$$\left(\frac{\partial S_{M1}}{\partial M}\right)=\left(\frac{1}{V(1~-~\alpha \Delta)}\right)\times P=\frac{S_{M1}}{M}$$

$$\left(\frac{\partial S_{M1}}{\partial V}\right) =\left(\frac{M}{V^2(1~-~\alpha \Delta)}\right)\times P=\frac{S_{M1}}{V}$$

$$\left(\frac{\partial S_{M1}}{\partial \alpha}\right)=\left(\frac{M\times\Delta\times V}{V(1~-~\alpha \Delta)^2}\right)\times P=\frac{S_{M1}\times\Delta\times V}{(1-\alpha \Delta)}$$

$$\left(\frac{\partial S_{M1}}{\partial \Delta}\right)=\left(\frac{M\times\alpha\times V}{V(1~-~\alpha \Delta)^2}\right)\times P=\frac{S_{M1}\times\alpha\times V}{(1-\alpha \Delta)}$$

$ \mathbf{\checkmark} $Incerteza Expandida ($ U(S_{M1}) $)

Aqui, a expressão (4.5.1.2) representa a incerteza expandida para a concentração mãe.

$$U(S_{M1}) = k\times u_c(S_{M1})~~(4.5.1.2)$$

em que k (fator de abrangência), que é o quantil da distribuição t-Student com $ \nu_{eff}(S_{M1}) $ graus de liberdade e confiança de 95%. Os graus de liberdade são dados por:

$$\nu_{eff}(S_{M1})=\frac{(u_c(S_{M1}))^4}{\frac{u^4(M)}{\infty}+\frac{u^4(V)}{\infty}+\frac{u^4(\alpha)}{\infty} + \frac{u^4(\Delta)}{\infty} +\frac{u^4(ResM)}{\infty}}$$

Para efetuar os cálculos da calibração, as informações necessárias são:

Fontes Valor Incerteza (U) Fator de Abrangência (k) Resolução (Res) Fator de Abrangência (k)
Massa (mg) 150 0,1 2,52 0,1 3,464101615
Volume (mL) 25 0,008 2,231
Coef.Expansão (ºC-1) 0,0001 0,000001 3,464101615
Dif.Temperatura (ºC) 0,5 0,005 3,464101615
Pureza 0,99 0 1,96

Incerteza Combinada ($ u_{c}(S_{M1}) $)

A incerteza combinada para a grandeza concentração mãe é dada por:

$ u_c(S_{M1})=\sqrt{\left(\frac{\partial S_{M1}}{\partial M}\right)^2 (u^2(M)+u^2(Res_M))+\left(\frac{\partial S_{M1}}{\partial V}\right)^2 u^2(V)+\left(\frac{\partial S_{M1}}{\partial \alpha}\right)^2 u^2(\alpha)+ \left(\frac{\partial S_{M1}}{\partial \Delta} \right)^2 u^2(\Delta)} $

$$=0,002121917$$

em que

$$\left(\frac{\partial S_{M1}}{\partial M}\right)=\left(\frac{1}{V(1~-~\alpha \Delta)}\right)\times P=\frac{S_{M1}}{M}=0,03960198$$

$$\left(\frac{\partial S_{M1}}{\partial V}\right)=\left(\frac{M}{V^2(1~-~\alpha \Delta)}\right)\times P=\frac{S_{M1}}{V}=0,237611881$$

$$\left(\frac{\partial S_{M1}}{\partial \alpha}\right)=\left(\frac{M\times\Delta\times V}{V(1~-~\alpha \Delta)^2}\right)\times P=\frac{S_{M1}\times\Delta\times V}{(1-\alpha \Delta)}=74,25742556$$

$$\left(\frac{\partial S_{M1}}{\partial \Delta}\right)=\left(\frac{M\times\alpha\times V}{V(1~-~\alpha \Delta)^2}\right)\times P=\frac{S_{M1}\times\alpha\times V}{(1-\alpha \Delta)}=0,014851485$$

Incerteza Expandida ($ U(S_{M1}) $)

$$U(S_{M1}) = k\times u_c(S_{M1})=1,95996563\times0,002121917=0,004158885$$

Os graus de liberdade são dados por:

$$\nu_{eff}(S_{M1}) =\frac{(u_c(S_{M1}))^4}{\frac{u^4(M)}{\infty}+\frac{u^4(V)}{\infty}+\frac{u^4(\alpha)}{\infty} + \frac{u^4(\Delta)}{\infty}+\frac{u^4(ResM)}{\infty}}\rightarrow\infty $$

Screenshot_66

Concentração da Solução Filha

$ \mathbf{\checkmark} $ Equação de Medição

A expressão (4.5.1.3) representa a equação de medição utilizada para a grandeza concentração filha, neste caso temos:

$$S_{F}~=~\frac{V_{p1}\times(1-\alpha_{p1}\Delta)\times S_{M1}}{V_b~(1~-~\alpha_b\Delta)}~~(4.5.1.3)$$

em que

  • $V_{p1}$: representa o Volume da pipeta da solução mãe do composto químico, mL. Temos que Vp1 tem distribuição normal com incerteza $ u(V_{p1}) $ e o fator de abrangência dado pelo certificado de calibração;
  • $V_{b1}$: representa o Volume do balão da solução mãe do composto, mL. Temos que Vb1 tem distribuição normal com incerteza $ u(V_{b1}) $ e o fator de abrangência dado pelo certificado de calibração;
  • $ \alpha_{p1} $: representa o coeficiente de expansão volumétrico da pipeta, ºC-1. Podemos supor que a $ \alpha_{p1} $ tem distribuição retangular (ou uniforme) no intervalo $ \left(-\frac{\alpha_{p1}}{2},~\frac{\alpha_{p1}}{2}\right) $,

em que $ \alpha_{p1} $ é o coeficiente de expansão volumétrico. Logo a incerteza devido ao coeficiente de expansão volumétrico será dada por:

$$u(\alpha_{p1}) = \frac{\alpha_{p1}}{2\sqrt{3}};$$

  • $ \alpha_{b1} $: representa o coeficiente de expansão volumétrico do balão, ºC-1. Podemos supor que a $ \alpha_{b1} $ tem distribuição retangular (ou uniforme) no intervalo $ \left(-\frac{\alpha_{b1}}{2},~\frac{\alpha_{b1}}{2}\right) $,

em que $ \alpha_{b1} $ é o coeficiente de expansão volumétrico. Logo a incerteza devido ao coeficiente de expansão volumétrico será dada por:

$$u(\alpha_{b1}) = \frac{\alpha_{b1}}{2\sqrt{3}};$$

  • $ \Delta $: representa a a diferença entre a temperatura final e inicial ($ \Delta~=~t_f~-~t_i $), ºC;
  • $S_{M1}$: representa a incerteza herdada da concentração mãe, mg/mL. Temos que SM1 tem distribuição normal com incerteza $ U(S_{M1}) $ e o fator de abrangência calculado no tópico anterior;

$ \mathbf{\checkmark} $ Incerteza Combinada ($ u_{c}(S_F) $)

A incerteza combinada para a grandeza concentração mãe é dada por:

$$u^2_c(S_F)=\left(\frac{\partial S_F}{\partial V_{p1}}\right)^2 u^2(V_{p1})+\left(\frac{\partial S_F}{\partial V_{b}}\right)^2 u^2(V_{b})+\left(\frac{\partial S_F}{\partial \alpha_{p1}}\right)^2 u^2(\alpha_{p1}) +\left(\frac{\partial S_F}{\partial \alpha_{b}}\right)^2 u^2(\alpha_{b})+ $$

$$+\left(\frac{\partial S_F}{\partial \Delta} \right)^2 u^2(\Delta) + \left(\frac{\partial S_F}{\partial S_{M1}} \right)^2 u^2(S_{M1})$$

em que

$$\left(\frac{\partial S_F}{\partial V_{p1}}\right)=\left(\frac{(1-\alpha_{p1}\Delta)\times S_{M1}}{V_b(1-\alpha_b \Delta)}\right)=\frac{S_F}{V_{p1}}$$

$$\left(\frac{\partial S_F}{\partial V_b}\right)=\frac{V_{p1}\times(1-\alpha_{p1}\Delta)\times S_{M1}}{V_b^2(1-\alpha_b \Delta)}=\frac{S_F}{V_b}$$

$$\left(\frac{\partial S_F}{\partial \alpha_{p1}}\right)=\left(\frac{V_{p1}\times\Delta\times S_{M1}}{V_b(1-\alpha_b \Delta)}\right)$$

$$\left(\frac{\partial S_F}{\partial \alpha_b}\right)=\left(\frac{V_{p1}\times(1-\alpha_{p1}\Delta)\times S_{M1}}{V_b(1-\alpha_b\Delta)^2}\right)\times\Delta=\frac{S_F\times\Delta}{(1-\alpha_b\Delta)}$$

$$\left(\frac{\partial S_F}{\partial \Delta}\right)=\left(\frac{V_{p1}\times S_{M1}\times(\alpha_b-\alpha_{p1})}{(1-\alpha_b \Delta)^2}\right)$$

$$\left(\frac{\partial S_F}{\partial S_{M1}}\right)=\frac{V_{p1}\times(1-\alpha_{p1}\Delta) }{V_b(1-\alpha_b\Delta)}=\frac{S_F}{S_{M1}}$$

em que, $ u^2(.) $ é a contribuição de cada fonte de incerteza.

