Proficiency Testing
On this page

Book

16 Proficiency Testing

50 min read

Statistical methods are increasingly being recognized as an important instrument for diagnosing and optimizing the management and operation of various systems, from human systems to the most complex physical systems. The applications of these methods have played an important role in solving many problems from industry to agriculture, including the socioeconomic, administrative and health sectors. Furthermore, much of the most advanced scientific research in various areas of knowledge increasingly depends on statistical methods.

Specifically in industry, we have observed that the use of extremely simple statistical techniques can help maintain product quality within a certain level. Most of the time, the statistical methods used contribute to increasing production or even reducing costs, without any additional investment or industrial expansion measures needing to be taken. Of the various statistical methods we can employ, we will develop a procedure to statistically evaluate proficiency testing programs.

Proficiency testing programs are widely used in interlaboratory comparisons in order to determine the performance of participating laboratories in carrying out specific tests (or calibrations) and to monitor the consistency and comparability of data. Inter-laboratory comparisons are conducted not only to analyze laboratories, but also to evaluate methods and standards, see ISO/IEC 17043:2010 [9].

Proficiency tests (PE) through inter-laboratory comparisons are commonly used by accreditation bodies (INMETRO), with the aim of evaluating and monitoring the measurement capacity of laboratories. They are internationally recognized as an important element in establishing mutual trust between laboratories in different countries and between national laboratory accreditation bodies. An interlaboratory comparison program involving several laboratories requires an adequate organizational structure and a strong commitment from participants to comply with the conditions pre-established in a guidance document.

16.1 Introduction

The Interlaboratory Comparisons were defined according to ISO/IEC 17043:2010 [9] as the organization, performance and evaluation of measurements or tests with the same or similar items by two or more laboratories according to predetermined conditions. Interlaboratory comparisons are conducted not only to analyze laboratories, but also to evaluate methods and standards. Below, we will mention some purposes for interlaboratory comparisons:

  1. Evaluate laboratory performance for specific testing, measurement, monitoring and ongoing performance of laboratories;
  2. Identify problems in laboratories and initiate improvement actions, such as:
    • inadequate testing or measurement procedures;
    • effectiveness of personnel training;
    • supervision or calibration of equipment.
  3. Establishment of the effectiveness and comparability of test or measurement methods;
  4. Providing additional confidence to laboratory customers;
  5. Identification of interlaboratory differences;
  6. Education of laboratories based on the results of such participant comparisons;
  7. Validation of uncertainty requests;
  8. Assessment of the performance characteristics of a method, often described as collaborative studies;
  9. Assignment of reference values ​​and materials, assessment of their suitability for use in specific test or measurement procedures;
  10. Support for demonstrations of measurement equivalence from National Institutes of Metrology through supplementary comparisons performed on behalf of the International Bureau of Weights and Measures (BIPM) and associated regional metrology organizations.

Proficiency testing involves the use of comparisons between laboratories to determine laboratory performance, as mentioned in items 1 to 7. Proficiency testing does not address items 8 to 10, as laboratory competence is assumed in these applications, but these applications can be used to provide independent manifestations of laboratory competence.

The need for confidence in laboratory performance is not only essential for laboratories and their customers, but also for other interested parties such as laboratory accreditation regulatory bodies (INMETRO) and other organizations that specify requirements for laboratories. The ISO/IEC 17011 [12] (2005) standard states that it is necessary for an accreditation body to manage the participation and performance of laboratories in proficiency tests.

There is an increasing need for proficiency testing for other conformity assessment activities, such as product inspection or certification. If relevant, interested parties should document their policies for participation in proficiency testing and such documented policies should be publicly available to laboratories and other interested parties. Issues that should be addressed in participation policies for specific proficiency testing schemes include:

  • Whether participation in specific proficiency testing programs is mandatory or voluntary;
  • The frequency of participation;
  • The criteria used by the interested party to judge satisfactory or unsatisfactory performance;
  • The obligation for laboratories to participate and monitor proficiency tests if performance is considered unsatisfactory;
  • How the results of proficiency tests will be used in performance assessment and subsequent decisions;
  • details of the interested party’s policy on preserving the confidentiality of participants.

Next we will describe the proficiency testing methods.

16.2 Proficiency Testing Methods

Proficiency Testing (PE) methods depend on the nature of the item or material under test, the testing method in use, and the number of participating laboratories. There are six distinct types of EP, discussed in ISO/IEC 17043:2010 [9], among them we mention only four, which are:

  • Sequential scheme: when one or more proficiency test items are distributed sequentially for testing or measurement and returned to the proficiency testing provider at regular intervals;
  • Simultaneous regime: when proficiency test items are distributed for simultaneous testing or measurement within a defined period of time;
  • Continuous regime: When proficiency test items are provided at regular intervals, for example, external quality assessment (EQA) programs.

In this content, we will describe in detail only the sequential scheme. In addition to monitoring the consistency and comparability of data obtained in trials, the EP program can contribute to improving the data collected by laboratories, see Richardson et al.11, for example.

The analysis of data obtained in an EP follows 4 basic steps, which are common for all types of programs, which are:

  1. Determine the Reference Value;
  2. Compare results;
  3. Determine performance statistics;
  4. Preliminary determination of proficiency, i.e. homogeneity and stability test.

Participant results should be compared with the values ​​or responses that most demonstrate competence with the method. Reference values ​​can be chosen to evaluate participants fairly and to encourage agreement between laboratories. There are a variety of common methods for determining the reference value. One of them is the reference value by the consensus value of the participants (average of measurements), or by the measurement result of a laboratory with international experience (via calibration). Still, there are other ways to determine the reference value described in ISO/IEC 17043:2010 [9]. These are two of the most common procedures used in routine PE accreditation programs.

In the case of consensus values, it is important to carry out an analysis of extreme values ​​(for more details see Grubbs Test), as their presence can harm the analysis. An observation considered an extreme value cannot be used, as it may under or overestimate the mean of the observations, which is one of the estimators of the consensus value.

After determining the reference value, the next step is to transform each participant’s measurements into performance statistics. The objective is to measure the difference between laboratory measurements and the reference value, in order to allow comparison with a defined criterion. Currently, the comparison is made directly and ISO/IEC 17043:2010 [9] proposes the normalized error ($E_n$), which is expressed by

$$E_n=\frac{\mid y - \mu_{x} \mid}{\sqrt{U^{2}_{lab} +U^{2}_{ref}}}$$

in which,

  • y is the measurement from the participating laboratory and
  • $μ_x$ is the average of the reference value.

The Uref and Ulab notations correspond to half the amplitude value of the 95% interval for μx and Y, respectively. These parameters are determined prior to analysis and are considered known values. The En score describes the difference between laboratory measurements and the average reference value relative to the variability involved in the difference.

In this case, each laboratory measurement is compared with the reference value via individual hypothesis testing. However, in several situations, we are interested in establishing multiple comparisons to evaluate the performance of the group of participating laboratories with respect to the reference value. However, this multiple comparison is not possible considering the score $E_n$.

Screenshot_1

Table 2.1: Laboratory data.

An important aspect of our model is the variance components. We use external information about the measurement system to estimate these components. In general, a measurement system has imperfections that incorporate errors into the result of a measurement. Traditionally, we understand that an error has two components, called random and systematic (see more details in MSA 4th edition). It is assumed that, after correction, the expected value of the error arising from this systematic effect is zero, however, with positive variance.

Take, for example, an EP in which the same item is sequentially distributed among participants, so that each participant performs several measurements at various points (levels or thresholds). An application for this is the case of an engine sequentially distributed among 10 laboratories, where each one performs 5 power measurements for each of the 18 rotation points. A way of presenting the data obtained can be seen in table (2.2), in which Yijk represents the (observed) value of the kth measurement, carried out at the jth level by the ith laboratory.

Screenshot_2

Table 2.2: Model in which the item is sequentially distributed among participants.

Again, ISO/IEC 17043:2010 [9] proposes the normalized error to evaluate the competence of laboratories participating in the EP, however, we know that the En score does not allow us to perform multiple comparisons. In our example, we would have to calculate 10*18 = 180 times the En score, as we now have 18 thresholds and 10 laboratories.

An important aspect for the development of a product is the quality of measurements and performance. For example, if a company develops an engine for the automobile industry, it is important that it knows how to maximize engine power and minimize cost and gas emissions. If this company has a problem related to engine power measurements, it will be difficult for it to know what the product’s real performance is. Consequently, this company may make wrong decisions during engine development, which, in general, has high costs. Therefore, it is important that the company knows how to appropriately define its measurement system.

16.2.1 Preliminary determination of proficiency

To guarantee the confidence necessary for the Proficiency Test, for example, in Engines. The artifact was developed based on the ILAC-G12:2000 standard. This standard describes the manufacturer’s quality system requirements and the procedures necessary to ensure the stability and consistency of the artifact.

