23.4 Accuracy
In this section, we will evaluate the accuracy for this we use section VI of RDC No. 166. In article 44, the accuracy of an analytical method is defined as the degree of agreement between the individual results, obtained by the method under study, in relation to a reference value accepted as true. The test preparation criteria are described in RDC No. 166 section VI.
| Analyte, % | Mass Fraction (C) | Unit | Average recovery, % |
|---|---|---|---|
| 100 | 1 | 100% | 98-102 |
| 10 | 10 $^{-1}$ | 10% | 98-102 |
| 1 | 10 $^{-2}$ | 1% | 97-103 |
| 0,1 | 10 $^{-3}$ | 0,1% | 95-105 |
| 0,01 | 10 $^{-4}$ | 100 ppm (mg\kg) | 90-107 |
| 0,001 | 10 $^{-5}$ | 10 ppm (mg\kg) | 80-110 |
| 0,0001 | 10 $^{-6}$ | 1 ppm (mg\kg) | 80-110 |
| 0,00001 | 10 $^{-7}$ | 100 ppb (μg\kg) | 80-110 |
| 0,000001 | 10 $^{-8}$ | 10 ppb (μg\kg) | 60-115 |
| 0,0000001 | 10 $^{-9}$ | 1 ppb (μg\kg) | 40-120 |
Table 23.4.1: Criteria for evaluating accuracy. Source: AOAC, 2016
The recovery measurement equation is given by:
$$\text{Rec}=\frac{C_o}{C_t}\times 100$$
in which
- Rec: is the recovery, in ($ \char37 $);
- is the concentration obtained in $ mg/mL; $
(there are missing elements in the equation above)
- is the theoretical concentration on $ mg/mL; $
(there are missing elements in the equation above)
Let us consider a simple random sample $ \text{Rec}_1,\text{Rec}_2,\ldots,\text{Rec}_n $, obtained from a population with normal distribution, with mean $ \mu=100 $ and variance $ \sigma^2 $ unknown. As in this case the variance is unknown, we will use the sample variance $ s^2 $ instead of $ \sigma^2 $. So, we have to
$$T=\frac{|\overline{\text{Rec}}-100|}{s/\sqrt{n}}\sim t_{(n-1)}\quad ~(1)$$
that is, the variable $ T $ has a Student t distribution with $ n-1 $ degrees of freedom.
So, when we set the significance level $ \alpha $, we obtain from the Student t distribution table with $ n-1 $ degrees of freedom, the value $ t_{((n-1),\alpha/2)} $, which satisfies
$$\mathbb{P}\left(-t_{((n-1),\alpha/2)}\leq T\leq t_{((n-1),\alpha/2)}\right)=1-\alpha$$
Analogously to the previous case, we obtain that
$$\mathbb{P}\left(-t_{((n-1),\alpha/2)}\leq \frac{\overline{\text{Rec}}-100}{s/\sqrt{n}}\leq t_{((n-1),\alpha/2)}\right)=1-\alpha$$
i.e.
$$\mathbb{P}\left(\overline{\text{Rec}}-t_{((n-1),\alpha/2)}\frac{s}{\sqrt{n}}\leq100\leq \overline{\text{Rec}}+t_{((n-1),\alpha/2)}\frac{s}{\sqrt{n}}\right)=1-\alpha.$$
Therefore, the interval with $ 100(1-\alpha)\char37 $ trustworthy for $ \mu=100 $, with unknown variance, will be given by
$$IC(\mu=100,1-\alpha)=\left(\overline{\text{Rec}}-t_{\alpha/2}\frac{s}{\sqrt{n}};\overline{\text{Rec}}+t_{\alpha/2}\frac{s}{\sqrt{n}}\right).$$
To facilitate the execution of the test, we can follow the steps:
- Establish the hypotheses:
We fix $\mu=\mu_0=100 $.
(there are missing elements in the equation above)
$\mu\neq\mu_0 \quad \text{(teste bilateral)} $;
(there are missing elements in the equation above)
-
Set the significance level $ \alpha $.
-
Determine the critical region.
As the test is two-sided, we determine the critical points $ -t_{\alpha/2} $ and $ t_{\alpha/2} $ such that $ \mathbb{P}[T \ > \ t_{\alpha/2}]=\mathbb{P}[T \ < -t_{\alpha/2}]=\alpha/2 $ from the Student t distribution with $ n-1 $ degrees of freedom.
- Calculate, under the null hypothesis, the value:
$$T_{\text{obs}}=\frac{|\overline{\text{Rec}}-100|}{\frac{s}{\sqrt{n}}}$$
in which
- $ \overline{\text{Rec}} $: average recovery value.