$ \mathbf{\checkmark} $ Incerteza Expandida ($ U(S_F) $)

A expressão (4.5.1.4) representa a incerteza expandida para a concentração filha.

$$U(S_F) = k\times u_c(S_F)$$

em que k (fator de abrangência) é o quantil da distribuição t-Student com $ \nu_{eff}(S_F) $ graus de liberdade e confiança de 95%. Como neste cálculo estamos trabalhando apenas com incertezas do tipo B, então o equivale a $ \nu_{eff}(S_F)=\infty $

Para a solução filha são preparadas 6 soluções, para nossa aplicação faremos para a solução filha 2. Para efetuar os cálculos da calibração, as informações necessárias são:

Fontes Valor Incerteza (U) Fator de Abrangência (k)
Volume do Balão (mL) 25 0,008 2,231
Volume da Pipeta 1 (mL) 0,5 0,0011 2,31
Coef.Expansão balão (ºC-1) 0,0001 0,000001 3,464102
Coef.Expansão pipeta 1 (ºC-1) 0,00012 0,0000012 3,464102
Dif.Temperatura (ºC) 0,5 0,005 3,464102
Solução Mãe 1 5,940297 0,004158886 1,959966357

Aplicando os valores da tabela obtemos os seguintes resultados:

Screenshot_67

Concentração da Solução Neta

Agora, vamos considerar o cálculo de incerteza para as concentrações netas.

$ \mathbf{\checkmark} $ Equação de Medição

A expressão (4.5.1.5) representa a equação de medição utilizada para a grandeza concentração filha, neste caso temos:

$$S_{N}=\frac{V_{\mu pf}(1-\alpha_{\mu pf} \Delta)}{V_{\mu pf}(1-\alpha_{\mu pf} \Delta)+V_{\mu ps}(1-\alpha_{\mu ps} \Delta)}\times S_{F}+\varepsilon~~(4.5.1.5)$$

em que

  • $ V_{\mu pf} $: representa o Volume da micro pipeta da solução filha, $ mL $. Temos que $ V_{\mu pf} $ tem distribuição normal com incerteza $ u(V_{\mu pf}) $ e o fator de abrangência dado pelo certificado;
  • $ V_{\mu ps} $: representa o Volume da micro pipeta do simulante, $ mL $. Temos que $ V_{\mu ps} $ tem distribuição normal com incerteza $ u(V_{\mu ps}) $ e o fator de abrangência dado pelo certificado;
  • $ \alpha_{\mu pf} $: representa o coeficiente de expansão volumétrico da micro pipeta, $ {}^\circ C^{-1} $. Podemos supor que a $ \alpha_{\mu pf} $ tem distribuição retangular (ou uniforme) no intervalo $ \left(-\frac{\alpha_{\mu pf}}{2},~\frac{\alpha_{\mu pf}}{2}\right) $, em que $ \alpha_{\mu pf} $ é o coeficiente de expansão volumétrico. Logo a incerteza devido ao coeficiente de expansão volumétrico será dada por:

$$u(\alpha_{\mu pf}) = \frac{\alpha_{\mu pf}}{2\sqrt{3}};$$

  • $ \alpha_{\mu ps} $: representa o coeficiente de expansão volumétrico da micro pipeta, $ {}^\circ C^{-1} $. Podemos supor que a $ \alpha_{\mu ps} $ tem distribuição retangular (ou uniforme) no intervalo $ \left(-\frac{\alpha_{\mu ps}}{2},~\frac{\alpha_{\mu ps}}{2}\right) $, em que $ \alpha_{\mu ps} $ é o coeficiente de expansão volumétrico. Logo a incerteza devido ao coeficiente de expansão volumétrico será dada por:

$$u(\alpha_{\mu ps}) = \frac{\alpha_{\mu ps}}{2\sqrt{3}};$$

  • $ \Delta $: representa a a diferença entre a temperatura final e inicial ($ \Delta~=~t_f~-~t_i $), $ {}^\circ C $;
  • $ S_F $: representa a incerteza herdada da concentração filha, mg/mL. Temos que $ S_F $ tem distribuição normal com incerteza $ U(S_F) $ e o fator de abrangência calculado na seção anterior;
  • $ \varepsilon $: representa a repetitividade. Podemos supor que a distribuição da média das medições tem distribuição normal. Logo, podemos estimar sua incerteza (desvio padrão da média) como:

$$u(\varepsilon) = \frac{s}{\sqrt{n}},$$

em que $ s^2 = \frac{1}{n-1} \displaystyle\sum_{i=1}^{n} (x_i -\overline{X})^2 $ é a variância amostral e $ \overline{X} = \frac{1}{n}\displaystyle \sum^{n_{i=1}} x_i $ é a média amostral;

$ \mathbf{\checkmark} $ Incerteza Combinada ($ u_{c}(S_N) $)

A incerteza combinada para a grandeza concentração neta é dada por:

$$u_c^2(S_N)=\left(\frac{\partial S_N}{\partial V_{\mu pf}}\right)^2 u^2(V_{\mu pf})+\left(\frac{\partial S_N}{\partial V_{\mu ps}}\right)^2 u^2(V_{\mu ps})+\left(\frac{\partial S_N}{\partial \alpha_{\mu pf}}\right)^2 u^2(\alpha_{\mu pf}) +\left(\frac{\partial S_N}{\partial \alpha_{\mu ps}}\right)^2 u^2(\alpha_{\mu ps}) +$$

$$+\left(\frac{\partial S_N}{\partial \Delta} \right)^2 u^2(\Delta) + \left(\frac{\partial S_N}{\partial S_F} \right)^2 u^2(S_F)+ u^2(\varepsilon)$$

em que,

$$\left(\frac{\partial S_N}{\partial V_{\mu pf}}\right)=\frac{V_{\mu ps}(1-\alpha_{\mu pf} \Delta)(1-\alpha_{\mu ps} \Delta)}{(V_{\mu pf}(1-\alpha_{\mu pf} \Delta)+V_{\mu ps}(1-\alpha_{\mu ps} \Delta))^2}\times S_{F}$$

$$\left(\frac{\partial S_N}{\partial V_{\mu ps}}\right)=\frac{V_{\mu pf}(1-\alpha_{\mu pf} \Delta)(1-\alpha_{\mu ps} \Delta)}{(V_{\mu pf}(1-\alpha_{\mu pf} \Delta)+V_{\mu ps}(1-\alpha_{\mu ps} \Delta))^2}\times S_{F}$$

$$\left(\frac{\partial S_N}{\partial \alpha_{\mu pf}}\right)=\frac{V_{\mu ps} V_{\mu pf}\Delta(1-\alpha_{\mu ps} \Delta)}{(V_{\mu pf}(1-\alpha_{\mu pf} \Delta)+V_{\mu ps}(1-\alpha_{\mu ps} \Delta))^2}\times S_{F}$$

$$\left(\frac{\partial S_N}{\partial \alpha_{\mu ps}}\right)=\frac{V_{\mu ps} V_{\mu pf}\Delta(1-\alpha_{\mu pf} \Delta)}{(V_{\mu pf}(1-\alpha_{\mu pf} \Delta)+V_{\mu ps}(1-\alpha_{\mu ps} \Delta))^2}\times S_{F}$$

$$\left(\frac{\partial S_N}{\partial \Delta}\right)=\frac{V_{\mu ps} V_{\mu pf}(\alpha_{\mu ps}-\alpha_{\mu pf})}{(V_{\mu pf}(1-\alpha_{\mu pf} \Delta)+V_{\mu ps}(1-\alpha_{\mu ps} \Delta))^2}\times S_{F}$$

$$\left(\frac{\partial S_N}{\partial S_F}\right)=\frac{V_{\mu pf}(1-\alpha_{\mu pf} \Delta)}{V_{\mu pf}(1-\alpha_{\mu pf} \Delta)+V_{\mu ps}(1-\alpha_{\mu ps} \Delta)}=\frac{S_N}{S_F}$$

em que, $ u^2(.) $ é a contribuição de cada fonte de incerteza.