Artifact consistency throughout the assay

The reference laboratory must test the artifact at the beginning and end of the proficiency test, according to the logistics described in the procedure. Criteria for homogeneity and adequate stability must be established and must be based on the effect that heterogeneity and instability will have on the assessment of participants' performance. Below, we describe some observations described in the ISO/IEC 17043:2010 standard [9]:

  • The requirements are intended to ensure that each participant receives the proficiency testing items so that they remain stable throughout the proficiency test.
  • Careful planning, manufacturing, and transportation are required to achieve a stable proficiency item; testing is often necessary to confirm it.
  • In some cases, it is not feasible for proficiency test items to be subjected to stability and homogeneity tests. Such cases may include, for example, when limited material is available to prepare proficiency test items.
  • In some cases, materials that are not sufficiently homogeneous or stable, in such cases they may still be useful as proficiency test items, provided that the uncertainties of the assigned values ​​or the evaluation of results take this into account (see B.3.1.3 and ISO 13528:2005 [19], annex B);
  • Homogeneity and stability considerations are discussed in ISO Guide 34 [17], ISO Guide 35 [18] and ISO 13528:2005 [19].

Procedures for assessing homogeneity and stability must be documented and carried out, where applicable, in accordance with appropriate statistical experiments. Whenever possible, the proficiency testing coordinator shall statistically use a random selection of a representative number of proficiency test items from the entire batch of a test material in order to evaluate the homogeneity of the material.

The homogeneity assessment will typically be performed after the proficiency test items (artifact) have been packaged prior to distribution to participants. The main objective of artifact consistency is to demonstrate that it is sufficiently stable to ensure that it will not undergo any significant changes throughout the period of proficiency testing, including storage and transportation conditions. When this is not possible, stability should be quantified and considered as an additional component of the measurement uncertainty associated with the value assigned to the proficiency test element, and/or taken into account in the assessment criteria.

When proficiency testing artifacts from previous rounds are retained for future use, the values ​​of properties to be determined in the proficiency testing program must be confirmed by the proficiency testing coordinator prior to distribution. In cases where homogeneity and stability testing is not feasible, the proficiency testing coordinator must demonstrate that the procedures used to collect, produce, package, and distribute the proficiency testing items are sufficient for the purpose of the proficiency testing.

Example

The reference laboratory tested the artifact at the beginning and end of the proficiency test, according to logistics described in the procedure. Table 2.1.1 presents the measurements of the Corrected Power variable by rotation range obtained in both tests.

Initial Round Initial Round Final Round Final Round
Range Average Standard Deviation Average Standard Deviation Trend
1200 8,93 0,062 8,99 0,046 -0,06
1600 12,61 0,035 12,69 0,08 -0,08
2000 16,05 0,046 16,18 0,037 -0,13
2400 19,55 0,078 19,73 0,081 -0,18
2800 24,87 0,086 25,12 0,14 -0,25
3000 27,04 0,072 27,26 0,12 -0,22
3200 28,53 0,068 28,81 0,092 -0,28
3600 31,74 0,111 31,91 0,084 -0,17
4000 34,22 0,171 34,48 0,171 -0,26
4400 37,35 0,14 37,68 0,186 -0,33
4800 41,33 0,166 41,74 0,218 -0,41
5200 44,54 0,241 44,67 0,187 -0,13
5400 46,53 0,277 46,69 0,151 -0,16
5600 47,62 0,208 47,93 0,169 -0,32
5800 48,68 0,261 49,01 0,254 -0,32
6000 49,59 0,203 49,61 0,235 -0,02
6200 50,02 0,365 50,19 0,211 -0,16
6400 50,24 0,318 50,61 0,178 -0,36

Table 2.1.1: Average measurements from the reference laboratory.

We calculate the artifact trend during rounds (start and end).

Rotation Average Trend $ u_c(\text{ref}) $ k U
1200 8,963 -0,063 0,073 2 0,147
1600 12,648 -0,077 0,102 2 0,205
2000 16,115 -0,126 0,13 2 0,259
2400 19,641 -0,182 0,16 2 0,32
2800 24,998 -0,248 0,205 2 0,41
3000 27,152 -0,216 0,22 2 0,439
3200 28,667 -0,282 0,233 2 0,465
3600 31,825 -0,169 0,255 2 0,51
4000 34,353 -0,258 0,28 2 0,559
4400 37,516 -0,333 0,306 2 0,611
4800 41,537 -0,411 0,34 2 0,681
5200 44,604 -0,132 0,359 2 0,718
5400 46,609 -0,157 0,375 2 0,751
5600 47,775 -0,316 0,386 2 0,771
5800 48,844 -0,323 0,397 2 0,795
6000 49,601 -0,02 0,398 2 0,795
6200 50,103 -0,164 0,407 2 0,814
6400 50,424 -0,365 0,41 2 0,821

Table 2.1.2: Summary with mean values of measurements, trend and combined and expanded uncertainty from the reference laboratory.

Table 2.1.3 presents the test associated with the normalized error (for more details see Performance Statistics). As the normalized error in all rotation ranges is less than 1, we conclude that the two rounds of measurements from the reference laboratory are compatible. Therefore, the artifact showed stable behavior throughout the proficiency test.

Rotation Lab trend Reference Value Trend U Normalized Error (En) W P-value
1200 -0,063 0 -0,063 0,14 0,45 4,053 0,044
1600 -0,077 0 -0,077 0,195 0,394 3,099 0,078
2000 -0,126 0 -0,126 0,247 0,51 5,212 0,022
2400 -0,182 0 -0,182 0,304 0,597 7,131 0,008
2800 -0,248 0 -0,248 0,39 0,635 8,059 0,005
3000 -0,216 0 -0,216 0,418 0,517 5,348 0,021
3200 -0,282 0 -0,282 0,443 0,637 8,104 0,004
3600 -0,169 0 -0,169 0,484 0,349 2,438 0,118
4000 -0,258 0 -0,258 0,532 0,485 4,705 0,03
4400 -0,333 0 -0,333 0,582 0,572 6,537 0,011
4800 -0,411 0 -0,411 0,648 0,633 8,02 0,005
5200 -0,132 0 -0,132 0,683 0,194 0,751 0,386
5400 -0,157 0 -0,157 0,714 0,221 0,973 0,324
5600 -0,316 0 -0,316 0,733 0,431 3,723 0,054
5800 -0,323 0 -0,323 0,756 0,428 3,658 0,056
6000 -0,02 0 -0,02 0,755 0,027 0,015 0,904
6200 -0,164 0 -0,164 0,774 0,212 0,895 0,344
6400 -0,365 0 -0,365 0,781 0,467 4,355 0,037

Table 2.1.3: Artifact homogeneity by rotation range.

fig1_homog_art

Table 2.1.2: Artifact homogeneity by rotation range.

16.2.2 Comparison of results

In this section we will describe three of the six distinct types of proficiency testing (PE), discussed in ISO/IEC 17043:2010 [9], among them we mention only four, which are:

  • Sequential scheme: where one or more proficiency test items are distributed sequentially for testing or measurement and returned to the proficiency testing provider at regular intervals;
  • Simultaneous regime: where proficiency test items are distributed for simultaneous testing or measurement within a defined period of time;
  • Continuous regime: where proficiency test items are provided at regular intervals, for example external quality assessment (EQA) programs.

Next we describe the sequential comparison scheme.

Sequential measurement comparison schemes

This is where a single test item is distributed sequentially among participants, with them sending the item to the next laboratory or coordinator for maintenance. This procedure is common in calibration standard comparisons, and measurement results, along with standard uncertainties, must be sent to the coordinator. Figure 2.2.1 illustrates how this comparison is made.

(image missing)

Figure 2.2.1: Sequential comparison scheme.

The results of individual measurements are compared with the value designated by the reference established in agreement with the participants. The coordinator must take into account the measurement uncertainty of each participant. It can be difficult to compare results across groups, for example, there may be relatively few participants with measurement capabilities that come close to each other.

Schemes involving sequential participation can take time (in some cases years) to complete. This causes a series of difficulties, such as

  • Ensure product stability;
  • Strict monitoring of circulation between participants and the time allowed for measuring individual participants;
  • Need to provide feedback on individual performance during system implementation rather than waiting until it is finished.

The artifacts or proficiency test items used in this type of proficiency test include, for example, measurement reference standards (resistors, micrometers, etc.), or in biological tests, histological slides with a confirmed diagnosis.

Trials that follow this comparison scheme but are limited to situations in which a single participant is tested independently are often called “measurement audits.” In some situations, the reference value for a proficiency testing item or artifact may be determined by consensus after all participants (in some situations a subset of participants) complete the comparison measurement.

Simultaneous participation schemes

Concurrent participation proficiency testing generally involves subsamples randomly selected from a source of material to be distributed simultaneously to participants for simultaneous testing.

In some systems, participants are required to collect samples, which are considered proficiency test items for analysis. Upon completion of the test, the results are returned to the proficiency testing coordinator and compared to the benchmark to give an indication of the performance of each participant and the group as a whole.

Examples of proficiency test items used in this type of comparison include foods, bodily fluids, agricultural products, water, soil, minerals, and other environmental materials. In some cases, separate portions of previously established reference materials are circulated. Educational advice or comments are typically part of the report returned to participants by the proficiency testing coordinator with the goal of promoting improvement in performance.

Known value schemes use reference values ​​that are determined independently of participants and involve the preparation of proficiency testing items with a number of known measures or characteristics.

Certified reference materials can also be used in these schemes, as well as their certified value and measurement uncertainty can be used directly. Direct comparison can also be made between a proficiency test item and a certified reference material under repeatability conditions. However, care must be taken to ensure that the certified reference material closely resembles a proficiency test item. Proficiency test items from previous proficiency testing rounds may be used in this type of scheme if the item demonstrates stability.

