Analytical Method Validation 6 of 8

Analytical Method Validation

23.5 Detection Limit

Chapter 5 of 7 4 min read

In this section, we will analyze the limit of detection (LD), according to the criteria of RDC No. 166. In it, the limit of detection is defined as the smallest amount of the analyte present in a sample that can be detected, however, not necessarily quantified, under the established experimental conditions. In other words, the detection limit is the lowest analyte concentration that can be reliably detected and distinct from zero (or the system noise level), but not necessarily quantified at the concentration at which a measured value is greater than the uncertainty associated with it.

I - for the sample blank, according to the following equation:

$$LD=X+t_{(n-1;1-\alpha)}s$$

in which

$ \bullet $ X is the average of the values of the whites in the sample; 5%

$ \bullet $ t is the quantile of the Student distribution, dependent on the sample size and confidence level, n is the number of samples,

$ \bullet $ α is the significance level and

$ \bullet $ s is the sample standard deviation of the whites in the sample.

II - for the blank sample with addition of the lowest acceptable concentration of the analyte, according to the following equation:

$$LD=0+t_{(n-1;1-\alpha)}s$$

where:

$ \bullet $ t is the quantile of the Student distribution, depending on the sample size and confidence level, and n is the number of samples,

$ \bullet $ α is the significance level and

$ \bullet $ s is the sample standard deviation of the sample blanks, with addition, with n-1 degrees of freedom.

LD can be expressed in response units ($ Y_{LD} $) and is normally taken as three times the noise level for techniques with continuous recording (e.g. chromatography). The criterion adopted by RDC No. 166 is 3.3. In this way we obtain:

$$Y_{LD}=Y_{\text{branco}}+3,3\times s_{\text{branco}} \tag{1}$$

which is equivalent to method I, but adopting a confidence level of approximately 99.5%. It is worth remembering that in this method, we must have a minimum of 10 blanks from independent samples. Alternatively, when the blank sample cannot produce any response, that is, voltammetry (electroanalytical methods). In this case, we use 10 independent sample blanks enriched in the lowest acceptable concentration of the analyte, and are then measured, $ Y_{LD}=3,3s, $ being the standard deviation of a set of measurements.

However, LDs expressed in sign units are terrible to handle. According to Gustavo González, it is more advisable to use analyte concentration levels. With this, $ Y_{LD} $ are converted to $ Z_{LD} $ using the calibration function:

$$Z_{LD}=\dfrac{Y_{LD}-\widehat{\beta}_0}{\widehat{\beta}_1}$$

Therefore, adopting the null intercept criterion and using equation (1) without $ Y_{\text{branco}} $, we get:

$$Z_{LD}=\frac{3,3 s_{\text{branco}}}{\widehat{\beta}_1}$$

Therefore, we obtain the third criterion of RDC No. 166 [1].

III - For determination based on analytical curve parameters, the detection limit can be calculated by:

$$LD=\frac{3,3s_{*}}{\widehat{\beta}_1}$$

Where:

  • $ \widehat{\beta}_1 $ is the slope of the calibration curve,
  • $ s_{*} $ is the standard deviation and can be obtained in 3 ways:

a) the standard deviation of the intercept with the Y axis of, at least, 3 calibration curves constructed containing analyte concentrations close to the supposed quantification limit;

b) the residual standard deviation of the regression line

c) from the calibration curve resulting from the analysis of an appropriate number of blank samples.

Example 1.6.1

Next, we present the data collected for the detection limit. For this, we use the same data collected from linearity.

Concentration Area
0,24 8597,852
0,24 8597,258
0,24 8596,783
0,24 8596,908
0,24 8597,301
0,24 8597,496
0,27 9607,39
0,27 9607,714
0,27 9607,443
0,27 9608,133
0,27 9607,176
0,27 9607,247
0,3 10617,69
0,3 10617,97
0,3 10617,99
0,3 10617,95
0,3 10617,8
0,3 10617,8
0,33 11627,84
0,33 11628,43
0,33 11628,01
0,33 11628,33
0,33 11628,34
0,33 11628,1
0,36 12637,93
0,36 12638,12
0,36 12638,54
0,36 12638,19
0,36 12638,63
0,36 12638,32

Table 23.5.1: Data for calculating the detection limit

Next, we calculate the detection limit using the Action Stat software and obtain the following results:

  1. Firstly, we present the detection threshold function in Action Stat.

(image missing)

  1. The next step is to fill in the detection limit window

(image missing)

Below, we present the results obtained by the Action Stat software.

(image missing)

(image missing)

With that, let’s calculate the detection limit:

$$LD=\frac{3,3\sigma_{\text{resíduos}}}{\widehat{\beta}_1}=\frac{3,3\times 0,278311917}{33675,66936}=2,72\times 10^{-5}$$

↑