$ \mathbf{\checkmark} $ Incerteza Expandida ($ U(S_N) $)

A expressão (4.5.1.6) representa a incerteza expandida relativa para a concentração neta.

$$U(S_N) = k\times u_c(S_N)~~(4.5.1.6)$$

em que k (fator de abrangência) é o quantil da distribuição t-Student com $ \nu_{eff}(S_N) $ graus de liberdade e confiança de 95%. Como neste cálculo estamos trabalhando apenas com incertezas do tipo B, então o equivale a $ \nu_{eff}(S_N)=\infty $

Para as soluções netas são preparadas 6 soluções, para nossa aplicação faremos para a solução neta 2. Para efetuar os cálculos da calibração, as informações necessárias são:

Screenshot_68

Aplicando os valores da tabela obtemos os seguintes resultados:

Screenshot_69

18.4.5.2 Cálculo de Incerteza devido ao Cromatógrafo

Incerteza Devido a Regressão

Vamos considerar a seguinte relação linear entre intensidade e concentração:

$$Y_{ij} = \beta_0 + \beta_1~x_{i} + \varepsilon_{ij}~~~~ i =1,\cdots, n;$$

em que,

  • $ Y_{ij} $: representa a i-ésima medição de intensidade referente a i-ésima Área;
  • $ X_{i} $: representa a Área referente a i-ésima Concentração;
  • $ \beta_0 $: representa o coeficiente linear;
  • $ \beta_1 $: representa o coeficiente angular;
  • $ \varepsilon_{ij} $: representa o j-ésimo erro cometido na medição da i-ésima Concentração. Consideramos que os $ \varepsilon_{ij}~\overset{i.i.d.}{\sim}~N(0,\sigma^2), $ ou seja, são independentes e identicamente distribuídos com distribuição Normal com média zero e variância $ \sigma^2. $

Pelo método dos mínimos quadrados, temos que

$$\hat{\beta_1}=\frac{S_{xy}}{S_{xx}}~~e~~\hat{\beta_0}=\overline{y}-\hat{\beta_1}\overline{x}$$

em que,

$ S_{xx} = \displaystyle\sum^n_{i=1} (x_i-\overline{x})^2 = \sum^n_{i=1} x_i^2- 2\overline{x}\sum^n_{i=1} x_i+n\overline{x}^2 = \sum^n_{i=1} x_i^2-n\overline{x}^2; $

$ S_{xy} =\displaystyle \sum^n_{i=1} (x_i-\overline{x})(y_i-\overline{y}) =\sum^n_{i=1}(x_i-\overline{x})y_i; $

$ \overline{y} = \displaystyle\frac{1}{n} \sum^n_{i=1} (y_i) $ representa a média das leituras de área;

$ \overline{x} = \displaystyle\frac{1}{n} \sum^n_{i=1} (x_i) $ representa a média das leituras de concentração.

Na prática, temos interesse em predizer o valor de Área ($ x $) dado uma observação em Concentração ($ y $). Então, dado $ y_0 $ observado, tomamos como estimativa

$$\hat{x_0} = \frac{y_0-\hat{\beta_0} }{\hat{\beta_1}}~.$$

A variância da estimativa $ \hat{x_0} $ é dada por (Veja Brown, 1993, pg. 26)

$$\widehat{Var(\hat{x_0})} = \frac{QME}{\hat{\beta_1}^2}~.$$

A tabela (4.5.2.1) apresenta os dados utilizados na regressão linear.

X Y
Concentracao Area
0 0
0 0
2,329493525 0,137554
2,329493525 0,129687
4,65898705 0,283453
4,65898705 0,258843
9,3179741 0,561361
9,3179741 0,514503
13,97696115 0,808883
13,97696115 0,796494
23,29493525 1,42358
23,29493525 1,41858

Tabela 4.5.2.1: Conjunto de dados para aplicação do modelo de regressão linear.

Aplicando o modelo de regressão linear (para mais detalhes acesse o conteúdo Análise de Regressão), obtemos as seguintes estimativas dos parâmetros:

$ \hat{\beta_0}=-0,013364664 $;

$ \hat{\beta_1}= 0,060596425 $;

$ \sigma $ é estimado pelo $ QME $ da seguinte forma:

Screenshot_71

Therefore, the estimate of uncertainty, for concentration, due to regression is:

$$u(Reg)=\sqrt{\frac{QME}{\hat{\beta_1}^2}}=\frac{\sigma_{{resíduos}}}{\hat{\beta_1}} =\frac{0,023}{0,0606} =0,38$$

Reg_Migracao_0

Reg_Migracao1_0

Reg_Migracao2_0

Uncertainty due to the Chromatograph

$ \mathbf{\checkmark} $ Measurement Equation

The expression (4.5.2.1) represents the measurement equation used for the chromatograph, in this case we have:

$$Crom~=~S+Reg~~~(4.5.2.1)$$

in which,

  • $ Crom $: represents the Chromatograph, $ mg/kg $;
  • $ S $: represents the concentration of net solution i, where $ i~=~1,\dots,6 $, $ mg/kg $. We have to $ S $ has a normal distribution with uncertainty $ u(S) $ and the coverage factor calculated in the Uncertainty module due to solutions;
  • $ Reg $: represents the uncertainty due to regression, $ mg/kg $. We have to $ Reg $ has a normal distribution with uncertainty $ u(Reg) $ and the coverage factor calculated in the previous topic.

$ \mathbf{\checkmark} $ Combined Uncertainty ($ u_{c}(Crom) $)

The combined uncertainty for the chromatograph is given by:

$$u_c^2(Crom)=\sqrt{u^2(S)+u^2(Reg)}$$

in which, $ u^2(.) $ is the contribution of each source of uncertainty.

$ \mathbf{\checkmark} $Expanded Uncertainty ($ U(Crom) $)

Here, the expression (4.5.2.2) represents the expanded uncertainty for the chromatograph.

$$U(Crom) = k\times u_c(Crom)~~(4.5.2.2)$$

in which $ k $ (coverage factor) is the quantile of the t-Student distribution with $ \nu_{eff}(Crom) $ degrees of freedom and 95% confidence. As in this calculation we are only working with type B uncertainties, the effective degrees of freedom are equivalent to $ \nu_{eff}(Crom)=\infty $

To carry out the calculations, the necessary information is:

Sources Value Uncertainty (U) Coverage Factor (k)
Neta Solution 1 25,626 0,010508869 1,96
Dev.Pad.Waste 0,022947469
Coef. of the Conc. (Beta1) 0,060596425
Simulator Type 1

Applying the values in the table we obtain the following results:

Screenshot_70

Final result of uncertainty due to chromatograph

We will calculate the Relative Uncertainty ($ U_{relativa} $) for concentration 2 the other points the calculation is analogous.

$$Conc_{2}=25,626$$

$$U_{relativa}(S_{2})=\frac{U(S_{2})}{Conc_{2}}= \frac{0,742300523}{25,626}=0,028966695$$