External quality assessment (EQA) programs

Most EQA programs are designed to provide information about the entire laboratory workflow path and not just the testing processes. These programs are ongoing schemes that include long-term monitoring of laboratory performance.

A typical feature of EQA programs is to provide education to participants and promote quality improvement. Consultancy and training comprise part of the report returned to participants to achieve this objective. Some EQA programs evaluate the performance of the pre- and post-analytical phases of testing, as well as the entire analytical phase.

In such programs, the nature of the proficiency test item may differ significantly from that used in traditional proficiency testing schemes. The “proficiency test” item could be a questionnaire or case study circulated by the EQA provider to each participant for feedback on specific responses. Alternatively, pre-analytic information may accompany the proficiency test item, which requires the participant to select an appropriate approach to testing and interpreting results, not just for and not just to perform the test.

In “review” sampling schemes, participants may be required to provide “proficiency testing” to the EQA provider. This may take the form of a processed specimen or specimen, e.g., stained slide or fixed tissue.

For laboratory data we can exemplify:

  • Test results;
  • Laboratory reports;
  • Quality assurance reports/control records.

For documentation, we can give an example:

  • Procedures;
  • Method verification criteria.

16.2.3 Determination of the reference value and its uncertainty

According to ISO/IEC 17043:2010 [9], some procedures can be used to establish the reference value and its uncertainty. In this standard, the most common reference values used are:

  1. Known values: results determined by specific formulation of the proficiency test item (e.g., manufacturing or dilution);
  2. Certified values: determined by definitive test or measurement methods (by quantitative tests);
  3. Values per reference laboratory: determined by analysis, measurement or comparison of the proficiency test item alongside a reference material or standard, traceable to an international standard;
  4. Consensus values of expert participants: specialists (which may, in some situations, be reference laboratories) must have proven competence in determining the measurand(s) under test, validated using methods known to be highly accurate and comparable to general methods;
  5. Participant Consensus Values: using statistical methods described in ISO 13528 [15] and the IUPAC International Harmonized Protocol, and with consideration of the effects of outliers.

In this study we will use procedures 3 and 4 to determine the reference value.

  • Consensus among participants (average of measurements);
  • Consensus between rounds from a laboratory with international experience (via calibration).

Consensus among Participants

Determined through consensus between laboratory measurements. Before carrying out this procedure, it is necessary to check the existence of extreme values. To do this, we can use the Extreme Value Test (Grubbs), described in the Extreme Value Test (Grubbs) content. With laboratory measurements not considered extreme values, we have to

$$VR=\displaystyle\frac{1}{p}\sum^{p}_{i=1}y_i$$

where:

  • $y_i$: Represents the average of measurements from the ith laboratory;
  • p: Represents the number of laboratories, whose measurements are not considered extreme values.

To calculate the reference value by consensus among participants, we proceed as follows:

$$U(\text{ref})=~2 \times \sqrt{u_{\text{ag}}^2(\text{labs})}$$

or if we consider the uncertainty due to the artifact (engine), we have

$$U(\text{ref})=~2 \times \sqrt{u_{\text{ag}}^2(\text{labs})+u^2(\text{motor})}$$

in which,

  • U(ref): Represents the expanded uncertainty of the reference;
  • u(motor): Represents the standard uncertainty associated with the intrinsic variability of the artifact (VR × 0.25%).

The grouped laboratory uncertainty is given by:

$$u_{\text{ag}}(\text{labs}) = \sqrt{\displaystyle\frac{1}{p^2}\sum^{p}_{i=1}u^2(\text{lab}_i)}$$

  • $u_{ag}$(labs): Represents the grouped uncertainty of the laboratory;
  • u($lab_i$): Represents the standard measurement uncertainty of the ith laboratory;

We can illustrate, as an example, the test in which we used the artifact (motor). Thus, the laboratory uncertainty is calculated as follows:

$$u(\text{lab}_i)=\sqrt{u^2(\text{cal})+u^2(\delta)}$$

  • u(cal): Represents the uncertainty of the bank calibration;
  • u(δ): Represents the uncertainty due to Repeatability;

Example 2.2.1

The laboratory in an EP tested the artifact, according to logistics described in the procedure. Table 2.2.1 presents the measurements of the Corrected Power variable in the rotation range of 2000 rpm.

laboratory Corrected Power laboratory Corrected Power
1 27 4 27,4
1 26,8 4 27,4
1 27 4 27,4
1 26,8 4 27,4
1 26,9 4 27,5
2 27,41 5 27,06
2 27,43 5 27,33
2 26,97 5 27,01
2 27,27 5 26,97
2 27,62 5 27,13
3 26,43 6 27,23
3 26,37 6 27,33
3 26,41 6 27,24
3 26,45 6 27,18
3 26,37 6 27,30

Table 2.2.1: Corrected Power Measurements in the 2000 rpm range.

Next, we will make a descriptive summary, for this we will use Action.

(image missing)

Figure 2.2.1: Descriptive summary output using Action.

In this example, we will take laboratory 2, the other cases are calculated in a similar way. Laboratory 1 presented an uncertainty due to bank calibration of u(cal)=0.164 kW and the uncertainty due to repeatability was calculated in the descriptive summary (see figure 2.2.1), which is given by the standard deviation of the mean (for more details access the measurement uncertainty content), that is, we have that uncertainty due to repeatability is u(δ)=0.045 kW.

Thus, the combined uncertainty of laboratory 1 is calculated as follows:

$$u(\text{lab}_1)=\sqrt{0,164^2+0,045^2}=0,17~kW$$

Therefore, the expanded uncertainty of laboratory 1 is calculated as follows:

$$U(\text{lab}_1)=0,17\times 2=0,34~kW$$

Below, in table 2.2.2 we present a summary of all laboratories.

Laboratory u(cal) u(δ) u($lab_i$) U($lab_i$) k
1 0,164 0,045 0,170 0,34 2
2 0,259 0,057 0,265 0,53 2
3 0,354 0,016 0,355 0,71 2
4 0,150 0,02 0,152 0,30 2
5 0,410 0,063 0,415 0,83 2
6 0,258 0,027 0,259 0,52 2

Table 2.2.2: Summary of uncertainty calculation for laboratories.

As the standard uncertainty associated with the intrinsic variability of the artifact (VR × 0.25%), then we have:

$$VR=\displaystyle\frac{1}{p}\sum^{p}_{i=1}y_i=\frac{26,9+\dots+27,3}{6}=27,1~kW$$

$$u(\text{motor})=\frac{0,25}{100}\times 27,1=0,068~kW$$

Now let’s calculate the pooled uncertainty.

$$u_{\text{ag}}(\text{labs}) = \sqrt{\displaystyle\frac{1}{p^2}\sum^{p}_{i=1}u^2(\text{lab}_i)}=\sqrt{\frac{1}{6^2}\sum^6_{i=1}0,17^2+\dots+0,259^2}=0,1163~kW$$

Therefore, the expanded uncertainty of the reference value is given by:

$$U(\text{ref})=~2 \times \sqrt{u_{\text{ag}}^2(\text{labs})+u^2(\text{motor})}=0,269~kW$$

Extreme Value Test (Grubbs)

This test is developed to check the presence of extreme values in sample observations. Extreme values can be considered as manifestations of the random variability inherent in the data, or a calculation error during data collection and even a hasty note by the operator.

There are several criteria for testing extreme values. For all of them we calculate a performance statistic and compare it with a critical value, based on the theory of random samples, to decide whether or not there is an observation considered an extreme value.

In the Grubbs test, we use the following statistic

$$Z~=~\frac{|x_i ~-~\overline{x} |}{s}$$

where:

  • $x_i$: It is an observation of the sample x1, x2, …, xn;
  • $ \overline{X} $: It is the sample mean;
  • s: It is the sample standard deviation.

This statistic tests the following hypotheses

Screenshot_6

(there are missing elements in the equation above)

We reject the hypothesis $ H_0 $, with significance level $ \alpha $, if $ Z> Z_c $, in which $ Z_c $ is a critical value based on the distribution of Z and is tabulated (see F. E. Grubbs (1969) [16]) for some values of $ \alpha $.

Example 2.2.2

The laboratory in an EP tested the artifact, according to logistics described in the procedure. Table 2.2.2 presents the measurements of the Corrected Power variable in the rotation range of 1200 rpm.

Screenshot_3

Table 2.2.2: Corrected Power Measurements in the 1200 rpm range.

Let’s calculate the mean and standard deviation:

$$\overline{X}=\frac{6,8+8,9+\dots+8,1}{6}=8,35$$

$$s=\sqrt{\frac{(6,8-8,38)^2+\dots+(8,1-8,35)^2}{6-1}}=0,8167$$

With that, let’s calculate the Grubbs test for point 1, using the following statistic

$$Z=\frac{|X_i-\overline{X}|}{s}=\frac{|X_1-\overline{X}|}{s}=\frac{|6,8-8,35|}{0,8167}=1,89788$$

How $ Z> 1,887 $ therefore, this measurement is an extreme value (outlier).