- $ s $: value of the sample standard deviation.
- $ n $: sample size.
- Criteria:
Two-sided test: if $ T_{\text{obs}} \ > \ t_{\alpha/2} $ or if $ T_{\text{obs}} \ < \ -t_{-\alpha/2} $, we reject $ H_0 $. Otherwise, we do not reject $ H_0 $.
- The p-value in the two-sided test is given by
$$\text{p-valor} = \mathbb{P}[|t| \ > \ |T_{\text{obs}}||H_0]=2\mathbb{P}[T \ > \ |T_{\text{obs}}| | H_0].$$
(image missing)
- As we saw previously, the confidence interval is given by
$$IC(\mu,1-\alpha)=\left(\overline{\text{Rec}}-t_{\alpha/2}\frac{s}{\sqrt{n}};\overline{\text{Rec}}+t_{\alpha/2}\frac{s}{\sqrt{n}}\right)$$
According to Gustavo González [2], the test statistics (equation 1) can be written as:
$$T=\frac{\overline{\text{Rec}}-100}{u(\text{Rec})}$$
in which
$$u(\text{Rec})=\sqrt{\left(\frac{\partial \text{Rec}}{\partial C_o}\right)^2u^2(C_o)+\left(\frac{\partial \text{Rec}}{\partial C_t}\right)^2u^2(C_t)+u^2(\varepsilon)}$$
in which
- Rec: is the recovery, in (%);
- is the concentration obtained in $ mg/mL. $ The uncertainty is given by the sample preparation (for more details see calculation of uncertainty due to solutions)
(image missing)
- is the theoretical concentration on $ mg/mL; $ Uncertainty is given by the reference solution certificate (ISO GUIDE);
(image missing)
- is the standard deviation of the mean given by $ s/\sqrt{n}. $
(image missing)
Furthermore, we have to:
$$\frac{\partial \text{Rec}}{\partial C_o}=\frac{1}{C_t}$$
$$\frac{\partial \text{Rec}}{\partial C_t}=-\frac{C_o}{C^2_t}$$
According to the LGC/VAM protocol described in the article by Gustavo González [2], if the degrees of freedom associated with the recovery uncertainty are known, T is compared with the two-tailed tabulated value $ t_{(\nu,1-\alpha)} $ for the number of degrees of freedom $ \nu $ with $ (1-\alpha)\char37 $ trustworthy. And if$ T\leq t_{tab}, $ the consensus recovery is not significantly different from 1. Alternatively, instead of the $ t_{tab}, $ We can use the coverage factor $ k $ for comparison. Typical values are $ k=2 $ or 3 for 95% or 99% confidence, respectively. So
Se $ \frac{|\overline{\text{Rec}}-100|}{u(\text{Rec})}\leq k $, recovery is not significantly different from 100;
Se $ \frac{|\overline{\text{Rec}}-100|}{u(\text{Rec})}> k $, the recovery is significantly different from 100 and the analytical result has to be corrected by $ \overline{\text{Rec}}. $
Another form of evaluation described in Gustavo González [2] is to evaluate the acceptable limits given by regulatory bodies.
Example 1.5.1
In this section, the relative expanded uncertainty was calculated, which corresponds to 0.29%. Below, we present the collected data.

For normal specificity we obtain the following measurements:
| Concentration |
|---|
| 0,182856676 |
| 0,201313318 |
| 0,190572332 |
| 0,181481821 |
| 0,181340121 |
| 0,189112134 |
| 0,190115447 |
| 0,176140016 |
| 0,19486329 |
| 0,183852917 |
| Sample | Mass (mg) | Pd Vol (mL) | Theoretical concentration (mg/mL) | Area (mAU*s) | Concentration obtained (mg/mL) | Recovery (%) |
|---|---|---|---|---|---|---|
| 1 | 728,68 | 0,00 | 0,1912 | 7447,27975 | 0,18716 | 97,88 |
For normal specificity we have:
$$\text{Rec}=\frac{C_o}{C_t}\times100=\frac{0,18716}{0,1912}\times100=97,88\char37\pm 3,9\char37$$
Next, we test the accuracy using the Action Stat software and obtain the following results:
- Firstly, we present the Accuracy function in Action Stat.
(image missing)
- The next step is to fill in the Accuracy window
(image missing)
- Finally, we obtain the following results:
(image missing)
(image missing)
(image missing)
Therefore, recovery is within the acceptance criteria.