18.4.5.3 Test

$ \mathbf{\checkmark} $ Measurement Equation

The expression (4.5.3.1) represents the measurement equation used to obtain the solution related to the test, in this case we have:

$$Sol_i = \frac{A_{P}}{A_{PI}~\beta_1}-\frac{\beta_0}{\beta_1}+\varepsilon\;~~~i=1,\dots,6~~(4.5.3.1)$$

in which,

  • $ Sol_i $: represents the solution i relative to the test, $ i=1,\dots,6 $, in $ mg/kg $;
  • $ A_{P} $: represents the area measurement of the primary standard in the Chromatograph (Crom). We have to $ A_{P} $ assumes normal distribution with uncertainty $ u(A_{P})=u(Crom_P) $ and coverage factor given in the Uncertainty module due to the Chromatograph;
  • $ A_{PI} $: represents the area measurement of the secondary standard in the Chromatograph (Crom). We have to $ A_{PI} $ assumes normal distribution with uncertainty $ u(A_{PI})=u(Crom_{PI}) $ and coverage factor given in the Uncertainty module due to the Chromatograph;
  • $ \varepsilon $: represents repetitiveness. We can assume that the distribution of the mean of measurements has a normal distribution. Therefore, we can estimate its uncertainty (standard deviation from the mean) as:

$$u(\varepsilon) = \frac{s}{\sqrt{n}},$$

Screenshot_73

$ \mathbf{\checkmark} $ Combined Uncertainty ($ u_{c}(Sol_i) $)

The combined uncertainty for the solution quantity relative to the test is given by:

Screenshot_74

in which,

$$\left(\frac{\partial Sol_i}{\partial A_{P}}\right)=\frac{1}{A_{PI}~\beta_1}$$

$$\left(\frac{\partial Sol_i}{\partial A_{PI}}\right)=\frac{A_{P}}{A^2_{PI}~\beta_1}$$

in which, $ u_c^2(.) $ is the contribution of each source of uncertainty.

$ \mathbf{\checkmark} $ Expanded Uncertainty ($ U(Sol_i) $)

Here, the expression (4.5.3.2) represents the expanded uncertainty for the test.

$$U(Sol_i)~=~k\times u_c(Sol_i)~~(4.5.3.2)$$

in which $ k $ (coverage factor) is the quantile of the t-Student distribution with $ \nu_{eff}(Sol_i) $ degrees of freedom and 95% confidence.

$$\nu_{eff}(Sol_i)=\frac{u^4(Sol_i)}{\frac{u^4(A_{P})}{\infty}+\frac{u^4(A_{PI})}{\infty}+\frac{u^4(\varepsilon)}{n-1}}$$

To carry out test calculations, the necessary information is:

Sources Value Uncertainty (U) Coverage Factor (k)
Primary pattern area 47,5558 0,90639002 1,96
Secondary pattern area 345,724 6,589328966 1,96
Repetitiveness 0,002727375 1
Linear coefficient (Intercept) -0,01336
Coef. Angular 0,060596

Applying the values in the table we obtain the following results:

Screenshot_72

Calculation of relative uncertainty

Be, $ \overline{Sol_i} $ is the average of the solutions relative to the test, for $ i=1,\dots, 6 $. For Concentration 2 we have:

$$U_{relativo}=\frac{U(Sol_2)}{\overline{Sol_2}}=\frac{0,061420392}{2,49518586}=0,024615558$$

Thus, the calculation of uncertainty in percentage:

$$U((porcentagem))=U_{relativo}\times100(porcentagem)=2,46(porcentagem)$$

18.4.6 Calculation of Uncertainty for Impedance

In this section, we will apply the uncertainty calculation to the impedance test, which can be defined as the total resistive load of an alternating current (AC) circuit. Impedance is composed of two effects:

  1. When a certain electronic component creates resistance and expends energy in the form of heat (Joule Effect), we call this effect resistance;
  2. When a given electronic component does not expend energy in the form of heat, we have the reactance.

Therefore, when resistance and reactance are present, we call it impedance.

18.4.6.1 Calculation of uncertainty for electric current

Consider a current test using a multimeter. 5 readings are taken in the unit of milli amperes (mA). We have that the multimeter’s instability is ± 0.2%.

The expression (4.6.1.1) represents the measurement equation used to obtain the current

$$I = M+ Res+Inst+\Delta\quad\text{(4.6.1.1)}$$

in which

  • I: Represents electric current;
  • M: Represents the voltage on the laboratory’s standard multimeter;
  • Res: Multimeter resolution;
  • Inst: Multimeter instability;
  • Δ: Represents the equipment reading.

Combined Uncertainty ($ u_c(I) $)

The combined uncertainty for the current is given by

$$u_c(I) = \sqrt{u^2(herd_{M}) + u^2(Res) + u^2(Inst) + u^2(\Delta)}.$$

Expanded Uncertainty ($ U(I) $)

The expression (4.6.1.2) represents the expanded uncertainty for the current.

$$U(I)=k \times u_c(I)\quad \text{(4.6.1.2)}$$

in which

k (coverage factor) is the quantile of the t-Student distribution with $ \nu_{eff}(I) $ degrees of freedom and 95% confidence. The degrees of freedom are given by

$$\nu_{eff}(I) = \frac{u^4_c(I)}{u^4(\Delta)}(n-1).$$

Table (4.6.2.3) presents the readings for calculating repeatability.

n Readings (mA)
1 500,41
2 500,32
3 501,75
4 498,06
5 496,71
Average 499,45
Standard deviation 2,025
Repeatability 0,9058

Table 4.6.2.1: Data for the current channel test.

The data required for this application are:

The uncertainty of the multimeter U($herd_M$)=1% of the reading with k=2;

The resolution of the multimeter res(M) = 0.1% of the reading;

The instability of the multimeter inst(M) = 0.4% of the reading;

The repeatability (%) is given by: $ u(\Delta)=\dfrac{s}{\sqrt{n}}=\dfrac{2,025}{\sqrt{5}} = 0,9058. $

The combined uncertainty for the current is given by

$$u_c(I)=\sqrt{u^2(herd_{M})+u^2(Res)+u^2(Inst)+u^2(\Delta)}=\sqrt{2,49^2+0,0144^2+0,58^2+0,9058^2}=$$

$$=2,722$$

Therefore, the expanded uncertainty is given by:

$$U(I)=k\times u_c(I)=2\times 2,722=5,5$$

where the effective degrees of freedom is calculated as follows:

$$\nu_{eff}(I)=\left(\frac{u_c(I)}{u(\Delta)}\right)^4(n-1)=\left(\frac{2,722}{0,9058}\right)^4\times 4=326,27$$

Screenshot_75

18.4.6.2 Calculation of uncertainty for electrical voltage

Consider a current test using a multimeter. 5 readings are taken in the unit of millivolt (mV). We have that the multimeter’s instability is ± 0.2%.

The expression (4.6.2.1) represents the measurement equation used to obtain the current

$$T = M+ Res+Inst+\Delta\quad\text{(4.6.2.1)}$$

in which

  • T: Represents electrical voltage;
  • M: Represents the voltage on the laboratory’s standard multimeter.
  • Ans: Multimeter resolution.
  • Inst: Multimeter instability.
  • Δ: Represents the equipment reading.

Combined Uncertainty ($ u_c(T) $)

The combined uncertainty for the current is given by

$$u_c(T) = \sqrt{u^2(herd_{M}) + u^2(Res) + u^2(Inst) + u^2(\Delta)}.$$

Expanded Uncertainty ($ U(T) $)

The expression (4.6.2.2) represents the expanded uncertainty for the electrical voltage.

$$U(T)=k \times u_c(T)\quad \text{(4.6.2.2)}$$

in which

k (coverage factor) is the quantile of the t-Student distribution with $ \nu_{eff}(T) $ degrees of freedom and 95% confidence. The degrees of freedom are given by

$$\nu_{eff}(T) = \frac{u^4_c(T)}{u^4(\Delta)}(n-1).$$

Table (4.6.2.3) presents the readings for calculating repeatability.

n Readings (mV)
1 97,761
2 102,488
3 100,87
4 98,941
5 100,51
Average 100,114
Standard deviation 1,822
Repeatability 0,815

The data required for this application are:

The uncertainty of the multimeter U($herd_M$)=1% of the reading with k=2;

The resolution of the multimeter res(M) = 0.1% of the reading;

The instability of the multimeter inst(M) = 0.4% of the reading;

Repeatability is given by: $ u(\Delta)=\dfrac{s}{\sqrt{n}}=\dfrac{1,822}{\sqrt{5}}= 0,815. $

The combined uncertainty for the current is given by

$$u_c(T)=\sqrt{u^2(herd_{M})+u^2(Res)+u^2(Inst)+ u^2(\Delta)}=\sqrt{0,501^2+0,0289^2+0,01155^2+0,815^2}=$$

$$=0,964$$

Therefore, the expanded uncertainty is given by:

$$U(I)=k\times u_c(I)=2,4\times 0,964=2,31$$

where the effective degrees of freedom is calculated as follows:

$$\nu_{eff}(T)=\left(\frac{u_c(T)}{u(\Delta)}\right)^4(n-1)=\left(\frac{0,964}{0,815}\right)^4\times 4=7,82$$

Screenshot_76

18.4.6.3 Calculation of Uncertainty in the Impedance Test

Consider an impedance test using the data in sections 18.4.6.1 (electrical current) and 4.6.2 (electrical voltage).