(image missing)

Consensus between Rounds from a laboratory with international experience

Determined through consensus between measurements from the two rounds of the Reference Laboratory. With this, we have to

$$VR~=~\displaystyle\frac{1}{p}\sum^{p}_{i=1}y_i$$

where:

  • $y_i$: Represents the average of measurements from the ith round;
  • p: Represents the number of rounds carried out by the Reference Laboratory (p=2).

$$U(\text{ref})=~2 \times \sqrt{u^2_{\text{ag}}(\text{Lref})+u^2(\text{motor})}$$

and

$$u_{\text{ag}}(\text{ref}) = \sqrt{\frac{u^2(\text{rod}_1)+u^2({\text{rod}_2})}{2}},$$

where:

  • U(ref): represents the expanded uncertainty of the reference;
  • $u_{ag}$(Lref): represents the grouped uncertainty of the two rounds of the reference laboratory;
  • u($rod_i$): represents the standard measurement uncertainty of the ith round, for i=1,2;
  • u(motor): represents the standard uncertainty associated with the intrinsic variability of the artifact (VR × 0.25%).

Example 2.2.3

The reference laboratory tested the artifact at the beginning and end of the EP in a sequential comparison scheme, according to logistics described in the procedure. Table 2.2.3 presents the measurements of the Corrected Power variable in the rotation range of 1200 rpm.

Screenshot_4

Table 2.2.3: Corrected Power Measurements in the 1200 rpm range for the Reference Laboratory.

After analyzing the consistency of the artifact (for more details see the section Preliminary determination of proficiency), we will determine the reference value and its uncertainty. To do this, we calculate as follows:

Screenshot_5

The reference laboratory presented an uncertainty due to bank calibration of u(cal)=0.165 kW and the uncertainty due to repeatability was presented in table 2.2.3. First, let’s calculate the pooled uncertainty of the reference laboratory.

$$u_{\text{ag}}(\text{Lref})=\sqrt{u^2(cal)+u^2(\delta_1)+u^2(\delta_2)}=\sqrt{0,165^2+0,0322^2+0,0179^2}=0,169~kW$$

in which

$$u(\delta_1)=\frac{s_1}{\sqrt{5}}=\frac{0,072}{\sqrt{5}}=0,0322~kW,\quad\text{e}\quad u(\delta_2)=\frac{s_2}{\sqrt{5}}=\frac{0,04}{\sqrt{5}}=0,0179~kW$$

Now, let’s calculate the uncertainty of the engine.

$$u(\text{motor})=\frac{0,25}{100}\times8,945=0,02236~kW$$

Thus, the uncertainty of the reference laboratory is given by:

$$U(\text{ref})=~2 \times \sqrt{u^2_{\text{ag}}(\text{Lref})+u^2(\text{motor})}=2\times\sqrt{0,169^2+0,02236^2}=0,56~kW$$

16.2.4 Performance statistics

To evaluate the performance of the results obtained, we will use the Normalized Error test suggested by ISO/IEC 17043:2010 [9], the trend test according to Leão, Aoki and Silva (2009) [14] and the Trend and Linearity test according to Leão, Aoki and Silva (2009) [14].

16.2.4.1 Normalized Error

The Normalized Error test tests the compatibility of laboratory measurement results with respect to the reference value. In this case, the ISO/IEC 17043:2010 [9] standard proposes the following statistics:

Normalized Error - $ E_n $

$$E_n=\frac{|\text{Res}_{\text{lab}}-VR|}{\sqrt{U^2(\text{lab})+U^2(\text{ref})}}$$

where:

  • $Res_{lab}$: It is the result of the participating laboratory;
  • VR: It is the reference value;
  • U(lab): is the expanded uncertainty of the participating laboratory;
  • U(ref): is the expanded uncertainty of the reference value;
  • Trend =$Res_{lab}$-VR

.

  • Compatible Laboratory: Laboratory whose calibration result for a given circulating artifact obtained a normalized error (En) less than or equal to unity.
  • Non-Compatible Laboratory: Laboratory whose calibration result for a given circulating artifact obtained a normalized error (En) greater than unity.

The Figure illustrates the graphical representation of the results of this comparison.

(image missing)

Example 2.4.1.1

The laboratory in an EP tested the artifact, according to logistics described in the procedure. Table 2.4.1.1 presents the data summary.

Laboratory Lab Average VR Trend $U_{lab}$ $U_{ref}$ U $E_n$
1 15,809 15,698 0,111 0,304 0,442 0,536 0,208
2 15,335 15,698 -0,363 0,272 0,442 0,519 0,700
3 15,778 15,698 0,080 0,115 0,442 0,457 0,175
4 15,918 15,698 0,220 0,334 0,442 0,554 0,397
5 15,885 15,698 0,188 0,801 0,442 0,915 0,205
6 16,019 15,698 0,321 0,370 0,442 0,577 0,557
7 15,461 15,698 -0,237 0,393 0,442 0,592 0,401
8 15,900 15,698 0,202 0,582 0,442 0,731 0,277

Table 2.4.1.1: Summary of EP in the 2000 rpm range.

Next, we will calculate the normalized error for laboratory 2. First, we will calculate the trend modulus.

$$|\text{Tendência}|=|\text{Res}_{\text{lab}_2}-VR|=|15,698-15,335|=0,363$$

and

$$U=\sqrt{U^2(\text{lab}_2)+U^2(\text{ref})}=\sqrt{0,272^2+0,442^2}=0,519$$

Thus, we calculated the normalized error En for laboratory 2.

$$E_n=\frac{|\text{Res}_{\text{lab}_2}-VR|}{\sqrt{U^2(\text{lab}_2)+U^2(\text{ref})}}=\frac{0,363}{0,519}=0,7$$

How $E_n$= 0,7 < 1, então concluímos que o laboratório 2 está compatível em relação ao valor de referência.

16.2.4.2 Teste de Tendência

O teste de tendência testa a compatibilidade dos resultados das medições do laboratório com respeito ao valor de referência. Neste caso, vamos utilizar o teste da razão de verossimilhança generalizada proposto por Leão, Aoki e Silva (2009) [14].

FIG_MSA_009

Para desenvolver o teste, foi considerado o seguinte modelo

$$Y_{ij}=\alpha_i+X+\epsilon_{ij},~j = 1,\ldots ,n_i~;~i=1,\ldots,k,~~(2.4.1)$$

em que $ Y_{ij} $ representa a j-ésima medição do i-ésimo laboratório, $ \alpha_i $ representa o parâmetro de tendência do i-ésimo laboratório em relação a $ X $ (valor de referência) e $ \epsilon_{ij} $ representa o erro de medição associado ao i-ésimo laboratório na j-ésima medição.

Com o modelo dado por (2.4.1), testamos a competência do laboratório conforme as seguintes hipóteses

Screenshot_7

(there are missing elements in the equation above)

Therefore, the statistic used for the test is

$$W~=~\hat{\alpha}^\top~(J^\top \Sigma^{-1} J)~\hat{\alpha}$$

in which

  • α: Represents the trend parameter estimator;
  • J: Represents an n x 1 vector of constants;
  • Σ: Represents an n x n matrix of constants. This matrix is ​​formed by functions of the combined uncertainties, both from the laboratory and from the reference value.

We have that the distribution of the statistics is

$$W\sim\chi^2_{1,\tau}~~ \text{(não central)}~(2.4.2)$$

in which $ \tau = \displaystyle\frac{1}{2} \alpha^\top(J^\top\Sigma^{-1}J)\alpha. $

Under $\alpha = 0 $, we have to $ \tau=0 $. Then, $ W\sim\chi^2_1 $ (central) and by the generalized likelihood ratio test, we reject $ H_0 $ if $ W \geq w_0 $, in which $ w_0 $ is a value obtained from the chi-square distribution with 1 degree of freedom and significance level $ \eta $. Therefore, if we do not reject the hypothesis $ H_0 $, we can say that the laboratory measurement is TRVG-Satisfactory.

We can also calculate the p-value, which is the smallest value for which we reject the hypothesis $ H_0 $. This is calculated by

$$P(W> W_{obs}).$$

Example 2.4.2.1

Consider the data from example 2.4.1.1. Table 2.4.2.1 presents the data necessary to calculate the W statistic.

Laboratory Trend $U_{lab}$ $U_{ref}$ $n_i$
1 0,111 0,304 0,442 5
2 -0,363 0,272 0,442 24
3 0,080 0,115 0,442 18
4 0,220 0,334 0,442 9
5 0,188 0,801 0,442 16
6 0,321 0,370 0,442 16
7 -0,237 0,393 0,442 28
8 0,202 0,582 0,442 16

Table 2.4.2.1: Data for calculating the W statistic.

Initially, to calculate the W statistic, we divide the expanded uncertainties by k. Let’s take lab 1:

$$u^2(\text{lab}_1)=\left(\frac{U(\text{lab}_1)}{k}\right)^2=\left(\frac{0,304}{2}\right)^2=0,023,\quad \text{e}\quad u^2(\text{ref})=\left(\frac{U(\text{ref})}{k}\right)^2=\left(\frac{0,442}{2}\right)^2=0,0488$$

Now, we divide the uncertainty of laboratory 1 by the number of laboratory measurements at the rotation point of 2000 rpm.

$$\hat{\sigma}^2_x=\frac{u^2(\text{lab}_1)}{n_1}=\frac{0,023}{5}=0,00461$$

With this, we have to

$$\hat{\sigma}^2_x+u^2(\text{ref})=0,00461+0,0488=0,053$$

Therefore, the W statistic is

$$W=\frac{\hat{\alpha}^2_1}{\hat{\sigma}^2_x+u^2(\text{ref})}=\frac{0,0123}{0,053}=0,233$$

As previously stated, $ W\sim\chi^2_1 $ (central). Thus, we calculate the p-value, which is the smallest value for which we reject the hypothesis $\alpha=0 $.

$$P(W> W_{obs})=0,63$$

Therefore, we do not reject the hypothesis $H_0$, that is, we have evidence that the trend is not significant at the 5% significance level.