Measurement Equation

The expression (4.6.3.1) represents the measurement equation used to obtain the impedance

$$R = \frac{V}{I}\quad (4.6.3.1)$$

in which

  • A: Represents impedance;
  • V: Represents voltage measurement;
  • I: Represents current measurement.

Combined Uncertainty (uc(R))

The combined uncertainty for the impedance is given by

$$u_c(R)=\sqrt{\left(\frac{\partial R}{\partial V}\right)^2 u^2(V)+\left(\frac{\partial R}{\partial I}\right)^2 u^2(I)}$$

$$=\sqrt{\left(\frac{1}{I}\right)^2 u^2(V) +\left(-\frac{V}{I^2}\right)^2 u^2(I)}.$$

Expanded Uncertainty (U(R))

The expression (4.6.3.2) represents the expanded uncertainty for the impedance.

$$U(R) = k \times u_c(R)\quad (4.6.3.2)$$

where k (coverage factor) is the quantile of the t-Student distribution with $ \nu_{eff}(R) $ degrees of freedom and 95% confidence.

Table (4.6.3.3) presents the summary for calculating impedance uncertainty.

n Readings (mV) Readings (mA) Impedance
1 97,761 500,41 0,195
2 102,488 500,32 0,205
3 100,87 501,75 0,201
4 98,941 498,06 0,199
5 100,51 496,71 0,202
Average 100,114 499,45 0,200
Standard deviation 1,822 2,0254 0,00362
Repeatability 0,815 0,9058 0,00162

The data required for this application are:

The uncertainty inherited from tension $ U(V) = 2,341 $ Go with $ k=2,429 $;

The uncertainty inherited from the current $ U(I) = 0,0055 $ The with $ k=2 $;

The combined uncertainty for the impedance is given by

$$u_c(R)=\sqrt{\left(\frac{1}{I}\right)^2 u^2(V) +\left(-\frac{V}{I^2}\right)^2 u^2(I)}=$$

$$=\sqrt{\left(\frac{1}{0,499}\right)^2 0,964^2+\left(-\frac{100,1}{0,499^2}\right)^2 0,0055^2}=$$

$$2,218~$$

Considering the effective degrees of freedom and a confidence of approximately 95%, we have from the t-Student distribution table that the coverage factor is k=2. Therefore, the expanded uncertainty for the impedance variable is given by

$$U=k\times u_c(R)=2\times 2,218=4,44~\Omega$$

Screenshot_77

18.4.7 Specific Surface Area Measurement Test

Description of the Test

In this section we describe the Specific Surface Area tests. Below, we present a flowchart of the uncertainty calculation.

(image missing)

18.4.7.1 Mass (Scale)

Mass (Scale)

This section is dedicated to the mass uncertainty calculation, which only takes into account balance and balance resolution uncertainties. Expression (4.7.1.1) presents the mass measurement equation (balance)

$$M_{B} = B + \text{Res}_B \quad \text{(4.7.1.1)}$$

in which

  • $M_B$: represents the mass (Scale), in $ (g); $
  • $B$: represents the scale in $ (g). $ We have to $ B $ has a normal distribution with uncertainty $ u(B) $ and coverage factor given by the scale calibration certificate;
  • $Res_B$: represents the resolution of the balance in $ (g). $ We have to $ u(Res_B) $ has a rectangular (or uniform) distribution over the interval $ \left(-\frac{{Res_B}}{2},\frac{{Res_B}}{2}\right). $ Therefore, the uncertainty due to the resolution will be given by:

$$u(Res_{B})=\frac{{Res}_{B}}{2\sqrt{3}}$$

Combined uncertainty ($ u_c(M_B) $):

The combined mass uncertainty is given by the expression (4.7.1.2)

$$u_c(M_B)=\sqrt{u^2(B)+u^2(\text{Res}_B)} \quad \text{(4.7.1.2)}.$$

Expanded Uncertainty ($ U(M_B) $):

The expanded mass uncertainty is given by the expression (4.7.1.3)

$$U(M_B) =u_c(M_B)\times k \quad \text{(4.7.1.3)}$$

in which $ k $ is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and confidence level of approximately 95.45%. As we do not have type A uncertainty, we adopt $ k=2. $

Application:

Next, we will calculate the uncertainty of the mass (balance), for this we need the data from the table presented below:

Sources Uncertainty (U) Coverage Factor (k)
Scales 0,0002 2
Balance Resolution 0,0001 2

The combined mass uncertainty is given by

$$u_c(M_B)=\sqrt{u^2(B)+u^2(\text{Res}_B)}=\sqrt{0,0001^2+(2,88675\times 10^{-5})^2}=0,000104083.$$

The expanded mass uncertainty is given by

$$U(M_B) =u_c(M_B)\times k=0,000104083\times 2=0,000208167$$

in which $ k $ is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and confidence level of approximately 95.45%. As we do not have type A uncertainty, we adopt $ k=2. $

Below, we present a summary of the results.

Screenshot_78

18.4.7.2 Mass of the degassed sample

Degassed sample mass

This section is dedicated to calculating the mass uncertainty of the degassed sample. The dry sample is obtained when we pass the sample through Vac Prep for 40 minutes at 200ºC, after this time we wait for the sample to cool. Below we present the measurement equation.

Measurement Equation:

The expression (4.7.2.1) represents the measurement equation used for the mass of the degassed sample, in this case we have:

$$M_{A} = M_D- M_T+ \varepsilon \quad \text{(4.7.2.1)} $$

in which

  • $M_A$: represents the mass of the sample, in $ (g); $
  • $M_T, M_D$ represent the masses of the tube and tube plus the degassed sample, in $ (g). $ We have to $ M_T $ and $ M_D $ have a normal distribution with uncertainty $ u(M_B) $ and coverage factor k obtained in section (4.7.1);
  • $ \varepsilon $: represents repetitiveness. We can assume that the distribution of the mean of measurements has a normal distribution. Therefore, we can estimate its uncertainty (standard deviation from the mean) as:

$$u(\epsilon) = \frac{s}{\sqrt{n}},$$

in which

$$s^2 = \frac{1}{n-1} \sum^{n}_{i=1} (x_i -\overline{X})^2$$

is the sampling variance and

$$\overline{X} = \frac{1}{n} \sum^{n_{i=1}} x_i$$

is the sample mean.

Combined Uncertainty:

The combined uncertainty of the sample mass is given by the expression (4.7.2.2)

$$u_c(M_A)=\sqrt{u^2(M_D)+u^2(M_T)+u^2(\varepsilon)} \quad \text{(4.7.2.2)}$$

Expanded Uncertainty:

The expanded uncertainty of the sample mass is presented in expression (4.7.2.3)

$$U(M_A) =u_c(M_A)\times k \quad \text{(4.7.2.3)}$$

in which $ k $ is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and confidence level of approximately 95%.

$$\nu_{eff}(M_A)=\frac{u^4(M_A)}{u^4(\varepsilon)}(n-1) $$

Application:

Next, we will calculate the uncertainty of the mass of the degassed sample, for this we need the data in the following table:

Sources Value Uncertainty (U) Coverage Factor (k)
Mass (Tara) 16,42 0,000208167 2,00
Degassed dough 17,10 0,000208167 2,00
Repetitiveness 0 1

The combined uncertainty of the sample mass is given by

$$u_c(M_A)=\sqrt{u^2(M_D)+u^2(M_T)+u^2(\varepsilon)}=\sqrt{0,000104083^2+0,000104083^2+0^2}= 0,000147196.$$

The expanded uncertainty of the sample mass is given by

$$U(M_A) =u_c(M_A)\times k=0,000147196\times 2=0,000294392$$

in which $ k $ is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and confidence level of approximately 95%. Since the repeatability is zero, we have to $ \nu_{eff}=\infty, $ consequently we have $ k=2. $ To calculate the mass of the degassed sample, simply apply equation (4.7.2.1)

$$M_{A} = M_D- M_T=17,1006-16,43=0,6696$$

Below, we present a summary of the results.