Below is a summary for all labs.

Trend $ \alpha_i $ $U_{lab}$ $U_{ref}$ $n_i$ $u^2(lab)$ $u^2(ref)$ $u^2(lab)/n_i$ $(u^2(lab)/n_i)+u^2(ref)$ W P-value
0,111 0,304 0,442 5 0,0231 0,0488 0,00461 0,0534 0,233 0,63
-0,363 0,272 0,442 24 0,0184 0,0488 0,00077 0,0496 2,659 0,1
0,080 0,115 0,442 18 0,0033 0,0488 0,00018 0,0490 0,131 0,72
0,220 0,334 0,442 9 0,0279 0,0488 0,00310 0,0519 0,933 0,33
0,188 0,801 0,442 16 0,1603 0,0488 0,01002 0,0588 0,598 0,44
0,321 0,370 0,442 16 0,0343 0,0488 0,00214 0,0510 2,022 0,16
-0,237 0,393 0,442 28 0,0387 0,0488 0,00138 0,0502 1,119 0,29
0,202 0,582 0,442 16 0,0846 0,0488 0,00529 0,0541 0,756 0,38

Table 2.4.2.2: Summary of statistical calculation $W_i$.

16.2.4.3 Trend and Linearity Test

The trend and linearity test, proposed by Leão, Aoki and Silva (2009) [14], tests the compatibility of laboratory measurement results with respect to the reference value, for all measurement points. In this case, we will use Wald statistics, according to Sen and Singer (2000) [13]. To develop the test, the following model was considered

$$y_{ijk}=\alpha_{i}+\beta_{i}x_{j}~\quad(2.4.3.1),$$

$$Y_{ijk}=y_{ijk}+e_{ijk}~,$$

where, i=1,…,p, j=1,…,m, and k=1,…,ni. Thus, we have that Yijk (observed value) is subject to measurement error and we assume that yijk satisfies the ultra-structural linear relationship with the

true value xj (unobservable). In model (2.4.3.1) we have:

αi: represents the additive trend of the ith laboratory;

βi: represents the multiplicative trend of the ith laboratory;

xj: represents the “true” value of the characteristic to be measured. We assume that

xj v.a.ind. ∼ $ N(\mu_{xj} , \sigma^2_{xj}) $;

$ \mu_{xj} $: represents the location parameter;

σxj: represents the scale parameter and is considered known.

ϵijk: represents the random error in carrying out the kth measurement, at the jth level by the ith laboratory. We assume that

eijk v.a.ind. ∼ N(0, σ2ij);

σij: represents the scale parameter and is considered known.

We also consider that xj is independent of eijk.

The objective of this analysis is to compare the additive (αi) and multiplicative (βi) trend of each laboratory (i) with the reference value, which in this case, results in α = 0 and β = 1. Here, we will consider the first of the p laboratories as being the reference laboratory, therefore, and considering the initial assumptions, we have that:

Screenshot_8

Considering the proposed model, we tested the laboratory’s competence by developing the following tests:

  • Hypothesis 1: We will test the following hypotheses

Screenshot_9

(there are missing elements in the equation above)

In this case, we are testing whether the laboratory’s additive trend can be considered null.

  • Hypothesis 2: We will test the following hypotheses

Screenshot_10

(there are missing elements in the equation above)

In this case, we are testing whether the laboratory’s multiplicative tendency can be considered equal to one.

  • Hypothesis 3: We will test the following hypotheses

Screenshot_11

(there are missing elements in the equation above)

In this case, we are testing whether the additive trend is zero and whether the multiplicative trend can be considered equal to one.

The statistics used to carry out the tests are:

  • Test 1:

$$Qw_{1i}=\frac{\widehat{\alpha}^2_i}{\upsilon_{\alpha_i \alpha_i}}\quad\text{para}~i=2,\cdots, p.$$

  • Test 2:

$$Qw_{2i}=\frac{(\widehat{\beta}_i - 1)^2}{\upsilon_{\beta_i\beta_i}}\quad\text{para}~i=2,\cdots, p.$$

  • Test 3:

Screenshot_12

in which

Screenshot_13

Screenshot_14

Regardless of the test to be performed, according to Sen and Singer (2000) [13] (theorem 5.6.3, page 240), we have that, under $H_0$ the statistics $ Qw\sim\chi^2_1. $ Therefore, we reject $H_0$ if Qw ≥ qw, where qw is a value obtained from the chi-square distribution with 1 degree of freedom and significance level $ \eta $. Therefore, if we do not reject the hypothesis $H_0$, we can say that the laboratory measurement is Qw-Satisfactory.

We can also calculate the p-value, which is the smallest value for which we reject the hypothesis $H_0$. This is calculated by

$$P(Q_w> Q_{w_{obs}}).$$

Trend and linearity graph

In an individual analysis, we will present a graph with the trends (tijk) of each laboratory, for the entire measurement range. Here, we define bias as the difference between laboratory measurements and the true engine value, i.e. tijk = Yijk − μxj. Furthermore, we illustrate on the graph the 99% confidence interval for the average trend, which is given by

Screenshot_15

Example 2.4.3.1

The laboratory in an EP tested the artifact, according to logistics described in the procedure. Let’s consider the study of trend and linearity for laboratory 1.

1200 2000 3000 3600 4400 5200 5600 6000 6400
8,89 15,84 26,84 31,41 37,19 44,35 47,49 49,92 50,92
8,83 15,80 26,61 31,31 37,12 44,35 47,33 49,76 50,74
8,86 15,80 26,86 31,40 37,24 44,28 47,56 49,98 50,89
8,85 15,79 26,85 31,40 37,17 44,31 47,60 49,90 50,84
8,88 15,81 26,92 31,50 37,32 44,40 47,79 49,96 50,94

Table 2.4.3.1: Corrected Power Measurements for each rotation range of laboratory 1.

1200 2000 3000 3600 4400 5200 5600 6000 6400
8,74 15,70 26,47 31,13 36,91 43,81 46,87 48,65 49,52

Table 2.4.3.2: Corrected Power Measurements for each rotation range of the reference laboratory.

laboratory 1200 2000 3000 3600 4400 5200 5600 6000 6400
1 0,085 0,152 0,258 0,302 0,357 0,426 0,457 0,479 0,489

Table 2.4.3.3: Combined uncertainty of corrected Power for each rotation range of laboratory 1.

1200 2000 3000 3600 4400 5200 5600 6000 6400
0,123 0,221 0,373 0,439 0,518 0,614 0,656 0,681 0,692

Table 2.4.3.4: Combined uncertainty of corrected Power for each rotation range of the reference laboratory.

Rotation $n_i$ Average Standard Deviation Amplitude
1200 5 8,86 0,025 0,062
2000 5 15,81 0,020 0,050
3000 5 26,81 0,121 0,311
3600 5 31,40 0,067 0,189
4400 5 37,21 0,075 0,199
5200 5 44,34 0,043 0,114
5600 5 47,55 0,165 0,452
6000 5 49,90 0,086 0,222
6400 5 50,87 0,082 0,205

Table 2.4.3.5: Descriptive Statistics for corrected power measurements for each rotation range in laboratory 1.

See below the results obtained by the Action software for an example.

(image missing)

Figure 2.4.3.1: Trend and Linearity test results obtained by the Action software.

(image missing)

Figure 2.4.3.2: Graph of the Trend and Linearity test obtained by the Action software.

From the results obtained in Figure 2.4.3.1 and Figure 2.4.3.2, we verified that laboratory 1 does not present satisfactory measurements, as in the joint test of additive and multiplicative trends (hypothesis 3) we rejected $H_0$, as the p-value=0.016 is below the significance level of 0.05. Furthermore, when we tested only the null additive trend hypothesis, we realized that we did not reject $H_0$, as the p-value is equal to 0.714. For the multiplicative trend, we reject the hypothesis $H_0$ multiplicative trend equal to 1, as the p-value is equal to 0.047 (considering a significance level of 0.05). Therefore, the laboratory needs to better evaluate its measurement system, as it presents significant linearity and a pattern of positive trends.

16.3 Applications of the Proficiency Test

In this section, we will show some applications of the Proficiency Test. All cases presented are real and were prepared by the client with the participation of the ESTATCAMP team.

16.3.1 Proficiency Test for Net Effective Engine Power

The purpose of a measurement system is to determine the value of a particular quantity to be measured, in our example, engine power. A measurement begins with an appropriate specification of the quantity to be measured, with a definition of the measurement method and procedure. Engine power (P) is the product of Torque (T), obtained by the engine, and the associated rotation (R). Furthermore, it is well known that engine power is influenced by altitude. Unfortunately, the laboratories that carry out engine testing are located in different places at different altitudes. Consequently, it is necessary to specify a correction factor (CF) in order to adjust laboratory measurements to a standard altitude (sea level). The measurement method, specified by the automobile industry, to measure engine power is the “load cell” dynamometer device. Finally, the measurement procedure describes how to use the dynamometer device to obtain engine power measurements, that is, it defines how to connect the engine with the device and the engine rotation value during measurements. All these definitions are described in the ABNT NBR ISO 1585 standard [10].

In EP, the engine is distributed sequentially among the participants, so that each laboratory, after carrying out engine power measurements at different rotation points, sends the engine to the next participant. Next, we will describe the specifications of the Artifact (Gasoline Engine):

  • Supplier: GM Powertrain;
  • Model: 1:0 MPFI;
  • Fuel: E25;
  • Compression Ratio: 9.4 : 1;
  • Quantity of Oil: 2.66 kg;
  • Oil Specification: Texaco Energy 2 - SAE 5W30 SL.

Participating laboratories are

  • Robert Bosch
  • Mahle Metal Light
  • Ford
  • Mauá Institute of Technology
  • Magneti Marelli
  • KSPG Automotive Brazil
  • Delphi
  • GM Powertrain

Renault

Each Laboratory must make 5 turns with this engine. The sources of uncertainty necessary for the EP are:

  • Uncertainty inherited from the test bench;
  • Repetitiveness;
  • Uncertainty due to the artifact.

An important point of our analysis is the use of variability intrinsic to the artifact. We define that this intrinsic variability is associated with engine technology and its production. For this, we adopted a standard deviation of 0.25% of the reading in corrected power (kW) associated with the artifact. For example, for a measurement of 50 kW, we associate a standard deviation of 0.002550kW = 0.125kW, due to the gasoline engine (artifact). Therefore, this intrinsic variability of the artifact is considered as a source of additional uncertainty.

Figure 3.1.1 illustrates the engine and its components, arranged inside the packaging used to transport them.

(image missing)

Figure 3.1.1: Detail of the engine and its components arranged inside the transport packaging.

Figure 3.1.2 illustrates the motor coupled to the dynamometric device in one of the participating laboratories.

(image missing)

Figure 3.1.2: Detail of the engine coupled to the dynamometric device.

(image missing)

Figure 3.1.3: Detail of the engine coupled to the dynamometric device.

(image missing)

After taking the measurements, each laboratory must calculate its uncertainty associated with the corrected power value. Next, we can use any of the applications for the methodology developed to evaluate the competence of laboratories participating in the EP. We can consider, for example, that the reference value is obtained by the consensus value between the participating laboratories.

16.3.1.1 Artifact Consistency Analysis

The proficiency test was carried out for the Corrected Power variable obtained in the test of a Gasoline engine, in accordance with NBR ISO 1585/1996 [10].

$${Potência}={Torque}\times{Rotação}$$

$${Potência Corrigida}={Potência}\times{Fator de Correção}$$

$${Fator de Correção}=\left(\frac{99}{P_{as}}\right)^{1,2}\times\left(\frac{T_{adm}}{298}\right)^{0,6}$$

in which,

  • $P_{as}$: Represents dry air pressure (kPa);
  • $T_{adm}$: Represents the inlet temperature (Kelvin).

Furthermore, the following rotation ranges were considered:

Rotation Ranges (rpm):

1200 2000 3000 3600 4400 5200 5600 6000 6400

One of the laboratories tested the artifact at the beginning and end of the proficiency test, according to the logistics described in the procedure. Below, we present a descriptive summary of the measurements of the initial and final runs by rotation range.

Screenshot_16

Figure 3.1.1.1: Descriptive statistics of the final and initial rounds.

(image missing)

Figure 3.1.1.1: Boxplot of the corrected power of the initial round by rotation range.

(image missing)

Figure 3.1.1.2: Boxplot of the corrected power of the final round by speed range.

(image missing)

Figure 3.1.1.3: Comparison between the Final and Initial Rounds.

For example, taking the 1200 Rotation range we have:

Trend

$${Tendência}=\overline{Rod}_i-\overline{Rod}_f=8,92-8,86=0,058$$

Rotation $ \overline{\text{Rod}}_i $ $ \overline{\text{Rod}}_f $ Trend k U(ref)
1200 8,92 8,86 0,058 2 0,17
2000 16,08 15,81 0,272 2 0,3
3000 26,99 26,81 0,173 2 0,52
3600 31,9 31,4 0,496 2 0,6
4400 37,68 37,21 0,474 2 0,71
5200 44,96 44,34 0,625 2 0,85
5600 48,38 47,55 0,821 2 0,93
6000 50,61 49,9 0,703 2 0,96
6400 51,81 50,87 0,941 2 0,98

Table 3.1.1.2: Comparison between the Final and Initial Rounds.

Table 3.1.1.3 presents the test associated with the normalized error. We have that the normalized error in all rotation ranges is less than 1. We conclude that the two rounds of laboratory measurements are compatible.

Therefore, the artifact showed stable behavior throughout the proficiency test.

Rotation Lab Trend VR Trend U $E_n$ W P-value
1200 0,058 0 0,058 0,170 0,341 0,465 0,495
2000 0,272 0 0,272 0,302 0,901 3,246 0,072
3000 0,173 0 0,173 0,520 0,332 0,442 0,506
3600 0,496 0 0,496 0,602 0,824 2,716 0,099
4400 0,474 0 0,474 0,711 0,667 1,780 0,182
5200 0,625 0 0,625 0,850 0,735 2,158 0,142
5600 0,821 0 0,821 0,925 0,888 3,152 0,076
6000 0,703 0 0,703 0,958 0,734 2,154 0,142
6400 0,941 0 0,941 0,983 0,957 3,661 0,056

Table 3.1.1.3: Artifact homogeneity by rotation range.

(image missing)

Figure 3.1.1.4: Comparison graph between the Final and Initial Rounds.

16.3.1.2 Compatibility test

In this section we present the graphs and the main descriptive statistics for the corrected power variable throughout the rotation range, then we will carry out compatibility tests between the laboratory results and the reference value. For the latter we will use the Normalized Error (for more details see Performance Statistics). In this application we will take the rotation ranges of 1200 and 2000 rpm and a consensus reference value has been defined (for more details see determination of the reference value and its uncertainty). Below we present the scatter plots of the average corrected power for each laboratory by speed range.

(image missing)

Figure 3.1.2.1: Average curves per laboratory.

(image missing)

Figure 3.1.2.2: Scatterplot by laboratory and for all rotation ranges.

Next, let’s take the 1200 and 2000 rpm points as an example, the other points are calculated in a similar way. In figures 3.1.2.3 and 3.1.2.4, we have scatter plots and a boxplot of the corrected power for the 1200 and 2000 rpm ranges. In table 3.1.2.1 we have the descriptive statistics of the data and, in table 3.1.2.2 we see the result of the extreme value test (see determination of the reference value and its uncertainty), then the calculation of the reference values: Consensus among Participants.

Rotation Minimum 1st Quartile Average Median Standard Deviation 3rd Quartile Maximum
1200 8,46 8,6 8,74 8,75 0,15 8,85 9,13
2000 15,05 15,45 15,7 15,76 0,28 15,94 16,35

Table 3.1.2.1: Descriptive statistics for the 1200 and 2000 rpm ranges.

(image missing)

Figure 3.1.2.3: Scatterplots and boxplot for the 1200 rotation range.

(image missing)

Figure 3.1.2.4: Scatterplots and boxplot for rotation range 2000.

Taking laboratory 1 as an example, we have:

$$Z~=~\frac{|x_i ~-~\bar{x} |}{s}=\frac{|8,86 ~-~8,78 |}{0,13}=0,65~({Rotação 1200})$$

$$Z~=~\frac{|x_i ~-~\bar{x} |}{s}=\frac{|15,81 ~-~15,76 |}{0,24}=0,19~({Rotação 2000})$$

As there are no extreme values for the 1200 rpm range, the reference value - VR is given by the average of measurements from all laboratories, with the respective expanded uncertainty, according to Table 3.1.2.2.

$$VR~=\frac{\displaystyle\sum^{8}_{i=1}y_i}{8}=\frac{8,86+\dots+8,71}{8}=8,78$$

$$U_{ref}=~2 \times \sqrt{\sum^{8}_{i=1}\frac{u^2(lab_i)}{8}}=2 \times \sqrt{\frac{0,085^2+\dots+0,159^2}{8}}=0,25$$

Therefore, for the rotation range of 1200 rpm we have$ VR=8,78~kW;~U=0,25~kW. $

Similarly, for the 2000 rpm range, the reference value - VR is given by the average of measurements from all laboratories, with the respective expanded uncertainty, according to Table 3.1.2.2.

$$VR~=\frac{\displaystyle\sum^{8}_{i=1}y_i}{8}=\frac{15,81+\dots+15,9}{8}=15,76$$

$$U_{ref}=~2 \times \sqrt{\sum^{8}_{i=1}\frac{u^2(lab_i)}{8}}=2 \times \sqrt{\frac{0,152^2+\dots+0,291^2}{8}}=0,44$$

Therefore, for the rotation range of 2000 rpm we have $ VR=15,76~kW;~U=0,44~kW. $

Screenshot_17

Table 3.1.2.2: Grubbs test for detecting outliers.