Screenshot_79

18.4.7.3 Uncertainty due to regression

Uncertainty due to regression

Let us consider the following linear relationship between $ 1/[Q(P_0/P-1)] $ and relative pressure:

$$Y_{ij} = \beta_0 + \beta_1~x_{i} + \varepsilon_{ij}~~~~ i =1,\cdots, n;$$

in which,

  • $ Y_{ij} $: represents the ith measurement of $ 1/[Q(P_0/P-1)] $ referring to the i-th relative pressure;
  • $ X_{i} $: represents the relative pressure referring to the i-th $ 1/[Q(P_0/P-1)] $;
  • $ \beta_0 $: represents the linear coefficient;
  • $ \beta_1 $: represents the angular coefficient;
  • $ \varepsilon_{ij} $: represents the j-th error made when measuring the ith $ 1/[Q(P_0/P-1)] $. We consider that the $ \varepsilon_{ij} $ are independent and identically distributed with normal distribution with zero mean and variance $ \sigma^2, $ i.e. $ \varepsilon_{ij}\overset{i.i.d.}{\sim}N(0,\sigma^2) $ .

Using the least squares method, we obtain that

$$\widehat{\beta_1}=\frac{S_{xy}}{S_{xx}}\quad \text{e} \quad \widehat{\beta}_0=\overline{y}-\widehat{\beta_1}\overline{x}$$

in which,

  • $ S_{xx} = \displaystyle\sum_{i=1}^{n} (x_i-\overline{x})^2 = \sum_{i=1}^{n}x_i^2- 2\overline{x}\sum_{i=1}^{n} x_i+n\overline{x}^2 = \sum_{i=1}^{n} x_i^2-n\overline{x}^2; $
  • $ S_{xy} = \displaystyle\sum_{i=1}^{n}(x_i-\overline{x})(y_i-\overline{y}) =\sum_{i=1}^{n}(x_i-\overline{x})y_i; $
  • $ \overline{y} = \frac{1}{n}\displaystyle \sum_{i=1}^{n}(y_i) $ represents the average of area readings;
  • $ \overline{x} = \frac{1}{n}\displaystyle \sum_{i=1}^{n}(x_i) $ represents the average of concentration readings.

Now let us deduce the uncertainty due to the volume of gas adsorbed when the surface of the solid is completely covered by a monolayer ($ Q_m $). To do this, from the equation of the straight line we obtain:

$$Y= \beta_0+\beta_1 X_i$$

$$\dfrac{1}{Q(P_0/P-1)} = \beta_0+\beta_1 \dfrac{P}{P_0} \quad \text{(4.7.3.1)}$$

$$\dfrac{1}{Q(P_0/P-1)} = \dfrac{1}{Q_m\times C}+\dfrac{C-1}{Q_m\times C} \dfrac{P}{P_0} \quad \text {(4.7.3.2)}$$

Relating (4.7.3.1) and (4.7.3.2) we obtain that

Screenshot_80

in addition,

$$\quad Q_m=\dfrac{1}{\beta_0+\beta_1}$$

Now, let’s use the delta method to calculate the variability of $ Q_m $ (for more details see Uncertainty due to calibration curve: Delta Method). We have to:

$$\text{Var}(Q_m) \approx \left(\dfrac{\partial Q_m}{\partial \beta_0}\right)^2 \text{Var}(\widehat{\beta}_0)+\left(\dfrac{\partial Q_m}{\partial \beta_1}\right)^2 \text{Var}(\widehat{\beta}_1) + 2\left( \dfrac{\partial Q_m}{\partial \beta_0} \dfrac{\partial Q_m}{\partial \beta_1}\right)\text{Cov}(\widehat{\beta}_0, \widehat{\beta}_1)=$$

$$ \approx \left(-\dfrac{1}{(\widehat{\beta}_0+\widehat{\beta}_1)^2}\right)^2 \text{Var}(\widehat{\beta}_0)+\left(-\dfrac{1}{(\widehat{\beta}_0+\widehat{\beta}_1)^2}\right)^2 \text{Var}(\widehat{\beta}_1)+2\left(-\dfrac{1}{(\widehat{\beta}_0+\widehat{\beta}_1)^2}\right)^2\text{Cov}(\widehat{\beta}_0, \widehat{\beta}_1)=$$

$$ \approx \left(-\dfrac{1}{(\widehat{\beta}_0+\widehat{\beta}_1)^2}\right)^2 [\text{Var}(\widehat{\beta}_0)+\text{Var}(\widehat{\beta}_1)+2\text{Cov}(\widehat{\beta}_0, \widehat{\beta}_1)]$$

Therefore, the uncertainty due to the volume of gas adsorbed when the surface of the solid is completely covered by a monolayer ($ Q_m $) is given by:

$$u(Q_m)=\sqrt{\left(-\dfrac{1}{(\widehat{\beta}_0+\widehat{\beta}_1)^2}\right)^2 [u^2(\widehat{\beta}_0)+u^2(\widehat{\beta}_1) +2\text{Cov}(\widehat{\beta}_0, \widehat{\beta}_1)} ]\quad \text{e} \quad Q_m=\dfrac{1}{(\widehat{\beta}_0+\widehat{\beta}_1)}$$

Application

Below is the data used in the linear regression. It is worth remembering that the response variable $ Y=1/Q(P_0/P-1) $ and explanatory variable is $ X=P/P_0. $ With this, we obtain the data for the calibration curve.

$ \bf X $ $ \bf Y $
0,050500102 0,008679
0,087812073 0,014308
0,125881379 0,019918
0,162958319 0,025366
0,201177794 0,031032

Below, we present the results obtained by the Action software.

(image missing)

(image missing)

Furthermore, using the Action software, we obtain that $ \text{Cov}(\widehat{\beta}_0, \widehat{\beta}_1)=-1,89784e^{-08} $, in this case, we will consider the covariance between the parameters to be null.

Sources Value Uncertainty (U) Coverage Factor (k)
Intercept 0,001248178 5.30388E-05 1
Angular Coefficient 0,148110327 0,000388617 1

Therefore, the uncertainty due to the volume of gas adsorbed when the surface of the solid is completely covered by a monolayer ($ Q_m $) is given by:

$$u(Q_m)=\sqrt{ \left(-\dfrac{1}{(\widehat{\beta}_0+\widehat{\beta}_1)^2}\right)^2 [u^2(\widehat{\beta}_0)+u^2(\widehat{\beta}_1)]}=$$

$$= \sqrt{\left(-\dfrac{1}{(0,001248178 + 0,148110327)^2}\right)^2 [(5,3\times 10^{-5})^2+(0,000388617)^2]}=$$

$$=0,017582036$$

and

$$Q_m=\dfrac{1}{(\widehat{\beta}_0+\widehat{\beta}_1)}=\dfrac{1}{(0,001248178 + 0,148110327)}=6,6953$$

Below, we present a summary of the results.

Screenshot_81

18.4.7.4 Specific surface area

Specific Surface Area

This section is dedicated to calculating uncertainty for Specified Surface Area. Below we present the measurement equation.

Measurement Equation

The expression (4.7.4.1) represents the measurement equation used for Specific Surface Area, in this case we have:

$$S_{\text{SE}}= \dfrac{Q_m\times L_{av}\times A_{ad}}{M_v} \quad \text{(4.7.4.1)}$$

in which

  • $ S_{SE} $ represents the specific surface area, in $ (m^2/g); $
  • $ Q_m $ represents the volume of gas adsorbed when the surface of the solid is completely covered by a monolayer, in $ (cm^3/g). $ We have to $ Q_m $ has a normal distribution with uncertainty $ u(Q_m) $ obtained in section (4.7.3);
  • $ L_{av} $ represents Avogadro’s number, whose value of this constant is 6.022$ \times 10^{23}. $ We have to $ L_{av} $ is a constant and consequently has no associated uncertainty;
  • $ A_{ad} $ represents the cross-sectional area of an adsorbate molecule, whose value of this constant is 0.162$ \times 10^{-18}m^2. $ We have to $ A_{ad} $ is a constant and consequently has no associated uncertainty;
  • $ M_v $ represents the volume occupied by 1 mole of the adsorbate molecule, whose value of this constant is 2.2414$ \times 10^{4}cm^3. $ We have to $ M_{v} $ is a constant and consequently has no associated uncertainty;

Combined Uncertainty

The combined uncertainty of the Specific Surface Area Determination test is given by the expression (4.7.4.2)

$$u_c(S_{SE})=\sqrt{\left(\dfrac{\partial S_{SE}}{\partial Q_m}\right)^2 u^2(Q_m)} \quad \text{(4.7.4.2)}$$

in which

$$\left(\dfrac{\partial S_{SE}}{\partial Q_m}\right)=\dfrac{L_{av}\times A_{ad}}{M_v}$$

Expanded Uncertainty

The expanded uncertainty is given by the expression (4.7.4.3)

$$U(S_{SE}) =u_c(S_{SE})\times k \quad \text{(4.7.4.3)}$$

in which $ k $ is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and confidence level of 95 (percentage). As we do not have type A uncertainty, we adopt $ k=2. $

Application

Next, we will calculate the uncertainty for Specific Surface Area. It is worth remembering that from the section (\ref{Regre}) we obtain the uncertainty of the volume of the adsorbed gas. In order to calculate the uncertainty due to specific surface area, we need the constants given in the table.

Sources Value Uncertainty (U) Coverage Factor (k)
Volume of adsorbed gas 6,695300041 0,035164071 2,00
Avogadro’s Number 6,022$ \times 10^{23} $
Cross-sectional area 1,62$ \times 10^{-19} $
1 mol volume 22414

Combined Uncertainty

The combined uncertainty for the Specific Surface Area is given by

$$u_c(S_{SE})=\sqrt{\left(\dfrac{\partial S_{SE}}{\partial Q_m}\right)^2 u^2(Q_m)}=$$

$$=\sqrt{\left(4,352476131\right)^2 0,017582036^2}=$$

$$= 0,07652539$$

in which

$$\left(\dfrac{\partial S_{SE}}{\partial Q_m}\right)=\dfrac{L_{av}\times A_{ad}}{M_v}=\dfrac{6,022\times 10^{23}\times 1,62\times 10^{-19}}{22414} =4,352476131$$

Expanded Uncertainty

The expanded uncertainty is given by

$$U(S_{SE}) =u_c(S_{SE})\times k=0,07652539 \times 2=0,15~m^2/g$$

in which $ k $ is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and confidence level of 95 (percentage). As we do not have type A uncertainty, we adopt $ k=2. $

The specific surface area is given by:

$$S_{\text{SE}}= \dfrac{Q_m\times L_{av}\times A_{ad}}{M_v} =\dfrac{6,695300041 \times 6,022\times 10^{23}\times 1,62\times 10^{-19}}{22414} =29,14$$

Below, we present a summary of the results.

Screenshot_82

18.4.7.5 Repeatability and Reproducibility Test (R&R)

Repeatability and Reproducibility Testing (R&R):

In this section, we describe the study of Repeatability and Reproducibility for the specific surface area determination test. In this study, we selected 2 operators, 3 parts and measured each part 6 times. The following table corresponds to the study planning.

Ask Analyst Result
2 1 53,96
3 2 125,95
3 1 127,19
2 1 53,19
2 2 52,93
1 2 29,66
3 1 126,19
3 1 126,78
1 2 29,65
1 1 29,64
2 1 53,75
3 2 126,24
3 2 126
2 2 52,69
1 1 30,09
1 1 29,83
2 2 52,84
1 2 29,35

Next, we present the results for the specific surface area determination test.

ANOVA
G.L. Sum of squares Medium Squares F-statistic Pr(>F)
Parts 2 30578,17781 15289,08891 224015,9547 2.9083E-32
Operators 1 1,94045 1,94045 28,43150183 0,000105824
Repetitiveness 14 0,9555 0,06825
TOTAL VARIATION AND/OR TOLERANCE
Standard deviation ($ m^2/g $) Tolerance (%)
Repetitiveness 0,261 9,80
Reproducibility 0,456 17,10
Operators 0,456 17,10
Parts 50,479 1892,98
Repeatability and reproducibility 0,526 19,71
Total 50,482 1893,08

From the results obtained, we have that the estimate of uncertainty due to repeatability and reproducibility is $ u(RR)=0,526~m^2/g. $ For more details see MSA 4th edition.

(image missing)

18.4.7.6 Specific Surface Area Determination Test

Specific Surface Area Determination Test

This section is dedicated to the uncertainty calculation for the Specific Surface Area Determination test. Below we present the measurement equation.

Measurement Equation

The expression (4.7.6.1) represents the measurement equation used for the Specific Surface Area Determination test, in this case we have:

$$S_{\text{BET}}=S_{\text{SE}}+\text{Pad}+RR \quad \text{(4.7.6.1)}$$

in which

  • $ S_{BET} $ represents the determination of the specific surface area (BET), in $ (m^2/g); $
  • $ S_{SE} $ represents the specific surface area, in $ (m^2/g). $ We have to $ S_{\text{SE}} $ has a normal distribution with uncertainty $ u(S_{\text{SE}}) $ obtained in section (4.7.4);
  • Pad: represents the standard certificate, in $ (m^2/g). $ We have that Pad has a normal distribution with uncertainty $ u(\text{Pad}) $ obtained via standard certificate;
  • $ RR $: represents the repeatability and reproducibility in (N): We have that RR has a normal distribution with uncertainty $ u(RR) $ obtained in the RR test (section 18.4.7.5);

Combined Uncertainty

The combined uncertainty of the Specific Surface Area Determination test is given by the expression (4.7.6.2)

$$u_c(S_{\text{BET}})=\sqrt{u^2(S_{\text{SE}})+u^2(\text{Pad})+u^2(RR)} \quad \text{(4.7.6.2)}$$

Expanded Uncertainty

The expanded uncertainty is given by the expression (4.7.6.3)

$$U(S_{\text{BET}}) =u_c(S_{\text{BET}})\times k \quad \text{(4.7.6.3)}$$

in which $ k $ is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and 95% confidence level. As we do not have type A uncertainty, we adopt $ k=2. $

Application

Next, we will calculate the uncertainty for the Specific Surface Area Test. Below we present the data to calculate the uncertainty.

Sources Value Uncertainty (U) Coverage Factor (k)
Specific Surface Area 29,14 0,07652539 1
Standard certificate 0,75 2
Repeatability and Reproducibility 0,525616041 1

The combined uncertainty of the Specific Surface Area Determination test is given by

$$u_c(S_{\text{BET}})=\sqrt{u^2(S_{\text{SE}})+u^2(\text{Pad})+u^2(RR)}=$$

$$=\sqrt{0,07652539^2+0,375^2+0,525616041^2}=$$

$$= 0,650194862$$

The expanded uncertainty is given by

$$U(S_{\text{BET}}) =u_c(S_{\text{BET}})\times k= 0,650194862\times 2=1,3 $$

in which $ k $ is the coverage factor and this is obtained from the t-Student distribution with $ \nu_{eff} $ degrees of freedom and 95% confidence level. As we do not have type A uncertainty, we adopt $ k=2. $

Below, we present a summary of the results of the specific surface area determination test.

Screenshot_83

18.5 Characteristic Parameters of a Measurement System

To characterize the metrological behavior of a measuring system, some metrological parameters are used. These parameters can be expressed in the form of a simple number, a range of values ​​or in the form of a graph. Here, the main parameters are described.

Nominal range

Nominal Range or Nominal Interval is the set of values comprised between two rounded or approximate extreme indications, obtained with a particular positioning of the controls of a measuring instrument or measuring system, used to designate this positioning. A nominal range of indications is generally expressed in terms of its smallest and largest value.

Examples:

  • Thermometer: 100ºC to 200ºC;
  • Pressure gauge: 0 bar to 20 bar;
  • Counter: 5 digits (i.e. 99999 pulses).