Taking laboratory 1 as an example and for the 1200 rpm range we have:

Normalized Error - $ E_n $

$$E_n~=~\frac{\mid Res_{lab}~-~VR\mid}{\sqrt{U^2_{lab}~+~U^2_{ref}}}=\frac{\mid 8,862~-~8,744\mid}{\sqrt{0,153^2~+~0,246^2}}=0,393$$

Laboratory Lab average. VR Trend $U_{lab}$ $U_{ref}$ U $E_n$ W P-value
1 8,86 8,78 0,081 0,170 0,246 0,299 0,272 0,398 0,528
2 8,62 8,78 -0,164 0,153 0,246 0,289 0,566 1,745 0,187
3 8,79 8,78 0,014 0,064 0,246 0,254 0,054 0,012 0,912
4 8,84 8,78 0,064 0,185 0,246 0,308 0,207 0,252 0,615
5 8,90 8,78 0,116 0,448 0,246 0,511 0,227 0,740 0,390
6 8,93 8,78 0,144 0,206 0,246 0,321 0,449 1,318 0,251
7 8,60 8,78 -0,181 0,218 0,246 0,329 0,550 2,103 0,147
8 8,71 8,78 -0,075 0,319 0,246 0,402 0,185 0,333 0,564

Table 3.1.2.3: EN comparison - range 1200.

(image missing)

Figure 3.1.2.5: Comparison Chart - range 1200.

Taking laboratory 1 as an example and for the 2000 rpm range we have:

Normalized Error - $ E_n $

$$E_n~=~\frac{\mid Res_{lab}~-~VR\mid}{\sqrt{U^2_{lab}~+~U^2_{ref}}}=\frac{\mid 15,809~-~15,698\mid}{\sqrt{0,304^2~+~0,442^2}}=0,208$$

Laboratory Lab average. VR Trend $U_{lab}$ $U_{ref}$ U $E_n$ W P-value
1 15,81 15,76 0,046 0,304 0,442 0,536 0,086 0,040 0,842
2 15,33 15,76 -0,428 0,272 0,442 0,519 0,826 3,701 0,054
3 15,78 15,76 0,015 0,115 0,442 0,457 0,032 0,004 0,947
4 15,92 15,76 0,155 0,334 0,442 0,554 0,279 0,461 0,497
5 15,89 15,76 0,122 0,801 0,442 0,915 0,134 0,254 0,614
6 16,02 15,76 0,256 0,370 0,442 0,577 0,444 1,284 0,257
7 15,46 15,76 -0,302 0,393 0,442 0,592 0,511 1,820 0,177
8 15,90 15,76 0,137 0,582 0,442 0,731 0,188 0,347 0,556

Table 3.1.2.4: EN comparison - 2000 range.

(image missing)

Figure 3.1.2.6: Comparison Chart - 2000 range.

Extrapolating the results to the other rotation ranges and given that the other ranges are compatible with the reference. Therefore, all laboratories are compatible with the reference for each speed range.

16.3.1.3 Trend and Linearity Analysis

In situations where laboratories measure several times at various points and the variable under analysis presents natural variations, the model with error in the variables was used (for more details see Performance Statistics). In this section we will take laboratories 1 and 2 as an example, the other laboratories are calculated in a similar way.

Considering the model proposed in the Trend and Linearity section, we tested the laboratory’s competence by developing the following tests:

  • Hypothesis 1: We will test the following hypotheses

Screenshot_18

(there are missing elements in the equation above)

In this case, we are testing whether the laboratory’s additive trend can be considered null.

  • Hypothesis 2: We will test the following hypotheses

Screenshot_19

(there are missing elements in the equation above)

In this case, we are testing whether the laboratory’s multiplicative tendency can be considered equal to one.

  • Hypothesis 3: We will test the following hypotheses

Screenshot_20

(there are missing elements in the equation above)

In this case, we are testing whether the additive trend is zero and whether the multiplicative trend can be considered equal to one.

Below we present a summary of the data from laboratories 1 and 2, to apply the trend and linearity.

Laboratory Rotation Minimum 1st Quartile Mean 3rd Quartile Maximum Standard Deviation Amplitude Frequency

Screenshot_21

Table 3.1.3.1: Descriptive Statistics for corrected power measurements for each speed range in laboratories 1 and 2.

1200 2000 3000 3600 4400 5200 5600 6000 6400
8,74 15,70 26,47 31,13 36,91 43,81 46,87 48,65 49,52

Table 2.3.3.2: Corrected Power Measurements for each rotation range of the reference laboratory.

laboratory 1200 2000 3000 3600 4400 5200 5600 6000 6400
1 0,085 0,152 0,258 0,302 0,357 0,426 0,457 0,479 0,489

Table 3.1.3.2: Combined uncertainty of corrected Power for each rotation range of laboratory 1.

1200 2000 3000 3600 4400 5200 5600 6000 6400
0,123 0,221 0,373 0,439 0,518 0,614 0,656 0,681 0,692

Table 3.1.3.3: Combined uncertainty of corrected Power for each rotation range of the reference laboratory.

See below the results obtained by the Action software for an example.

(image missing)

Figure 3.1.3.1: Trend and Linearity test results obtained by the Action software.

(image missing)

Figure 3.1.3.2: Graph of the Trend and Linearity test obtained by the Action software.

From the results obtained in Figure 2.3.3.1 and Figure 2.3.3.2, we verified that laboratory 1 does not present satisfactory measurements, as in the joint test of additive and multiplicative trends (hypothesis 3) we rejected $H_0$, as the p-value=0.016 is below the significance level of 0.05. Furthermore, when we tested only the null additive trend hypothesis, we realized that we did not reject $H_0$, as the p-value is equal to 0.714. For the multiplicative trend, we reject the hypothesis $H_0$ multiplicative trend equal to 1, as the p-value is equal to 0.047 (considering a significance level of 0.05). Therefore, the laboratory needs to better evaluate its measurement system, as it presents significant linearity and a pattern of positive trends.

Now for laboratory 2 we obtain the following results:

(image missing)

Figure 3.1.3.3: Trend and Linearity test results obtained by the Action software.

(image missing)

Figure 3.1.3.4: Graph of the Trend and Linearity test obtained by the Action software.

From the results obtained in Figure 2.3.3.3 and Figure 2.3.3.4, we verified that laboratory 2, like laboratory 1, does not present satisfactory measurements, as in the joint test of additive and multiplicative trends (hypothesis 3) we rejected $H_0$, as the p-value=0.016 is below the significance level of 0.05. Furthermore, when we tested only the null additive trend hypothesis, we realized that we did not reject $H_0$, as the p-value is equal to 0.95. For the multiplicative trend, we reject the hypothesis $H_0$ multiplicative trend equal to 1, as the p-value is equal to 0.022 (considering a significance level of 0.05). Therefore, the laboratory needs to better evaluate its measurement system, since it presents significant linearity and, unlike laboratory 1, it presents a pattern of negative trends.

16.3.2 Proficiency Test for glassware committee

In this EP, the same item is sequentially distributed among participants, so that each participant performs several measurements at the same point (level or threshold). A 50 mL volumetric flask is sequentially distributed among 6 laboratories, where each one performs 10 volume measurements. One way of presenting the data obtained can be seen in table 3.2.1, where Yij represents the j-th measurement carried out by the ith laboratory.

Laboratories:

1 2 … k
$Y_{11}$ $Y_{21}$ … $Y_{k1}$
$Y_{12}$ $Y_{22}$ … $Y_{k2}$
$ \vdots $ $ \vdots $ $ \vdots $ $ \vdots $
$Y_{1n1}$ $Y_{1n2}$ … $Y_{knk}$

Table 3.2.1: Laboratory data.

Due to the fact that all laboratories measure the same item, we need a model that accounts for possible dependency between all measurements.

In this EP, a 50 mL volumetric flask, illustrated in figure 3.2.1, is distributed sequentially to 6 participating laboratories. Participants perform 10 measurements of the volume of the volumetric flask.

The procedure consists of filling the flask with water and obtaining the volume value using the measurement equation defined by:

$$V=\left(\frac{m_{H_2O}}{\rho_{H_2O}-\rho_{\text{Ar}}}\right)\left(1-\frac{\rho_{\text{Ar}}}{\rho_{mass}}\right)\left[1-\omega(t_{H_2O}-t_{\text{ref}})\right]+\varepsilon~~(3.2.1)$$

in which

  • $V$: represents the volume of the volumetric flask;
  • $t_{ref}$: represents the reference water temperature (20.0ºC);
  • $t_{H2O}$: represents the water temperature during the test (19.9ºC);
  • $m_{H2O}$: represents the mass of water (49.852 g);
  • $ρ_{H2O}$: represents the specific mass of water (0.9983 g/cm3);
  • $ρ_{Ar}$: represents the specific mass of air (0.00110852 g/cm3);
  • $ρ_{mass}$: represents the mass of the standard mass (7.8 g/cm3);
  • $ω$: represents a correction factor.
  • $ε$: represents repeatability.

(image missing)

Figure 3.2.1: Illustration of the 50 mL volumetric flask.

The six laboratories must calculate the combined uncertainty using the measurement equation (3.2.1), as described in ISO GUM (2008). Thus, the combined uncertainty for the volume is given by:

$$u^2_c(V)=\left(\frac{\partial V}{\partial m_{H_2O}}\right)^2 u^2(m_{H_2O})+\left( \frac{\partial V}{\partial\rho_{H_2O}}\right)^2 u^2(\rho_{H_2O})+\left( \frac{\partial V}{\partial t_{H_2O}} \right)^2 u^2(t_{H_2O})+$$

$$+\left(\frac{\partial V}{\partial \rho_{\text{Ar}}} \right)^2 u^2(\rho_{\text{Ar}})+\left(\frac{\partial V}{\partial \rho_{\text{mass}}}\right)^2 u^2(\rho_{\text{mass}})+u^2(\varepsilon).$$

Below, we present the information necessary to calculate the combined volume uncertainty.

Equipment Uncertainty
Scales 0.0004 g
Barometer 0.76 mbar
Thermometer 1 0.05ºC
Thermometer 2 0.3ºC
Thermohygrograph 4,00%
Density meter 0.0003 g/mL
Pasta 0.03 mg

Table 3.2.2: Information for calculating uncertainty.

With the information obtained in table 3.2.2, we calculated the combined and expanded uncertainty of the 6 laboratories (for more details see the Measurement Uncertainty content). The results of the proficiency test for the committee are summarized in table 3.2.3.

Note L1 L2 L3 L4 L5 L6
1 49,927 49,993 49,984 49,987 49,966 49,88
2 49,937 49,994 49,984 49,987 49,965 49,904
3 49,904 49,994 49,979 49,99 49,966 49,917
4 49,913 49,994 49,977 49,987 49,967 49,915
5 49,924 49,995 49,985 49,989 49,966 49,914
6 49,928 49,995 49,987 49,987 49,969 49,907
7 49,93 49,995 49,988 49,989 49,965 49,871
8 49,931 49,995 49,986 49,985 49,965 49,897
9 49,925 49,995 49,987 49,989 49,967 49,895
10 49,91 49,995 49,986 49,984 49,968 49,918
Average 49,923 49,994 49,984 49,987 49,966 49,902
$u_c$(lab) 0,004 0,014 0,008 0,003 0,009 0,01
U(lab) 0,008 0,028 0,016 0,006 0,018 0,02

Table 3.2.3: Volume measurements (mL): EP of glassware.

From figure 3.2.2 we observe that laboratories L1 and L6 present measurements that are slightly different from the other laboratories.

(image missing)

Figure 3.2.2: Boxplot of the glassware EP.

The reference value for this EP are determined by a calibration process. In our application, laboratory L5 is recognized as having competence, which exceeds the competence of other laboratories, to calibrate the item of interest (volumetric flask), as it is a member of the RBC (Brazilian Calibration Network) and is also a member of international EP programs. For this reason, we will consider its mean and standard deviation as the reference, that is, $ \overline{X_{\text{ref}}} = 49,9664 $, $ u_{\text{ref}} = 0,009 $ and $ U_{ref} = 0,018. $

To test the consistency of the group of laboratories we have the hypotheses

Screenshot_22

(there are missing elements in the equation above)

and we consider $ C={\bf I}_k $ and $ {\bf a}={\bf 0}_k $, with this, we can test the following joint hypotheses:

Screenshot_23

(there are missing elements in the equation above)

If we reject $ H_0 $, we concluded that the group of laboratories is not consistent. In this case, we start to develop individual tests with respect to the reference value. So, let’s test the following hypotheses:

Screenshot_24

(there are missing elements in the equation above)

for $ i=1,\ldots,k $ (in this case, we will consider in (3.2.2) that $ C $ is a row vector ($ 1 \times k $) with $ 1 $ in the i-th position and zero in the others, and $ a = 0 $). If we don’t reject $ H_0 $, we conclude that the measurements from the ith laboratory are TRVG-satisfactory.

Next, the first step is to identify the occurrence of outliers, for this, we will use the Grubbs test (for more details see Extreme value test (Grubbs)). Table 3.2.4 summarizes the results obtained by the Grubbs test.

Average 49,958 Standard Deviation 0,0395
alpha 0,05 Critical Value 1,72
Laboratory Average Z Status
L1 49,923 0,83 ok
L2 49,995 0,85 ok
L3 49,984 0,61 ok
L4 49,987 0,68 ok
L6 49,902 1,32 ok

Table 3.2.4: Grubbs Test (Extreme Value Test).

In table 3.2.4, we present the trend estimates, the test results via the W test statistic for individual laboratories, as well as the En test score.

Test Laboratory Lab average. VR Trend $U_{lab}$ $U_{ref}$ U $E_n$ W P-value
L1 x L5 L1 49,923 49,966 -0,0431 0,016 0,018 0,024 1,8 21,3 4.02E-06
L2 x L5 L2 49,995 49,966 0,0285 0,028 0,018 0,033 0,9 8,1 0,0045
L3 x L5 L3 49,984 49,966 0,0183 0,016 0,018 0,024 0,8 3,8 0,05
L4 x L5 L4 49,987 49,966 0,0214 0,006 0,018 0,019 1 5,6 0,018
L6 x L5 L6 49,902 49,966 -0,0642 0,02 0,018 0,027 2,4 45,3 1.70E-11

Table 3.2.4: Trend and tests for the competence of the five laboratories.

Considering α= 0.01 (test significance level) to decide between the hypotheses, we reject the hypothesis that the laboratories' trends are all null and thus, we conclude that the group of laboratories is not consistent. Furthermore, it follows from the multiple comparison tests that only measurements from laboratories L3 and L4 are TRVG-satisfactory (p-values ​​0.05 and 0.018, respectively). However, it follows from the En score that the measurements from laboratories L2, L3 and L4 are satisfactory (En ≤ 1). Note that the L2 lab measurements are En-satisfactory, but they are not TRVG-satisfactory. This is a consequence of the magnitude of their combined standard deviation (s=0.014), which is the largest standard deviation among participants.

(image missing)

Figure 3.2.3: Laboratory compatibility comparison graph.

Terms and definitions

In this section we present some terms and definitions described in the ABNT NBR ISO/IEC 17043:2010 standard [9].

  1. Coordinator: one or more people with responsibility for organizing and managing all activities involved in the operation of a proficiency testing program;
  2. Interlaboratory comparison: organization, performance and evaluation of measurements or tests with the same or similar items by two or more laboratories according to predetermined conditions;
  3. Outlier: observation of a set of data that appears to be incompatible with the rest of the set; NOTE: An outlier may come from a different population, or be the result of an incorrect entry or other gross error.
  4. Participant: laboratory, organization, or individual that receives proficiency testing items and submits the results for review by the proficiency testing coordinator; NOTE: In some cases, the participant may be an inspection body.
  5. Proficiency Tests: evaluation of participants' performance in relation to pre-established criteria through interlaboratory comparisons;
  6. Proficiency Test Item: product sample, artifact, reference material, part of equipment, standard measurement, set of data or other information used for the proficiency test;
  7. Proficiency testing coordinator: organization that assumes responsibility for all tasks in the development and operation of a proficiency testing program;
  8. Proficiency Testing Program: proficiency testing designed and operated for one or more cycles of a specified area of testing, measurement, calibration or inspection; NOTE: A proficiency testing program may encompass a particular type of test, calibration, inspection, or a series of tests, calibrations, or verifications of proficiency test items.

Bibliographic References

[1] WEEC DOC. 19-1990: Guidelines for the Expression of the Uncertainty of Measurement in Calibration.

[2] WEEC DOC. 15-1987: International Measurements Audits.

[3] WEEC: International Inter-Laboratory Comparisons - A Tool for Gaining Mutual Confidence in Calibration and Testing Results - K. Brinkmann.

[4] Basic draft “Interlaboratory Comparisons” DICLA - GT-2.

[5] NIT-DICLA-026 (2008), ``Requirements on the participation of testing and calibration laboratories in proficiency testing activities''.

[6] EAL -P7 - 1996 - EAL Interlaboratory Comparisons.

[7] ISO GUM (2008), Measurement data evaluation - Guide for expressing measurement uncertainty, 1st Brazilian edition.

[8] DOQ-CGCRE-005 - Guidelines for the Organization of Interlaboratory Comparisons by DICLA Technical Committees, Ver. 00, September 2002.

[9] ISO/IEC 17043:2010. Conformity assessment-general requirements for Proficiency Testing, 2010.

[10] ABNT NBR ISO 1585 - Road Vehicles - Engine test code - Effective Net Power, Rio de Janeiro, 1996.

[11] Richardson, H., Wood, D., Whitby, J., Lannigan, R., and Fleming, C. (1996). Quality Improvement of Diagnostic Microbiology Through a Peer-group Proficiency Assessment Program: a 20 year Experience in Ontario. Archives of Pathology and Laboratory Medicine, 120, 445-455.

[12] ISO/IEC 17011:2005. Conformity assessment - General requirements for accreditation bodies accreditation of conformity assessment bodies, 2005.

[13] Sen, P. K. and Singer J. M. (2000) - Large Sample Methods in Statistics: An introduction with applications, Chapman & Hall.

[14] Leão, D. P. Jr., Aoki, R. and Silva, G. F., (2009) - Statistical Analysis of Bias and Linearity.

[15] ISO 13528:2005, Statistical methods for use in proficiency testing by interlaboratory comparisons.

[16] F. E. Grubbs (1969) - Procedures for Detecting Outlying Observations in Samples, Technometrics, TCTMTA, Vol. 11, No. 4, pp. 1-21.

[17] ISO Guide 34, General requirements for the competence of reference material producers.

[18] ISO Guide 35, Reference materials — General and statistical principles for certification.

[19] ISO 13528:2005, Statistical methods for use in proficiency testing by interlaboratory comparisons;.