When the lower limit is zero, the nominal range can be defined solely in terms of the upper limit.

Example:

The nominal range of 0V to 100V is expressed as “100V”.

Nominal range width

The amplitude of the nominal range is the difference, in modules, between the two limits of a nominal range from -10V to +10V. The amplitude of the nominal range is 20V. In some areas, the difference between the highest and lowest value is called range.

Measuring range

The measuring range, also called working range, is the set of values of a measurand for which we assume that the error of a measuring instrument is maintained within specified limits.

Example:

In a feeler clock the measuring range is $ -~50\mu m $ to $ +50\mu m $.

The measuring range is less than or at most equal to the nominal range. The measuring range value can be obtained through:

  • From the measurement system operating manual (SM);
  • Of signs engraved on the scale;
  • Technical standard specifications;
  • From calibration reports.

Scale division

It is the part of a scale between any two successive marks.

Length of a division

It is the distance between any two successive marks, measured along the scale length line. The length of a division is expressed in units of length, whatever the unit measured or the unit marked on the scale.

Value of a division

The value of a division is the difference between the scale values corresponding to two successive marks.

Stipulated operating conditions

Operating condition that must be met during a measurement for a measuring instrument or measuring system to function as designed. They specify the value ranges for the measured quantity and the influencing quantities.

Operating limit conditions

Extreme operating condition that a measuring instrument or measuring system must withstand without damage and without degradation of its specified metrological properties when subsequently operated under its stipulated operating conditions. This condition may differ for storage, transport and operation. They can also comprise limit values ​​for the measured quantity and for the influencing quantities.

Reference operating conditions

Reference conditions are the usual conditions prescribed for testing the performance of a measuring instrument or a measuring system or for intercomparing measurement results. These conditions generally include reference values ​​or reference ranges for the influencing quantities that affect the measuring instrument. These also specify the ranges of values ​​of the measurand and the influencing quantities. In IEC 60050-300, item 311-06-02, the term “reference condition” refers to an operating condition in which the specified instrumental measurement uncertainty is as small as possible.

Response characteristic

It is the relationship between a stimulus and the corresponding response, under defined conditions.

Example:

The electromotive force (emf) of a thermocouple as a function of temperature. The relationship may be expressed in the form of a mathematical equation, a numerical table or a graph.

Sensitivity

Sensitivity is characterized by the quotient between the variation in an indication of a measuring system and the corresponding variation in the value of the measured quantity. In instruments with a pointer indicator, sensitivity is commonly established as the relationship between the displacement of the tip of the pointer in (mm) and the unit value of the quantity to be measured.

Mobility threshold

It is the largest variation in the stimulus (value of a measured quantity) that does not produce a detectable variation in the response of a measuring instrument, with the variation in the input signal being slow and uniform. The mobility threshold (also called threshold) may depend, for example, on noise (internal and external) or friction. It may also depend on the value of the stimulus.

Resolution

Resolution is the smallest variation of the measured quantity that causes a noticeable variation in the corresponding indication. The resolution may depend, for example, on noise (internal or external) or friction, or even on the value of the measured quantity. The resolution assessment is carried out depending on the type of instrument:

a) For digital display devices, the resolution is the variation in the indication when the least significant digit varies by one unit.

b) In measuring systems with an analogue display device, the resolution is a function of the limitations of the person taking the reading, the quality of the indicator and the need for more or less accurate readings.

Stability

Stability is the ability of a measuring instrument to maintain its metrological characteristics constant over time. Stability can be quantified in several ways. Example: For the time in which the metrological characteristic varies from a certain value, or in terms of variation of a characteristic in a certain period of time.

Discretion

It characterizes the ability of a measuring instrument to not change the measured value.

Example:

A balance is a discreet instrument for measuring mass, as the measuring system does not change the mass value. A resistance thermometer that heats the environment in which the temperature is being measured is not discrete.

Drift

Drift is the variation in the indication over time, whether continuous or incremental, due to variations in the metrological properties of a measuring instrument. Drift is not related to a variation in the measured quantity, nor to a variation in any identified influence quantity.

Response time

It is the interval between the instant in which the stimulus (input value of a measuring instrument or a measuring system) is subjected to a sudden variation between two specified constant values and the instant in which the corresponding indication remains within specified limits around its final value in a steady state.

Accuracy of a measuring instrument

Accuracy, also called accuracy, is the ability of a measuring instrument to give responses close to a true value, or it can be defined as the degree of agreement between a measured value and a true value of a measurand. Accuracy is a qualitative concept and should not be confused with precision.

Accuracy class

Class of measuring instruments or measuring systems that satisfy established metrological requirements, intended to maintain measurement errors or instrumental measurement uncertainties within specified limits under specified operating conditions. This concept is applied to materialized measures. An accuracy class is usually indicated by a number or symbol adopted by convention and called a class index.

Example:

Class 0 standard block game.

Indication error on a measuring instrument

This error is determined by the difference between the indication of a measuring instrument and a true value of the corresponding input quantity. In practice, since a true value cannot be determined, a conventional true value is used. This concept of error is mainly applied when the instrument is compared to a reference standard. For a materialized measurement, the error is characterized between the indication and the value assigned to it.

Maximum permissible errors

Maximum permissible error or maximum permissible error is an extreme value of measurement error, with respect to a known reference value, admitted by specifications or regulations for a given measurement, measuring instrument or measuring system. This is also called permissible error limits or maximum tolerated error.

Trend

The bias or fairness error is the systematic error in the indication of a measuring instrument (also called: bias of a measuring instrument, erreur de justesse) defined as the average of repeated indications and a reference value. The bias of a measuring instrument is typically estimated by averaging the indication errors over an appropriate number of repeated measurements.

Trend exemption

Ability of a measuring instrument to give indications free from systematic errors.

Repetitiveness

Repeatability is the ability of a measuring instrument to provide very close indications, in repeated applications of the same measurand, under the same measurement conditions. These conditions include:

  • Reduction to a minimum of variables due to the observer;
  • Same measurement procedure;
  • Same evaluator;
  • Same measuring equipment, being used under the same conditions;
  • Same location;
  • Repetitions in a short period of time.

Repeatability can be expressed quantitatively in terms of the dispersion characteristics of the indications.

Hysteresis

The hysteresis of a measuring instrument is a measurement error, which occurs when there is a difference between the measurement, for a given value of the measured value when it was reached by increasing values, and the measurement when reached by decreasing values ​​of the measured value. This value may be different if the loading and unloading cycle is complete or partial. Hysteresis is a very typical phenomenon in mechanical instruments, with the source of error mainly being gaps and deformations associated with friction.

Bibliographic References

[1] CIPM (1981) INC, BIPM Proc. Verb. Com. Int. Poids et Mesures, vol. 18, Metrologia (1992), 43-44 (English).

[2] ISO GUM (2008), Measurement data evaluation - Guide for expressing measurement uncertainty, 1st Brazilian edition.

[3] NIS 3003, The Expression of Uncertainty and Confidence in Measurement for Calibrations, Edition 8, May 1995.

[4] Brazilian Version of Reference Document EA-4/02, Expression of Uncertainty in Measurement, Volumes I and II.

[5] Montgomery, D. C. (1997), Design and Analysis of Experiments, Fourth Edition - Ed. John Wiley & Sons.

[6] Ripley, B. D. (1987), Stochastic Simulation, Wiley: New York.

[7] Wichmann, B. A. and Hill, I. D. (1982), “Algorithm AS 183: An Efficient and Portable Pseudo-random Number Generator”, Applied Statistics, vol 31, p 188-190.

[8] ABNT NBR ISO 10012. Measurement management systems – requirements for measurement processes and measurement equipment. Rio de Janeiro: ABNT – Brazilian Association of Technical Standards, 2004.

[9] ABNT NBR ISO/IEC 17025. General requirements for testing and calibration laboratory competence. Rio de Janeiro: ABNT – Brazilian Association of Technical Standards, 1999.

[10] D. Deaver. Guardbanding and the world of iso guide 25 is there only one way? Fluke Corporation, NCSL, 1998.

[11] S.A. Khanam and E. Morse. Test uncertainty & test uncertainty ratio (tur).

[12] Grubbs, Frank E (1969), Procedures for detecting outlying observations in samples, Taylor & Francis Group, vol 11, p 1-21.